← Product Code [KOC](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC) · K121599

# CRIT-LINE CLIP (CLIC) (K121599)

_Fresenius Medical Care-North America · KOC · Nov 9, 2012 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC/K121599

## Device Facts

- **Applicant:** Fresenius Medical Care-North America
- **Product Code:** [KOC](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC.md)
- **Decision Date:** Nov 9, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5820
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Crit-Line Clip Monitor (CLiC) is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. CLiC measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

## Device Story

CLiC Monitor is a miniaturized, non-invasive device for real-time monitoring of hematocrit, oxygen saturation, and blood volume changes during dialysis. It utilizes an optical sensor clip (LED array and photodetector) that attaches to a sterile, single-use cuvette installed in the arterial line of a dialysis tubing set. The device processes light absorption and scattering data to calculate blood parameters. Unlike the predicate, the CLiC miniaturizes the microprocessor and electronics directly into the sensor clip housing and replaces the multi-conductor cable with a USB interface. It connects to a host computer (Windows OS) which provides power, processes data, and displays results. Clinicians or dialysis technicians use the output to adjust fluid removal rates during treatment, aiming to maximize fluid removal while minimizing adverse symptoms like cramping, nausea, and vomiting.

## Clinical Evidence

No clinical data. Evidence consists of in vitro bench testing, system verification, and validation testing to confirm performance against specifications. Safety and effectiveness were supported by adherence to IEC 60601-1 and IEC 60601-1-2 standards and a risk analysis per ISO 14971. Performance testing demonstrated that the CLiC Monitor's accuracy for hematocrit and oxygen saturation is equivalent to the predicate device specifications.

## Technological Characteristics

Compact measurement device; sensor clip housing with LED illuminators and photodetector arrays; USB-powered via host computer (Windows OS). Uses light absorption/scattering principle. Connectivity: USB peripheral. Sterilization: Single-use sterile cuvette (K935958). Safety standards: UL 60601-1/IEC 60601-1, EMC 60601-1-2.

## Regulatory Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

## Predicate Devices

- Crit-Line MI Monitor ([K972470](/device/K972470.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K121599
PAGE 1 OF 5

**NOV. 9 2012**

# 510k Summary

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.

#### A. Submitter's Information

| Name:                | Fresenius Medical Care North America |
|----------------------|--------------------------------------|
| Address:             | 920 Winter Street                    |
|                      | Waltham, MA 02451-1457               |
| Phone:               | (781) 699-4479                       |
| Fax:                 | (781) 699-9635                       |
| Contact Person:      | Denise Oppermann, Senior Director    |
|                      | Regulatory Affairs - Devices         |
|                      | Renal Therapies Group                |
| Date of Preparation: | 31 May 2012                          |

#### B. Device Name

Trade Name: Classification Name

Classification Number Product Code/Classification Panel: CLiC Monitor (Crit-Line Clip Monitor) Non-invasive Hematocrit, Oxygen Saturation, Percent change in blood volume Monitor Class II per § 876.5820 78 MQS/Gastroenterology/Urology Panel

# Legally Marketed Predicate Device (unmodified device) C.

Crit-Line MI Monitor (K972470).

{1}------------------------------------------------

K121599
PAGE 2 OF 5

Image /page/1/Picture/1 description: The image shows the Fresenius Medical Care logo. The logo consists of a stylized graphic on the left and the text "FRESENIUS MEDICAL CARE" on the right. The graphic is made up of three chevron-like shapes stacked on top of each other, pointing downwards. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

#### D. Device Description

The intended use of the Crit-Line III (CLM III) Monitor is as a continuous real-time monitor for non-invasive hematocrit, oxygen saturation, and percent change in blood volume measurement. The CLM III is used primarily as a monitor for dialysis treatments.

The CLM III uses the principle of light absorption and scattering through the blood under test to measure oxygen saturation (O2 SAT) and hematocrit (HCT). The HCT values are then used to calculate the related Blood Volume (BV) percentage change relative to the starting BV based on the beginning HCT.

The CLM III consists of a microprocessor-controlled main circuit board located within the display housing. The display housing contains the sensor clip driver measurementcalculation circuitry and display screen circuitry. The sensor clip with optical system (LED array and a photodetector array) is tethered to the main housing and circuit board by a multi-conductor cable. The cable connects to a single-use, sterile cuvette (K935958) installed in-line with the arterial line of the dialysis tubing set.

Calculations performed on the main circuit board within the display housing yield the hematocrit, oxygen saturation and the related value of blood volume measurement results and are displayed on the LCD display screen.

Modifications to the previously cleared Crit-Line III Monitor include:

#### 3. Miniaturization/Re-packaging

The CLiC Monitor incorporates all hardware/electronic components - miniaturized and contained within the sensor clip housing, for measuring indicated parameters (HCT O2 SAT). The primary modification of the device involves integrating a miniaturized microprocessor with the required electronic circuitry and software to drive the existing optical system (LED and photo-detector array) within the plastic sensor clip housing. The CLiC utilizes the same, previously cleared Crit-Line technology (K972470), sensor clip and optical system. The operating principles and fundamental measurement methods of the CLiC are the same as the predicate device.

{2}------------------------------------------------

K121599
PAGE 3 OF 5

Image /page/2/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a stylized symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three chevron-like shapes stacked on top of each other, with the top two pointing upwards and the bottom one pointing downwards. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

#### 4. Distributed Process Technology

A standard USB cable replaces the tethering multi-conductor cable used in the CLM III Monitor. USB-compatible hardware and software are incorporated in the CLiC housing. The CLiC can be hosted on a medical grade computer system (Windows OS) that can interface with, power a USB peripheral and operate supporting drivers. CLiC interlocks with a singleuse, sterile cuvette (K935958) which is located in-line with the arterial line of the dialysis tubing set for indicated measurements. The lightwave measurements are performed autonomously by the CLiC. Blood parameters are then calculated and transmitted to the host computer display microprocessor in text format by means of the USB interface. The accompanying display software installed on the host computer system facilitates the display of measured parameters: hematocrit, oxygen saturation and blood volume changes on the display screen.

#### E. Indications for Use

The Crit-Line Clip Monitor (CLiC) is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. The CLiC measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

### Technological Characteristics F.

The CLiC Monitor is a compact measurement device which can be used with any medical grade computer system (Windows OS) that can interface with and power a USB peripheral. With the supporting display software, the Windows-based computer system provides the user interface.

The CLiC is a modification to the Crit-line III with equivalent indications for use, to noninvasively measure hematocrit, oxygen saturation and percent change in blood volume. Both devices non-invasively measure blood parameters using multiple wavelengths of light

{3}------------------------------------------------

**FRESENIUS**
**MEDICAL CARE**

which are scattered and absorbed, respectively, by different constituents in the blood. The CLiC Monitor has the same operating principle and fundamental scientific technology as the predicate device. It is also comparable to the predicate device in terms of key safety, effectiveness, and quality assurance features.

The CLiC Monitor is designed and developed using product specifications, construction and manufacturing methods functionally equivalent to the predicate device. The plastic sensor clip housing and optical system (LED illuminators and photodetectors) are same the as the predicate device (K972470).

The modifications described in this submission have not altered nor do they have the potential to alter the fundamental scientific technology of the device.

FMCNA has concluded that the CLiC Monitor is of comparable type and substantially equivalent to the currently marketed Crit-line III, complies with the same or equivalent standards and has the same intended use. In addition, the CLiC Monitor is certified to comply with safety requirements of UL 60601-1/IEC 60601-1, and EMC standard 60601-1-2.

The following technical specifications of the modified device remain the same as the unmodified device:

- . LED illuminators and Photodetector arrays
- . Safety system
- System performance ●
- Environmental Requirements .
- Transportation and Storage condition ●
- User Interface (except USB port and display software on host computer system) .
- Hardware (except miniaturization) .
- Environmental Design .
- . Alarms
- Accuracy and Controls ●
- . Protection against excessive temperature or other hazards
- . Manufacturing Location and manufacturing processes (testing, shipping, installation and service).

Safety and effectiveness of the CLiC Monitor is confirmed by:

- . System verification and validation testing - to verify performance to specifications, applicable referenced FDA regulations, and user requirements.

{4}------------------------------------------------

- . Adherence to FDA-recognized industry and international standards (IEC 60601-1 and IEC 60601-1-2) - Electrical Safety and Electromagnetic Compatibility testing was conducted to ensure the safety and effectiveness of the device after the proposed modifications.
A risk analysis, per ISO 14971, has been completed and potential hazards associated with the modifications are identified and mitigations are verified wherever applicable. All potential risks were deemed acceptable after mitigation.

## G. Performance Data

The performance of the modified device was evaluated according to existing procedures, protocols, declared performance standards and guidelines of the quality system regulation (21 CFR 820). Design verification and validation tests were conducted to ensure that the modifications described in this submission did not affect the essential performance of the device and the device functions as intended.

The CLiC Monitor accuracy is equivalent to the predicate device specifications for both hematocrit and oxygen saturation measurements. Because blood volume calculations are based on hematocrit and the CLiC Monitor matches the predicate specifications, the percentage blood volume comparisons are also equivalent.

### H. Conclusion

The CLiC Monitor does not pose any new potential safety risks and performs the same measurement functions as the predicate device.

Test results demonstrated that the modified CLiC Monitor functions as intended and met pre-determined acceptance criteria. Results of performance testing (in vitro bench testing), safety testing and usability testing indicate that the modified CLiC device is substantially equivalent to the named predicate device and remains safe and effective for its intended use.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is an abstract image of an eagle with its wings spread.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002

November 9, 2012

Fresenius Medical Care - North America % Ms. Denise Oppermann Senior Director, Regulatory Affairs Devices 920 Winter Street WALTHAM MA 02451

Re: K121599

Trade/Device Name: Crit-Line Clip (CLIC) Monitor Regulation Number: 21 CFR\$ 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: KOC Dated: October 10, 2012 Received: October 11, 2012

Dear Ms. Oppermann:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{6}------------------------------------------------

Page 2 - Ms. Denise Oppermann

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

**Herbert R. Lerner**

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the Fresenius Medical Care logo. The logo consists of a stylized symbol on the left and the text "FRESENIUS MEDICAL CARE" on the right. The symbol is made up of three stacked chevrons, each pointing downwards. The text is in bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.

510(k) Number (if known):

K121599

Device Name:

Crit-Line Clip (CLiC) Monitor

# Indications for Use:

The Crit-Line Clip Monitor (CLiC) is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume. CLiC measures hematocrit, percent change in blood volume and oxygen saturation in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.

XPrescription Use (Per 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K121599

Page 75 of 209

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC/K121599](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KOC/K121599)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
