← Product Code [KOD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD) · K180354

# UroGen Ureteral Catheter (K180354)

_Urogen Pharma, Ltd. · KOD · Oct 30, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD/K180354

## Device Facts

- **Applicant:** Urogen Pharma, Ltd.
- **Product Code:** [KOD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD.md)
- **Decision Date:** Oct 30, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The UroGen Ureteral Catheter is indicated for use by physicians for facilitating access to the urinary tract through a retrograde route and may be used in conjunction with a guidewire or for the injection of gels or fluids into the urinary tract.

## Device Story

Single-use ureteral catheter; facilitates retrograde access to upper urinary tract via standard endoscopic technique. Used with guidewires (up to 0.038 inches) or for delivery of gels/fluids. Operated by physicians in clinical settings. Features Luer lock glued to catheter body to withstand higher pressures during fluid delivery. Supplied sterile; intended for short-term use (<1 hour). Benefits include reliable access and fluid delivery capability for urological procedures.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance validated via biocompatibility (ISO 10993: cytotoxicity, sensitization, irritation) and mechanical bench testing (flow rate, kinking, tensile force, connector security, Luer hub performance, leakage, and instillation force).

## Technological Characteristics

Single-use ureteral catheter. Materials evaluated per ISO 10993. Features glued Luer lock hub for high-pressure fluid delivery. Compatible with guidewires up to 0.038 inches. Standards: ISO 10555, ASTM F623-99, EN 13868, ISO 594, EN 1616, EN 1618, ANSI/AAMI/ISO 11135. Sterilized via validated process.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- Backstop Catheter ([K110491](/device/K110491.md))

## Reference Devices

- Ureteral Catheter, Boston Scientific ([K830840](/device/K830840.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 30, 2018

UroGen Pharma Ltd. % James G. Ottinger, R.Ph. VP Regulatory Affairs, UroGen Pharma UroGen Pharma 499 Park Avenue, 12th Floor, Suite 1200 New York, NY 10022

Re: K180354

> Trade/Device Name: UroGen Ureteral Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: KOD Dated: September 26, 2018 Received: September 28, 2018

Dear James G. Ottinger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Timothy Martin -S 2018.10.30 09:14:32 -04'00'

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K180354

Device Name UroGen Ureteral Catheter

Indications for Use (Describe)

The UroGen Ureteral Catheter is indicated for use by physicians for facilitating access to the urinary tract through a retrograde route and may be used in conjunction with a guidewire or for the injection of gels or fluids into the urinary tract.

| Type of Use (Select one or both, as applicable) | <span> <span style="font-family: DejaVu Sans, sans-serif;">✔</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> <span> <span style="font-family: DejaVu Sans, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)             </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

*PRAStaff@fda.hhs.gov*

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of

information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY [as required by section 807.92(c)] UroGen Ureteral Catheter 510(k) Number K180354

#### SUBMITTER 1.

### Applicant's Name:

UroGen Pharma Ltd. 9 Ha`Ta`asiya Street Ra'anana, Israel. 4365007 Tel: +972-9-770-7600 Fax: +972-77-4171410

### Contact Person:

James G. Ottinger, R.Ph. VP Regulatory Affairs, UroGen Pharma 499 Park Avenue, 12th Floor, Suite 1200 NY, NY, 10022 USA Phone: 610-745-1230 jim.ottinger@urogen.com

### Date Prepared:

October 30, 2018

#### 2. DEVICE

### Trade Name:

UroGen Ureteral Catheter

### Common or Usual Name:

Ureteral Catheter

### Classification:

| Classification Name:  | Urological catheter and accessories |
|-----------------------|-------------------------------------|
| Regulatory Class:     | Class II per 21 CFR 876.5130        |
| Product Code:         | KOD                                 |
| Classification Panel: | Gastroenterology-Urology            |

#### 3. PREDICATE DEVICES

### Primary predicate:

- Backstop Catheter, by Pluromed, Inc., product code KOD; cleared under K110491 . Reference devices:
- . Ureteral Catheter, by Boston Scientific, product code: GBM; cleared under K830840

{4}------------------------------------------------

#### 4. DEVICE DESCRIPTION

The device is a single use ureteral catheter, designed to assist in access to the upper urinary tract using standard endoscopic technique for drainage and delivery of gels or fluids. The catheter is indicated for use by physicians for facilitating access to the urinary tract through a retrograde route and may be used in conjunction with a guidewire or for the injection of gels or fluids into the urinary tract.

The catheter is supplied sterile in a Tyvek pouch with a sterilization indicator. The catheter is inserted into the body for a typical duration of less than 1 hour. The catheter is placed over a guidewire of up to 0.038 inches in diameter that is prepositioned through the urological tract.

#### 5. INDICATIONS FOR USE

The UroGen Ureteral Catheter is indicated for use by physicians for facilitating access to the urinary tract through a retrograde route and may be used in conjunction with a guidewire or for the injection of gels or fluids into the urinary tract.

### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The UroGen Ureteral Catheter has similar technological characteristics as its predicate devices. The main difference between the UroGen Ureteral Catheter and other FDA cleared ureteral catheters is that in the UroGen Catheter, the Luer lock is glued to the catheter body. This allows the UroGen Ureteral Catheter to withstand higher pressures associated with the delivery of fluids with varying viscosities. This minor technological difference was tested to ensure the device functions as intended and does not raise new safety or effectiveness concerns.

#### 7. PERFORMANCE DATA

Below is a list of the tests that have been performed and successfully completed for the UroGen Ureteral Catheter.

### Biocompatibility:

Testing was conducted in accordance with FDA guidance: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The Catheter underwent the following GLP biocompatibility testing per ISO 10993: Cytotoxicity, Sensitization and Irritation.

# Bench: Test (Standard)

- Flow rate (ISO 10555, ASTM F623-99) -
- Kinking (EN 13868) -
- Peak Tensile Force (ISO 10555-1, EN1618) -
- Connector security (EN 1616) -
- Luer Hub Performance (ISO 594) -
- Hub Liquid Leakage (ISO 594) -

{5}------------------------------------------------

- Sterilization, Shelf Life and Packaging (ANSI/AAMI/ ISO 11135, -AAMI/ISO 11138, ISO 10993 and EN 1422)
- Instillation force test (Internal Standard) -

# Conclusion:

The UroGen Ureteral Catheter has similar indications for use as the Backstop predicate device and is substantially equivalent in technological and performance characteristics to the predicate device.

Consequently, it is concluded that the UroGen Ureteral Catheter is as safe and effective as its predicate device without raising any new safety and/or effectiveness concerns. Any differences have been addressed by bench testing and validation and therefore negate any safety or effectiveness concerns.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD/K180354](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD/K180354)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
