Enteral Feeding Sets

K240039 · Hangzhou Primecare Medical Co., Ltd. · KNT · Apr 17, 2024 · Gastroenterology, Urology

Device Facts

Record IDK240039
Device NameEnteral Feeding Sets
ApplicantHangzhou Primecare Medical Co., Ltd.
Product CodeKNT · Gastroenterology, Urology
Decision DateApr 17, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5980
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets include a bag to contain the feeding solution and/or a EnPlus spike to connect to a pre-filled container. The device is used for infants, children, adolescents and adults.

Device Story

Enteral Feeding Sets consist of tubing, drip chambers, roller clamps, and either a formula bag or an EnPlus spike for connection to pre-filled containers. Used in clinical or home settings to deliver liquid nutrition to patients via feeding tubes. Operated by clinicians or patients/caregivers. Gravity-fed or pump-driven flow; user controls rate via roller clamp or pump settings. Provides a closed or semi-closed system for nutrition administration. Benefits include reliable, controlled delivery of enteral nutrition, reducing risk of misconnection via ISO 80369-3 compliant connectors.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including metrology, leakage, tensile strength, flow rate, and biocompatibility (cytotoxicity, sensitization, irritation) per ISO 10993 series.

Technological Characteristics

Materials: PVC, ABS, Silicone, LDPE, paper. Components: EnFit connectors (ISO 80369-3:2016), ENPlus spikes (ISO 18250-3:2018), tubing, roller clamps, drip chambers. Non-sterile, single-use, disposable. Connectivity: ISO 80369-3 compliant small-bore connectors. No software or electronic components.

Indications for Use

Indicated for infants, children, adolescents, and adults requiring enteral nutrition delivery via feeding tube using gravity or pump-assisted systems.

Regulatory Classification

Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. April 17, 2024 Hangzhou Primecare Medical Co., Ltd. Li Xuedong Regulatory Director Room 408-409, Zancheng Center West, Shangcheng District Hang zhou, Zhe jiang 310008 China Re: K240039 Trade/Device Name: Enteral Feeding Sets Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube And Accessories Regulatory Class: Class II Product Code: PIF, KNT, Dated: March 18, 2024 Received: March 18, 2024 #### Dear Li Xuedong: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {1}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. {2}------------------------------------------------ 3 Page For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Sivakami Venkatachalam -S for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K240039 Device Name Enteral Feeding Sets #### Indications for Use (Describe) Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets include a bag to contain the feeding solution and/or a EnPlus spike to connect to a pre-filled container. The device is used for infants, children, adolescents and adults. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif font, also in blue. A curved line underlines the word "MEDICAL". ## 510(k) Summary - K240039 Device Trade Name: Enteral Feeding Sets Common Name: Enteral Feeding Sets Classification Name: Gastrointestinal tube and accessories Product Code: KNT, PIF Regulation: 876.5980 Regulatory Class: Class II #### Submitter/Manufacturer: Hangzhou Primecare Medical Co., Ltd. Room 408-409, Zancheng Center West, Shangcheng District, Hangzhou Zhejiang 310008 China Tel: 86-571-81958620 Regulatory Contact: Li Xuedong, Regulatory Director ## Predicate Device Device Trade Name: CONOD Enteral Feeding Sets 510(k) Number: K181276 Classification: Class II, 876.5980, Gastrointestinal tube and accessories Product Code: KNT, PIF ## Device Description The Enteral Feeding Sets have four models, Enteral Feeding Pump bag Set, Enteral Feeding Pump EnPlus spike Set, Enteral Feeding Gravity bag Set, Enteral Feeding Gravity EnPlus spike Set. The Enteral Feeding Pump bag Set and Enteral Feeding Gravity bag Set both have a structure of a formula bag with 100ml graduations and a protective closure cap. The difference is the former is droved by pump and the latter is droved by gravity. Enteral Feeding Pump EnPlus spike Set and Enteral Feeding Gravity EnPlus spike Set feature a piercing spike (with or without screw cap) used to connect the set to a prefilled container of enteral feeding solution. It is a single-use, non-sterile device. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below that, the word "MEDICAL" is written in a smaller, sans-serif font, also in blue. A curved line underlines the word "PrimeCare". # Intended Use / Indications for Use Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets include a bag to contain the feeding solution and/or a EnPlus spike to connect to a pre-filled container. The device is used for infants, children, adolescents and adults. ## Technological Comparison - > The subject device has the same intended use as the predicate device, dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution, and uses the same technology. - > The subject device has the same technological characteristics and provide the same principle of operation as the predicate device. - > The differences: - The predicate device has only 1200ml volume bags, the subject device adds different volumes (1200ml,1000ml,500ml,100ml), Material and intended used are the same. - The subject device adds one universal screw cap(with or without) to the spike set to extend the connectivity for different pre-filled formula containers. Material of the screw cap is exactly same as ENPlus spike. - The subject device adds a new model that replaces the built-in enteral bag with ENPlus spike and screw cap so that it can connects to the pre-filled nutrition containers instead of having to prepare formula by the user. the ENPlus spike and screw cap introduced in this set is the same as those used in Enteral feeding pump EnPlus Spike sets. - The subject device removes the EnFit transitional connector. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a stylized, cursive font in blue. Below "PrimeCare" is the word "MEDICAL" in smaller, block letters, also in blue. A curved line underlines the word "MEDICAL". #### Shown below is a side by side comparison of the subject device with the predicate device. A | Device<br>Characteristic | Subject device | Predicate device<br>(K181276) | Results | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Device name | Enteral Feeding Sets | CONOD Enteral Feeding Sets | same | | Indications for<br>Use | Enteral Feeding Sets are intended to dispense liquid nutrition<br>(feeding solutions) at a user controlled rate. These enteral<br>feeding sets interface with a patient's feeding tube and may use<br>gravity or an enteral feeding pump to dispense feeding solution.<br>The enteral feeding sets include a bag to contain the feeding<br>solution and/or a EnPlus spike to connect to a pre-filled<br>container. The device is used for infants, children, adolescents<br>and adults. | CONOD Enteral Feeding Sets are intended to dispense<br>liquid nutrition (feeding solutions) at a user controlled rate.<br>These enteral feeding sets interface with a patient's<br>feeding tube and may use gravity or an enteral feeding<br>pump to dispense feeding solution. The enteral feeding<br>sets include a bag to contain the feeding solution and/or a<br>piercing spike to connect to a pre-filled container. The<br>device is used for infants, children, adolescents and adults. | same | | Product code | PIF, KNT | PIF, KNT | same | | Classification<br>regulation | 21 CFR 876.5980 | 21 CFR 876.5980 | same | | Device<br>configurations | ● Enteral Feeding Pump bag Set (1200ml、1000ml、<br>500ml、100ml)<br>■ 1200ml/1000ml/500ml/100ml Enteral Bag with<br>closure cap<br>■ PVC tubing<br>■ Roller clamp<br>■ Drip chamber<br>■ Silicone tube<br>■ Silicone tube connector<br>■ Warning label<br>■ EnFit connector<br>■ Protection cap for the EnFit connector | ● CONOD Feeding Pump Set (1200ml)<br>■ 1200ml Enteral Bag with closure cap<br>■ PVC tubing<br>■ Roller clamp<br>■ Drip chamber<br>■ Silicone tube<br>■ Silicone tube connector<br>■ Warning label<br>■ EnFit connector<br>■ EnFit transitional connector<br>■ Protection cap for the transition connector | Similar<br>1.Adding different bag volume<br>.Material and intended used are the<br>same.<br>2. Remove the EnFit<br>transitional connector. | | | ● Enteral Feeding Pump EnPlus spike Set<br>■ Universal screw cap (with or without)<br>■ Protection cap for the spike<br>■ ENPlus spike | ● CONOD Safety spike plus Pump Set<br>■ Protection cap for the spike<br>■ ENPlus spike<br>■ PVC tubing<br>■ Roller clamp | Similar<br>1.Adding one universal screw cap<br>to the spike set to extend the | | PVC tubing Roller clamp Drip chamber Silicone tube Silicone tube connector Warning label EnFit connector Protection cap for the EnFit connector | Drip chamber Silicone tube Silicone tube connector Warning label EnFit connector EnFit transitional connector Protection cap for the transition connector | connectivity for different pre-<br>filled formula containers.<br><br>2.Material of the screw cap is<br>exactly same as ENPlus spike.<br><br>3.Remove the EnFit transitional<br>connector. | | | Enteral Feeding Gravity bag Set (1200ml、1000ml、<br>500ml、100ml) 1200ml/1000ml/500ml/100ml Enteral Bag with<br>closure cap PVC tubing Roller clamp Drip chamber Warning label EnFit connector Protection cap for the EnFit connector | CONOD Gravity Feeding Set (1200ml) 1200ml Enteral Bag with closure cap PVC tubing Roller clamp Drip chamber Warning label EnFit connector EnFit transitional connector Protection cap for the transition connector | Similar<br><br>1.Adding different bag volume<br><br>Material and intended used are the<br>same.<br><br>2.Remove the EnFit transitional<br>connector. | | | Enteral Feeding Gravity EnPlus spike Set Universal screw cap (with or without) Protection cap for the spike ENPlus spike PVC tubing Roller clamp Drip chamber Warning label EnFit connector Protection cap for the EnFit connector | CONOD Gravity Feeding Set (1200ml) 1200ml Enteral Bag with closure cap PVC tubing Roller clamp Drip chamber Warning label EnFit connector EnFit transitional connector Protection cap for the transition connector | Different<br><br>1.Replacing the built-in enteral<br>bag with ENPlus spike and screw<br>cap so that it can connects to the<br>pre-filled nutrition containers<br>instead of having to prepare<br>formula by the user. the ENPlus<br>spike and screw cap introduced in<br>this set is the same as those used in<br>Enteral feeding pump EnPlus<br>Spike sets | | | | | | 2.Remove the EnFit transitional<br>connector. | | Design feature | Non DEHP material sets featuring ISO 80369-3:2016 connector | Non DEHP material sets featuring ISO 80369-3:2016 connector | same | | Materials | PVC、ABS、Silicone、LDPE、paper | PVC、ABS、Silicone、LDPE、paper | same | | Prescription use or not | Prescription use | Prescription use | same | | Disposable or not | Disposable for single use | Disposable for single use | same | | Intended use time | No longer than 24 hours | No longer than 24 hours | same | | Sterile or not | Non-sterile | Non-sterile | same | | Shelf-life | 5 years | 5 years | same | | Available tube lengths | Enteral Feeding Pump bag Set: 60cm+157cm<br>Enteral Feeding Pump EnPlus spike Set: 60cm+157cm<br>Enteral Feeding Gravity bag Set: 170cm<br>Enteral Feeding Gravity EnPlus spike Set: 170cm | Pump set: 60cm+157cm<br>Spike set: 60cm+157cm<br>Gravity set: 170cm | Similar | {7}------------------------------------------------ Image /page/7/Picture/2 description: The image contains the logo for PrimeCare Medical. The logo is in blue and consists of the word "PrimeCare" in a script font, with the word "MEDICAL" in smaller, block letters underneath. A curved line underlines the word "MEDICAL". {8}------------------------------------------------ Image /page/8/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a stylized blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif font, also in blue. A curved blue line underlines the word "MEDICAL". {9}------------------------------------------------ Image /page/9/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif font, also in blue. A curved line underlines the word "MEDICAL". # Summary of Non-Clinical Testing The non-clinical bench performance testing of the Enteral Feeding Sets is tested according to ISO 80369-3:2016(5-123), ISO20695:2020(9-138), ISO18250-3:2018(5-139), IEC62366-1:2020(5-129), ISO 10993 series of biological standards, Packaging test related standards and the guidance: Safety Considerations to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications. ## Table 1: Metrology (include EnFit connector, ENPlus spike & screw cap) | No. | Items | Standards | Acceptance criteria | |-----|------------------------------|------------------|-----------------------| | 1 | Dimensions (EnFit connector) | ISO 80369-3:2016 | Meet the requirements | | 2 | Dimensions (ENPlus spike) | ISO 18250-3:2018 | Meet the requirements | | 3 | Dimensions (screw cap) | ISO 18250-3:2018 | Meet the requirements | ## Table 2: Appearance inspection | No. | Items | Standards | Acceptance criteria | |-----|---------------------------------------------------------------------------------------------|----------------|---------------------| | 1 | Deformed parts, Cracking and<br>burr, Incomplete seal or<br>breakage of bag,<br>Kinked tube | Study in-house | None | | 2 | Bag Printing | Study in-house | Present | ## Table 3: EnFit connector performance testing | No. | Items | Standards | Acceptance criteria | |-----|-------------------------------------------------|------------------|----------------------------------------------------| | 1 | Fluid leakage<br>(Leakage by pressure<br>decay) | ISO 80369-3:2016 | shall not leak by more than<br>0.005 Pa.m³/s | | 2 | Stress cracking | ISO 80369-3:2016 | No signs of leakage | | 3 | Resistance to<br>separation from axial<br>load | ISO 80369-3:2016 | shall not separate from the<br>reference connector | | 4 | Resistance to<br>separation from<br>unscrewing | ISO 80369-3:2016 | shall not separate from the<br>reference connector | {10}------------------------------------------------ Image /page/10/Picture/1 description: The image contains the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif font, also in blue. A curved line underlines the word "MEDICAL". | 5 | Resistance to<br>overriding | ISO 80369-3:2016 | shall not override the threads<br>or lugs of the reference<br>connector | |---|--------------------------------|------------------|-----------------------------------------------------------------------------------------------------------------------| | 6 | Disconnection by<br>unscrewing | ISO 80369-3:2016 | shall separate from the<br>reference connector with an<br>applied unscrewing<br>torque of no greater than 0.35<br>N·m | # Table 4: ENPlus spike & screw cap performance testing | No. | Items | Standards | Acceptance criteria | |-----|---------------------------------------------|------------------|-------------------------------------------------------------------------| | 1 | Positive pressure liquid<br>leakage | ISO 18250-3:2018 | shall show no signs of<br>leakage | | 2 | Subatmospheric-pressure<br>air leakage | ISO 18250-3:2018 | shall not leak by more than<br>0.005 Pa.m3/s | | 3 | Stress cracking | ISO 18250-3:2018 | No signs of leakage | | 4 | Resistance to separation<br>from axial load | ISO 18250-3:2018 | shall not separate from the<br>reference connector. | | 5 | Resistance to separation<br>from unscrewing | ISO 18250-3:2018 | shall not separate from the<br>reference connector | | 6 | Resistance to overriding | ISO 18250-3:2018 | shall not override the<br>threads or lugs of the<br>reference connector | # Table 5: Enteral Feeding Sets performance testing | No. | Items | Standards | Acceptance criteria | |-----|------------------|----------------------------------|--------------------------------------------------------------------------------------------------| | 1 | Tensile strength | ISO 20695-2020 | shall withstand a tensile<br>force of 15 N before<br>breaking, becoming<br>detached, or cracking | | 2 | Leakage | ISO 20695-2020 | shall not show signs of<br>leakage sufficient to<br>form a falling drop of<br>water | | 3 | Flow rate | ISO 20695-2020<br>Study in-house | Meet the requirements | {11}------------------------------------------------ Image /page/11/Picture/1 description: The image shows the logo for PrimeCare Medical. The word "PrimeCare" is written in a cursive, blue font. Below "PrimeCare" is the word "MEDICAL" in a smaller, sans-serif, blue font. There is a curved line underneath the word "MEDICAL". ## Hangzhou Primecare Medical Co., Ltd. Table 6: Packaging performance testing | No. | Items | Standards | Acceptance criteria | |-----|--------------------|------------------------------------|------------------------------------------------------| | 1 | Visual inspection | ASTMF1886/F1886M-16 | No defect of the integrity<br>of seals must be found | | 2 | Seal strength test | ASTM F88/F88M-21<br>Study in-house | Should not be less than<br>12 N/15mm | ## Table 7: biological test | No. | Items | Standards | Acceptance criteria | |-----|------------------------------|-------------------|---------------------| | 1 | Cytotoxicity | ISO 10993-5:2009 | Non-Cytotoxic | | 2 | Sensitization | ISO 10993-10:2021 | Non-Sensitizer | | 3 | Intracutaneous<br>Irritation | ISO 10993-23:2021 | Non-Irritant | ## Table 8: Usability Study | No. | Items | Standards | Acceptance criteria | |-----|-----------------|------------------|-----------------------| | 1 | Usability Study | IEC 62366-1:2020 | Meet the requirements | ## Summary of Clinical Testing Clinical testing was not required for device evaluation. ## Conclusion The conclusions drawn from the nonclinical testing for the subject device, Enteral Feeding Sets, demonstrate that the device is as safe, as effective, and performs as well as the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...