← Product Code [FPD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD) · K100099

# ORAL/ENTERAL SYRINGES (K100099)

_Philips Children'S Medical Ventures · FPD · Jun 14, 2010 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K100099

## Device Facts

- **Applicant:** Philips Children'S Medical Ventures
- **Product Code:** [FPD](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD.md)
- **Decision Date:** Jun 14, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric, 3rd-Party Reviewed

## Indications for Use

The Oral/Enteral Syringes is intended for the delivery of liquid medication, formula, and breast milk.

## Device Story

Oral/Enteral Syringes facilitate delivery of liquid medication, formula, and breast milk to neonates and infants. Device features specialized safety connectors preventing connection to IV systems (non-Luer compatible). Operated by trained caregivers in hospital or home settings. Syringe barrels, plungers, and end caps made of Polypropylene; pistons made of Silicone. Disposable, single-patient use design ensures hygiene and prevents cross-contamination. Safety connectors mitigate risk of inadvertent enteral-to-IV misconnections.

## Clinical Evidence

Bench testing only. Testing included volume measurement accuracy, enteral feeding extension set interface, size and material inspection, biocompatibility, sterility, reliability, labeling, environmental operating, and storage testing. Results met all acceptance criteria.

## Technological Characteristics

Materials: Polypropylene (barrels, plungers, end caps), Silicone (pistons). Design: Manual syringe with safety connectors incompatible with Luer lock/slip fittings. Single-patient use, disposable. No energy source. No software.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Vygon Corp. Nutrisafe 2 System Transparent Syringes ([K060944](/device/K060944.md))
- Respironics, Inc. Enteral Only Extension Set ([K082654](/device/K082654.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K100099 Page 1 of 2

Oral/Enteral Syringes

Premarket Notification 510(k) Tab 5 – 510(K) Summary

TAB 5

### 510(K) SUMMARY

# JUN 1 4 2010

| Date of Submission                                      | 11/9/09                                                                                                                                                                                                                         |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact / Address<br>of Manufacturing facility | Matthew Young<br>Manager, Quality Assurance and Regulatory Affairs<br>Children's Medical Ventures<br>191 Wyngate Drive<br>Monroeville, PA 15146<br>Phone: 724-387-4012<br>Fax: 412-380-8850<br>Email: matthew.young@philips.com |
| Proprietary Name                                        | Oral/Enteral Syringes                                                                                                                                                                                                           |
| Common/Usual Name                                       | Gastrointestinal tubes and accessories                                                                                                                                                                                          |
| Classification Reference                                | 21 CFR 876.5980                                                                                                                                                                                                                 |
| Classification                                          | Class II                                                                                                                                                                                                                        |
| Appropriate Classification Panel                        | Gastroenterology / Urology                                                                                                                                                                                                      |
| Product Code                                            | FPD                                                                                                                                                                                                                             |
| Predicate Devices                                       | Vygon Corp. Nutrisafe 2 System Transparent Syringes<br>(K060944)<br>Respironics, Inc. Enteral Only Extension Set (K082654)                                                                                                      |

### Intended Use/Indications for Use

The Oral/Enteral Syringes is intended for the delivery of liquid medication, formula, and breast milk.

# Patient Population/Environment of Use

The patient population is neonates and infants. The environments for use are hospital and home environments by trained caregivers only. The Oral/Enteral Only Syringes is disposable and for single patient use only.

{1}------------------------------------------------

K1000 Y9 Page 2 of 2

Premarket Notification 510(k) Tab 5 – 510(K) Summary

I/Enteral Syringes

# Substantial Equivalence

This traditional 510(k) submittal demonstrates that the Oral/Enteral Syringes are substantially equivalent to the Respironics, Inc. Enteral Only Extension Set (K082654) and the Vygon Corp. Nutrisafe 2 System Transparent Syringes (K060944).

Design verification tests were performed on the Oral/Enteral Syringes as a result of the risk analysis and product requirements. The following tests were conducted:

- . Volume Measurement Accuracy Testing
- . Enteral Feeding Extension Set Interface Testing
- . Size and Material Inspection
- . Biocompatibility Testing
- . Sterility Testing
- Reliability Inspection .
- . Labeling Inspection
- Environmental Operating Testing .
- . Storage Testing
- . ISTA Testing

Testing Results demonstrated that the Oral/Enteral Syringes meet the required acceptance criteria. Children's Medical Ventures has determined that the Oral/Enteral Syringes are substantially equivalent to the predicate devices.

#### Device Description

The Oral/Enteral Syringes are intended for the delivery of liquid medication, formula, and breast milk. The syringes incorporate safety connectors that mate with the Respironics Enteral Only Extension Set, but will not mate with Luer lock or Luer slip fittings. The safety connectors prevent inadvertent connection of the enteral system to an IV system or an IV system to the enteral system. The Oral/Enteral Syringe barrels, plungers and end caps are comprised of Polypropylene, and the pistons are comprised of Silicone.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an emblem that features a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus has a staff with two snakes winding around it, and a pair of wings at the top.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G60 Silver Spring, MD 20993-0002

Philips Children's Medical Ventures c/o Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

#### JUN 1 4 2010

Re: K100099

Trade/Device Name: Oral/Enteral Syringes Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: FPD Dated: June 3, 2010 Received: June 4, 2010

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{3}------------------------------------------------

Page 2 -

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address ... http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive. Abdominal, and Radiological Devices office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

K100077

Device Name:

Oral/Enteral Syringes

Intended Use/Indications for Use:

Oral/Enteral Syringes are intended for the delivery of liguid medication, formula, and breast milk.

#### Patient Population/Environment of Use:

The patient population is neonates and infants. The environments for use are hospital and home environments by trained caregivers only. The Oral/Enteral Syringes are disposable and for single patient use only.

Prescription Use × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K100099](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FPD/K100099)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
