← Product Code [KDI](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI) · K030050

# TORAYSULFONE DIALYZER, BS-L SERIES (K030050)

_Toray Industries (America), Inc. · KDI · Sep 29, 2003 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K030050

## Device Facts

- **Applicant:** Toray Industries (America), Inc.
- **Product Code:** [KDI](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI.md)
- **Decision Date:** Sep 29, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Toraysulfone Dialyzer, BS-L Series is indicated for use in hemodialysis treatment of patients with acute or chronic renal failure.

## Device Story

Toraysulfone Dialyzer, BS-L Series is a high-permeability hemodialyzer; consists of polysulfone hollow fiber membranes housed in polystyrene casing with polystyrene headers. Device requires use with dialysis machines equipped with ultrafiltration controllers. Available in three sizes based on membrane surface area. Intended for single-use; gamma-ray sterilized. Used in clinical settings for hemodialysis treatment of patients with renal failure.

## Clinical Evidence

Bench testing conducted in accordance with FDA guidance; all samples met acceptance criteria. Clinical performance characteristics are supported by published materials included in the notification.

## Technological Characteristics

Polysulfone hollow fiber membrane; polystyrene casing and headers. High permeability hemodialysis system. Gamma-ray sterilized. Single-use. Dimensions vary by surface area. No software or electronic components.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- Toraysulfone™ Dialyzer ([K002512](/device/K002512.md))
- Fresenius Optiflux 200NR Hemodialyzer ([K002277](/device/K002277.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

SEP 2 9 2003

K030050
Page 1 of 1

## 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR & 807.92.

Submitter's Name: Toray Industries (America), Inc. 600 Third Avenue, 5th floor New York, NY 10016-1902 Telephone: (212) 697-8150 Contact person: Mr. Koji Hagimoto, Director, Medical and Pharmaccutical

Date of Summary: January 2, 2003

Device Name: Toraysulfone™ Dialyzer, BS-L Series

Device Classification Name: High permeability hemodialysis system (78 KDI); 21 CFR, Part 876.5860

Legally Marketed Device to which Equivalence is Claimed: The legally marketed predicate device is the Toraysulfone™ Dialyzer (K002512), determined to be substantially equivalent to a legally marketed (preAmendment) device on October 15, 2001: and the Fresenius Optiflux 200NR Hemodialyzer (K002277), determined to be substantially equivalent to a legally marketed (preAmendment) device on August 25, 2000.

Device Description: The Toraysulfone Dialyzer, BS-L Series consists of highly permeable polysulfone hollow fibers housed in a polystyrene casing and secured at each end with polystyrene headers. High permeability dialyzers must be used only in conjunction with dialysis machines equipped with an ultrafiltration controller. The Toraysulfone Dialyzer, BS-L Series is provided in a range of three sizes based on the surface area of the hollow fiber membrane. The device is gammaray sterilized and intended for single use only.

Intended Use: The Toraysulfone Dialyzer, BS-L Series is indicated for use in hemodialysis treatment of patients with acute or chronic renal failure.

Descriptive Summary of Technological Characteristics and Those of Predicate Device: The material of the membrane (hollow fiber) of the Toraysulfone Dialyzer, BS-L Series is identical to that of the legally marketed Toraysulfone dialyzer. The features and capabilities of the predicate and proposed devices are identical: they also have similar flow and clearance data.

## Performance Data:

Bench Testing: Applicable testing was conducted on the Toraysulfone Dialyzer, BS-L Series in accordance with the FDA guidance. All samples met the acceptance criteria. The test results establish that the Toraysulfone Dialyzer, BS-L Scries possesses performance characteristics that make it acceptable for its intended use.

Clinical Evaluations: Some performance characteristics of the Toraysulfone Dialyzer, BS-L Series are described in published materials included in this 510(k) Notification.

Conclusion: The information and data provided in this 510(k) Notification establish that the Toraysulfone Dialyzer, BS-L Series is substantially equivalent to the legally marketed predicate devices.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

SEP 2 9 2003

Toray Industries (America), Inc. c/o Ms. Lisa Jones Regulatory Affairs Consultant Devices for the Future 540 College Street BELLAIRE, TX 77401

Re: K030050

Trade/Device Name: Toraysulfone™ Dialyzer BS-L Series Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: 78 KDI Dated: June 30, 2003 Received: July 1, 2003

Dear Ms. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

KC30050

January 2, 2003

Page 1 of 1

**510(k) Number:** K030050

Device Name: Toraysulfone™ Dialyzer, BS-L Series

Indications for Use: The Toraysulfone Dialyzer, BS-L Series, is indicated for use in hemodialysis treatment of patients with acute or chronic renal failure.

(Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-the-Counter Use

Qamil A. digman

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K030050](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/KDI/K030050)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
