The Cook® Cystostomy Catheter Set is used for temporary suprapubic urinary diversion and drainage for less than or equal to four weeks.
Device Story
The Cook® Cystostomy Catheter Set is a sterile, single-use kit designed for temporary suprapubic urinary diversion and drainage. The set includes a single-lumen silicone catheter (8-12 French), a trocar stylet with access sheath (10-14 French), a drainage adapter with stopcock, and a silicone catheter support. The device is used by clinicians in a clinical setting to establish and maintain urinary drainage. The catheter is inserted into the bladder to facilitate urine diversion. The system is intended for short-term use (≤ 4 weeks). The device provides a mechanical pathway for drainage, assisting in patient management of urinary retention or diversion needs.
Clinical Evidence
Bench testing only. Performance data included biocompatibility, tensile strength, leakage and flow testing, and dimensional/compatibility verification to confirm the device meets design input requirements.
Technological Characteristics
Single-lumen silicone catheter (8-12 Fr, 54 cm length); trocar stylet with access sheath (10-14 Fr, 11-11.2 cm length); drainage adapter with stopcock; silicone catheter support. Sterile, single-use, three-year shelf life.
Indications for Use
Indicated for patients requiring temporary suprapubic urinary diversion and drainage for a duration of four weeks or less.
Regulatory Classification
Identification
A suprapubic urological catheter and accessories is a flexible tubular device that is inserted through the abdominal wall into the urinary bladder with the aid of a trocar and cannula. The device is used to pass fluids to and from the urinary tract. This generic type of device includes the suprapubic catheter and tube, Malecot catheter, catheter punch instrument, suprapubic drainage tube, and the suprapubic cannula and trocar.
Predicate Devices
All Silicone Supra-Pubia Catheter, Models 4880/4890 (K014002)
Universa® Percutaneous Drainage Catheter Set (Malecot Suprapubic Set) (K140085)
Related Devices
K970969 — KENDALL CURITY SUPRAPUBIC CATHETER TRAY, KIT, & REPLACEMENT SET · The Kendal Co. · Apr 23, 1997
K120127 — SWAN VALLEY MEDICAL TRANURETHRAL SUPRAPUBIC ENDO-CYSTOSTOMY (T-SPEC) SURGICAL SYSTEM · Swan Valley Medical, Incorporated · Oct 26, 2012
K132890 — SUPRAPUBIC CATHETER AND INTRODUCER SET · Mediplus , Ltd. · Jan 17, 2014
K240514 — AMT Low-Profile Suprapubic Catheter & Drainage Set · Applied Medical Technology, Inc. · Nov 1, 2024
K014002 — ALL SILICONE SUPRA-PUBIA CATHETER, MODELS 4880/4890 · Fortune Medical Instrument Corp. · Mar 4, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a blue square and the agency's name in a sans-serif font.
January 18, 2019
Cook Incorporated Danlu Chen Regulatory Affairs Specialist 750 Daniels Way, P.0. Box 489 Bloomington, IN 47402
K181097 Re:
Trade/Device Name: Cook® Cystostomy Catheter Set Regulation Number: 21 CFR§ 876.5090 Regulation Name: Suprapubic Urological Catheter and Accessories Regulatory Class: II Product Code: KOB Dated: December 19, 2018 Received: December 20, 2018
Dear Danlu Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801: medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
B-4
See PRA Statement below.
510(k) Number (if known) K181097
Device Name Cook® Cystostomy Catheter Set
Indications for Use (Describe)
The Cook® Cystostomy Catheter Set is used for temporary suprapubic urinary diversion and drainage for less than or equal to four weeks.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) Premarket Notification Cook® Cystostomy Catheter Set Cook Incorporated January 16, 2019
## 2.0 510(k) Summary
# Cook® Cystostomy Catheter Set 21 CFR §876.5090 Date Prepared: January 16, 2019
## Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|---------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Contact: | Jiayin Furnish<br>Andrew Breidenbach |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone: | (812) 335-3575 |
| Contact Fax: | (812) 332-0281 |
| Device Information: | |
#### Trade Name: Cook® Cystostomy Catheter Set Device Common Name: Suprapubic Urological Catheter and Accessories Classification Name: Suprapubic Urological Catheter and Accessories 21 CFR §876.5090 Regulation Number: Product Code: KOB Device Classification: Class II Review Panel: Gastroenterology/Urology
## Predicate Devices:
All Silicone Supra-Pubia Catheter, Models 4880/4890 (K014002)
Universa® Percutaneous Drainage Catheter Set (Malecot Suprapubic Set) (K140085)
## Device Description:
The Cook® Cystostomy Catheter Set is composed of a single-lumen silicone catheter available in 8 to 12 French with a working length of 54 centimeters, a trocar stylet with access sheath having an outer diameter of 10 to 14 French with a length of either 11 or 11.2 centimeters, a drainage
{4}------------------------------------------------
Traditional 510(k) Premarket Notification Cook® Cystostomy Catheter Set Cook Incorporated January 16, 2019
adapter with stopcock available in 12 to 16 gauge and a soft low-profile silicone catheter support with pull tie. The set is supplied sterile, intended for one-time use only, and packaged in a peelopen pouch with a three-year shelf life.
# Indications for Use:
The Cook® Cystostomy Catheter Set is used for temporary suprapubic urinary diversion and drainage for less than or equal to four weeks.
# Comparison to Primary Predicate Device:
The subject device, Cook® Cystostomy Catheter Set, and the predicate device, All Silicone Supra-Pubia Catheter, Models 4880/4890 (K014002), are similar in intended uses, fundamental technological characteristics, and methods of operation. Differences between the characteristics of the subject device and the predicate device include:
- Indications for Use .
- . Design Features
- Dimensions ●
Differences between the characteristics of the subject device and the predicate device are supported by testing. The subject device raises no new questions of safety and effectiveness as compared to the predicate device.
# Comparison to Secondary Predicate Device:
The subject device, Cook® Cystostomy Catheter Set, and the secondary predicate device, Universa® Percutaneous Drainage Catheter Set (Malecot Suprapubic Set) (K140085), are similar in intended uses, fundamental technological characteristics, and methods of operation. Differences between the characteristics of the subject device and the predicate device include:
- Indications for Use
- Design Features
- . Dimensions
Differences between the characteristics of the subject device and the secondary predicate device are supported by testing. The subject device raises no new questions of safety and effectiveness as compared to the predicate device.
{5}------------------------------------------------
Traditional 510(k) Premarket Notification Cook® Cystostomy Catheter Set Cook Incorporated January 16, 2019
## Performance Data:
The following testing was performed to demonstrate that the Cook® Cystostomy Catheter Set met applicable design requirements.
- . Biocompatibility
- . Tensile Strength
- . Leakage and Flow
- Dimensional and Compatibility .
Results of the testing support a conclusion that the subject device Cook® Cystostomy Catheter Set met the design input requirements based on the intended use and raises no new questions of safety and effectiveness as compared to the predicate device
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.