← Product Code [FIP](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP) · K251851

# DIASAFEplusUS (F00013010) (K251851)

_Fresenius Medical Care Renal Therapies Group, LLC · FIP · Oct 9, 2025 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP/K251851

## Device Facts

- **Applicant:** Fresenius Medical Care Renal Therapies Group, LLC
- **Product Code:** [FIP](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP.md)
- **Decision Date:** Oct 9, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5665
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The DIASAFE®plusUS is intended for the preparation of ultrapure dialysis fluid (for the 2008-series machines and 5008X machines) and sterile, non-pyrogenic substitution fluid (for the 5008X machines) from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system). The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively. The DIASAFE®plusUS can only be used with Fresenius Medical Care hemodialysis machines fitted for use with the DIASAFE®plusUS.

## Device Story

DIASAFEplusUS is a non-sterile fluid filter for hemodialysis machines; produces ultrapure dialysis fluid and sterile, non-pyrogenic substitution fluid. Input: pretreated water/dialysis fluid. Operation: fluid forced through polysulfone-polyvinylpyrrolidone capillary fibers; microorganisms/large molecules trapped in fibers; flushed/disinfected during routine machine cycles. 2008-series machines use one filter; 5008X machines use two filters (one for dialysis fluid, second for substitution fluid). Used in acute/chronic hemodialysis/hemodiafiltration environments. Installed via DIAFIX Lock System. Output: filtered fluid for hemodialyzer or infusion. Benefits: reduces microbial contaminants/endotoxins in dialysate; enables on-line HDF treatments.

## Clinical Evidence

No clinical studies were performed. Device is identical to the predicate in design and materials; bench testing and biocompatibility data from the predicate apply.

## Technological Characteristics

Filter housing: polypropylene; capillary fibers: polysulfone-polyvinylpyrrolidone blend; potting mass: polyurethane; O-rings/seals: silicone. Filtration rate: ≥ 300 mL/hr/mmHg. Surface area: 2.1 m². Max pressure: 2 bar. Max life: 90 days. Non-electrical, mechanical filtration.

## Regulatory Identification

A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- DIASAFE®plusUS ([K243505](/device/K243505.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 9, 2025

Fresenius Medical Care Renal Therapies Group, LLC
Timothy Groves
Regulatory Affairs - Senior Lead
920 Winter Street
Waltham, MA 02451

Re: K251851
Trade/Device Name: DIASAFEplusUS (F00013010)
Regulation Number: 21 CFR§ 876.5665
Regulation Name: Water Purification System For Hemodialysis
Regulatory Class: II
Product Code: FIP
Dated: June 16, 2025
Received: September 16, 2025

Dear Timothy Groves:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251851 - Timothy Groves
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

{2}

K251851 - Timothy Groves
Page 3

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Maura Rooney -S

Maura Rooney
Assistant Director
DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices
OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DIASAFEplusUS
Page 8 of 36

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K251851 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  DIASAFEplusUS (F00013010)  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The DIASAFE®plusUS is intended for the preparation of ultrapure dialysis fluid (for the 2008-series
machines and 5008X machines) and sterile, non-pyrogenic substitution fluid (for the 5008X machines) from
pretreated water and is not intended to provide primary purification. Attention must still be given to the
chemical and microbiological quality of water and concentrates and the maintenance of supply systems
(e.g. RO system, central delivery system). The microbiological quality (microbial count [CFU/mL] and
endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL,
respectively. The DIASAFE®plusUS can only be used with Fresenius Medical Care hemodialysis machines
fitted for use with the DIASAFE®plusUS.  |   |   |
|  Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |

{4}

FRESENIUS MEDICAL CARE
510(k) Summary DIASAFE®plusus Traditional 510(k)
K251851 Page 1 of 4

# 1. 510(K) SUMMARY

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.

## 1.1. Submitter's Information

Name: Fresenius Medical Care Renal Therapies Group, LLC
Address: 920 Winter Street
Waltham, MA 02451-1457
Phone: (781) 460-1087
Fax: (781) 699-9635
Contact Person: Timothy Groves, Senior Lead
Preparation Date: 16 June 2025

## 1.2. Device Name

Trade Name: DIASAFEplusUS (F00013010)
Common Name: Fluid filter
Regulation Name: Water purification system for hemodialysis
Regulatory Class: Class II per 21 CFR § 876.5665
Product Code: FIP
Product Code Name: Subsystem, water purification
FDA Review Panel: Gastroenterology/Urology

## 1.3. Legally Marketed Predicate Device

The legally marketed primary predicate device is the DIASAFE®plusUS filter cleared under K243505. The predicate has not been subject to a design-related recall.

## 1.4. Device Description

### 1.4.1. Device Identification

The DIASAFE®plusUS (F00013010) is the subject of this Traditional 510(k).

### 1.4.2. Device Characteristics

The DIASAFE®plusUS is a non-sterile dialysis fluid filter that produces ultrapure dialysis fluid (for the 2008-series machines and 5008X machines) and sterile, non-pyrogenic substitution fluid (for the 5008X machines) for use in chronic and acute hemodialysis and hemodiafiltration. The DIASAFE®plusUS reduces microbial contaminants including endotoxin in the dialysate during treatments. The DIASAFE®plusUS is available in one (1) configuration and is compatible for use with 2008 series and 5008X hemodialysis machines. It is installed on the hemodialysis machines

{5}

#

FRESENIUS MEDICAL CARE

510(k) Summary DIASAFE®plusus Traditional 510(k)

K251851 Page 2 of 4

using the DIAFIX™ Lock System. The DIAFIX™ Lock System is a standard feature on the hemodialysis machines and is installed during machine production.

# 1.4.3. Environment of Use

The DIASAFE®plusus is used in environments where acute and chronic hemodialysis and hemodiafiltration are performed.

# 1.4.4. Brief Written Description of the Device

The 2008 series hemodialysis machines use 1 DIASAFE®plusus. Mixed dialysis fluid is forced through an open filter port across the fibers of the DIASAFE®plusus. A closed bypass valve at the other end forces the dialysis fluid across the fiber membrane. Dialysis fluid passes through the porous membrane and into the filtrate compartment where it is allowed to flow through the uncapped side ports and into the hemodialyzer. Microorganisms and molecules too large to pass through the membrane are trapped in the fibers until they are flushed out, and/or disinfected during a routine cleaning cycle.

The 5008X hemodialysis system uses two (2) DIASAFE®plusus for preparation of the dialysis and substitution fluid which can be used for up to 100 on-line HDF treatments. The dialysis fluid is filtered by the first DIASAFE®plusus. A second DIASAFE®plusus is used to filter the ultrafiltered dialysis fluid a second time to prepare the substitution fluid. The substitution fluid can be infused in the post-dilution mode (i.e., after the blood has passed through the dialyzer), in the pre-dilution mode (i.e., before the blood enters the dialyzer), or combined.

# 1.4.5. Materials of Use

The DIASAFE®plusus is classified as an externally communicating, blood path indirect, long-term contact (&gt; 30 days, Category C) duration, Class II device in accordance with FDA guidance Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (08 September 2023).

The DIASAFE®plusus components are composed of the materials listed in Table 1.

Table 1: DIASAFE®plusus Component Materials

|  Component | Material  |
| --- | --- |
|  Capillary Fibers (fiber bundle) | Polysulfone-polyvinylpyrrolidone blend  |
|  Filter Housing | Polypropylene  |
|  PU-Resin (potting mass) | Polyurethane  |
|  O-ring | Silicone  |
|  Welded Header (cap/flange) | Polypropylene  |
|  Plastic Tabs (tear-off brackets) | Silicone  |
|  Sealing Disc-ring | Silicone  |

# 1.4.6. Key Performance Specifications/Characteristics

Performance specifications for the DIASAFE®plusus are provided in Table 2.

{6}

FRESENIUS MEDICAL CARE
510(k) Summary
DIASAFE®plusUS Traditional 510(k)
K251851
Page 3 of 4

Table 2: DIASAFE®plusUS Features
|  Features | Specifications  |
| --- | --- |
|  Filtration Rate | ≥ 300 mL/hr/mmHg
≥ 3.75 L/min/bar  |
|  Maximum Filtration Pressure | 2 bar  |
|  Surface Area | 2.1 m²  |
|  Connection System | DIAFIX™ Lock System  |
|  Dialysis Fluid Purity | Ultrapure dialysis fluid:
< 0.03 EU/mL
< 0.1 CFU/mL
Supports generation of sterile, non-pyrogenic substitution fluid  |
|  Maximum Use Life | 90 days  |

## 1.5. Intended Use

Purification of dialysis fluid (for the 2008-series machines and 5008X machines) and online hemodiafiltration (HDF) fluid (for the 5008X machines).

## 1.6. Indications for Use

The DIASAFE®plusUS is intended for the preparation of ultrapure dialysis fluid (for the 2008-series machines and 5008X machines) and sterile, non-pyrogenic substitution fluid (for the 5008X machines) from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system).

The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be &lt; 200 CFU/mL and &lt; 2 EU/mL, respectively.

The DIASAFE®plusUS can only be used with Fresenius Medical Care hemodialysis machines fitted for use with the DIASAFE®plusUS.

## 1.7. Comparison of Technological Characteristics with the Predicate Device

The proposed DIASAFE®plusUS has the same principle of operation, materials, design specifications, and performance requirements as the predicate device (K243505). There are no changes to the DIASAFE®plusUS that impact performance. The changes being proposed are limited to labeling.

## 1.8. Performance Data

The proposed DIASAFE®plusUS is identical to the predicate DIASAFE®plusUS with respect to manufacturing, design, and materials. The proposed labeling changes do not impact the performance of the dialyzers.

{7}

FRESENIUS MEDICAL CARE
510(k) Summary DIASAFE®plus™ Traditional 510(k)
K251851 Page 4 of 4

## 1.8.1. Biocompatibility Testing

The proposed DIASAFE®plus™ is identical to the predicate DIASAFE®plus™ with respect to manufacturing, design, and materials. The proposed labeling changes do not impact the biological safety of the dialyzers.

## 1.8.2. Human Factors Validation Testing

The DIASAFE®plus™ was found to be safe and effective for its intended users, uses, and use environments.

## 1.8.3. Electrical Safety and Electromagnetic Compatibility (EMC)

Not applicable. The DIASAFE®plus™ is not an electrical mechanical device.

## 1.8.4. Software Verification and Validation Testing

Not applicable. The DIASAFE®plus™ does not contain software.

## 1.8.5. Animal Studies

No animal studies were performed.

## 1.8.6. Clinical Studies

No clinical studies were performed.

## 1.9. Conclusion

The intended use, indications for use, principle of operation, technological characteristics, design, materials, and performance requirements are substantially equivalent to those of the predicate device. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the DIASAFE®plus™ is safe and effective for its intended use.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP/K251851](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP/K251851)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
