← Product Code [FPD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD) · K120815

# INNERSENSE ESOPHAGEAL TEMPERATURE/FEEDING TUBE (K120815)

_Philips Medical Systems · FPD · Mar 6, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K120815

## Device Facts

- **Applicant:** Philips Medical Systems
- **Product Code:** [FPD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD.md)
- **Decision Date:** Mar 6, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The InnerSense Esophageal Temperature/Feeding Tube is a dual function, single-use sterile device which simultaneously provides continuous monitoring of esophageal temperature and delivers oral medications and/or enteral nutrition in neonatal and pediatric patients oro/nasogastrically, as directed by a physician for up to 30 days. The InnerSense device can be used solely for the purpose of monitoring esophageal temperature in situations where invasive monitoring is indicated.

## Device Story

Dual-function, single-use, sterile nasogastric tube; provides continuous esophageal temperature monitoring and enteral feeding/medication delivery. Used in neonatal and pediatric patients; placed by healthcare providers. Temperature probe auto-locates in distal esophagus upon insertion. Device connects to external monitoring equipment to display temperature data; assists clinicians in managing patient thermal status and nutrition. Benefits include reduced need for multiple invasive devices by combining monitoring and feeding functions.

## Clinical Evidence

No clinical data. Bench testing only, including functional performance verification (tensile strength, leakage, fluid extraction) and temperature sensor accuracy/stability testing. Compliance with IEC 60601-1, IEC 60601-1-2, EN 12470-4, and ISO 80369-1 standards confirmed.

## Technological Characteristics

Dual-lumen, single-use, sterile nasogastric tube. Materials evaluated per ISO 10993 (cytotoxicity, sensitization, irritation, systemic toxicity). Temperature sensing via electrical thermometer probe. Dimensions/form factor designed for neonatal/pediatric nasogastric placement. Connectivity: interfaces with external monitoring equipment. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993 series, EN 12470-4, ISO 80369-1.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K120815
Page 1 of 4

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

MAR 0 6 2013

# InnerSense Esophageal Temperature / Feeding Tube

# Submitter's Name and Address

| Submitter's Name:     | Philips Medical Systems                |
|-----------------------|----------------------------------------|
| Division:             | PCCI - Medical Consumables and Sensors |
| Address:              | 3000 Minuteman Road                    |
| City, State, and Zip: | Andover, MA 01810                      |

# Contact Person / Submission Correspondent

| Name:      | Peter Schipelliti                  |
|------------|------------------------------------|
| Title:     | Sr. Quality and Regulatory Manager |
| Telephone: | (978) 659-4744                     |
| Facsimile: | (978) 659-7712                     |
| E-mail:    | peter.schipelliti@philips.com      |

# Date of Summary

Date:

March 5, 2013

# Manufacturers' Information; Establishment Registration Number.

| Establishment name:            | Philips Medical Systems                  |
|--------------------------------|------------------------------------------|
| Address:                       | 3000 Minuteman Road<br>Andover, MA 01810 |
| Establishment Registration No. | 1218950                                  |

### New Device Details

| Proprietary or Trade Name: | InnerSense Esophageal Temperature / Feeding<br>Tube |
|----------------------------|-----------------------------------------------------|
| Common Name:               | Feeding Tube / Temperature Probe                    |
| Device Class:              | Class II                                            |
| Device Procode:            | FPD                                                 |
| Device CFR:                | 21 CFR 876.5980                                     |
| Classification Panel:      | Gastroenterology/Urology                            |

{1}------------------------------------------------

K120815
Page 3 of 4

#### Indication for Use

The InnerSense Esophageal Temperature/Feeding Tube is a dual function, single-use sterile device which simultaneously provides continuous monitoring of esophageal temperature and delivers oral medications and/or enteral nutrition in neonatal and pediatric patients oro/nasogastrically, as directed by a physician for up to 30 days. The InnerSense device can be used solely for the purpose of monitoring esophageal temperature in situations where invasive monitoring is indicated.

#### Product Description

Philips' InnerSense Esophageal Temperature/Feeding Tube is a disposable, single-use, sterile device designed for nasogastric placement in neonatal and pediatric patients. It is used to continuously monitor esophageal temperature, deliver oral medications, and/or provide enteral feeding to a patient for up to 30 days. It should be used only under direct supervision of a licensed physician or healthcare provider, according to the hospital standard of care. Note: the temperature probe is auto-located in the distal esophagus when device is placed in accordance with the Insertion steps provided in the Application section of the IFU.

#### Device Testing / Performance Data

Extensive verification of functional performance has been performed and successfully completed. The proposed device was subjected to numerous verification tests for both the feeding tube attributes and the temperature sensor attributes. Feeding tube related performance tests included tensile strength / pull cycle testing, resistance to liquid leakage under pressure and fluid extraction. Temperature sensor related tests included temperature accuracy and stability. Testing included the standards identified below:

| Standards                                                                                                                                                                               |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1: 1988 + A1:1991 + A2:1995- Medical Electrical Equipment Part 1: General<br>Requirements for Safety                                                                          |
| IEC 60601-1-2: 2001+ A1:2004- Medical Electrical Equipment - Part 1-2: General Requirements for<br>Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests |
| ISO 10993-4: 2002- Biological evaluation of medical devices- Part 4: Selection of Tests for<br>Interactions with Blood                                                                  |
| ISO 10993-5: 2009- Biological evaluation of medical devices - Part 5: Tests for In Vitro Cytotoxicity                                                                                   |
| ISO 19003-6:2007- Biological evaluation of medical devices – Part 6: Tests for Local Effects After<br>Implantation                                                                      |
| ISO 10993-10:2010- Biological evaluation of medical devices - Part 10: Tests for Irritation and<br>Delayed-Type Hypersensitivity                                                        |
| ISO 10993-11: 2006- Biological evaluation of medical devices - Part 11: Tests for Systemic Toxicity                                                                                     |
| ISO 10993-17: 2002- Biological evaluation of medical devices- Part 17: Establishment of allowable<br>Limits for Leachable Substances                                                    |
| ISO 10993-18: 2005- Biological evaluation of medical devices- Part 18: Chemical Characterization of<br>Materials                                                                        |
| EN 12470-4: 2000+ A1:2009- Clinical Thermometers- Part 4: Performance of Electrical<br>Thermometers for Continuous Measurement                                                          |
| ISO 80369-1:2010- Small-bore Connectors for Liquids and Gases in Healthcare Applications - Part 1:<br>General requirements                                                              |

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

# March 6, 2013

Philips Medical Systems % Mr. Peter Schipelliti Senior Quality and Regulatory Manager 3000 Minuteman Road ANDOVER MA 01810

Re: K120815

Trade/Device Name: InnerSense Esophageal Temperature / Feeding Tube Regulation Number: 21 CFR& 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: FPD Dated: February 19, 2013 Received: February 22, 2013

Dear Mr. Schipelliti:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and

listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of

{3}------------------------------------------------

#### Page 2 - Mr. Peter Schipelliti

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050-

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/6 description: The image shows the name "Herbert P. Lerner -S" in a bold, sans-serif font. The letters "P" and "L" are stylized and appear to be part of a logo or emblem. The text is horizontally oriented and appears to be a signature or title.

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# 4.0 Indications for Use Statement

# DHILIDS

510(k) Number: K120815 ---

Device Name: InnerSense Esophageal Temperature / Feeding Tube

Indications for Use:

The InnerSense Esophageal Temperature/Feeding Tube is a dual function, single-use sterile device which simultaneously provides continuous monitoring of esophageal temperature and delivers oral medications and/or enteral nutrition in neonatal and pediatric patients oro/nasogastrically, as directed by a physician for up to 30 days. The InnerSense device can be used solely for the purpose of monitoring esophageal temperature in situations where invasive monitoring is indicated.

. X Prescription Use (21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

HerbertPLerner-S

vision Sign-Off) n of Reproductive. Gastro-Renal, and k) Number

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K120815](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K120815)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
