← Product Code [KPI](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPI) · K243079

# NeuroTrac® MyoPlus Pro (MYO120U) (K243079)

_Verity Medical , Ltd. · KPI · Jun 26, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPI/K243079

## Device Facts

- **Applicant:** Verity Medical , Ltd.
- **Product Code:** [KPI](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPI.md)
- **Decision Date:** Jun 26, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5320
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

STIM Electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in adult women. EMG Relaxation muscle training and muscle re-education. ETS Treatment of stress, urge or mixed urinary incontinence by assessing EMG activity of the pelvic floor and strengthening pelvic floor muscles using electrical stimulation. This device is intended to be used by adult woman (aged above 22 years).

## Device Story

NeuroTrac® MyoPlus Pro is a single-channel pelvic floor muscle stimulator; assists patients with urinary incontinence to increase muscle control; overcomes pelvic muscle dysfunction. Inputs: EMG signals from pelvic floor via vaginal probe. Operation: device functions as electromyograph (EMG) and electrical stimulator (STIM); features EMG-triggered stimulation (ETS) mode. Output: symmetrical, rectangular, biphasic electrical stimulation pulses delivered via vaginal probe. Used in clinic or home; operated by physician or patient. Output allows healthcare providers to monitor muscle activity and provide targeted rehabilitation; benefits patient by strengthening pelvic floor muscles and improving urinary control.

## Clinical Evidence

Bench testing only. Testing included electrical safety (IEC 60601-1:2020), electromagnetic compatibility (IEC 60601-1-2:2020), software verification/validation, and cybersecurity risk management. Performance metrics (net charge, phase charge, power density) were calculated and met all pre-determined acceptance criteria.

## Technological Characteristics

Single-channel pelvic floor stimulator; constant current; biphasic, symmetrical, rectangular waveform. Frequency 2-100 Hz. Power: 4xAAA alkaline batteries (6V). Applied parts: vaginal probe (cleared under K122194) and skin electrodes. Standards: IEC 60601-1:2020, IEC 60601-1-2:2020.

## Regulatory Identification

A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).

## Predicate Devices

- NuTrac Pelvator, model PEL 200 ([K083704](/device/K083704.md))
- Medline DeNovo 4Pro Electrical Stimulation Device ([K201290](/device/K201290.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

June 26, 2025

Verity Medical Ltd.
Nigel Verity
Managing Director
Churchtown House
Tagoat, Co. Wexford Y35 XY44
IRELAND

Re: K243079
Trade/Device Name: NeuroTrac® MyoPlus Pro (MYO120U)
Regulation Number: 21 CFR 876.5320
Regulation Name: Nonimplanted Electrical Continence Device
Regulatory Class: II
Product Code: KPI, HCC
Received: May 20, 2025

Dear Nigel Verity:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K243079 - Nigel Verity
Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See

{2}

K243079 - Nigel Verity
Page 3

the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica K. Nguyen -S

Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K243079  |   |
|  Device Name NeuroTrac® MyoPlus Pro (MYO120U)  |   |
|  Indications for Use (Describe) STIM Electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in adult women. EMG Relaxation muscle training and muscle re-education. ETS Treatment of stress, urge or mixed urinary incontinence by assessing EMG activity of the pelvic floor and strengthening pelvic floor muscles using electrical stimulation. This device is intended to be used by adult woman (aged above 22 years).  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

V M V E R I T Y M E D I C A L L T I E
K243079
Page 1 of 4

# TRADITIONAL 510(K) SUMMARY

## 1. SUBMITTER INFORMATION

Applicant
Verity Medical Ltd.
Churchtown House
Tagoat, Co. Wexford
Ireland Y35 XY44

Applicant Contact
Nigel C. Verity, Managing Director
E-mail: nigel@veritymedical.com
Phone: +353 (0) 53 913 2433

Date Prepared
June 23, 2025

## 2. DEVICE NAME

Trade Name of the Device
NeuroTrac® MyoPlus Pro (MYO120U)

Classification Name:
Nonimplanted electrical continence device

Classification Regulation:
21 CFR 876.5320

Device Class:
II

Product Code:
KPI, HCC

Panel:
Gastroenterology/Urology

## 3. PREDICATE DEVICES

|   | 510(k)# | Trade Name  |
| --- | --- | --- |
|  Primary Predicate | K083704 | NuTrac Pelvator, model PEL 200  |
|  Secondary Predicate | K201290 | Medline DeNovo 4Pro
Electrical Stimulation Device  |

Predicate devices were not subjected to any design related recall.

## 4. DEVICE DESCRIPTION:

The NeuroTrac® MyoPlus Pro is a single-channel pelvic floor muscle stimulator (STIM) designed to assist patients with urinary incontinence to increase muscle control and overcome pelvic muscle dysfunction. It can also act as an electromyograph (EMG) and features an EMG-triggered stimulation (ETS) mode. The stimulation waveform is symmetrical, rectangular and biphasic. Stimulation is delivered using a vaginal probe. The vaginal probe included with the subject device was cleared under K122194. The subject device may either be operated directly by the prescribing physician, or by the patient as instructed by the physician.

{5}

V M V E R I T Y

K243079

Page 2 of 4

# 5. INDICATIONS FOR USE:

## STIM

Electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in adult women.

## EMG

Relaxation muscle training and muscle re-education.

## ETS

Treatment of stress, urge or mixed urinary incontinence by assessing EMG activity of the pelvic floor and strengthening pelvic floor muscles using electrical stimulation.

This device is intended to be used by adult woman (aged above 22 years).

# 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES

|  Device & Predicate Device(s): | K243079 (Subject) | K083704 (Primary Predicate) | K201290 (Secondary Predicate)  |
| --- | --- | --- | --- |
|  Device name (Sponsor) | NeuroTrac® MyoPlus Pro (Verity Medical) | NuTrac Pelvator (Verity Scientific) | Medline DeNovo 4Pro Electrical Stimulation Device (Medline Industries)  |
|  Indication for use | STIM
Electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in adult women.
EMG
Relaxation muscle training and muscle re-education.
ETS
Treatment of stress, urge or mixed urinary incontinence by assessing EMG activity of the pelvic floor and strengthening pelvic floor muscles using electrical stimulation.
This device is intended to be used by adult woman (aged above 22 years). | The "nutrac pelvator" pelvic muscle trainer is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women. | For EMG mode: - Relaxation muscle training and muscle re-education
For EMG Triggered Stimulation (ETS) mode (nonimplanted electrical continence device only):- Acute and ongoing treatment of stress, urge or mixed urinary incontinence and where the following results may improve urinary control: Inhibition of the detrusor muscles through reflexive mechanisms and strengthening of pelvic floor muscles-Incontinence treatment for assessing EMG activity of the pelvic floor and accessory muscles (abdominal or gluteal)  |
|  Prescription Only | Yes | Yes | Yes  |

{6}

V M V E R I T Y

K243079
Page 3 of 4

|  Home Use | Yes | Yes | Yes  |
| --- | --- | --- | --- |
|  Power Source(s) | 4xAAA Alkaline, 6V | 9v alkaline battery | 4xAA niMH, 4.8V Rechargeable battery Pack  |
|  Program duration | 15-45 mins (based on the programs) Adjustable: 1-99 mins | 15-45 mins (based on the programs) Adjustable: 1-99 mins | 5-60 mins (based on the programs) Adjustable: 1-99 mins  |
|  No. of Output Channels | For stimulation: 1 For EMG: 2 | 2 | 2  |
|  Applied Parts | Vaginal Probe and Skin electrode | Vaginal Probe | Vaginal Probe and Skin electrode  |
|  Current or Voltage Regulated | Constant current | Constant current | Constant current  |
|  Waveform | Biphasic | Biphasic | Biphasic  |
|  Shape | Symmetrical rectangular | Symmetrical rectangular | Symmetrical rectangular  |
|  Maximum Output Voltage (V) | 45 V @ 500 Ω 103 V @ 2 kΩ 170 V @ 10 kΩ | 45 V @ 500 Ω 103 V @ 2 kΩ 170 V @ 10 kΩ | 40V @ 500 Ω 70V @ 2 kΩ 70V @ 10 kΩ  |
|  Frequency | 2 Hz – 100 Hz | 2 Hz – 100 Hz | 2 Hz – 100 Hz  |
|  Net Charge (μC per pulse) | 0 [μC] @ 500Ω | 0 [μC] @ 500Ω | 0 [μC] @ 500Ω  |
|  Maximum Output Current (mA) | 90 mA @ 500 Ω 50 mA @ 2 kΩ 19 mA @ 10 kΩ | 90 mA @ 500 Ω 50 mA @ 2 kΩ 19 mA @ 10 kΩ | 90 mA @ 500 Ω 35 mA @ 2 kΩ 7mA @ 10 kΩ  |
|  Maximum Current Density (mA/cm²) | 12.73 mA/cm² (with electrode surface area of 7.07 cm²) | 14.1 mA/ cm² (with electrode surface area of 6.4 cm²) | 1.42mA /cm²(with electrode surface area of 19 cm²)  |
|  Maximum Power Density @ 500 Ω | 37.81mW/cm² for electrode surface area of 7.07 cm² | 41.8 mW/ cm² for electrode of 6.4 cm² | 19[mW/ cm²] for electrode of 19 cm²  |

The sponsor used the primary predicate to support the STIM mode of the subject device and used the secondary predicate to support the EMG and ETS modes of operation of the subject. As evidenced by the above table, the subject and the predicate devices have the same intended use but differ in technological characteristics. Performance testing was conducted on the subject device, and it was established that the differences in technological characteristics do not raise different questions of safety or effectiveness.

# 7. SUMMARY OF NONCLINICAL TESTING:

Below is a list of the tests that were performed and successfully completed for the subject device per the specified guidance and standards:

- Electrical Safety testing according to IEC 60601-1: 2020 - Medical electrical equipment -

{7}

VERITY MECHANICS
K243079
Page 4 of 4

## Basic safety and essential performance

- Electromagnetic Compatibility testing according to IEC 60601-1-2: 2020 - General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests
- Software Verification and Validation Testing according to FDA’s Guidance “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices”
- Cybersecurity risk management activities according to FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions, issued September 27, 2023”

Additionally, the values of net charge, maximum phase charge and maximum power density of the subject device were calculated under 500Ω load from the output stimulation waveforms to establish performance of the subject device.

All pre-determined acceptance criteria were met.

## 8. CONCLUSION

Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPI/K243079](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KPI/K243079)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
