← Product Code [KDI](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI) · K100334

# DIACAP POLYSULFONE HIFLO 23 HEMODILAYZER , MODEL 72036811 (K100334)

_B. Braun Avitum AG · KDI · Apr 22, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K100334

## Device Facts

- **Applicant:** B. Braun Avitum AG
- **Product Code:** [KDI](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI.md)
- **Decision Date:** Apr 22, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Diacap® Polysulfone HiFlo 23 Hemodialyzer is designed for single use in acute and chronic hemodialysis.

## Device Story

Diacap Polysulfone HiFlo 23 Hemodialyzer is a high-permeability dialyzer for acute and chronic hemodialysis. Device consists of polysulfone hollow fiber capillary membranes housed in polycarbonate with polyurethane potting resin and silicone O-ring. It functions as an extracorporeal blood filter to remove waste products and excess fluid from patient blood during dialysis. Intended for single-use in clinical settings; operated by trained healthcare professionals. Output is the filtered blood returned to the patient. Benefits include effective solute clearance and fluid management for patients with renal failure. Design is an extension of existing B. Braun polysulfone dialyzer line, featuring an increased surface area of 2.3m².

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

High-permeability hemodialyzer; polysulfone hollow fiber capillary membranes; polycarbonate housing; polyurethane (PUR) potting resin; silicone O-ring; 2.3m² surface area; single-use; sterile.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- Diacap® LO PS (10, 12, 15) and Diacap® HI PS (10, 12, 15, 18, 20) Hemodialyzers ([K071518](/device/K071518.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

B. Braun Avitum AG Special 510(k) Premarket Notification Diacap Polysulfone HiFlo 23 Hemodialyzer

#### 510(k) SUMMARY ഗ്

# APR 9 8 2010

B. Braun Avitum AG APPLICANT: Schwarzenberger Weg 73-79 34212 Melsungen, Germany Establishment Registration Number: 3002879653

SUBMITTER:

B. Braun Medical Inc. 901 Marcon Boulevard Allentown, PA 18109-9341 610-266-0500

#### CONTACT:

Angela J. Caravella Sr. Regulatory Affairs Analyst Phone: 610-596-2966 Fax: 610-596-2502 E-mail: Angela.Caravella@bbraun.com

DEVICE NAME:

COMMON OR USUAL NAME:

DEVICE

High Permeability Hemodialysis System

Diacap® Polysulfone HiFlo 23 Hemodialyzer

Class II per 21 CFR § 876.5860 Product Code KDI

## CURRENTLY MARKETED DEVICE (PREDICATE):

CLASSIFICATION:

DESCRIPTION:

Diacap® LO PS (10, 12, 15) and Diacap® HI PS (10, 12, 15, 18, 20) Hemodialyzers, (K071518)

The Diacap® Polysulfone HiFlo 23 Hemodialyzer is a high permeability dialyzer intended to be used in acute and chronic hemodialysis. It is composed of polysulfone hollow fiber capillary membranes in a polycarbonate housing with a polyurethane potting resin and a silicone O-ring. The proposed Diacap HiFlo 23 hemodialyzer will be available in one model, the 23, which represents the effective surface area of 2.3m². The dialyzer is intended for single use and is provided sterile.

## INTENDED USE:

The Diacap® Polysulfone HiFlo 23 Hemodialyzer is designed for single use in acute and chronic hemodialysis.

18

{1}------------------------------------------------

B. Braun Avitum AG Special 510(k) Premarket Notification Diacap Polysulfone HiFlo 23 Hemodialyzer

### SUBSTANTIAL EQUIVALENCE:

Diacap® Polysulfone HiFlo 23 Hemodialyzer has the same intended use and utilizes the same fundamental technology as the predicate device, the currently marketed Diacap LO PS (10, 12, 15) and Diacap HI PS (10, 12, 15, 18, 20) Hemodialyzers, (K071518). The Diacap HiFlo 23 Hemodialyzer is a modified version of the existing Diacap LO PS and HI PS dialyzers and can be considered an extension to the line of B. Braun Polysulfone Dialyzers.

10033.

The technological characteristics of the Diacap Polysulfone HiFlo 23 Hemodialyzer are equivalent to those of the existing Diacap polysulfone dialyzers. The Diacap Polysulfone HiFlo 23 Hemodialyzer is similar to the predicate device in material composition. Changes in the design of the proposed dialyzer include an increased surface area and two alternate materials have been added: polyurethane (PUR) potting resin and an alternate silicone material qualified for the dialyzer O-ring.

These changes do not pose a significant impact upon the fundamental technology of the Diacap polysulfone capillary dialyzers, as identified within the Risk Analysis. Verification and validation testing for the proposed dialyzer have been completed and all acceptance criteria have been met. The testing demonstrated that there are no differences between the predicate and the proposed device that raise new issues of safety or effectiveness.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter. The text is written in all capital letters.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

JUN -3 2010

B.Braun Avitum AG % Ms. Angela J. Caravella Sr. Regulatory Affairs Analyst B. Braun Medical, Inc. 901 Marcon Boulevard ALLENTOWN PA 18109-9341

Re: K100334

Diacap® Polysulfone HiFlo 23 Hemodialyzer Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: March 30, 2010 Received: April 1, 2010

Dear Ms. Caravella:

This letter corrects our substantially equivalent letter of April 22, 2010. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth

{3}------------------------------------------------

Page 2 - Ms. Angela J. Caravella

in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Herbert Kemen

Janine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

B. Braun Avitum AG Special SIO(k) Premarket Notification Diacap Polysulfone HiFlo 23 Hemodialyzer

| 4. INDICATIONS FOR USE STATEMENT                                                                            |                                           |
|-------------------------------------------------------------------------------------------------------------|-------------------------------------------|
|                                                                                                             | Page 1 of 1                               |
| 510(k) Number (if known):                                                                                   | K100334                                   |
| Device Name:                                                                                                | Diacap® Polysulfone HiFlo 23 Hemodialyzer |
| Indications For Use:                                                                                        |                                           |
| The Diacap® Polysulfone HiFlo 23 Hemodialyzer is designed for single use in acute and chronic hemodialysis. |                                           |
| Prescription Use X (Per 21 CFR 801.109)                                                                     | OR Over-The-Counter Use                   |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 00334 510(k) Number.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K100334](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K100334)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
