← Product Code [KPM](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPM) · K003451

# DENVER PAK W/ DOUBLE-VALVE ASCITES SHUNT,DENVER PAK W/ SINGLE-VALVED ASCITES SHUNT, DENVER PAK, DOUBLED-VALVED ASCITES T (K003451)

_Denver Biomedicals, Inc. · KPM · Jan 2, 2001 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPM/K003451

## Device Facts

- **Applicant:** Denver Biomedicals, Inc.
- **Product Code:** [KPM](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPM.md)
- **Decision Date:** Jan 2, 2001
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5955
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Regulatory Identification

A peritoneo-venous shunt is an implanted device that consists of a catheter and a pressure activated one-way valve. The catheter is implanted with one end in the peritoneal cavity and the other in a large vein. This device enables ascitic fluid in the peritoneal cavity to flow into the venous system for the treatment of intractable ascites.

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(2) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” and
(3) Backflow specification and testing to prevent reflux of blood into the shunt.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPM/K003451](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPM/K003451)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).