Rusch FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter

K181979 · Teleflexmedical, Inc. · KOD · May 24, 2019 · Gastroenterology, Urology

Device Facts

Record IDK181979
Device NameRusch FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter
ApplicantTeleflexmedical, Inc.
Product CodeKOD · Gastroenterology, Urology
Decision DateMay 24, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5130
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The subject device is a tubular device that is inserted through the urethra to pass urine from the bladder.

Device Story

Single-use, sterile, 18 Fr. Coudé hydrophilic intermittent catheter; constructed of PVC with polished eyes. Integrated package includes catheter, protective sleeve for touchless insertion, and 0.9% sterile saline sachet. User breaks sachet to hydrate hydrophilic coating before insertion. Used in clinical or home settings by patients or healthcare providers. Facilitates urine drainage from bladder; touchless sleeve reduces contamination risk. Benefits include ease of use, reduced friction during insertion, and sterile, ready-to-use configuration.

Clinical Evidence

Bench testing only. Testing included visual inspection, connector security (EN 1616), pull tests (EN 1617/1618), flow rate (EN 1618/ASTM F623), coating presence, friction testing, biocompatibility (ISO 10993-1/5/10), and sterilization validation (ISO 11135/11137). All tests passed.

Technological Characteristics

Materials: PVC shaft and funnel. Dimensions: 18 Fr. gauge, 40 cm length. Hydrophilic coating. Sterilization: Ethylene Oxide (catheter), Gamma irradiation (saline). Standards: ASTM F623:2013, EN 1616:1997, EN 1617:1997, EN 1618:1997, ISO 10993-1:2009. Coudé tip configuration.

Indications for Use

Indicated for adult patients requiring intermittent catheterization to pass urine from the bladder. Contraindicated in patients with insurmountable urethral obstructions, urethral injuries, or urethral inflammation.

Regulatory Classification

Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. May 24, 2019 Teleflex Medical, Inc. Sirisha Kommana Regulatory Affairs Specialist 3015 Carrington Mill Blvd., Suite 600 North Morrisville, NC 27560 Re: K181979 > Trade/Device Name: Rusch FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: GBM Dated: April 24, 2019 Received: April 26, 2019 Dear Sirisha Kommana: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. {1}------------------------------------------------ You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Glenn B. Bell, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration #### Indications for Use 510(k) Number (if known) K181979 Device Name Rusch FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter Indications for Use (Describe) The subject device is a tubular device that is inserted through the urethra to pass urine from the bladder. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) ption Use (Part 21 CFR 601 Subpart D) Over -- The -- Counter Use (21 CFR 801 Subpart C) Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) SUMMARY ### A. Name, Address, Phone and Fax Number of Applicant Teleflex Medical, Inc. 3015 Carrington Mill Blvd Morrisville, NC 27560 Phone: 919-433-4831 919-361-3939 Fax: ### B. Contact Person Sirisha Kommana Requlatory Affairs Specialist, Respiratory Division ### C. Date Prepared May 23, 2019 #### D. Device Name | Trade Name: | Rusch FloCath Quick™ 18 Fr. Coudé Hydrophilic Intermittent<br>Catheter | |-----------------------|------------------------------------------------------------------------| | Common Name: | Catheter, Urethral | | Product Code: | GBM | | Regulation Number: | CFR 876.5130 | | Regulation Name: | Urological Catheter and Accessories | | Classification: | II | | Classification Panel: | Gastroenterology/Urology | #### E. Predicate Device This submission demonstrates substantial equivalence to the predicate device ### Rusch® FloCath - K000070 ### F. Device Description The Teleflex Rusch®FloCath Quick™ 18 French (Fr.) Coudé Hydrophilic Intermittent Catheter is single use, disposable and sterile. The catheter is made of clear polyvinylchloride (PVC) with vertically cut and softly rounded, polished eyes and has a hydrophilic coating. It is composed of a shaft with a proximal funnel and a tip. The funnel is made of PVC. The shaft size is an 18 French gauge and is identified on the packaging. The funnel is color coded red to facilitate size identification. The catheter is 40 cm in overall length. {4}------------------------------------------------ The FloCath Quick is an integrated package that contains the FloCath catheter with a blue protective catheter sleeve and a packet of sterile 0.9% saline. Before opening, the sterile saline pouch is broken to hydrate the catheter. The protective catheter sleeve allows for touchless insertion. The sleeve can also be pushed back and used as an extension to the catheter. The FloCath Quick packaging has a packaging hole to facilitate opening the catheter. Either the packaging hole or the adhesive patch on the catheter pack can be used to hang up the catheter in an appropriate, convenient location while the catheter is hydrated. ### G. Indications for Use The Subject device is a tubular device that is inserted through the urethra to pass urine from the bladder. #### H. Contraindications - Insurmountable urethral obstructions - Urethral injuries - Urethral inflammation #### l. Substantial Equivalence The proposed Teleflex Rusch FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter is substantially equivalent to the predicate devices with respect to intended use, technology and construction. The differences between the predicates and the proposed devices are minor and any risks have been mitigated through testing. Table 036-1 summarizes the differences between the proposed and predicate devices. The proposed device is substantially equivalent to the predicate device: {5}------------------------------------------------ | Features | Teleflex Medical<br>Rusch FloCath Quick<br>Urological Catheter<br>(Subject Device) | Rusch FloCath<br>(Predicate Device-<br>K000070) | | |---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------| | Classification<br>Name | Catheter Urethral | Catheter Urological | | | Product Code | GBM | KOD | | | Class | II | II | | | Regulation<br>Number | 876.5130 | 876.5130 | | | Single Use | Yes | Yes | | | Patient Population<br>for 18 Fr. | Adult | Adult | | | Size Range | 18 Fr. | 5-30 Fr. | | | 18 Fr. Overall<br>Length | 40 cm | 40 cm | | | Color Indicating<br>Size for 18 Fr. | Red | Red | | | Sterile | Yes | Yes | | | Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide or<br>Gamma irradiation. | | | Device Description | The FloCath Quick is<br>a sterile, ready-to-use,<br>flexible PVC catheter<br>for intermittent<br>catheterization. The<br>catheter is hydrophilic<br>coated and has an<br>integrated pouch<br>containing a 0.9%<br>sterile saline solution<br>and a handling<br>sheath. The FloCath<br>Quick is a Coudé, 18<br>Fr. Catheter. | The Rusch FloCath<br>Catheter consists of<br>a tubular PVC shaft<br>with attached<br>drainage funnel.<br>The catheters is<br>designed with either<br>Nelaton, Olive or<br>Tiemann tip. This<br>device shaft may be<br>uncoated or<br>Hydrogel/ Hydrophilic<br>coated. | | | | | | | | Shaft | Tubular | Tubular | | | Shaft Material | PVC | PVC | | | Funnel Material | PVC | PVC | | | Intended Use | The FloCath Quick<br>catheter is a tubular<br>device that is inserted<br>through the urethra<br>and is used to pass<br>fluids from urinary<br>tract. | FloCath catheter is a<br>tubular device that is<br>inserted through the<br>urethra and is used<br>to pass fluids to or<br>from urinary tracts. | | | Indication for Use | The subject device is<br>a tubular device that is<br>inserted through the<br>urethra to pass urine<br>from the bladder. | FloCath catheter is a<br>tubular device that is<br>inserted through the<br>urethra and is used<br>to pass fluids to or<br>from urinary tracts. | | | Contraindications | Insurmountable<br>urethral obstructions Urethral injuries Urethral<br>inflammation | None | | | Coating | Hydrophilic | Uncoated or<br>Hydrogel/<br>Hydrophilic Coating | | | Materials<br>Evaluation | The materials used in<br>the construction of | The materials used in<br>the construction of | | | | Section 036 – Updated 510(k) Summary | | Hydrophilic Intermittent Catheter | | | FloCath Quick were<br>evaluated per ISO<br>10993-1:2009 | predicate were<br>evaluated per ISO<br>10993-5:1993,<br>10993-10:1995 | | | Principle of<br>Operation - Short<br>Description of Use | Squeeze saline<br>sachet.<br>Move packaging up<br>and down 2 or 3 times<br>to ensure coating of<br>the catheter.<br>Hang and let soak for<br>30 seconds.<br>Peel package open.<br>Insert catheter using<br>blue protective<br>catheter sleeve. | Tear peel back<br>wrapper about 5cm.<br>Pour water into<br>catheter funnel.<br>Takes about 30<br>seconds for catheter<br>to activate.<br>Remove catheter<br>from the package by<br>the funnel and insert<br>the catheter. | | | Eyelets | Yes | Yes | | | Liquid for Wetting | 0.9% Sterile saline<br>solution | Tap water | | | Primary Packaging | Paper and film peel<br>back | Paper and film peel<br>back | | | Secondary<br>Packaging | Corrugated board | Corrugated board | | | Shipper Case | Corrugated board | Corrugated board | | | Shelf-life | 5 Years | 5 Years | | | Performance<br>Standards | ASTM F623:2013<br>EN 1616:1997,<br>EN 1617:1997, | ASTM F623-89 | | ### Table 36-1. Comparison of Predicate vs. Proposed Device {6}------------------------------------------------ ### Rusch FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter ### Section 036 – Updated 510(k) Summary {7}------------------------------------------------ ### Rusch FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter {8}------------------------------------------------ #### J. Comparison to the Predicate Table 36-1 illustrates the similarities and differences between the subject and predicate device (K000070). The basic technological and operating principles are the same for both devices. Both the predicate and subject device have same intended use. Both the subject and predicate device are intended for similar patient populations, adult. Both the subject and predicate devices are disposable, sterile, single use devices. As evidenced by comparison Table 7-1, above, the proposed FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter is substantially equivalent to the predicate device. There are no significant differences that would affect safety and efficacy. #### K. Performance Data The bench testing performed verifies that the performance of the subject device FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter is substantially equivalent in terms of critical performance characteristics to the predicate device. These tests include: | Testing Performed | Reference to Standard | Results | |--------------------------|-----------------------------------------------------------|---------| | Visual Inspection | NA | Pass | | Connector Security Test | EN 1616:1997 | Pass | | Instron Pull Test | EN 1617:1997<br>EN 1618:1997 | Pass | | Flow Rate | EN 1618:1997<br>ASTM F623:2013 | Pass | | Coating Presence | NA | Pass | | Friction Test | NA | Pass | | Biocompatibility Testing | ISO 10993-1:2009<br>ISO 10993-5:2009<br>ISO 10993-10:2010 | Pass | | Sterilization | ISO 11135-1: 2014<br>ISO 11137-1: 2006 | Pass | The proposed devices were tested to the requirements of EN 1616:1997, EN1617:1997, EN 1618:1997. In addition, testing was performed to ensure compliance to ASTM F623:2013. Cytotoxicity, Sensitization, Irritation, were performed to demonstrate biocompatibility of the patient contacting materials. {9}------------------------------------------------ The proposed device is sterilized using Ethylene Oxide method and the saline sachet is sterilized using Gamma Irradiation method. The respective sterilization validations performed are ETO overkill method and the dose audit study. Overall, the results are comparable to the predicate and support a determination of substantial equivalence. ## L. Conclusion The FloCath Quick 18 Fr. Coudé Hydrophilic Intermittent Catheter has the same intended use and technological characteristics as its predicates. Test results demonstrate that the proposed device meets its intended use and is as safe, as effective, and performs as well as or better than the legally marketed predicate device. It is for these reasons that the subject device can be found substantially equivalent.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...