← Product Code [KOD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD) · K032710

# CLOSED SYSTEM INTERMITTENT CATHETERIZATION KIT (K032710)

_Apogee Medical, Inc. · KOD · Jan 9, 2004 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD/K032710

## Device Facts

- **Applicant:** Apogee Medical, Inc.
- **Product Code:** [KOD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD.md)
- **Decision Date:** Jan 9, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Apogee Closed System Intermittent Catheterization Kit is intended to be used to drain urine from the patient's bladder into a collection bag.

## Device Story

Apogee Closed System Intermittent Catheterization Kit; single-use, disposable, pre-lubricated catheter; includes sterile collection bag and introducer. Device allows insertion and advancement of catheter into bladder without direct hand contact, reducing contamination risk. Used by patients or caregivers in clinical or home settings when bladder drainage into commode/bedpan is impractical. Output is urine collection in bag. Benefits include reduced risk of contamination during intermittent catheterization.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Single-use, disposable, pre-lubricated catheter; includes collection bag and introducer tip for non-contact insertion. Mechanical device; no energy source or software.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- Mentor Self-Cath Closed System ([K003873](/device/K003873.md))
- Hollister InCare Pre-Lubricated Intermittent Catheters ([K013483](/device/K013483.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JAN - 9 2004

# 510(k) SUMMARY

K632710 Assigned 510(k) # is _________________________________________________________________________________________________________________________________________________________

# Sponsor's Name and Address and Contact Person:

Apogee Medical, Inc. 90 Weathers Street Youngsville, NC 27596

Contact Person: Diane N. Peper, RA/QA Telephone: (919) 570-9605 Facsimile: (919) 570-9611 email address: dpeper@apogeemed.com

# Date Summary Prepared: August 22, 2003

# Device Name and Classification:

Common Name: Urological Catheter Classification Name: Gastrology and Urology Product Code: KOD

#### Manufacturer:

Apogee Medical, Inc. 90 Weathers Street Youngsville, NC 27596

# Name of Predicate Device(s):

K003873 Mentor Self-Cath Closed System Hollister InCare Pre-Lubricated Intermittent Catheters K013483

#### Device Description:

The Apogee Closed System Intermittent Catheterization Kit is a single-use, disposable, pre-lubricated catheter contained in a sterile collection bag. The Closed System Intermittent Catheter is intended to be user to drain urine from the bladder. When it is not practical or feasible for the patient to drain the bladder into a commode or bedpan, the urine may be drained into the collection bag of the Closed System Intermittent Catheterization Kit. The Closed System Intermittent Catheterization Kit is designed with a collection bag and introducer which provide the mechanism for inserting and advancing the catheter to the bladder without direct hand contact to the catheter, thereby, reducing possibility of contamination.

#### Statement of Intended Use

The Apogee Closed System Intermittent Catheterization Kit is intended to brain urine from the patient's bladder into a collection bag.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling three stylized human figures or lines, possibly representing people or services provided by the department.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN - 9 2004

Ms. Diane N. Peper Regulatory Affairs/Quality Assurance Manager Apogee Medical, Inc. 90 Weathers Street YOUNGSVILLE NC 27596

Re: K032710

Trade/Device Name: Closed System Intermittent Catheterization Kit Regulation Number: 21 CFR §876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: Il Product Code: 78 KOD Dated: October 20, 2003 Received: November 7, 2003

Dear Ms. Peper:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for dos based or to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may , and is provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that I'DA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act that + Drederal statutes and regulations administered by other Federal agencies. You must or any I educal the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin marketing of substantial equivalence of your device to a legally premarket notheation. The PDA inding vices in the your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific article to your do to be of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the promote the regulation entitled, "Misbranding Other of Compliance at (501) 57 + 185 (21CFR Part 807.97) you may obtain. Other general by relection on your responsibilities under the Act may be obtained from the Division of Small Information on your responsional and Consumer Assistance at its toll-free number (800) 638-2041 or Manufacturers, International and Section: http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

{3}------------------------------------------------

Page 1 of 1

510(k) Number (if known): Ko327 | O

Device Name: Closed System Intermittent Catheterization Kit

Indications For Use:

The Apogee Closed System Intermittent Catheterization Kit is intended to be used to drain urine from the patient's bladder into a collection bag.

(PLEASE DO NOT WRITE BELOW THIS INE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Dianne Brogdon

(Optional Format 3-10-98)

Division of Reproductive, Abdominal, and Radiological Device 510(k) Number .

Prescription Use
(Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD/K032710](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOD/K032710)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
