H20NLY COMPLETE WATER PURIFICATION SYSTEM FOR HEMODIALYSIS
Applicant
Serv-A-Pure Co.
Product Code
FIP · Gastroenterology, Urology
Decision Date
May 3, 2000
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5665
Device Class
Class 2
Indications for Use
H2Only Water Purification System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device, Federal law restricts this device to sale by or on the order of a physician.
Device Story
System purifies water for hemodialysis; removes organic/inorganic substances and microbial contaminants. Input: feed water. Process: multi-stage treatment including temperature blending, booster pumps, sediment/cartridge filtration, activated carbon filtration (chlorine/chloramine removal), water softening (calcium/magnesium exchange), and dual-stage reverse osmosis (RO). RO permeate stored in tank with sub-micron vent filter; recirculated through distribution loop. Optional components: UV disinfection, deionization exchange tanks, ultrafilters. Used in clinical dialysis facilities; operated by trained staff. Output: purified water meeting AAMI standards for dialysate preparation. Remote audible/visual alarms at nurses' station monitor system status. Benefits: ensures water quality for patient safety during hemodialysis; twin RO design improves reliability and reduces downtime.
Clinical Evidence
No clinical data provided. Bench testing only; system performance validated by ability to meet AAMI Hemodialysis Water Quality Standards for dissolved inorganics and contaminants.
Technological Characteristics
System components: temperature blending valve, booster pumps, 5.0 micron cartridge filters, backwashing carbon filters, water softener, reverse osmosis membranes, storage tank with sub-micron vent filter, UV disinfection, deionization tanks, ultrafilters. Monitoring: pressure gauges, flow meters, temperature-compensated water quality monitors. Connectivity: remote audible/visual alarm system. Materials: standard industrial water treatment media. Operation: multi-stage filtration and RO. Sterilization: sanitization mode with spray array for storage tank.
Indications for Use
Indicated for use in hemodialysis treatment to purify water used to dilute dialysate concentrate. No specific patient age or gender restrictions provided.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
U.S.Filter Company Water Treatment System for Hemodialysis (K980182)
Osmonics Inc. Water Treatment System for Hemodialysis (K931595)
Better Water Water Treatment System for Hemodialysis (K920186)
Submission Summary (Full Text)
{0}------------------------------------------------
K993520 Page 1 of 5
MAY - 3 2000
# Section 5 Summary of Safety & Effectiveness
H2Only Complete Water Purification System for Hemodialysis
> submitted by Serv-A-Pure Company 1101 Columbus Ave. Bay City, MI 48708 517/892-7745 Fax 517/892-1092
Contact: Richard Herzberger September 16, 1999
Common or Classification Name: Complete Water Purification System for Hemodialysis
Proprietary Name: H2Only Complete Water Purification System for Hemodialysis
#### Claiming substantial equivalence to:
| U.S.Filter Company, (800) 466-7873<br>Water Treatment System for Hemodialysis | K980182 |
|-------------------------------------------------------------------------------|---------|
| Osmonics Inc., (612) 933-2277<br>Water Treatment System for Hemodialysis | K931595 |
| Better Water, (615) 355-6063<br>Water Treatment System for Hemodialysis | K920186 |
Intended Use: H2Only Water Purification System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device, Federal law restricts this device to sale by or on the order of a physician.
#### Device Description
The H2Only Water Purification System for Hemodialysis is a complete water purification system and consists of either one (1) or two (2) reverse osmosis units (RO). Our recommendation to use an RO individually is based upon feed water quality and the demand of water to be used. Pretreatment includes activated carbon filtration media to remove organics, namely chlorine and chloramines. In order to optimize the performance of the RO system, the following items, are utilized:
1) a water break tank or back flow preventer as required;
{1}------------------------------------------------
- booster pumps as required to increase the feed pressure to the treatment equipment; 2)
- a water softener to remove scale-forming minerals; 3)
- cartridge filters to remove suspended solids; 4)
- a temperature blending valve to produce the desired water temperature. 5)
In order to compensate for fluctuations in demand, an air-tight storage tank with a 0.2 micron vent filter, level control and internal spray array is utilized to store RO product water until needed. The first RO system feeds the storage tank, the second RO system is fed from the storage tank and its product water is directly fed to the distribution plumbing. What ever water is not used in the distribution loop is returned to the storage tank, thus a constant recirculation flow is maintained in the distribution plumbing loop.
- when only one RO system is utilized, Ultraviolet disinfection, Deionization Exchange 6) Tanks and Ultrafilters are also incorporated into the system design.
By utilizing this twin RO system concept a few components can be eliminated from the system design. These components are: Deionization backup tanks, ultraviolet sanitizer, ultrafilters and repressurization pumps. The elimination of these additional components improves system reliability and reduces overall system operational costs . Also by using the twin RO design the system may be safely operated on either RO system feeding the distribution loop, this feature provides system backup and eliminates down time.
When feed water quality demands or when requested by the facility physician, optional devices can include the following components:
- cartridge filtration for the removal of silt or feed water sediment; 1)
- 2) automatic backwashing multimedia fitter for the removal of suspended solids;
- 3) ultraviolet disinfection to reduce bacteria either in the influent water supply or on the product supply;
- 4) a booster pump to increase influent water pressure sufficient to provide adequate flow and pressure to the water treatment system.
Water treatment components are recommended based on the demand of water required, a feed water analysis, and the ability to produce water which meets AAMI standards. H2Only Water Systems will always recommend a water purfication system which will exceed Hemodialysis Water Quality Standards as set forth by AAMI.
{2}------------------------------------------------
### Summary of Technological Characteristics
H2Only Water Purification Systems designs, recommends and assembles the components of the system. H2Only Water Systems is not the manufacturer of several of the components and is not the manufacturer of component parts which are assembled by H2Only Water Systems.
Although, H2Only Water Purification Systems recommends the water treatment system and components, the dialysis facility's physician has ultimate authority of the type of equipment installed for water purification. The recommended system consists of the following components. A complete system and the intended purposes of each components are described below. Our design is of equivalent to that of the predicate devices.
Hot and cold filtered water supplies are mixed together using a temperature blending valve to the desired 77 degrees. Sufficient water pressure must be supplied in order to operate the water treatment effectively. If sufficient pressure is not available, a booster pump is installed to supply the required flow and pressure of feed water. If silt or fine sediment are present in the feed water supply, optional filtration is used, 5.0 micron filter cartridges can be used to remove the fine sediment and silt. Next, the water flows through dual exchange carbon filters to remove the chlorine, chloramines and other organics from the water supply. The carbon filters are sized to provide adequate empty bed contact time of 3-5 minutes for chiorine and 6-10 minutes for chloramines. Water is then directed to the water softener where calcium and magnesium ions are exchanged for non-scaling sodium ions. Water is next directed to a multi cartridge 5.0 micron filter housing where sediment is removed.
The "pre-treated" water then enters into the reverse osmosis membrane where dissolved minerals are rejected from the product stream. The product or permeate water is where dissolved minerals are rejected from this water stream. The product water, permeate, then enters Into a storage tank. The storage tank is complete with a sub micron vent filter to prevent airbome bacteria from entering into the storage tank. Stored product water is then used to feed the second RO system and its product water is directly fed to the distribution plumbing. What ever water is not used in the distribution loop is returned to the storage tank, thus a constant recirculation flow is maintained in the distribution plumbing loop.
{3}------------------------------------------------
## Identification of Components
Temperature Blending Valve - is used to blend hot and cold water to achieve a temperature of 77 degrees. Tempered water increases the efficiency of the reverse osmosis unit.
Break tank with air gap or back flow preventer is used to prevent a water back flow condition from occurring. Either one meets local plumbing codes.
Booster Pump - (optional) is used to increase the influent water pressure. The reverse osmosis unit, as well as the other pretreatment equipment, requires sufficient feed water pressure to operate.
Multi-media Filter - (optional) is used to remove suspended solids from the feed water supply. Excessive suspended solids can cause malfunction of pretreatment components. i.e. media plugging, moving parts binding etc.
Water Softener - is used to remove calcium and magnesium bicarbonate from the influent water supply. Calcium and magnesium bicarbonate will cause the reverse osmosis membrane to scale, thus reducing production of treated water.
Backwashing Carbon Filters - are used to remove organics, namely chlorine and chloramines, from the influent water supply. Chlorine will attack and destroy the reverse osmosis membrane, thus reducing the quality of the treated water. Also, AAMI recommends that chlorine and chloramines be removed from the treated water supply. By occasional backwashing of the carbon beds, the carbon media is not allowed to become compacted in the tank and the possibility of channeling is greatly reduced. If a carbon bed channels, the effectiveness of the media is greatly reduced.
Reverse Osmosis Unit #1 - rejects inorganics, dissolved minerals, suspended solids and microbiological contaminants from the product water stream. In order to meet AAMI standards for dissolved inorganics, a reverse osmosis unit is utilized.
Storage Tank - stores reverse osmosis product water and acts as a integral part of the dialysis water loop. The storage tank utilized in this design is a cone bottom on a stand to insure complete drainage when sanitizing. The reverse osmosis unit produces water at a steady rate which is sometimes slower than demand, the storage tank will allow for fluctuations in this demand. The storage tank is equipped with level sensors to turn on the reverse osmosis unit when water is required, and to turn off when the tank is full. A vent filter removes airborne bacteria while allowing the tank to breathe. A spray array is utilized in the sanitization mode so that the interior of the tank is continually rinsed.
{4}------------------------------------------------
Reverse Osmosis Unit #2 - Stored product water is then used to feed the second RO system and its product water is directly fed to the distribution plumbing. What ever water is not used in the distribution loop is returned to the storage tank, thus a constant recirculation flow is maintained in the distribution plumbing loop. The water produced meets or exceeds AAMI standards for dissolved inorganics.
Remote Audible / Visual Alarm - is mounted at the nurses station and activates when an alarm / alert condition occurs in the RO system located in the water treatment room. The RO system may be shut down from this location, but must be restarted from the RO controller.
Deionizer Exchange Tanks - are sometimes incorporated into the system design as a water polish and back up to the single RO system. DI is used to remove inorganic water contaminants and dissolved gases from the water supply. By using water quality monitoring devices, this method assures treated water will meet AAMI recommended standards for inorganic removal. Whenever deionization exchange tanks are used in a water purification system for hemodialysis, sub micron membrane post filtration is used to remove any suspended solids and bacteria. Also, a temperature-compensated quality-indicating device is installed on all systems utilizing deionization exchange tanks. For additional safety, a remote audible/visual alarm is installed at the nurses' station to indicate any problems that may occur in the water treatment room.
Ultraviolet (UV) Disinfection Units are normally installed when Deionization exchange tanks are utilized in the system. Ultraviolet is a safe, clean and effective means to rid the water of bacteria, mold, virus and algae with out the use of heat or chemicals. A UV intensity monitor is also used any time a UV disinfection unit is installed in a water system.
Ultra Filters & Sub micron Post Filters - are used to remove any suspended solids, "free floating" bacteria, and inhibit pyrogens from entering the water used in dialysis. Post filters must be used in conjunction with storage tanks, DI and UV.
Monitoring Devices - are used to assure the proper performance of the water treatment equipment. Pressure gauges, flow meters, temperature-compensated water quality monitors, test ports, temperature gauge, water meter, etc. are installed at specific points throughout the system to measure each component's performance.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The symbol is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. The text is written in all capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 3 2000
Mr. Richard C. Herzberger H2Only Company Serve-A-Pure Company 1101 Columbus Avenue Bay City, MI 48708
Re: K993520 H2Only Complete Water Purification System for Hemodialysis Dated: February 11, 2000 Received: February 14, 2000 Regulatory Class: Il 21 CFR §876.5665/Procode: 78 FIP
Dear Mr. Herzberger:
We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{6}------------------------------------------------
Page 1 of
510(k) Number (if known): K993520
Device Name: H2Only Complete Water Purification System for Hemodialysis
Indications For Use:
H2Only Water Purification System for Hemodialysis is intended for use in hemodialysis treatment. It is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. When used as a medical device, Federal law restricts this device to sale by or on the order of a physician.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K993520
BS
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.