KENDALL CURITY SUPRAPUBIC CATHETER TRAY, KIT, & REPLACEMENT SET
Device Facts
| Record ID | K970969 |
|---|---|
| Device Name | KENDALL CURITY SUPRAPUBIC CATHETER TRAY, KIT, & REPLACEMENT SET |
| Applicant | The Kendal Co. |
| Product Code | KOB · Gastroenterology, Urology |
| Decision Date | Apr 23, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5090 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The Kendall CURITY Suprapubic Catheter Trays and kits are intended to provide urinary bladder drainage or irrigation by percutaneous placement of a suprapubic catheter. This is a useful method of controlling urinary bladder drainage postoperatively.
Device Story
Sterile, single-use suprapubic catheter trays and kits; provide bladder access for irrigation and urine drainage. Device consists of 100% silicone two-way balloon catheters (10, 12, 14 French) with drainage and inflation lumens; packaged with various accessories. Used in clinical settings for postoperative bladder management. Healthcare providers perform percutaneous placement to facilitate drainage; output is effective bladder management. Benefits include controlled urinary drainage and irrigation access.
Clinical Evidence
No clinical data. Bench testing only: biocompatibility per ISO-10993; functional/mechanical testing including flow rates, balloon burst volume, valve retention, bond strength, and load testing.
Technological Characteristics
100% silicone two-way balloon catheters; 10, 12, 14 French sizes. Sterile, single-use. Biocompatibility per ISO-10993. Mechanical testing: flow rates, balloon burst volume, valve retention, funnel/shaft bond strength, static/dynamic load, water loss.
Indications for Use
Indicated for patients requiring urinary bladder drainage or irrigation via percutaneous suprapubic catheter placement, particularly for postoperative bladder drainage control.
Regulatory Classification
Identification
A suprapubic urological catheter and accessories is a flexible tubular device that is inserted through the abdominal wall into the urinary bladder with the aid of a trocar and cannula. The device is used to pass fluids to and from the urinary tract. This generic type of device includes the suprapubic catheter and tube, Malecot catheter, catheter punch instrument, suprapubic drainage tube, and the suprapubic cannula and trocar.
Predicate Devices
- Kendall CURITY Suprapubic Drainage System (K842899)
Related Devices
- K013174 — PROGES FOLYSIL SILICONE FOLEY CATHETER, STRAIGHT CYLINDRICAL TIP, SILICONE FEMALE FOLEY CATHETER, STRAIGHT CYLINDRICAL, · Porges S.A. · Jan 4, 2002
- K182463 — Silicone Foley Catheter · Biosensors International Pte, Ltd. · Nov 6, 2019
- K041983 — OPTION-VM URINARY CATHETER/ADPATER, MODELS MV39016, MVA39016 · Opticon Medical · Nov 17, 2004
- K132890 — SUPRAPUBIC CATHETER AND INTRODUCER SET · Mediplus , Ltd. · Jan 17, 2014
- K172807 — Silicone Foley Catheter for single use · Changzhou Rongxin Medicine Minimal Invasion Technology · Jun 4, 2018