CRE FIXED WIRE BALLON DILATION CATHETER

K122924 · Boston Scientific Corp · KNQ · Oct 24, 2012 · Gastroenterology, Urology

Device Facts

Record IDK122924
Device NameCRE FIXED WIRE BALLON DILATION CATHETER
ApplicantBoston Scientific Corp
Product CodeKNQ · Gastroenterology, Urology
Decision DateOct 24, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.5365
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The CRE™ Balloon Dilatation Catheter is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the esophagus.

Device Story

CRE™ Balloon Dilatation Catheter is an esophageal dilator designed for use through an endoscope with a 2.8mm or larger working channel. The device features a balloon capable of 3 distinct, progressively larger diameters via controlled radial expansion. A stainless steel core wire runs the length of the shaft and remains in tension within the balloon to facilitate catheter removal from the endoscope post-dilation. Used by clinicians in endoscopic procedures to mechanically dilate esophageal strictures, the device provides a therapeutic intervention to improve esophageal patency. The current iteration incorporates a minor material formulation change to the flexible tip compared to the predicate.

Clinical Evidence

No clinical data. Evidence consists of bench testing including biocompatibility (cytotoxicity, sensitization, intracutaneous reactivity), chemical analysis, and tensile strength testing. All components and full devices met required specifications.

Technological Characteristics

Esophageal balloon dilator; stainless steel core wire; flexible tip with modified material formulation. Dimensions: compatible with 2.8mm+ endoscope working channels. Biocompatibility per AAMI/ANSI/ISO 10993-1:2009. Mechanical principle: controlled radial expansion of balloon. Non-powered, mechanical device.

Indications for Use

Indicated for adult and adolescent populations requiring endoscopic dilation of esophageal strictures.

Regulatory Classification

Identification

An esophageal dilator is a device that consists of a cylindrical instrument that may be hollow and weighted with mercury or a metal olive-shaped weight that slides on a guide, such as a string or wire and is used to dilate a stricture of the esophagus. This generic type of device includes esophageal or gastrointestinal bougies and the esophageal dilator (metal olive).

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K122924 pg 1 of 2 SECTION 5 510(k) SUMMARY #### 510(k) SUMMARY #### 1. Submitter: Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4560 Fax: 508-683-5939 Contact: Janis F. Taranto, M.S., RAC Regulatory Affairs Specialist Date Prepared: September 21, 2012 #### 2. Proposed Device: Trade Name: CRE™ Balloon Dilatation Catheter Classification Name: Dilator, Esophageal Regulation Number: 876.5365 Product Code: KNQ Classification: Class II #### 3. Predicate Device: Trade Name: CRETM Balloon Dilatation Catheter Manufacturer and Clearance Number: Boston Scientific Corporation, K971320 Classification Name: Dilator, Esophageal Regulation Number: 876. 5365 Product Code: KNQ Classification: Class II #### 4. Proposed Device Description: The proposed CRE™ Balloon Dilatation Catheter is capable of 3 distinct and progressively larger size diameters via controlled radial expansion. Specific balloon sizes are printed on each package and hub label. It is designed to be used through an endoscope having a 2.8mm or larger working channel. A stainless steel core wire extends through the entire length of the shaft. The wire is in tension within the balloon, which facilitates the removal of the catheter from the endoscope following balloon dilatation. The proposed device incorporates a minor material formulation change for the flexible tip of the device. #### 5. Indication for Use: The CRE™ Balloon Dilatation Catheter is intended for use in adult and adolescent populations · to endoscopically dilate strictures of the esophagus. Special 510(k) Premarket Notification, CRE Fixed Wire Balloon Dilatation Catheter {1}------------------------------------------------ #### 6. Technological Characteristics: The proposed CRE™ Balloon Dilatation Catheter is identical in design and manufacturing processes to the predicate CRE™ Balloon Dilatation Catheter (K971320) while incorporating a minor formulation change to the flexible tip. #### 7. Performance Data: Biocompatibility of the proposed device was confirmed via AAMI/ANSVISO 10993-1: 2009, and included Cytotoxicity, Sensitization Intracutaneous Reactivity (Irritation), In-virro testing has been performed and all components, subassemblies, and/or full devices met the required specifications for the completed tests including: chemical analysis, biocompatibility and tensile strength. #### 8. Conclusion: Boston Scientific Corporation has demonstrated that the proposed Boston Scientific CRETM Balloon Dilatation Catheter with minor device modifications is substantially equivalent to the predicate CRETM Balloon Dilatation Catheter (K971320). 1015 {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized symbol resembling a bird or wing design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 OCT 2 4 2012 Ms. Janis F. Taranto Regulatory Affairs Specialist Boston Scientific Corporation 100 Boston Scientific Way MARLBOROUGH MA 01752 Re: K122924 Trade/Device Name: CRE™ Balloon Dilatation Catheter Regulation Number: 21 CFR& 876.5365 Regulation Name: Esophageal dilator Regulatory Class: II Product Code: KNQ Dated: September 21, 2012 Received: September 24, 2012 Dear Ms. Taranto: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ ### Page 2 - device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 Jod attr.cl/www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Sincerely yours, Benjamin K. Twitchell Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # KI2292 4 #### SECTION 4 INDICATIONS FOR USE STATEMENT To Be Decemine 510(k) Number (if known): Device Name: Indications for Use: CRE™ Balloon Dilatation Catheter The CRE™ Balloon Dilatation Catheter is indicated for use in adult and adolescent populations to endoscopically dilate strictures of the esophagus. Prescription Use __ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 807 Subpart C) 000013 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Herbert Semler ivision Sign-Off ivision of Reproductive. Gastro-R urological Devices 510(k) Number Special 510(k) Premarket Notification, CRE Fixed Wire Balloon Dilatation Catheter
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...