GABRIEL BLUE TUBE MAGNETICALLY GUIDED ENTERAL FEEDING TUBE, MODEL GFT-111
K072787 · Syncro Medical Innovations, Inc. · KNT · Oct 31, 2007 · Gastroenterology, Urology
Device Facts
Record ID
K072787
Device Name
GABRIEL BLUE TUBE MAGNETICALLY GUIDED ENTERAL FEEDING TUBE, MODEL GFT-111
Applicant
Syncro Medical Innovations, Inc.
Product Code
KNT · Gastroenterology, Urology
Decision Date
Oct 31, 2007
Decision
SESE
Submission Type
Special
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The device is intended for direct placement into the small bowel. The tube functions as a conduit to facilitate enteral feeding, and may be used in pediatric, adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with a functioning gut who require short- to moderate-term feeding support, such as post-trauma patients, post-surgical patients, burn patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom malnutrition exists, or may result, secondary to an underlying disease or condition. The external steering magnet functions as a guidance tool to assist in the safe, rapid placement of the feeding tube into the small bowel.
Device Story
Gabriel Blue Tube is an enteral feeding tube (11 Fr, 50" length) designed for small bowel placement. Device includes an internal stylet with a magnetic tip. An external steering magnet is used by a clinician to guide the tube tip into the duodenum. A reed switch and light indicator signal when the internal magnet is captured by the external magnet, facilitating safe and rapid placement. The tube acts as a conduit for enteral nutrition. Used in clinical settings for patients requiring short- to moderate-term feeding support. Benefits include reduced risk of tube misplacement and improved efficiency of placement.
Clinical Evidence
Bench testing only. Performance tests included tensile, flow, flexibility, and biocompatibility testing to demonstrate substantial equivalence.
Technological Characteristics
11 Fr, 50" (127cm) enteral feeding tube. Features internal stylet with magnetic tip and external steering magnet. Includes reed switch and light indicator for magnetic capture. Materials and design are identical to the predicate device.
Indications for Use
Indicated for pediatric, adult, or elderly patients with a functioning gut requiring short- to moderate-term enteral feeding support due to inability to consume adequate oral diet. Includes post-trauma, post-surgical, burn, and high-risk patients prone to tube misplacement or malnutrition.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
Gabriel Blue Tube Magnetically Guided Feeding Tube (formerly MagnaFlow® Magnetically Guided Enteral Feeding Tube) (K021991)
Submission Summary (Full Text)
{0}------------------------------------------------
K072787 R. lot 2
Special 510(k)
| Submitter: | Syncro Medical Innovations, Inc.<br>20 West Federal<br>Suite M-5B<br>Youngstown, OH 44503 | JC7 3 1 200 |
|---------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Submission<br>Correspondent: | William G. McLain<br>President and Principal Consultant<br>Keystone Regulatory Services, LLC.<br>Phone: 717-656-9656<br>Fax: 717-656-3434<br>Email: bill.mclain@keystoneregualtory.com | |
| Date summary prepared: | September 28, 2007 | |
| Device trade name: | Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube | |
| Device common name: | Feeding Tube | |
| Device classification name: | Tube, Feeding<br>78 KNT at 21 CFR Part 876.5980 | |
| Legally marketed devices<br>to which the device is<br>substantially equivalent: | Gabriel Blue Tube Magnetically Guided Feeding Tube (Formerly named<br>the MagnaFlow® Magnetically Guided Enteral Feeding Tube), K021991 | |
| Description of the device: | The Gabriel Blue Tube serves as a conduit through which enteral feeding<br>solutions are directly infused into the patient's small bowel. The device will<br>be marketed in one length (50" [127cm]) and one French size (11 Fr). The<br>feeding tube contains a stylet which has magnets attached to its tip. An<br>external magnet is used to capture the tip and guide the device to the<br>duodenum. A reed switch and light indicate when the magnet has been<br>captured and can be guided to its final location. | |
| Intended use of the device: | The device is intended for direct placement into the small bowel. The tube<br>functions as a conduit to facilitate enteral feeding, and may be used in<br>pediatric, adult or elderly patients who cannot consume an adequate diet<br>orally. Small bowel feeding may be indicated for patients with a<br>functioning gut who require short- to moderate-term feeding support, such<br>as post-trauma patients, post-surgical patients, burn patients, general<br>trauma patients, high-risk patients prone to tube misplacement<br>complications, and patients in whom malnutrition exists, or may result,<br>secondary to an underlying disease or condition.<br>The external steering magnet functions as a guidance tool to assist in the<br>safe, rapid placement of the feeding tube into the small bowel. | |
| Technological<br>characteristics: | The proposed device has the same technological characteristics as the<br>predicate device(s). | |
# SECTION 5. 510(K) SUMMARY
{1}------------------------------------------------
K072787pc. dat2
:
.
| Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube | | Special 510(k) |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Performance tests: | Tests were performed to demonstrate substantial equivalence in the following areas:<br>• Tensile<br>• Flow<br>• Flexibility<br>• Biocompatibility | |
| Conclusions: | The results of the laboratory tests demonstrate that the device is as safe and effective as the legally marketed predicate devices. | |
:
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## OCT 3 1 2007
Syncro Medical Innovations, Inc. c/o Mr. William McLain President and Principal Consultant Keystone Regulatory Services, LLC 342 E. Main Street, Suite 207 LEOLA PA 17540
Re: K072787
Trade/Device Name: Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: September 28, 2007 Received: October 1, 2007
Dear Mr. McLain:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in intentate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Fedral Food, Or to and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal In addin
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Nancy Brogdon
Nancv C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K07278/B/of/
Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube
Special 510(k)
### SECTION 6. INDICATIONS FOR USE STATEMENT
510(k) Number:
K072787
Device Name:
Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube
Indications for Use: The Gabriel Blue Tube Magnetically Guided Enteral Feeding Tube is intended for direct placement into the small
bowel. The tube functions as a conduit to facilitate enteral feeding, and may be used in pediatric, adult or elderly patients who cannot consume an adequate diet orally. Small bowel feeding may be indicated for patients with a functioning gut who require short- to moderate-term feeding support, such as post-trauma patients, post-surgical patients, burn patients, general trauma patients, high-risk patients prone to tube misplacement complications, and patients in whom mainutrition exists, or may result, secondary to an underlying disease or condition.
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert Remmer
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _
Syncro Medical Innovations, Inc.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.