ENvizion Medical ENvue; ENvizion Medical Enteral Feeding Tube
K182915 · Envizion Medical , Ltd. · KNT · Feb 13, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K182915
Device Name
ENvizion Medical ENvue; ENvizion Medical Enteral Feeding Tube
Applicant
Envizion Medical , Ltd.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Feb 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Indications for Use
The ENvizion Medical ENvue System is designed to aid qualified operators in the placement of the ENvizion Medical Enteral Feeding Tube of 10 Fr and 12 Fr into the stomach or small intestine of adult patients requiring enteral feeding. The ENvizion Medical ENvue System is intended as an adjunct to current placement practices for assisting clinical practitioners who place feeding tubes. The ENvizion Medical Enteral Feeding Tube (EFT) has been specifically designed for use with the ENvue System and is intended for placement in the stomach or small intestine. It is intended for use in adult patients who require intermittent or continuous feeding via the oro/nasoenteric route. The EFT is intended only to be used with a feeding pump and is not compatible with gravity-based feeding bags.
Device Story
ENvue System is an electro-mechanical device with embedded software designed to assist clinicians (physicians, technicians, nutritionists) in placing enteral feeding tubes. System tracks the ENvizion Medical Enteral Feeding Tube (EFT) as it progresses through the oro/nasoenteric route into the stomach or small intestine. Principle of operation involves electromagnetic (EM) field sensing technology; system uses a field generator and multiple EM receivers, including one integrated into the tube's distal tip. Real-time placement pathway is displayed on a computer screen for the operator. Once placement is confirmed, the ENvue is disconnected from the EFT. The EFT is a single-use, polyurethane, radiopaque, multi-lumen tube compatible with feeding pumps via ENFit connectors. Used in hospitals and healthcare facilities. Provides visual guidance to improve placement accuracy; reduces reliance on traditional methods; benefits patients by facilitating safe, efficient enteral access.
Clinical Evidence
Prospective, single-center, non-randomized clinical study of 58 subjects (54 placements). Primary endpoints focused on safety and effectiveness of tube guidance. Results showed no device guidance-related adverse events and no serious adverse events. One minor manufacturing nonconformance identified. Bench testing included biocompatibility, tensile strength, flow rate, and simulated use.
Technological Characteristics
Polyurethane radiopaque tube; multi-lumen design; 10 Fr and 12 Fr sizes. Electromagnetic field sensing technology for tracking. Connectivity: standalone system with display. Standards: ISO 10993-1 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 80369-3 (ENFit connectors), BS/EN 1615/1618.
Indications for Use
Indicated for adult patients requiring enteral feeding via oro/nasoenteric route into stomach or small intestine. Used as an adjunct to clinical practice for feeding tube placement.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
K191387 — ENvizion Medical ENVue SYSTEM, ENvizion Medical Enteral Feeding Tube · Envizion Medical , Ltd. · Jun 14, 2019
K203133 — ENvizion Medical ENvue ENvizion Medical Enteral Feeding Tube · Envizion Medical , Ltd. · Mar 24, 2021
K080679 — MODIFICATION TO CORTRAK ENTERAL ACCESS DEVICE, MODEL 20-0900 · Viasys Medsystems · Jun 27, 2008
K191340 — CORTRAK* 2 Equilateral Enteral Access System · Avanos Medical, Inc. · Jun 4, 2019
K240965 — CORTRAK* 2 Enteral Access System (20-0950) · Avanos Medical, Inc. · Jun 6, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 13, 2019
ENvizion Medical Ltd. % Clay Anselmo Principle Consultant Shriner & Associates 429 Whitepine Creek Road Trout Creek, MT 59874
Re: K182915
Trade/Device Name: ENvizion Medical ENvizion Medical Enteral Feeding Tube Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT Dated: January 18, 2019 Received: January 23, 2019
Dear Clay Anselmo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or (301) 796-7100).
Sincerely,
# Mark J. Antonino -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182915
#### Device Name
ENvizion Medical ENvue; ENvizion Medical Enteral Feeding Tube
#### Indications for Use (Describe)
The ENvizion Medical ENvue System is designed to aid qualified operators in the ENvizion Medical Enteral Feeding Tube of 10 Fr and 12 Fr into the stomach or small intestine of adult patients requiring enteral feeding. The ENvizion Medical ENvue System is intended as an adjunct to current practices for assisting clinical practitioners who place feeding tubes.
The ENvizion Medical Enteral Feeding Tube (EFT) has been specifically designed for use with the ENvue System and is intended for placement in the stomach or small intestine. It is intended for use in adult patients who require intermittent or continuous feeding via the oro/nasoenteric route. The EFT is intended only to be used with a feeding pump and is not compatible with gravity-based feeding bags.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
#### Introduction:
This document contains the 510(k) Summary for the ENvizion Medical ENvue and ENvizion Medical Enteral Feeding Tube. The content of this summary is based on the requirements set forth in 21 CFR 807.92(c).
#### SUBMITTER INFORMATION
| Applicant / Manufacturer<br>Name and Address | ENvizion Medical Ltd.<br>7 Haarad Street<br>Tel Aviv, 6971060<br>Israel<br>Phone +972 72-2288240 |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) contact person | Clay Anselmo<br>Principal Quality and Regulatory Consultant<br>Shriner & Associates<br>429 Whitepine Creek Road<br>Trout Creek, MT 59874<br>Clay.anselmo@yahoo.com<br>(303) 907-2955 |
| Date prepared | 10/11/2018 |
| DEVICE IDENTIFICATION | |
| Trade Names | ENvizion Medical ENvue<br>ENvizion Medical Enteral Feeding Tube |
| Regulation Name | Gastrointestinal tube and accessories |
Tubes, Gastrointestinal (and Accessories)
21 CFR Part 876.5980
Class II
KNT
Product Code
Classification
Regulation Number
#### PREDICATE DEVICE
Product Code Device Name
| Trade names | CORTRAK Enteral Access Device and CORFLO<br>Nasoenteric Feeding Tubes with<br>Electromagnetic Transmitting Stylet |
|---------------|-------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K080679 (CORTRAK Enteral Access Device)<br>K821906 (CORFLO NasoEnteric Feeding Tube) |
{4}------------------------------------------------
#### DEVICE DESCRIPTION
The ENvizion Medical ENvue System is an electro-mechanical device with embedded software designed to aid in the placement of the ENvizion Medical Enteral Feeding Tube, which is an enteral feeding tube placed into the stomach or small intestine of patients requiring enteral feeding. The ENvue System tracks the Enteral Feeding Tube (EFT) as it progresses through the oro/nasoenteric route down the esophagus, into the stomach and to the small intestine anatomy and displays the placement pathway in real time during placement is completed, the user disconnects the ENvue from the EFT. The EFT connects to a feeding pump using ENFit connections.
#### INDICATIONS FOR USE
The ENvizion Medical ENvue System is designed to aid qualified operators in the placement of the ENvizion Medical Enteral Feeding Tube of 10 Fr and 12 Fr into the stomach or small intestine of adult patients requiring enteral feeding. The ENvizion Medical ENvue System is intended as an adjunct to current placement practices for assisting clinical practitioners who place feeding tubes.
The ENvizion Medical Enteral Feeding Tube (EFT) has been specifically designed for use with the ENvue System and is intended for placement in the stomach or small intestine. It is intended for use in adult patients who require intermittent or continuous feeding via the oro/nasoenteric route. The EFT is intended only to be used with a feeding pump and is not compatible with gravity-based feeding bags.
#### TECHNOLOGICAL CHARACTERISTICS COMPARISON
Substantial Equivalence: The ENvizion Medical ENvue is substantially equivalent to the CORTRAK Enteral Access Device and CORTRAK CORFLO NasoEnteric feeding tubes with transmitting stylet
The 510(k) Substantial Equivalence Decision-making Process (detailed) from the 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)] was followed as described below:
- . The ENvue device has the same intended use and similar indications for use asthe Predicate device.
- The ENvue device uses the same fundamental technology as the Predicate device and very similar detailed technological solutions as follows: Both products use electromagnetic fields and receivers to provide placement pathway of the tube through the patient's naso/oro enteral anatomy.
- The small differences between the ENvue device and the Predicate do not raise new types of questions of safety or effectiveness
- The biocompatibility of both products' patient contact materials complies with ISO O 10993-1 in accordance with FDA guidance related to the application of this standard.
- Both products have certification to medical device safety standards for 60601-1 O Electrical Safety and 60601-1-2 EMC
For a more detailed comparison, refer to the Substantial Equivalence table below.
### PERFORMANCE DATA
There are no known performance standards for this device.
The ENvizion Medical ENvue and Enteral Feeding Tube were verified and validated in accordance with 820.30. The following tests were completed to demonstrate substantial equivalence and that any technological differences do not raise new or different questions of safety and effectiveness.
{5}------------------------------------------------
The device successfully passed all of the testing and the results demonstrate the device is safe, effective, and performs as well or better than the predicate device.
- Biocompatibility
- Dimensional Inspection ●
- Liquid Leakage Testing
- Visual Inspection ●
- Tensile Strength Testing
- Flow Rate Testing
- Tubing Stiffness Comparison
- Simulated Use
- . Packaging Validation
- Shelf Life Validation
- Gastric Fluid Compatibility
- Software Validation
- IEC 60601-1 Electrical Safety
- IEC 60601-1-2 Electromagnetic Compatibility (EMC)
- Summative Usability
- . Comparative Testing
- MR Compatibility Testing
- Human Clinical Study
To provide confirmatory evidence of the safety and effectiveness of the ENvue device, two clinical studies were executed and the data pooled from 58 subjects. Each of the studies was a prospective, single center, non-randomized clinical trial in subjects that required placement of a enteral feeding tube as a result of their current medical condition. The study included placements through both the oro and naso routes.
An analysis of the data gathered from 54 placements during the clinical study showed no device guidance-related adverse events and no serious adverse events of any kind occurred during the study. There were thirteen alleged device malfunctions; twelve of the thirteen tube placements occurred were determined not to be adverse events per the case report forms. One incident was identified as an adverse event; however, upon further reviewed was determined not to be a serious adverse event and was identified as a manufacturing nonconformance. A review of the CRF's and associated procedural data showed the device performed safely and effectively without any guidance related incidents.
The device has been designed in conformance to the following voluntary recognized consensus standards:
- o BS/EN 1615:2000 – Enteral feeding catheters and enteral giving sets for single use and their connectors. Design and testing.
- BS/EN 1618:1997 Catheters other than intravascular catheters. Test methods for ● common properties.
{6}------------------------------------------------
- ANSI/AAMI/ISO 10993-1:2009(R) 2013 – Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- IEC 60601-1, Ed. 3: Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance; Corrigendum 1(2006): Corrigendum 2 (2007)
- IEC 60601-1-2 Ed 4.0: Medical Electrical Equipment Part 1-2: General Requirements for ● Basic Safety and Essential Performance – Collateral Standard: Electromagnetic Compatibility - Requirements and Tests
- IEC 62304:2015, Medical Device Software Software Lifecycle Processes ●
- IEC 62366-1:2015: Medical devices -- Part 1: Application of usability engineering to ● medical devices
- ISO 80369-3: 2016: Small-bore connectors for liquids and gases in healthcare applications -- Part 3: Connectors for enteral applications
### SUBSTANTIAL EQUIVALENCE COMPARISON
| Characteristic | CORPAK CORTRAK<br>(Predicate Device) | ENvizion Medical™ ENvue<br>(Subject Device) | Comparison<br>to Predicate |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K080679 (CORTRAK 2 Enteral<br>Access Device)<br>K821906* (CORFLO Feeding<br>Tube) | K182915 | N/A |
| Regulation Number | 21 CFR 876.5980 | 21 CFR 876.5980 | Identical |
| Classification Name | Gastrointestinal tube and<br>accessories. | Gastrointestinal tube and<br>accessories. | Identical |
| Product<br>Classification Code | KNT | KNT | Identical |
| Regulatory Class | Class 2 | Class 2 | Identical |
| Intended Use | Aids qualified operators in the<br>placement of the of<br>nasoenteral feeding tubes<br>into the stomach or small<br>intestine of patients requiring<br>enteral feeding. | Aids qualified operators in the<br>placement of the of enteral<br>feeding tubes into the<br>stomach or small intestine of<br>patients requiring enteral<br>feeding. | Identical |
| Indications for Use | The CORTRAK System is an<br>electrical device designed to<br>aid qualified operators in the<br>placement of the VIASYS<br>MedSystems feeding tubes of<br>8 FR or greater into the<br>stomach or small bowel of<br>patients requiring enteral<br>feeding. Prior to commencing<br>the delivery of food,<br>confirmation or correct tube<br>placement by acceptable<br>hospital protocol is required. | The ENvizion Medical ENvue<br>System is designed to aid<br>qualified operators in the<br>placement of the ENvizion<br>Medical Enteral Feeding Tube<br>of 10 Fr and 12 Fr into the<br>stomach or small intestine of<br>adult patients requiring<br>enteral feeding. The ENvizion<br>Medical ENvue System is<br>intended as an adjunct to<br>current placement practices<br>for assisting clinical<br>practitioners who place<br>feeding tubes. | Substantially<br>Equivalent<br>Differences<br>do not raise<br>new or<br>different<br>questions<br>regarding<br>safety or<br>effectiveness |
| Characteristic | CORPAK CORTRAK<br>(Predicate Device) | ENvizion Medical™ ENvue<br>(Subject Device) | Comparison<br>to Predicate |
| Operating Principle<br>and design | Rechargeable battery<br>powered electromagnetic<br>(EM) system sensing<br>technology to track and<br>display path of feeding tube<br>using an EM Tracking System,<br>Computer and Display.<br><br>The EM Transmitter is<br>integrated into the feeding<br>tube distal tip and the EM<br>Receiver is placed externally<br>on the Xiphoid Process. | The ENVIZION MEDICAL<br>Enteral Feeding Tube (EFT)<br>has been specifically designed<br>for use with the ENvue<br>System and is intended for<br>placement in the stomach or<br>small intestine. It is intended<br>for use in adult patients who<br>require intermittent or<br>continuous feeding via the<br>oro/nasoenteric route. The<br>EFT is intended only to be<br>used with a feeding pump and<br>is not compatible with<br>gravity-based feeding bags.<br><br>Rechargeable battery<br>powered electromagnetic<br>(EM) system sensing<br>technology to track and<br>display path of feeding tube<br>using an EM Tracking System,<br>Computer and Display.<br><br>The EM Transmitter is the<br>Field Generator and the<br>system uses multiple EM<br>Receivers including one<br>integrated in the tube distal<br>tip. | Substantially<br>Equivalent<br><br>Differences<br>do not raise<br>new or<br>different<br>questions<br>regarding<br>safety or<br>effectiveness |
| | A single-use polyurethane<br>radiopaque tube and tip (for<br>X-ray visualization).<br>Nutrition is administered with<br>the polymeric tubing<br>providing a fluid path<br>between the nutritional<br>supplement source (e.g.<br>feeding bag or feeding pump)<br>and the stomach or small<br>intestine of the patient. | A single-use polyurethane<br>radiopaque tube and tip (for<br>X-ray visualization). Nutrition<br>is administered with the<br>polymeric tubing providing a<br>fluid path between the<br>nutritional supplement source<br>(i.e. feeding pump) and the<br>stomach or small intestine of<br>the patient. | |
| Tube Type | Single lumen | Multi Lumen | Substantially<br>equivalent<br><br>Differences<br>do not raise<br>new or<br>different<br>questions |
| Characteristic | CORPAK CORTRAK<br>(Predicate Device) | ENvizion Medical™ ENvue<br>(Subject Device) | Comparison<br>to Predicate |
| Tube Outer<br>Diameter | 8 to 12 Fr | 10 and 12Fr | Substantially<br>equivalent<br>Differences<br>do not raise<br>new or<br>different<br>questions<br>regarding<br>safety or<br>effectiveness |
| Tube Usable Length | 36 to 55 in<br>91 to 140 cm | 36 to 55 in<br>91 to 140 cm | Identical |
| Patient contacting<br>tubing material | Polyurethane | Polyurethane | Identical |
| Biocompatibility | ISO 10993 | FDA application of ISO 10993 | Substantially<br>Equivalent |
| Feeding Connector | 80369-3 Connector - ENFit | 80369-3 Connector - ENFit | Identical |
| Sterilization | Non-sterile | Non-sterile | Identical |
| Target User | Intended for use by<br>physicians, technicians and<br>nutritionists. | Intended for use by<br>physicians, technicians and<br>nutritionists. | Identical |
| Use Environment | Hospitals and other<br>healthcare facilities | Hospitals and other<br>healthcare facilities | Identical |
| Access / Anatomical<br>Site | Nasoenteric | Oro / Nasoenteric | Substantially<br>Equivalent |
| Energy Type | Electromagnetic Field | Electromagnetic Field | Identical |
| Patient Population | Adults | Adults | Identical |
{7}------------------------------------------------
{8}------------------------------------------------
* Note: Indications for Use statement for K821906 are not currently available.
The ENvizion Medical ENvue System and Enteral Feeding Tube is substantially equivalent with respect to the indication for use, technological characteristics, target user, and use environment to the following legally marked Predicate devices:
- Corpak CORTRAK Enteral Access Device (K080679)
- . Corpak CORTRAK CORFLO feeding tube (K821906)
#### CONCLUSION
The ENvizion Medical ENvue and Enteral Feeding Tube are substantially equivalent to the CORTRAK Enteral Access Device and CORTRAK CORFLO NasoEnteric Feeding Tube with Transmitting Stylet.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.