← Product Code [FIZ](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIZ) · K020909

# HDM99 (K020909)

_Ibp Instruments GmbH · FIZ · May 22, 2002 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIZ/K020909

## Device Facts

- **Applicant:** Ibp Instruments GmbH
- **Product Code:** [FIZ](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIZ.md)
- **Decision Date:** May 22, 2002
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5820
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The HDM99 may be used by hemodialysis personnel to test the conductivity, temperature, pressure, pH and flow of the dialysate solution used with hemodialysis delivering systems. The HDM99 may also be used to test the conductivity/temperature and pH of acid and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications. The HDM99 may also be used to test the voltage and alternating signals in hemodialysis delivering systems.

## Device Story

HDM99 is a handheld multitest meter for hemodialysis systems; operated by hemodialysis personnel in clinical settings. Inputs include physical measurements of dialysate solution (conductivity, temperature, pressure, pH, flow), concentrates, water, and electrical signals (voltage, alternating signals). Device processes these inputs to provide quantitative readings on a graphic LCD. Features a waterproof membrane keypad for user input and an RS232 interface for PC data collection/visualization. Assists clinicians in verifying proper function of hemodialysis delivery systems; ensures accuracy of dialysate composition and system performance; supports patient safety by confirming equipment parameters are within operational specifications.

## Clinical Evidence

Bench testing only. No clinical data provided. Safety and effectiveness supported by adherence to standards: EN 1441 (Risk Analysis), IEC 60601-1-4 (Safety), IEC 601-1 (Safety), and ISO/DIS 14971 (Risk Management).

## Technological Characteristics

Multitest meter for conductivity, temperature, pressure, pH, and flow. Features graphic LCD, waterproof membrane keypad (22 keys), and RS232 interface for PC connectivity. Powered by internal electronics. Software-controlled measurement processing. Tested to IEC 60601-1 and IEC 60601-1-4 standards.

## Regulatory Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

## Predicate Devices

- Neo2 meter ([K992431](/device/K992431.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K020909

Image /page/0/Picture/1 description: The image shows a logo with the letters "IBP" in a stylized font. To the left of the letters is a black graphic with diagonal lines. The letters "IBP" are outlined in black and have a rounded, cartoonish appearance. The logo appears to be for a company or organization with the initials "IBP".

# MAY 2 2 2002

510(k) Summery - HDM99

Page : 1 of : 2

# 510(k) Summery HDM99

### 1. Submitted by:

IBP Instruments GmbH Sutelstr. 7A 30659 Hannover Germany Phone +49 511 651647 +49 511 652283 Fax eMail pfingstmann@ibpmt.com Internet: www.ibpmt.com

- 2. Contact Person: Dipl. Ing Werner Pfingstmann
- 3. Date Prepared: November 15, 2001

## 4. Product Classification:

Device Name: HDM99 Common Name: Conductivity/Temperature/Pressure/pH multitest meter Classification Name: Meter, Conductivity, non-remote 78FIZ

#### 5. Predicate Device:

Neo2 meter - 510(k) Number: K992431 Automata Instrumentation, Inc. 7830 East Redfield Road #12 Scottsdale, Arizona 85260 USA

### 6. Device Description

The HDM99 was developed in 1990. The first version was called HDM90. 1996 we added the Flow measurement. From this time on the device was called HDM96.

1999 we increased accuracy and software functionality. The device was renamed to HDM99.

The HDM99 has a large, easy readable Graphic Liquid Crystal Display. It comes with a waterproof key membrane. All selections are done with 22 key's including On and Off.

The software is designed to be logical and easy to use for the user either safe and precise in measurement.

The HDM99 has a RS232-Interface for the possibility to collect and visualize data on a PC.

The software is designed to be logical and easy to use for the user either safe and precise in measurement.

#### 7. Indication for use:

The HDM99 may be used by hemodialysis personnel to test the conductivity, temperature, pressure, pH and flow of the dialysate solution used with hemodialysis delivering systems.

The HDM99 may also be used to test the conductivity/temperature and pH of acid and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications.

The HDM99 may also be used to test the voltage and alternating signals in hemodialysis delivering systems.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a logo with the letters "IBP" in a stylized font. The "I" is formed by a black square with a white diagonal line running through it. The letters "B" and "P" are outlined in black and filled with white. The logo appears to be for a company or organization with the initials "IBP".

## 510(k) Summery - HDM99

The indication for use for the HDM99 Dialysis Meter is similar to the predicate device Neo2 meter, K992431

- 8. Test:
The following standards were used in testing the HDM99 Dialysis Meter: - EN 1441 - Risk Analysis - Medical Devices (1997)

- IEC 60601-1-4: 1996 Medical Electrical Equipment, Part 1: General Requirements for Safety;
- IEC 601-1:1988 Medical Electrical Equipment, Part 1: General Requirements for Safety
- ISO/DIS 14971: Feb. 1999 Medical devices - Risk Management

#### 9. Conclusion:

It is concluded that the proposed HDM99 Dialysis Meter is safe and effective for the intended use and is substantially equivalent to the predicate device.

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.'

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally aligned. The words are stacked on a single line and are easily readable.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 2 2002

IBP Instruments GMBH c/o Mr. Mark Job TPR Project Manager TÜV Product Service 1775 Old Highway 8 NEW BRIGHTON MN 55112-1891 Re: K020909

Trade/Device Name: HDM 99 (non-remote conductivity meter) Regulation Number: 21 CFR 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II

Product Code: 78 FIZ Dated: May 6, 2002 Received: May 7, 2002

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy Crogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# HDM99 - Premarket Notification 510(k)

## Indications for Use Statement

#### 510(k) Number

## Device Name: HDM99

#### Indication for Use:

The HDM99 may be used by hemodialysis personnel to test the conductivity, temperature, pressure, pH and flow of the dialysate solution used with hemodialysis delivering systems.

The HDM99 may also be used to test the conductivity/temperature and pH of acid and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications.

The HDM99 may also be used to test the voltage and alternating signals in hemodialysis delivering systems.

#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-the-Counter Use

Varind A. de Lyon

OR

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIZ/K020909](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIZ/K020909)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
