← Product Code [KDI](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI) · K031378

# FRESENIUS ULTRAFLUX HEMODIALYZER (K031378)

_Fresenius Medical Care North America · KDI · May 30, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K031378

## Device Facts

- **Applicant:** Fresenius Medical Care North America
- **Product Code:** [KDI](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI.md)
- **Decision Date:** May 30, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Fresenius Ultraflux dialyzer is designed for single use acute and chronic hemodialysis.

## Device Story

The Fresenius Ultraflux is a single-use hemodialyzer used for acute and chronic hemodialysis. It functions as a semi-permeable membrane filter to remove waste products and excess fluid from the blood of patients with renal failure. The device is used in clinical settings by trained healthcare professionals. It processes blood and dialysate flows to facilitate solute clearance and ultrafiltration. The output is the treated blood returned to the patient, which helps maintain electrolyte balance and fluid homeostasis. The device is designed for use with standard hemodialysis machines.

## Clinical Evidence

Bench testing only. In vitro performance testing conducted to evaluate priming volume, bloodside and dialysate-side pressure drops, ultrafiltration coefficient, and clearance of urea, creatinine, Vitamin B12, phosphate, and lysozyme. No clinical testing was performed.

## Technological Characteristics

Single-use hemodialyzer. High permeability membrane. Materials, design, and construction are equivalent to the predicate Optiflux 200NR. Operates via passive diffusion and convection (ultrafiltration) during hemodialysis. No software or electronic components.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- Fresenius Optiflux 200NR ([K002277](/device/K002277.md))
- Gambro Polyflux 17S ([K982414](/device/K982414.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a black and white illustration of a funnel. The funnel is composed of four sections that are stacked on top of each other. Each section is a trapezoid shape, with the widest part at the top and the narrowest part at the bottom.

K031378

Ray 1/3

# Fresenius Medical Care

### Fresenius Ultraflux Single Use Hemodialyzer "Special" 510(k) Premarket Notification

## Summary of Safety and Effectiveness

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.

#### A. Submitter's Information:

| Name:                | Fresenius Medical Care North America              |
|----------------------|---------------------------------------------------|
| Address:             | 95 Hayden Ave<br>Lexington, MA 02420              |
| Phone:               | 1-781-402-9068                                    |
| Fax:                 | (781) 402-9635                                    |
| Contact Person:      | Arthur Eilinsfeld, Director of Regulatory Affairs |
| Date of Preparation: | 29 April 2003                                     |

#### B. Device Name:

| Common Name:                       | Dialyzer, High Permeability with or without<br>Sealed Dialysate System |
|------------------------------------|------------------------------------------------------------------------|
| Product Code/Classification Panel: | 78KDI/Gastroenterology-Urology                                         |
| Classification:                    | Class II per §876.5860                                                 |

#### C. Predicate Device

The predicate devices for the Fresenius Ultraflux Hemodialyzers are:

- Fresenius Optiflux 200NR - #K002277 (8/25/00);
- . Gambro Polyflux 17S - #K982414 (3/26/99).

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the text "Page 2/3" written in black ink on a white background. The word "Page" is written in cursive, while the numbers "2/3" are written in a more standard font. The slash between the 2 and 3 is underlined.

Image /page/1/Picture/1 description: The image shows a geometric figure that resembles a downward-pointing pyramid or funnel. The figure is composed of four stacked, truncated triangles, each smaller than the one above it. The triangles are outlined with thick, black lines, creating a layered effect.

K031378

# Fresenius Medical Care

Fresenius Ultraflux Single Use Hemodialyzer "Special" 510(k) Premarket Notification

## Summary of Safety and Effectiveness

#### D. Indications for Use:

The Fresenius Ultraflux dialyzer is designed for single use acute and chronic hemodialysis.

#### E. Substantial Equivalence:

#### 1. Is the product a device?

YES - The Ultraflux dialyzers are devices pursuant to 21 CFR §201 [321] (h).

#### 2. Does the new device have the same intended use?

YES - The intended use for the Ultraflux is identical to that for the Optiflux 200NR and is as follows:

#### Intended Use for Ultraflux

The Fresenius Ultraflux dialyzer is designed for single use acute and chronic hemodialysis.

#### Intended Use for Optiflux 200NR

Optiflux 200NR dialyzers are designed for single use acute and chronic hemodialysis.

#### 3. Does the device have technological characteristics that raise new types of safety or effectiveness questions?

NO - The Ultraflux are equivalent in materials, sterilization and indications for use to the currently manufactured Optiflux 200NR. The technological characteristics of the Ultraflux are equivalent to those of the Optiflux 200NR and Gambro Polyflux 17S and raise no new types of safety or effectiveness questions.

#### 4. Does descriptive or performance information demonstrate equivalence?

YES - Fresenius Medical Care North America believes that the information provided in this submission clearly describes the Ultraflux and demonstrates that it is substantially equivalent to the Optiflux 200NR.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "Page 3/3" in a handwritten style. The word "Page" is written in cursive, followed by the numbers "3/3". This indicates that it is the last page of a document. The text is in black ink on a white background.

Image /page/2/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a stylized triangle shape with horizontal lines inside, positioned above the text "Fresenius Medical Care". The text is in a simple, sans-serif font and is left-aligned.

Ko 3 1378

Fresenius Ultraflux Single Use Hemodialyzer "Special" 510(k) Premarket Notification

## Summary of Safety and Effectiveness

#### F. Safety Summary

The Ultraflux hemodialyzers are substantially equivalent in construction, design, materials, and intended use to the commercially available Fresenius Optiflux hemodialyzers. In addition, testing of Ultraflux hemodialyzers indicates that the dialyzers are safe and effective for their intended use.

In vitro testing was performed to determine the following: priming volume, bloodside and dialysate-side pressure drops, ultrafiltration coefficient, urea clearance, creatinine clearance, Vitamin B12 clearance, phosphate and lysozyme clearance. No clinical testing was performed.

#### G. General Safety and Effectiveness Concerns

The device labeling contains a Package Insert, which includes indications for use, cautions and warnings, as well as the general operating instructions required br proper use of the device. This information promotes safe and effective use of the dialyzer.

Arthur Eilinfeld

ur Eilinsfeld Director of Regulatory Affairs

4/29/03
Date

Date

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.

MAY 3 0 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Nichole Riek Regulatory Affairs Supervisor Fresenius Medical Care North America Two Ledgemont Center 95 Hayden Avenue LEXINGTON MA 02173

Re: K031378

Trade/Device Name: Fresenius Ultraflux Single Use Hemodialyzer Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: 78 KDI Dated: April 29, 2003 Received: May 1, 2003

Dear Ms. Riek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy Crogdon

Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Fresenius Medical Care. The logo consists of an inverted pyramid shape made up of several horizontal lines, with the company name written below it. The text "Fresenius Medical Care" is written in a simple, sans-serif font. The logo is black and white.

### Fresenius Ultraflux Single Use Hemodialyzer "Special" 510(k) Premarket Notification

## Indications for Use Statement

Device Name:

Fresenius Ultraflux Hemodialyzers

#### Indications for Use:

The Fresenius Ultraflux dialyzer is designed for single use acute and chronic hemodialysis.

PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

David W. Sezmore

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number

0000001

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K031378](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K031378)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
