← Product Code [KNT](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KNT) · K061432

# HOSPIRA ENTERAL FEEDING SET, MODEL 20640 (K061432)

_Hospira, Inc. · KNT · Sep 11, 2006 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KNT/K061432

## Device Facts

- **Applicant:** Hospira, Inc.
- **Product Code:** [KNT](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KNT.md)
- **Decision Date:** Sep 11, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The HOSPIRA * Enteral Feeding Sets is intended to deliver, via gravity or a pamp, liquid nuttition formulas to an enteral access device (a feeding tube):

## Device Story

Hospira Enteral Feeding Sets consist of non-DEHP plasticized polyvinyl tubing, non-PVC enteral feeding bags, fluid shut-off devices, sight chambers, and infusion cassette technology. Device delivers liquid nutrition formulas to enteral feeding tubes via gravity or pump. Used in clinical settings to support patient nutrition. Output is the controlled flow of nutritional fluid into the patient's feeding tube. Benefits include non-DEHP and non-PVC material composition to reduce potential exposure to plasticizers.

## Clinical Evidence

No clinical data. Substantial equivalence is based on functional claims and material comparisons to predicate devices.

## Technological Characteristics

Components include non-DEHP plasticized polyvinyl tubing, non-PVC enteral feeding bags, fluid shut-off devices, sight chambers, and infusion cassette technology. Materials are non-DEHP and non-PVC. Device is intended for gravity or pump-assisted delivery of nutritional fluids.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Enteral Pump Set with Integral Container ([K810996](/device/K810996.md))
- Enteral Pump Set with Integral Container ([K052052](/device/K052052.md))
- Enteral Feeding Set with Cassette ([K865059](/device/K865059.md))
- Enteral Feeding Set with Cassette ([K052059](/device/K052059.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K06 1432
1 of 2

# SEP 1 1 2006

## 510(k) Summary

HOSPIRA. Incorporated Name of Submitter: 275 North Field Drive Lake Forest, Illinois 60045 Owner/Operator #: 9063339

#### Manufacturer and Establishment Registration Number:

Manufacturer:

HOSPIRA LTD. Carretera Sanchez, km 18 1/2 Parque Industrial Itabo, S.A. Haina, San Cristobal, Dominican Republic

Establishment Registration #: 9613251

Proprietary or Trade Name of Proposed Device: Hospira® Enteral Feeding Sets

Common Name: Enteral Bag w/Pump Set

### Device Classification, Pancode and ProCode: Class II, 80-KNT

Performance Standards: No performance standards have been established under Section 514 of the Food. Drag and Cosmetic Act for Infusion pumps. Tubes, gastrointestinal (and accessories) are regulated withm 21 CFR 880.5860.

#### Intended Use / Indications for Use:

The HOSPIRA * Enteral Feeding Sets is intended to deliver, via gravity or a pamp, liquid nuttition formulas to an enteral access device (a feeding tube):

### Proposed Device Description:

The HOSPIRA * Enteral Feeding Sets will consist of variations of the following components: non-DETIP Enteral feeding bags, non-DEHP plasticized polyvinyl tubing, fluid shut off devices, non-DEHP sight chambers, cassette technology and adapters for enteral feeding catheters.

#### Summary of Substantial Equivalence

The HOSPIRA " Enteral Feeding Sets as described in this substantially equivalent to the primary predicate device Enteral Pump Set with Integral Container (K810996 & K052052) with respect to the following:

Similarities:

- Both devices are mended to temporarily hold and deliver nutritional fluids from a container to a 】 patient's enteral access site
- The operating principles are the same for both devices 2)
- Both primary labeling and device bags clearly state not for I.V. or Epidural Use 3)

{1}------------------------------------------------

Difference:

1) The HOSPIRA * Enteral Feeding Sets will be comprised of all non-DEHP materials. 1) The HOSPIRA® Enteral Feeding Sets will contain a non-PVC enteral feeding bag.

The HOSPIRA® Enteral Feeding Sets as described in this substantially equivalent to the I he HUSFINA Timeral Pecaling Bets as assette (K865059 & K052059 & K052052) with respect to the following:

#### Similarities:

- 1) Both use the same infusion cassette technology
2) The operating principles are the same for both devices.

#### Difference:

- I ) The predicate is cleared for I.V. use, the subject is for enteral use
3) The HOSPIRA® Enteral Feeding Sets will contain a non-PVC enteral feeding bag.

r) The HOSPIRA® Enteral Feeding Sets will contain a non-PVC enteral feeding bag.
4) The HOSPIRA® Enteral Feeding Sets will be comprised of all non-DEHP materials.

## Statement of Safety and Effectiveness

The HOSPIRA " Enteral Feeding Sets meets the functional claims and intended use as described in the I he IrOST trent - Enterate and effective in terms of substantial equivalence as the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle-like symbol with three curved lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

## SEP 1 1 2006

Ms. Diane Rennpferd Sr. Associate, Global Regulatory Affairs - Devices Hospira, Inc. Department 0389, Bldg. H2 275 North Field Drive LAKE FOREST IL 60045

K061432 Re:

> Trade/Device Name: HOSPIRA® Enteral Feeding Sets Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KNT Dated: July 20, 2006 Received: July 21, 2006

Dear Ms. Rennpferd:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image is a black and white circular logo. The logo commemorates the centennial of the FDA from 1906-2006. The logo has the letters FDA in a bold font in the center with the word Centennial underneath. There are three stars below the word Centennial.

oting Public

{3}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known)

HOSPIRA® Enteral Feeding Sets Device Name:

Indications for Use:

The Hospira® Enteral Feeding Sets is intended to deliver, via gravity or a pump, liquid nutrition formulas to an enteral access device (a feeding tube).

Prescription Use X (Part 21 CFR 801 Subpart D)

ANDIOR

Over-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David Le. Byrom

(Division S Division of and Radiol 510(k) Num Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KNT/K061432](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KNT/K061432)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
