PROSYS BEDSIDE DRAINAGE BAG (STERILE) MODEL 650167
Device Facts
| Record ID | K960949 |
|---|---|
| Device Name | PROSYS BEDSIDE DRAINAGE BAG (STERILE) MODEL 650167 |
| Applicant | Bristol-Myers Squibb |
| Product Code | KNX · Gastroenterology, Urology |
| Decision Date | Apr 19, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.5250 |
| Device Class | Class 2 |
Indications for Use
The ProSys™ Bedside Drainage Bag (Sterile) is indicated for use in the collection of free flowing urinary drainage.
Device Story
Sterile bedside urinary drainage bag; 2000 ml capacity; 55-inch inlet tube with universal connector; integrated anti-reflux valve to prevent urine backflow; integrated sample port for urine collection; T-tap drainage valve. Used in clinical settings for urine containment. Device collects urine via gravity drainage from indwelling or external catheters. Benefits include infection risk reduction via anti-reflux mechanism and ease of sampling.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (non-irritating, non-sensitizing, non-cytotoxic), resistance to bursting, resistance to impact, leakage, anti-reflux valve performance, tensile strength, and weld integrity.
Technological Characteristics
2000 ml capacity sterile drainage bag; PVC/plastic construction; gravity-fed drainage; T-tap valve; anti-reflux valve; universal connector; integrated sample port. Sterilized product.
Indications for Use
Indicated for patients requiring collection of free flowing urinary drainage via indwelling or external catheters.
Regulatory Classification
Identification
A urine collector and accessories is a device intended to collect urine. The device and accessories consist of tubing, a suitable receptacle, connectors, mechanical supports, and may include a means to prevent the backflow of urine or ascent of infection. The two kinds of urine collectors are:(1) A urine collector and accessories intended to be connected to an indwelling catheter, which includes the urinary drainage collection kit and the closed urine drainage system and drainage bag; and (2) A urine collector and accessories not intended to be connected to an indwelling catheter, which includes the corrugated rubber sheath, pediatric urine collector, leg bag for external use, urosheath type incontinence device, and the paste-on device for incontinence.
Special Controls
*Classification* —(1)*Class II (special controls) for a urine collector and accessories intended to be connected to an indwelling catheter.* The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.(2) *Class I (general controls).* For a urine collector and accessories not intended to be connected to an indwelling catheter, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9. If the device is not labeled or otherwise represented as sterile, it is exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.
Predicate Devices
- Bard® Urinary Drainage Bag with Anti-reflux Chamber
Reference Devices
- ProSys™
- Bard® Dispoz-A-Bag®
- Flip-Flo™