Colonic Plus Regular, Small, and Straight Shape Hydrokit
K180800 · Colonic Plus · KPL · Jun 1, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K180800
Device Name
Colonic Plus Regular, Small, and Straight Shape Hydrokit
Applicant
Colonic Plus
Product Code
KPL · Gastroenterology, Urology
Decision Date
Jun 1, 2018
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5220
Device Class
Class 2
Indications for Use
The use of this device is restricted to colon cleansing when medically indicated, such as before radiological or endoscopic examination.
Device Story
Colonic Plus Hydrokits (Regular, Small, Straight) are plastic, single-use accessories for colonic irrigation systems; used to instill water into the colon via a nozzle inserted 2-3 inches into the rectum to evacuate lower colon contents. Operated by clinicians in a clinical setting. Device components include a speculum, obturator, plastic water line, EVA corrugated/segmented waste hose, and surgical lubricant. The device facilitates bowel preparation for medical procedures. Benefits include effective colon cleansing for diagnostic or therapeutic access.
Clinical Evidence
No clinical studies submitted. Substantial equivalence established via bench testing and biocompatibility assessment.
Technological Characteristics
Plastic, single-use colonic irrigation accessory. Components: speculum, obturator, PVC water line, EVA corrugated/segmented waste hose, surgical lubricant. Non-sterile. Mechanical operation via manual insertion and connection to irrigation system. Biocompatibility testing performed for cytotoxicity, irritation, and sensitization.
Indications for Use
Indicated for colon cleansing when medically indicated, such as before radiological or endoscopic examination.
Regulatory Classification
Identification
A colonic irrigation system is a device intended to instill water into the colon through a nozzle inserted into the rectum to cleanse (evacuate) the contents of the lower colon. The system is designed to allow evacuation of the contents of the colon during the administration of the colonic irrigation. The device consists of a container for fluid connected to the nozzle via tubing and includes a system which enables the pressure, temperature, or flow of water through the nozzle to be controlled. The device may include a console-type toilet and necessary fittings to allow the device to be connected to water and sewer pipes. The device may use electrical power to heat the water. The device does not include the enema kit (§ 876.5210).
Predicate Devices
Transcom Colon Hydrotherapy Models HC-1 and HC-1 Classic And Hydrokit (K131852)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 1, 2018
Colonic Plus Rosa Alvarez Owner 73-925 Hwy. 111, Ste. L Palm Desert, CA 92260
Re: K180800
Trade/Device Name: Colonic Plus Regular, Small, and Straight Shape Hydrokits Regulation Number: 21 CFR& 876.5220 Regulation Name: Colonic Irrigation System Regulatory Class: II Product Code: KPL Dated: March 30, 2018 Received: April 4, 2018
Dear Rosa Alvarez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180800
#### Device Name
Colonic Plus Regular, Small, and Straight Shape Hydrokits
Indications for Use (Describe)
The use of this device is restricted to colon cleansing when medically indicated, such as before radiological or endoscopic examination.
Type of Use (*Select one or both, as applicable*)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Colonic Plus | Technical Evaluation<br>Documentation | # Document<br>Colonic Plus Hydrokits |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Section 05: | | |
| Traditional 510(k) Summary | | |
| DATE OF SUBMISSION: | 03/20/2018 | |
| SUBMITTER NAME: | Colonic Plus | |
| SUBMITTER ADDRESS: | 73-925 HWY 111, STE L<br>Palm Desert, CA 92260 | |
| CONTACT: | Rosa Alvarez | |
| TELEPHONE: | 760-340-0720 | |
| Fax: | 760-340-0189 | |
| E-mail: | Rosa.alvarez@yahoo.com | |
| DEVICE TRADE NAME: | Colonic Plus Regular, Small and Straight Shape Hydrokits | |
| COMMON NAME: | COLONIC IRRIGATION SYSTEM | |
| CLASSIFICATION NAME: | COLONIC IRRIGATION SYSTEM (21 CFR 876.5220) | |
| Procode: | KPL | |
| Regulatory Class: | II | |
| PREDICATE DEVICE: | Transcendencias Comerciales, S.L. K131852- Hydrokit<br>Trade Name: Transcom Colon Hydrotherapy Models HC-1<br>and HC-1 Classic And Hydrokit | |
### Device Description: Colonic Plus Hydrokits
The Colonic Plus Regular, Small and Straight shape Hydrokits is a plastic, non-sterile, single use product colonic irrigation system that is used as an accessory with lower bowel Colonic irrigation systems intended to instill water into the colon through a nozzle inserted 2" 3" into the rectum to evacuate the contents of the lower colon.
## Technical Characteristics Compared to the Predicate Device:
The proposed Colonic Plus Regular, Small and Straight shape Hydrokits (See Figues 3A-3C) are substantially equivalent in intended use, indications for use, design and material used in the Transcoms Hydrokit manufactured by Transcendenicas Comerciales, S.L.(K131852) (See figure 3D). The only differences between the new device and the predicate device is that the ring on the obturator is crescent and on the predicate device is round and that the tip of the obturator is slightly wider on the new device. Lastly on the Colonic Plus Straight shape speculum, it has no curve and is slightly thinner than the predicate device.
- > See diagrams below
{5}------------------------------------------------
# Proposed Devices:
Image /page/4/Figure/2 description: The image shows three different types of Colonic Plus Hydrokits, labeled as Figure 3A, 3B, and 3C. Figure 3A is the Regular Hydrokit, Figure 3B is the Small Hydrokit, and Figure 3C is the Straight Shape Hydrokit. The image uses arrows to point out features of each kit, such as the crescent ring, slightly wider insertion tip, and the absence of a bulb/curve in the straight shape kit. The straight shape kit is also noted to be slightly thinner.
# Predicate Devices:
Image /page/4/Picture/4 description: The image shows a medical device inside of a clear plastic bag. The device is white and has a round ring on one end. An arrow points to the round ring and the text "Round ring" is next to the arrow.
Figure 3D: Transcom Hydrokit (K131852)
{6}------------------------------------------------
Proposed Kit Components
Image /page/5/Picture/2 description: The image shows a collection of medical tubing and equipment on a dark surface. There are two corrugated tubes, one straight tube coiled with a label, and a plastic syringe-like device with a handle. The corrugated tubes appear to be connected by a metal piece in the center. The tubes are all clear and made of plastic.
Figure 4A: Colonic Plus Hydrokit (K163040)
# Predicate device Kit Components
Image /page/5/Picture/5 description: The image shows a collection of medical supplies against a black background. There is a long, corrugated tube coiled on the left side of the image. To the right of the tube, there is a smaller, clear tube coiled in a circle, along with a small white packet. Below these items, there is a clear, cylindrical object and a long, thin tool with a loop at one end.
### Figure 4B: Transcom (K131852)
Kit Components packaged with the proposed disposable Colonic Plus Regular, Small and straight shape Hydrokit (plastic water line, EVA corrugated and segmented waste hose, & surgical lubricant) are substantially equivalent to those included with the Transcom Hydrokit manufactured by Transcendencias Comerciales, S.L. (K131852) (See figure 4A above) Surgical Lubricant: Dynarex Sterile Lubricating Jelly K092488
# INTENDED USE COMPARED TO THE PREDICATES DEVICE:
The device is intended for use to cleanse the colon when medically indicated, such as before radiological or endoscopic examination. It has the same intended use as predicate devices Transcom Hydrokit Manufactured by Trascendencias Comerciales, S.L. (K131852).
{7}------------------------------------------------
# SUMMARY DISCUSSION OF NON-CLINICAL DATA:
The Colonic Plus Regular, Small, and Straight Shape Hydrokits are substantially equivalent in indications for use, intended use, design and material as that of the Transcoms Hydrokit (K131852). The water pressure leakage test and the Tensile & Compression tests performed on the finalized Colonic Plus Hydrokits were passed according to our acceptance criteria and were substantial equivalent to the predicate device. We also submitted our finalized Colonic Plus hydrokits (Speculum, Obturator, and PVC water line) to American Preclinical Services for Biocompatability Testing on Cytotoxicity, Irritation and Sensitization which have all passed.
## SUMMARY DISCUSSION OF CLINICAL DATA:
No clinical studies are submitted to support this premarket notification.
### CONCLUSION:
We believe the intended use, the indications for use, the design, materials used, the functionality and the operation of the speculum kit is essentially the same as the predicate devices: Transcoms Hydrokit (K131852). Hence, substantial equivalence of the The Colonic Plus Regular, Small and Straight shape Hydrokits with the legally marketed device may be established.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.