K160795 · Robling Medical · GBM · Jun 16, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K160795
Device Name
United Urologics Intermittent Catheter
Applicant
Robling Medical
Product Code
GBM · Gastroenterology, Urology
Decision Date
Jun 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
United Urologics Intermittent Catheter is intended to be used as a urinary incontinence device designed to drain urine from the bladder.
Device Story
Sterile, single-use, disposable intermittent catheter; flexible tube with atraumatic tip and side drainage holes; optional funnel or flared end for connection to universal adapter/collection bag. Used in healthcare facilities or home settings by patients or caregivers to drain urine from bladder. Facilitates bladder emptying via central lumen. Benefits patient by providing non-invasive method for managing urinary incontinence.
Clinical Evidence
No clinical data. Bench testing only. Device evaluated for biocompatibility (cytotoxicity, sensitization, irritation) per ISO 10993-1:2003 and functional performance per BS EN 1616:1997 and BS EN 1618:1997.
Technological Characteristics
Materials: Polyvinylchloride (PVC) or non-PVC with natural rubber latex. Design: Flexible tube, atraumatic closed tip, side drainage holes, funnel/flared connector. Sterilization: Ethylene oxide per ANSI/AAMI/ISO 11135-1:2007. Standards: BS EN 1616:1997, BS EN 1618:1997.
Indications for Use
Indicated for patients requiring urinary bladder drainage due to incontinence.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name encircling a symbol. The symbol is three stylized human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 16, 2016
Robling Medical, Inc. Diane Peper Director of Quality and Regulatory Affairs 90 Weathers Street Youngsville, NC 27596
Re: K160795 Trade/Device Name: United Urologics Intermittent Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: Class II Product Code: GBM Dated: May 24, 2016 Received: Mav 26, 2016
Dear Diane Peper,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Herbert P. Lerner -S**
Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K160795
Device Name United Urologics Intermittent Catheter
Indications for Use (Describe)
United Urologics Intermittent Catheter is intended to be used as a urinary incontinence device designed to drain urine from the bladder.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary (K160795)
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
| 1.0 SUBMITTER: | Robling Medical, Inc.<br>90 Weathers Street<br>Youngsville NC 27596<br>Telephone: 919-570-9605<br>Fax: 919-570-9611<br>Contact: Diane N. Peper, Director RA/QA<br>Date Prepared: May 24th, 2016 |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2.0 DEVICE NAME: | Trade Name: United Urologics Intermittent Catheter<br>Common Name: Urological Catheter<br>Classification: Class II, 21 CFR Part 876.5130, Product Code<br>GBM |
| 3.0 PREDICATE DEVICE: | Hollister Inc.<br>Apogee Intermittent Catheter (K992137)<br>Classification: Class II, 21 CFR Part 876.5130, Product Code<br>GBM |
## 4.0 DEVICE DESCRIPTION
The United Urologics Intermittent Catheter is a sterile, single use, disposable, urinary incontinence device designed to drain urine from the bladder. The device is manufactured of a flexible tube with a smooth radius, atraumatic tip to facilitate introduction into the urethra. The catheter is manufactured with drainage holes in the side of the tubing closest to the tipped end of the catheter in order to allow drainage of urine from the bladder through the central lumen of the catheter tube. The catheter may be provided with a funnel bonded to the end of the catheter tube opposite the tipped end. During use, the funnel or flared end of the catheter may be attached to a universal adapter to facilitate drainage of the urine into a standard urinary collection bag (the collection bag and universal adapter are not part of this device).
The device may be used by the patient or caregiver in a healthcare facility, hospital, and home or public restroom facilitate drainage of urine from the bladder.
#### INDICATIONS FOR USE 5.0
The United Urologics Intermittent Catheter is a urinary incontinence device designed to drain urine from the bladder.
#### TECHNOLOGICAL CHARACTERISTICS 6.0
{4}------------------------------------------------
Both the subject device and the predicate device are sterile, single use, disposable Intermittent Catheters have been designed and tested in accordance with BS EN 1616:1997 (plus A1:1999)- Sterile urethral catheters for single use and BS EN 1618:1997 - Catheters other than intravascular catheters - test methods for common properties.
Both the subject device and the predicate device use the same materials and processes to manufacture, package and sterilize the product.
Both the subject device and the predicate device have the same technological characteristics:
· The catheters that are not made with natural rubber latex. use Polyvinylchloride tubing for the catheter shaft.
- The Non-PVC catheters contain natural rubber latex
- Both catheters have smooth, atraumatic closed tips
- · Both have drainage holes in the side to allow for urine drainage
- The PVC catheters have polyvinylchloride funnels/connectors bonded on the end of the catheter
• The Non-PVC catheters have a flared end, to facilitate connection to a universal adapter
- · Both subject and predicate devices are provided sterile
#### PERFORMANCE CHARACTERISTICS 7.0
### Functionality
There are no FDA recognized performance standards for urethral catheters under product code GBM that are not Folev catheters.
The following International standards were utilized in designing and testing the United Urologics Intermittent Catheters:
BS EN 1616:1997 (plus A1:1999)- Sterile urethral catheters for single use
BS EN 1618:1997 – Catheters other than intravascular catheters – test methods for common properties.
### Biocompatibility
The United Urologics Intermittent Catheter is characterized as a mucosal membrane contacting for a period of less than 24 hours. Therefore, the fully assembled, sterile device was tested required by Annex A of ISO 10993-1:2003, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process. Testing for cytotoxicity, sensitization and irritation was completed.
### Sterilization
The United Urologics Intermittent Catheter is an Ethylene oxide sterilized device. Sterilization validation performed on the device complies with ANSI/AAMI/ISO 11135-1: 2007 – Sterilization of heath care products -- Ethylene oxide -- Part 1: Requirements for development, validation. and routine control of a sterilization process for medical devices.