← Product Code [FPD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD) · K180236

# VR Medical Feeding Tube and Enteral Feeding Extension Set (EFES) (K180236)

_Vr Medical Technology Co., Ltd. · FPD · Jan 18, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K180236

## Device Facts

- **Applicant:** Vr Medical Technology Co., Ltd.
- **Product Code:** [FPD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD.md)
- **Decision Date:** Jan 18, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The VR Medical Feeding Tube is intended for nasogastric delivery of various types of liquid nutritional media through the gastro intestinal tract of neonatal and small pediatric patients, and is not intended for use bevond 30 days. The Enteral Feeding Extension Set (EFES) is intended for use as an extension set for nasogastric / oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors/small bore connectors which are intended to reduce the misconnection between the following applications: I.V. system and Enteral Feeding System. The EFES is indicated for use in neonatal and pediatric patients in connection with an enteral feeding tube to provide nutrition via nasal or oral gastric placements.

## Device Story

Device consists of sterile, single-use, disposable enteral feeding tubes and extension sets (EFES); tubes available in polyurethane or PVC, 4-12 Fr, 40-110 cm; EFES available in various lengths with slide clamps and Y-ports. Tubes feature radiopaque stripes for X-ray visualization, French size markings, and tethered plugs. EFES incorporates ENFit small-bore connectors (ISO 80369-3) to prevent inadvertent connection to intravenous systems. Used in hospital, sub-acute, and home settings by clinicians. Provides access for liquid nutrition delivery; safety connectors mitigate risk of misconnection errors. Benefits include reduced risk of feeding-to-IV misconnection and confirmed tube placement via radiopacity.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing included connector compatibility (ISO 80369-3, ANSI/AAMI ID54), tensile strength of tubing and bonds (EN1615), liquid leakage, flow rate, radiopacity, and connector incapability tests. Biocompatibility testing performed per ISO 10993-1 for mucosal membrane contact (prolonged duration for tubes, limited duration for EFES), including cytotoxicity, sensitization, irritation, systemic toxicity, and pyrogenicity.

## Technological Characteristics

Materials: Polyurethane (PU) and Polyvinyl Chloride (PVC). Single lumen catheter. Dimensions: 4-12 Fr, 40-110 cm. Connectivity: ENFit small-bore connectors (ISO 80369-3). Sterilization: ETO. Features: Radiopaque stripe, slide clamps, tethered plugs. Biocompatibility: ISO 10993-1 compliant.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Kentec Enteral Feeding Tube ([K100526](/device/K100526.md))
- Covidien Argyle PVC Enteral Feeding Tubes ([K150084](/device/K150084.md))
- Kentec Enteral Feeding Tube Extension Set ([K120272](/device/K120272.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 18, 2019

VR Medical Technology Co., Ltd % Paul E. Dryden Consultant VR Medical c/o ProMedic, LLC 131 Bay Point Dr. NE St. Petersburg, FL 33704

Re: K180236

> Trade/Device Name: VR Medical Feeding Tube and Enteral Feeding Extension Set (EFES) Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: FPD, KNT, PIF Dated: December 7, 2018 Received: December 10, 2018

Dear Paul E. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Daniel G. Walter Jr -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K180236

#### Device Name

VR Medical Feeding Tube and Enteral Feeding Extension Set (EFES)

# Indications for Use (Describe)

The VR Medical Feeding Tube is intended for nasogastric delivery of various types of liquid nutritional media through the gastro intestinal tract of neonatal and small pediatric patients, and is not intended for use bevond 30 days.

The Enteral Feeding Extension Set (EFES) is intended for use as an extension set for nasogastric / oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors/small bore connectors which are intended to reduce the misconnection between the following applications: I.V. system and Enteral Feeding System.

The EFES is indicated for use in neonatal and pediatric patients in connection with an enteral feeding tube to provide nutrition via nasal or oral gastric placements.

Type of Use (Select one or both, as applicable)

XX Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) 

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo for VR Medical. The logo features the letters "VR" in a stylized font, with a green plus sign above the "R". Below the letters, the words "VR Medical" are written in a simple font. A curved green line is below the letters.

| Date Prepared:             | 11-Jan-2019                                                                                                                                         |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                 | VR Medical Technology Company, Ltd.<br>90 Gao Xin Road<br>ZhouZhuang<br>Kunshan Jiangsu, China 215325<br>T - +86 512 5720 0500                      |
| Official Contact:          | Lee Pan, Ph.D., CEO                                                                                                                                 |
| Proprietary or Trade Name: | VR Medical Feeding Tube and<br>Enteral Feeding Extension Set (EFES)                                                                                 |
| Common/Usual Name:         | Tubes, Gastrointestinal (And Accessories)                                                                                                           |
| Classification Name:       | Gastrointestinal Tube and Accessories                                                                                                               |
| Classification Details:    | 21CFR 876.5980<br>FPD, KNT, PIF<br>Class II                                                                                                         |
| Predicate Devices:         | K100526 - Kentec Enteral Feeding Tube<br>K150084 – Covidien Argyle PVC Enteral Feeding Tubes<br>K120272 - Kentec Enteral Feeding Tube Extension Set |

# Device Description:

#### Enteral Feeding Tube

The VR Medical Feeding Tube is intended for nasogastric delivery of various types of liquid nutritional media through the qastro intestinal tract of neonatal, pediatric and is not intended for use beyond 30 days.

They are sterile, individually packaged. The orange color coding provides easy visual recognition of the enteral connection. They are designed with a radiopague stripe for X-ray visualization to confirm proper feeding tube placement. They are marked with French size and markings to assist tube placement or check for migration. A tethered plug for connector closure is incorporated when not in use.

#### Enteral Feeding Extension Sets

The Enteral Feeding Extension Sets ("EFES") are intended to provide access from the feeding tube to a syringe or nutritional source accepting a connector for enteral applications. They are available in multiple configurations: with and without an in-line Y port and lengths.

The EFES are sterile disposable device for single patient use only. The EFES is designed to connect existing feeding tubes (nasogastric, gastric, etc.) to various delivery enteral syringes as well as to help minimize the potential for inadvertent delivery of enteral feedings through the intravenous route. i.e., the device cannot be connected to a luer connector

The EFES consists of flexible tubing with an orange strip for easy quick identification of enteral feeding lines as well as a "For Enteral Feeding Only" tag and a slide clamp to provide the additional safety assurance for connection errors.

#### Indications for Use:

The VR Medical Feeding Tube is intended for nasogastric delivery of various types of liquid nutritional media through the qastro intestinal tract of neonatal and small pediatric patients, and is not intended for use beyond 30 days.

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows a logo for VR Medical. The logo features the letters "V" and "R" in a stylized, interconnected design, with a green medical cross positioned above the "V". A curved, green line underlines the letters, adding a sense of motion or support. Below the logo, the words "VR Medical" are written in a simple, sans-serif font.

The Enteral Feeding Extension Set (EFES) is intended for use as an extension set for nasogastric / oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors/small bore connectors which are intended to reduce the misconnection between the following applications: I.V. system and Enteral Feeding System.

The EFES is indicated for use in neonatal and pediatric patients in connection with an enteral feeding tube to provide nutrition via nasal or oral gastric placements.

# Patient Population:

Neonate and pediatric patients.

#### Environments of use:

Hospital, sub-acute care and home settings.

#### Comparison to Predicates

#### Enteral Feeding Tube Comparison

The VR Medical Feeding Tube and Enteral Feeding Extension Set are substantially equivalent to the predicate devices, based upon the similar indications for use and similar technological characteristics.

Table 1 - Substantial Equivalence Comparison Table - VR Medical Feeding Tube and Enteral Feeding Extension Set to the predicate devices summarizes the comparison of the VR Medical subject devices to the predicate devices.

# Table 1 - Substantial Equivalence Comparison Table – VR Medical Feeding Tube to the Predicate Enteral Feeding Tube Devices

| Characteristics                                        | VR Medical Feeding Tube                                                                                                                                                                 | Kentec Enteral PU Feeding Tube (K100526)                                                                                                                                                | Arygle PVC Feeding Tube (K150084)                                                                                                                                                                                                                                                     |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                                    | The device is intended for nasogastric or orogastric delivery of various types of liquid nutritional media through the gastro intestinal tract of neonatal and small pediatric patients | The device is intended for nasogastric or orogastric delivery of various types of liquid nutritional media through the gastro intestinal tract of neonatal and small pediatric patients | The device is intended for enteral feeding to deliver enteral nutrition, liquid or medication to patient from an enteral feeding syringe or feeding set designed with a connector for enteral application. The device is intended for pediatric patients who require enteral feeding. |
| Regulation                                             | 21CFR 876.5980                                                                                                                                                                          | 21CFR 876.5980                                                                                                                                                                          | 21CFR 876.5980                                                                                                                                                                                                                                                                        |
| Number and Product Code                                | Product Code: FPD                                                                                                                                                                       | Product Code: FPD                                                                                                                                                                       | Product Code: PIF                                                                                                                                                                                                                                                                     |
| Single-Patient Use                                     | Yes                                                                                                                                                                                     | Yes                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                   |
| Disposable                                             | Yes                                                                                                                                                                                     | Yes                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                   |
| Sterile                                                | Yes                                                                                                                                                                                     | Yes                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                   |
| Use Environment                                        | Hospital or environments where placement of a feeding tube is required.                                                                                                                 | Hospital or environments where placement of a feeding tube is required.                                                                                                                 | Hospital or environments where placement of a feeding tube is required.                                                                                                                                                                                                               |
| Patient Population                                     | Neonatal and pediatric patients                                                                                                                                                         | Neonatal and pediatric patients                                                                                                                                                         | Neonatal and pediatric patients                                                                                                                                                                                                                                                       |
| Feeding Tube                                           | Polyurethane (PU)                                                                                                                                                                       | Polyurethane (PU)                                                                                                                                                                       | NA                                                                                                                                                                                                                                                                                    |
|                                                        | Polyvinyl Chloride (PVC)                                                                                                                                                                | No                                                                                                                                                                                      | Polyvinyl Chloride (PVC)                                                                                                                                                                                                                                                              |
| Small Bore (ENFit) Connector, Cap and Y Connector Base | Yes                                                                                                                                                                                     | Yes<br>ENFit not part of submission                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                   |

#### INDICATIONS FOR USE SUBSTANTIAL EQUIVALENCY COMPARISON

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo for VR Medical. The logo features a stylized blue "VR" with a green medical cross above the "V". Below the "VR" is the text "VR Medical" in a smaller font size. The bottom of the logo has a green curved line.

| TECHNICAL CHARACTERISTICS FOR SUBSTANTIAL EQUIVALENCY COMPARISON |                                                                                                                                                                                                |                                                                                                        |                                                             |
|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| A single lumen catheter<br>Sizes                                 | French sizes - (4 – 12 Fr)<br>Lengths - (40 - 110 cm)                                                                                                                                          | French sizes – (4 - 12 Fr)<br>Lengths - (40 - 110 cm)                                                  | French sizes - (3.5- 10 Fr)<br>Lengths - (31 to 107 cm)     |
| Oral Connector                                                   | Yes                                                                                                                                                                                            | Yes                                                                                                    | Yes                                                         |
| ENFit Connector                                                  | Yes                                                                                                                                                                                            | Not in the original file                                                                               | Yes                                                         |
| Cap                                                              | Yes                                                                                                                                                                                            | Yes                                                                                                    | Yes                                                         |
| Provided Sterile                                                 | Yes (ETO Sterilization)                                                                                                                                                                        | Yes (ETO Sterilization)                                                                                | Yes                                                         |
| Connector Design<br>Compliant to ISO<br>80369-3                  | ENFIT Connector                                                                                                                                                                                | ENFIT Connector                                                                                        | ENFIT Connector                                             |
| Performance<br>Testing per ISO<br>80369-3                        | Connector performance                                                                                                                                                                          | Connector performance                                                                                  | Yes                                                         |
| Performance<br>Testing Per<br>EN1615                             |                                                                                                                                                                                                |                                                                                                        | Yes                                                         |
| Biocompatibility ISO<br>10993-1                                  | Surface Contacting. Mucosal<br>membrane, Prolonged duration<br>Cytotoxicity, Sensitization, Irritation.<br>Acute systemic and Sub-acute<br>toxicity, Implantation (14 day) and<br>Pyrogenicity | Surface Contacting. Mucosal<br>membrane, Prolonged duration<br>Cytotoxicity, Sensitization, Irritation | Surface Contacting, Mucosal<br>membrane, Prolonged duration |

# Substantial Equivalence Discussion for Enteral Feeding Tubes

The VR Medical Feeding Tube and the predicate devices have similar indications for use.

Both the subject and predicate devices are prescription, sterile, disposable and single use devices intended for enteral use to provide nutrition/liquids via nasal or oral gastric placement.

Both the subject and predicate devices have similar use environment and patient populations. Both are intended to be used at Hospital or environments where placement of a feeding tube is required for neonatal and pediatric patients.

#### Comparison of the Enteral Feeding Tube to the Predicate Device for Component and Materials

The subject devices have similar materials and components and construction.

#### Comparison of the Enteral Feeding Tube to the Predicate Device for Technological Characteristics

Both the VR Medical Enteral Feeding Tube and Predicate Feeding Tube devices have a single lumen catheter with connector that is designed to be compatible only with enteral access devices or accessories. Both the subject and predicate devices are sterile, individual packaged, and available in various sizes.

The performance of the subject device has been established in performance testing conducted in accordance with ISO 80369-3 and EN1615.

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the logo for VR Medical. The logo features the letters "VR" in a stylized, interconnected design, with the "V" and "R" in a dark blue color. Above the "VR" is a green medical cross symbol, and below the letters is the text "VR Medical" in a smaller font size.

# Enteral Feeding Extension Set Substantial Equivalence Comparison

The Enteral Feeding Extension Sets are substantially equivalent to the predicate device, based upon the same intended use and indications for use, and similar technological characteristics.

Table 2 summarizes the comparison of the VR Medical subject devices to the predicate device in terms of intended use and technological characteristics.

# Table 2 - Substantial Equivalence Comparison Table -Enteral Feeding Extension Sets to the Predicate

| Comparison Table - VR Medical Enteral Feeding Extension Sets to the Predicate |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics                                                               | VR Medical Enteral Feeding Tube Extension Set                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Kentec Enteral Feeding Tube Extension Set (K120272)                                                                                                                                                                                                                                                                            |
| Indications for Use                                                           | The Enteral Feeding Extension Set (EFES) is intended for use as an extension set for nasogastric / oral gastric or gastric tube enteral feeding tubes, incorporating safety connectors/small bore connectors which are intended to reduce the misconnection between the following applications: I.V. system and Enteral Feeding System.<br>The EFES is indicated for use in neonatal and pediatric patients in connection with an enteral feeding tube to provide nutrition via nasal or oralgastric placements. | The Ameritus® Enteral Feeding Extension Set is intended for use as an extension set for nasogastric/oralgastric or gastric tube enteral feeding tubes, incorporating safety connectors which help mitigate the risk of accidental misconnection with an IV system to the enteral system or the enteral system to an IV system. |
| Single-Patient Use                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                            |
| Disposable                                                                    | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                            |
| Sterile                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                            |
| Use Environment                                                               | Hospital or environments where placement of a feeding tube is required.                                                                                                                                                                                                                                                                                                                                                                                                                                          | Hospital or environments where placement of a feeding tube is required.                                                                                                                                                                                                                                                        |
| Patient Population                                                            | Neonatal and pediatric patients                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Neonatal and pediatric patients                                                                                                                                                                                                                                                                                                |
| COMPONENT AND MATERIALS FOR SUBSTANTIAL EQUIVALENCY COMPARISON                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                |
| Characteristics                                                               | VR Medical Enteral Feeding Tube Extension Set                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Kentec Enteral Feeding Tube Extension Set (K120272)                                                                                                                                                                                                                                                                            |
| Tubing                                                                        | Polyvinyl Chloride (PVC)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Polyvinyl Chloride (PVC)                                                                                                                                                                                                                                                                                                       |
| Lengths                                                                       | 22" to 60"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 22" to 60"                                                                                                                                                                                                                                                                                                                     |
| Connectors                                                                    | Male / Female                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Male / Female                                                                                                                                                                                                                                                                                                                  |
| Slides                                                                        | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Yes                                                                                                                                                                                                                                                                                                                            |
| Performance<br>Testing per ISO<br>80369-3                                     | Connector Compatibility<br>Connector Performance                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                            |
| Performance<br>Testing Per<br>EN1615                                          | Tensile Property Test of tubing<br>Tensile Property Test of bonding<br>Liquid Leakage Test<br>Internal Capacity Test<br>Flow Rate Test                                                                                                                                                                                                                                                                                                                                                                           | EN 1615                                                                                                                                                                                                                                                                                                                        |
| Biocompatibility ISO<br>10993-1                                               | Surface Contacting. Mucosal membrane, Prolonged duration<br>Cytotoxicity, Sensitization, Irritation. Acute systemic and Sub-acute toxicity, Implantation (14 day) and                                                                                                                                                                                                                                                                                                                                            | Externally Communicating, Mucosal membrane,<br>Prolonged duration<br>Cytotoxicity ,Sensitization, Irritation                                                                                                                                                                                                                   |

{7}------------------------------------------------

Image /page/7/Picture/2 description: The image shows a logo for VR Medical. The logo features the letters "V" and "R" in a stylized, connected design, with a green medical cross positioned above the "R". A curved green line appears beneath the letters, giving the impression of an arc or a base. Below the logo, the words "VR Medical" are written in a smaller, sans-serif font.

The VR Medical Enteral Feeding Tube Extension Sets and the predicate devices both have similar indications for use.

Both the subject and predicate devices have similar use environment and patient populations. Both are intended to be used at Hospital or environments where placement of a feeding tube is required for neonatal and pediatric patients.

# Comparison of the Enteral Feeding Extension Set to the Predicate Device for Component and Materials

The subject devices have similar materials and components and construction.

# Comparison of the Enteral Feeding Extension Set to the Predicate Device for Technological Characteristics

Both the Enteral Feeding Tube Extension Sets and Predicate are tubing with connectors that allow connection between the solutions and the Feeding Tube. Both the subject and predicate devices are supplied sterile, individual packaged, and available in various lengths and configurations.

The performance of the subject device has been established in performance testing conducted in accordance with ISO 80369-3 and EN 1615.

# Summary of Performance Testing

Non-clinical Bench Testing

Enteral Feeding Tube and Extension Set have been tested using the following standards:

ISO 80369-3. First Edition. Small-bore Connectors for Liquids and Gases in Healthcare Applications - Part 3: Connectors for Enteral Applications

ISO 80369-20, First Edition, Small-bore Connectors for Liquids and Gases in Healthcare Applications - Part 20: Common Test Methods

ISO 80369-1, Small-bore Connectors for Liquids and Gases in Healthcare Applications - Part 1: General requirements

EN1615: 2000, Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing

ANSI/AAMI ID54:1996/(R) 2012, "Enteral feeding set adapters and connectors"

EN1618:1997 Catheters other than intravascular catheters-Test methods for common properties

The following performance testing of the Enteral Feeding Tube (EFT) was conducted.

- a. Connector Compatibility Test in accordance with section 4.2.1 of EN1615: 2000 Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and Testing and ANSI/AAMI ID54: 1996/(R) 2012 Enteral feeding set adapters and connectors.
- b. Tensile Property Test of tubing in accordance with section 4.2.2.1 of EN1615: 2000 Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing.
- ﻥ Tensile Property Test of bonding between connectors and tubing in accordance with section 4.2.2.1 of EN1615: 2000 Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing.

{8}------------------------------------------------

Image /page/8/Picture/2 description: The image shows a logo for VR Medical. The logo features a stylized letter "R" in blue, with a green medical cross positioned above and to the left of the "R". Below the "R", the text "VR Medical" is written in a smaller font size, also in blue. The logo has a modern and professional look, likely representing a healthcare or medical-related company.

- d. Liquid Leakage Test in accordance with section 4.3.2 of EN1615: 2000 Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing.
- e. Tensile Property test of Assembled connection of connectors in accordance with section 4.3.1 of EN1615: 2000 Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing.
- Connector Performance test in accordance with ISO 80369-3: 2016 Small-bore f. connectors for liguids and gases in healthcare applications- Part3: Connectors for enteral applications.
- Printing Firmness Test in accordance with ASTM D3359-09 Standard Test Methods g. for Measuring Adhesion by Tape Test.
- Priming Volume Test h.
- Flow Rate Test using Enfamil solution i.
- Radiopaque Capacity Test j.
- k. Connector Incapability test with specific connectors which defined in ISO 80369-1: 2010 Section 5.1 & Annex B.

The following performance testing of the Enteral Feeding Extension Sets (EFES) was conducted.

- Connector Compatibility Test in accordance with section 4.1.2 of EN1615: 2000 a. Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing and ANSI/AAMI ID54: 1996/(R) 2012 Enteral feeding set adapters and connectors.
- Connector Performance test in accordance with ISO 80369-3: 2016 Small-bore ﻒ connectors for liquids and gases in healthcare applications- Part3: Connectors for enteral applications.
- Tensile Property Test of tubing in accordance with section 4.1.1 of EN1615: 2000 C. Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing.
- ರ. Tensile Property Test of bonding between connectors and tubing in accordance with section 4.1.1 of EN1615: 2000 Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing.
- Liquid Leakage Test in accordance with section 4.1.3 & 4.3.2 of EN1615: 2000 e. Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing.
- Tensile Property test of Assembled connection of connectors in accordance with f. section 4.3.1 of EN1615: 2000 Enteral feeding catheters and enteral giving sets for single use and their connectors-Design and testing.
- Priming Volume Test g.
- Flow Rate Test using Enfamil solution h.
- Connector Incapability test with specific connectors which defined in ISO 80369-1: i. 2010 Section 5.1 & Annex B.

#### Biocompatibility

The biocompatibility evaluation for the Enteral Feeding Tube and Enteral Feeding Extension Set was conducted in accordance with the FDA Draft Guidance on the Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing".

The VR Medical Enteral Feeding Tubes are a

- Surface Contacting, mucosal membrane, prolonged duration (> 24 hours, < 30 days) ●
The Enteral Feeding Extension Set (EFES) are

- Externally Communicating, Tissue, Limited duration ( < 24 hours)

{9}------------------------------------------------

Image /page/9/Picture/2 description: The image contains a logo for VR Medical. The logo features the letters 'V' and 'R' in a stylized, interconnected design, with a green plus sign positioned above the 'V'. Below the logo, the words 'VR Medical' are written in a simple, sans-serif font. The logo has a modern and professional appearance.

ISO 10993-1 would require cytotoxicity, sensitization, and irritation, Acute systemic and Subacute toxicity, Implantation (14 day) and Pyrogenicity.

# Substantial Equivalence Conclusion -

Based on the results of the comparison of intended use and technological characteristics, the VR Medical Feeding Tube and Feeding Extension Set are substantially equivalent to their respective predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K180236](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K180236)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
