The Ready-to-use Hydrophilic Catheter is indicated for intermittent catheterization of the urethra for those individuals who are unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder -voiding episode. The catheter is inserted into urethra to reach the bladder allowing urine to drain.
Device Story
Single-use, disposable polyurethane catheter; pre-lubricated with polyvinylpyrrolidone (PVP) coating that binds water to create a smooth film. Available in Nelaton, Tapered (Tiemann), and Olive tip configurations; includes insertion aid (sleeve) for no-touch handling. Tiemann models feature guide stripes/ridges for orientation. Used by patients for intermittent bladder drainage; inserted into urethra to reach bladder. Device provides mechanical drainage of urine; benefits include reduced friction during insertion and simplified, hygienic handling via no-touch sleeve.
Clinical Evidence
Bench testing only. Performance validated per ISO 20696 (strength, connector security, flow rate, kink stability, peak tensile force), ASTM D 1894 (coefficient of friction), and internal methods (coating adhesion). Biocompatibility per ISO 10993-1. Sterilization validated per ISO 11137. Shelf life validated per ASTM F1980. All tests met acceptance criteria.
Technological Characteristics
Polyurethane catheter with PVP hydrophilic coating; sterile water hydration. Tip types: Nelaton, Tapered, Olive. Includes insertion sleeve. Sterilized per ISO 11137. Standards: ISO 20696 (performance), ASTM D 1894 (friction), ISO 10993-1 (biocompatibility), ASTM F1980 (shelf life).
Indications for Use
Indicated for intermittent urethral catheterization in patients unable to promote natural urine flow or those with significant residual urine volume following bladder voiding.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
November 27, 2019
Hangzhou Bever Medical Devices Co., Ltd. Ms. Allyson Zhou Management Representative Building 2, No. 1-1, Houmuqiao, Yongle Village, Cangqian St., Yuhang District Hangzhou, Zhejiang Province 311121 CHINA
Re: K192468
> Trade/Device Name: Male, Nelaton-tip Ready-to-Use Hydrophilic Catheter Female, Nelaton-tip Ready-to-Use Hydrophilic Catheter Pediatric, Nelaton-tip Ready-to-Use Hydrophilic Catheter Intermittent Catheter Male, Tapered-Tip Tiemann Ready-to-Use Hydrophilic Catheter Male, Olive-Tip Tiemann Ready-to-Use Hydrophilic Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: GBM, EZD Dated: September 4, 2019 Received: September 9, 2019
Dear Ms. Zhou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecological, and Urological Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K192468
#### Device Name
Ready-to-Use Hydrophilic Catheter
#### Indications for Use (Describe)
The Ready-to-use Hydrophilic Catheter is indicated for intermittent catheterization of the urethra for those individuals who are unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder -voiding episode. The catheter is inserted into urethra to reach the bladder allowing urine to drain.
Type of Use (Select one or both, as applicable)
Z Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved:: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for B. Ever. The logo consists of two vertical lines followed by a stylized letter B. The word "EVER" is written in capital letters to the right of the B. The logo is in black and white.
#### Hangzhou Beve al Devices Co., Ltd.
Add: Building 2, No. 1-1, Houmuqiao, Yongle Village, Canggian Street, Yuhang District 311121, Hangzhou, China Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515
# 510(k) Summary
Date of summary: November 22, 2019
### 1. Submitter (Owner) of 510 (k):
Hangzhou Bever Medical Devices Co., Ltd. Building 2, No. 1-1, Houmuqiao, Yongle Village, Cangqian Street, Yuhang District 311121, Hangzhou, China Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515 Registration Number: 3008729910
### 2. Contact person:
Allyson Zhou Management Representative Tel: +86-571-8861 6630 Fax: +86-0571-8861 6515 Email: allyson@bevermedical.com
#### 3. De vice Name :
Common Name: Catheter, Urethral Trade Name: Ready-to-use Hydrophilic Catheter Classification Name: Urological catheter and accessories (21 CFR 876.5130) Product Code: GBM Regulation Class: II
#### 4. Legally Marketed predicate(s):
Predicate Device: SpeediCath - K023254
Reference Device(s): LoFric® Primo™ - K122078 Self Cath - K100878
# 5. Device Description
The Ready-to-Use Hydrophilic Catheter is a single use, disposable polyurethane catheter. It is coated and placed in the water, packed and sealed in a foil bag and The catheter is prelubricated with a coating containing sterilized. polyvinylpyrrolidone, which binds the water molecules to the surface of the catheter creating a smooth and even lubricating film.
{4}------------------------------------------------
#### K192468, pg. 2 of 4
The Ready-to-Use Hydrophilic Catheter is available for men, women and children, in three different tip configurations of Nelaton (straight and rounded), Tapered (curved and tapered) and Olive (curved and olive), in single or combination with an insertion aid (a sleeve) which provides an easy grip, allowing for insertion without touching. There are two polished drainage eyelets on the catheter in various configurations and types.
The Tiemann catheter has a bended tip along with a guide stripe (in the shaft) and/or aligned with a raised ridge (on the funnel), which will help ensure the catheter tip is still correctly oriented when approaching the bladder.
### 6. Indications for Use
The Ready-to-Use Hydrophilic Catheter is indicated for intermittent catheterization of the urethra for those individuals who are unable to promote a natural urine flow or for those individuals who have a significant volume of residual urine following a natural bladder - voiding episode. The catheter is inserted into urethra to reach the bladder allowing urine to drain.
# 7. Technological Characteristics
| Characteristic | Subject device | Predicate device | Reference device | |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Ready-to-Use<br>Hydrophilic Catheter | SpeediCath<br>K023254 | LoFric® Primo™<br>K122078 | Self Cath<br>K100878 |
| Indication for<br>use | Indicated for<br>intermittent<br>catheterization of the<br>urethra for those<br>individuals who are<br>unable to promote a<br>natural urine flow or<br>for those individuals<br>who have a<br>significant volume of<br>residual urine<br>following a natural<br>bladder - voiding<br>episode. The catheter<br>is inserted into the<br>urethra to reach the<br>bladder allowing<br>urine to drain. | Indicated for use by<br>patients with<br>chronic urine<br>retention and<br>patients with a post<br>void residual<br>volume (PVR) due<br>to neurogenic and<br>non-neurogenic<br>voiding<br>dysfunction. The<br>catheter is inserted<br>into the urethra to<br>reach the bladder<br>allowing urine to<br>drain. | Intended for<br>Intermittent<br>catheterization<br>of the urethra. | Indicated for<br>those individuals<br>unable to<br>promote a natural<br>urine flow or for<br>those individuals<br>who have a<br>significant<br>volume of<br>residual urine<br>following a<br>natural bladder-<br>voiding episode. |
The table below summarizes the technological characteristics of Ready-to-Use Hydrophilic Catheter (subject device) as compared to the legally marketed predicates.
{5}------------------------------------------------
| Size | Female Fr 6, 8, 10,<br>12, 14, 16, 18<br>Male Fr 6, 8, 10, 12,<br>14, 16, 18<br>Tiemann Fr 8, 10,<br>12, 14, 16, 18<br>Pediatric Fr 6, 8, 10 | Female Fr 6, 8, 10,<br>12, 14, 16<br>Male Fr 8, 10, 12,<br>14, 16, 18<br>Tiemann Fr 10, 12,<br>14<br>Pediatric Fr 6, 8, 10<br>Boy Fr 6, 8, 10, 12 | Nelaton 8in<br>Fr 8, 10, 12, 14,<br>16, 18<br>Nelaton 16in<br>Fr 8, 10, 12, 14,<br>16, 18<br>Tiemann 16in<br>Fr 10, 12, 14,<br>16, 18<br>Pediatric 8in<br>Fr 6, 8, 10 | Female Fr 8, 10,<br>12, 14,<br>Male Fr 8, 10, 12,<br>14, 16, 18<br>Tiemann Fr 6, 8,<br>10, 12, 14, 16, 18<br>Pediatric Fr 5, 6,<br>8, 10 |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>composition | Polyurethane<br>catheter coated with<br>polyvinylpyrroli-<br>done, placed in<br>water, in single or<br>combination with an<br>insertion aid. | Polyurethane<br>catheter coated<br>with polyvinylpyr-<br>rolidone, placed in a<br>saline solution<br>containing poly-<br>vinylpyrrolidone. | Plastic catheter<br>coated with<br>polyvinylpyrro-<br>lidone, packed<br>with the sterile<br>water. | Polyvinyl<br>chloride catheter<br>(without coating) |
| Condition of<br>use | Singe Use | Singe Use | Singe Use | Singe Use |
| Coating | PVP(polyvinylpyr-<br>rolidone) Based<br>Coating | PVP(polyvinylpyr-<br>rolidone) Based<br>Coating | PVP(polyvi-<br>nylpyrrolidone)<br>Based Coating | No coating |
| Prelubricated | Yes-by water<br>hydration | Yes-by saline<br>solution hydration | Yes-by water<br>hydration | No |
| Lubricating<br>solution | Sterile water | Sterile saline<br>solution | Sterile water | No |
| Ready to use | Yes | Yes | Requires<br>bursting of<br>water packet<br>prior to use | No |
| No touch<br>design | Yes- contains an<br>insertion aid (sleeve) | No | Yes- by using<br>exterior<br>packaging | No |
| Tip<br>configuration | Nelaton tip, Tapered<br>tip and Olive tip | Nelaton tip and<br>Tapered tip | Nelaton tip and<br>Tapered tip | Nelaton tip,<br>Tapered tip and<br>Olive tip |
| Guide stripe<br>in the shaft | Yes-<br>Tiemann Catheter | No | No | Yes-<br>Tiemann Catheter |
| Drainage<br>Eyelets | Polished and<br>staggered | Polished and<br>staggered | Polished and<br>staggered | Polished and<br>staggered |
| End Design | Funnel | Funnel | Funnel | Funnel |
| Duration of<br>use | For intermittent use | For intermittent use | For intermittent use | For intermittent use |
| Sterile | Yes | Yes | Yes | Yes |
| Packaging | Peel Pack | Peel Pack | Peel Pack | Peel Pack |
{6}------------------------------------------------
#### 8. Summary of Non-Clinical Testing
#### Performance testing for Ready-to-Use Hydrophilic Catheter was conducted according to applicable sections of voluntary standards:
a) Biocompatibility testing according to ISO 10993-1:2009 and FDA Guidance "Use of International Standard ISO 10993-1" (2016) was completed.
b) Bench testing was completed per ISO 20696, ASTM D 1894 and internal test methods.
Performance testing was conducted according to applicable sections of standards in order to document the following properties of the Ready-to-Use hydrophilic catheter:
- Strength was checked by the test method in Annex A of ISO 20696.
- Connector security was checked by the test method in Annex B of ISO 20696.
- Flow rate was checked by the test method in Annex E of ISO 20696.
- Kink stability was checked by the test method in Annex G of ISO 20696.
- Peak tensile force was checked by the test method in Annex H of ISO 20696.
- Coefficient of friction was checked by the test method in ASTM D 1894.
- Coating adhesion was checked by the test method in BEVER internal methods.
All tests passed.
c) Sterilization validation was conducted according to AAMI/ANSI/ISO 11137-1, AAMI/ANSI/ISO 11137-2 and AAMI/ANSI/ISO 11137-3.
d) Accelerated Aged (in compliance with ASTM F1980) Shelf life testing was completed.
All tests met the pre-determined acceptance criteria.
#### 9. Conclusions
The subject device has the same intended use and similar technological characteristics to the currently-marketed predicate devices. The subject device is substantially equivalent to the currently-marketed predicate devices. Laboratory and safety testing conducted on the product has provided scientific evidence that this subject device is as safe and effective as the predicate device for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.