The RenaPure filter is a bacterial and endotoxin retentive filter intended for use in a centralized loop as the final step of a water purification cascade to provide standard dialysis water to machine servicing multiple patients. This filter is not intended to service as the sole means of water purification and therefore must be used in conjunction with other water treatment equipment.
Device Story
RenaPure is a 20-inch cartridge-style filter for hemodialysis water purification systems; installed in a centralized water loop as a final treatment step. Operates in dead-end mode to remove bacteria and endotoxins from water. Input is water from a purification cascade; output is purified water for dialysis machines. Utilizes dual-layered 0.22-micron polyethersulfone (PES) membranes with positive charge modification for endotoxin removal via charge attraction. Designed for use in clinical settings; operated by facility technicians or maintenance staff. Provides benefit by ensuring water quality meets standards for hemodialysis, reducing patient exposure to contaminants. Compatible with hot water and chemical disinfection routines.
Clinical Evidence
Bench testing only. Performance assessment included flow/pressure drop measurements, endotoxin retention, bacterial retention, compatibility with hot water and chemical disinfection, housing fit tests, and compatibility with deionized water. Biocompatibility testing was performed for an externally communicating, indirect blood contact device, including exhaustive extraction/leachable analysis and toxicological risk assessment following simulated disinfection cycles. All tests passed.
Technological Characteristics
20-inch cartridge filter; polypropylene casing; dual-layered 0.22-micron polyethersulfone (PES) pleated membranes; polypropylene screen layers; charge-modified membrane surface; industry-standard 222 header design; silicone O-rings; non-sterile. Operates via size exclusion and charge attraction. Designed for hot water and chemical disinfection compatibility.
Indications for Use
Indicated for use in centralized hemodialysis water loops as a final purification step to remove bacteria and endotoxins from water supplying multiple dialysis machines.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two main elements: the Department of Health & Human Services logo on the left, and the FDA text and emblem on the right. The FDA part of the logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 19, 2023
Evoqua Water Technologies LLC Robert Dudek Product Compliance Manager 558 Clark Road Tewksbury, Massachusetts 01876
## Re: K231410
Trade/Device Name: RenaPure Endotoxin Retentive Filter Regulation Number: 21 CFR 876.5665 Regulation Name: Water purification systems for hemodialysis Regulatory Class: Class II Product Code: FIP Dated: May 1, 2023 Received: May 15, 2023
## Dear Robert Dudek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Gema Gonzalez -S
Gema Gonzalez Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Submission Number (if known)
Device Name
RenaPure Endotoxin Retentive Filter
Indications for Use (Describe)
The RenaPure filter is a bacterial and endotoxin retentive filter intended for use in a centralized loop as the final step of a water purification cascade to provide standard dialysis water to machine servicing multiple patients. This filter is not intended to service as the sole means of water purification and therefore must be used in conjunction with other water treatment equipment.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitter | Evoqua Water Technologies LLC<br>558 Clark Road<br>Tewksbury MA, 01876<br>Owner/Operator :10057016 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Robert Dudek<br>Manager, Product Compliance<br>robert.dudek@evoqua.com<br>(978) 614-7359 |
| Date Prepared | May 10, 2023 |
| Device Name | RenaPure® Endotoxin Retentive Filter |
| Trade Name | RenaPure® Filter |
| Classification Name | Water purification system for hemodialysis, 21 CFR 876.5665 |
| Device Class | Class II |
| Certification Panel | Gastroenterology/Urology |
| Product Code | FIP |
| Predicate Device | Gambro Posiclear Filter - K061782 |
| Device Description | The RenaPure filter is a 20" cartridge style filter in a polypropylene casing that<br>contains dual-layered, 0.22 micron polyethersulfone (PES) pleated<br>membranes which are separated by polypropylene screen layers. The PES<br>membrane is charge modified with a positive charge coating, similar to the<br>predicate device, that aids in removal of endotoxin by charge attraction. Filter<br>configurations are based on an industry standard 222 header design with<br>silicone O-rings at the open end of the filter and flat or fin end caps at the<br>closed end. The filter cartridge is installed in a durable filter housing and<br>operates in a dead-end mode. The filter is designed to remove bacteria and<br>endotoxin from water used in hemodialysis with similar water flow<br>characteristics as the predicate device. The filter provided non-sterile. |
| Indications for Use | The RenaPure filter is a bacterial and endotoxin retentive filter intended for<br>use in a centralized loop as the final step of a water purification cascade to<br>provide standard dialysis water to machine servicing multiple patients. This<br>filter is not intended to service as the sole means of water purification and<br>therefore must be used in conjunction with other water treatment equipment. |
| Technological<br>Characteristics | Retention of bacteria by size exclusion (membrane pore size) and retention of<br>endotoxin by charge attraction (charge modified membrane) is achieved in<br>the same way for both the RenaPure filter and the predicate device. Materials<br>of construction are similar to the predicate device, with the exception that a<br>polyethersulfone base membrane is used, this providing better heat stability<br>for hot water disinfection routines performed during use of the device. |
| Non-clinical<br>Performance<br>Assessment | Performance (bench) testing has been performed which included (a) |
| | Flow/Pressure Drop measurements, (b) Endotoxin Retention, (c) Bacterial |
| | Retention, (d) Compatibility to Hot Water and Chemical Disinfections, (e)<br>Housing Fit tests, and (f) Compatibility with Deionized water. Testing<br>demonstrated that the RenaPure filter met the intended design requirements<br>and will operate as a drop-in replacement to the predicate device. |
| Biocompatibility | A series of biocompatibility tests was performed on the RenaPure filter based |
| | on it being used as an externally communicating, indirect blood contact |
| | device. This testing also included an exhaustive extraction test to identify |
| | extractables/leachables from the RenaPure filter followed by a toxicological |
| | risk assessment. Additional biocompatibility tests were performed after |
| | exposing the RenaPure to simulated hot water disinfection and chemical |
| | disinfection routines. All biocompatibility testing passed and the risk of any |
| | toxic leachates from the RenaPure filter were considered negligible. |
| | Conclusion |
## 510(k) Summary: RenaPure® Endotoxin Retentive Filter
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Evoqua Water Technologies. The logo consists of a stylized water drop symbol on the left, followed by the word "evoqua" in a sans-serif font. Below the word "evoqua" is the phrase "WATER TECHNOLOGIES" in a smaller font size.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.