The Ultrapure & Industrial Services Deionization Tanks are exchangeable/rechargeable mix bed tanks intended to remove ions from the water to a sufficient level to allow safe treatment of Hemodialysis patients. These deionization tanks are not to be used alone, but are intended to be a part of a larger water treatment system employing adequate pre-treatment and post treatment. Upon, exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or new resin altogether. The Ultrapure & Industrial Services Carbon Exchange Tanks are activated carbon tanks intended to remove chlorine and chioramines from the water to allow safe treatment of Hemodialysis patients. These carbon tanks are not to be used alone, but are intended to be part of a larger water treatment system employing adequate pre-treatment and post-treatment. Upon exhaustion, these tanks will be replaced with other tanks containing new activated carbon. All Ultrapure & Industrial Services Deionizer and Carbon Exchange Tank Service for hemodialysis are intended to be used in a hemodialysis facility according to ANSI/AAMI-RD62:2006 standards to supply purified water for use in hemodialysis.
Device Story
Device consists of Fiberglass Reinforced Polypropylene (FRP) tanks containing either mixed bed ion exchange resin or activated carbon (GAC). Used in hemodialysis facilities to purify feed water; DI tanks exchange H+ and OH- ions for cations/anions; Carbon tanks remove chlorine/chloramines via adsorption. Operated by facility staff; tanks are replaced upon exhaustion. System includes worker/polisher arrangement with resistivity monitoring (1 megohm/cm) and audible/visual alarms. Output is purified water meeting AAMI RD62:2006 standards. Benefits include removal of contaminants to ensure safe water for hemodialysis treatment.
Clinical Evidence
Bench testing only. System performance validated by monitoring chlorine, chloramines, and resistivity levels. Six water samples (three raw, three treated) analyzed by three independent laboratories (AmeriWater, Culligan, Oxidor) confirmed treated water met AAMI RD62:2006 standards for chemical components and TOC reduction.
Technological Characteristics
FRP tanks with seamless polyethylene inner shell. DI tanks use mixed bed ion exchange resin; Carbon tanks use 12x40 mesh coal-based GAC (iodine number >= 1000). Connectivity includes resistivity monitor with digital readout and remote alarm. Components meet NSF, ASME, and ANSI/AAMI RD62-2006 standards. Non-mechanical, passive filtration/ion exchange process.
Indications for Use
Indicated for use in hemodialysis facilities to provide purified water for hemodialysis patients by removing ions, chlorine, and chloramines as part of a larger water treatment system.
Regulatory Classification
Identification
A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.
Special Controls
*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
AmeriWater Dialysis and Carbon Exchange Tanks (K991519)
{0}------------------------------------------------
## UC ILTRAPURE & IND
page 1 of 9
#### APR 2 5 2013
1
#### 510 (K) Summary
#### Ref: K123282
#### Mixed Bed Deionization (DI) and Carbon Exchange Tanks
Revision Date: February 25, 2013
- 1. Prepared/Submitted by: Ultrapure & Industrial Services, LLC (UIS) 4429 Mint Way, Dallas, TX 75236
Contact Person: Wasif Asghar General Manager Telephone: (972) 432.9951
- 2. Devise Name: Deionization and Carbon Exchange Tanks for Dialysis
- 3. Devise Classifications: Water Purification Component Class II Medical Device 21 CFR 876.5665, Product Code - FIP
- 4. Predicate Devise: AmeriWater Dialysis and Carbon Exchange Tanks (K991519)
UIS Model Numbers:
| Deionization Tank Model No. | Carbon Tank Model No. |
|-----------------------------|-----------------------|
| UIS-618-MB-D | UIS-618-GAC |
| UIS-818-MB-D | UIS-818-GAC |
| UIS-844-MB-D | UIS-844-GAC |
| UIS-1047-MB-D | UIS-1047-GAC |
| UIS-1447-MB-D | UIS-1447-GAC |
{1}------------------------------------------------
# ULTRAPURE & IN
K123282 page 2 of 9
2
#### 5. Device Description:
Deionizers: Ultrapure & Industrial Services Mixed Bed Deionization Exchange Tanks (DI) are Fiberglass Reinforced Polypropylene (FRP) tanks filled with mixed bed deionization resin. The tank sizes are common for the Dialysis industry with similar inlet and outlet fittings, PVC or Noryl heads and tank distributors. The DI Exchange Tanks are dedicated for ion exchange resin only. Our tanks are designed to deliver Association for the Advancement of Medical Instrumentation (AAMI) standard water through an ion exchange process to remove contaminantes from water being fed to the system. The DI system exchanges hydrogen ion (H+) for cations and hydroxyl ions (OH-) for anions in the feed water. The hydrogen and hydroxyl ions then combine to form pure water (H2O). The tanks are based on the AmeriWater Dialysis Deionizer Exchange Tanks K991519.
Similar to predicate AmeriWater, Ultra-Pure segregated aseptic exchange deionizer service in an on-line worker/polisher arrangement with a temperature compensated 1 megobm/cm resistance light monitor with audible alarm between the worker/polisher and a 0-19 megohm/cm temperatura compensated digital readout resistance meter with adjustable remote audio and visual alarm. Upon exhaustion, these tanks are replaced with other Deionization Tanks containing newly regenerated resin or with new resin altogether.
| UIS Deionizer Exchange Tank Models | | |
|------------------------------------|------------|-------------|
| Model Number | Cubic Feet | Flow in GPM |
| UIS-618-D | 0.27 | 0.5 |
| UIS-818-D | 0.47 | 1.0 |
| UIS-844-D | 1.25 | 2.5 |
| UIS-1047-D | 1.90 | 3.8 |
| UIS-1447-D | 3.60 | 7.2 |
Divert to drain is now required by AAMI. When deionizer water resistivity drops below 1 megohm the deionized water dumps to drain.
{2}------------------------------------------------
# ULTRAPURE & UDDISTRIAL SERVICES,
Carbon: Ultrapure & Industrial Services Carbon Exchange Tanks are Fiberglass Reinforced Polypropylene (FRP) tanks filled with new activated carbon. This carbón filtration, often referred to as granular activated carbón (GAC), will remove chlorine and chloramine that are almost always present in the source water through a chemical process knows as "adsorption". The tank sizes are common for the Dialysis industry with similar inlet and outlet fittings; PVC or Noryl heads and tank distributors. The Carbon Exchange Tanks are dedicated for carbon only. Ultrapure & Industrial Services always recommends that two tanks be installed in series, with the first tank providing the primary purification and the second tank serving as backup. Same recommendation is offered by the predicate. The tanks are based on the AmeriWater Dialysis Carbon Exchange Tanks K991519.
K123282 page 3 of 9
31
Dialysis segregated exchange process carbon service utilizes new virgin coal based granular activated carbon (GAC) of 12X40 mesh size with an iodine number of 1000 or greater.
The tanks are disinfected with chlorine and cleaned before delivery to the dialysis application. Batch codes and dates are used to insure quality and freshness of the tanks. The tanks are re-bedded under our FDA/OS system with batch control of capacity, quality and microbiological control. Ultrapure & Industrial Services process includes a use of medical-grade carbon only followed by a 24-hour soak to fully wet the carbon followed by complete back-washing to remove any contaminants and provide a more natural pH product.
| UIS Carbon Exchange Tank Models | | |
|---------------------------------|------------|-------------|
| Model Number | Cubic Feet | Flow in GPM |
| UIS-618-GAC | 0.27 | 0.5 |
| UIS-818-GAC | 0.47 | 1.0 |
| UIS-844-GAC | 1.25 | 2.5 |
| UIS-1047-GAC | 1.90 | 3.8 |
| UIS-1447-GAC | 3.60 | 7.2 |
{3}------------------------------------------------
ULTRAPURE & IN
#### 6. Indication of Use:
- A. The Ultrapure & Industrial Services (UIS) Deionization Tanks are 0 exchangeable/rechargeable mix bed tanks intended to remove ions from the water to a sufficient level to allow safe treatment of Hemodialysis patients. These deionization tanks are not to be used alone, but are intended to be a part of a larger water treatment system employing adequate pre-treatment and post treatment. Upon, exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or new resin altogether.
| Deionizer Tank Models | |
|-----------------------|------------|
| Model Number | Cubic Feet |
| UIS-618-MB-D | 0.27 |
| UIS-818-MB-D | 0.48 |
| UIS-844-MB-D | 1.25 |
| UIS-1047-MB-D | 1.90 |
| UIS-MB-1447-D | 3.60 |
- B. The Ultrapure & Industrial Services Carbon Exchange Tanks are activated carbon tanks intended to remove chlorine and chloramines from the water to allow safe treatment of Hemodialysis patients. These carbon tanks are not be used alone, but are intended to be part of a larger water treatment system employing adequate pretreatment and post-treatment. Upon exhaustion, these tanks will be replaced with other tanks containing new activated carbon.
| Carbon Tank Models | | |
|--------------------|------------|--|
| Model Number | Cubic Feet | |
| UIS-618-GAC | 0.27 | |
| UIS-818-GAC | 0.47 | |
| UIS-844-GAC | 1.25 | |
| UIS-1047-GAC | 1.90 | |
| UIS-1447-GAC | 3.60 | |
All Ultrapure & Industrial Services Deionizer and Carbon Exchange Tank Service for hemodialysis are intended to be used in a hemodialysis facility according to ANSI/AAMI-RD62:2006 standards to supply purified water for use in hemodialysis.
Ultrapure & Industrial Services, LLC 4429 Mint Way, Dallas, Texas 75236 Telephone: 972.432.9951 Fax: 972.432.9954 www.ultrapure.com
ব
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Ultrapure & Industrial Services, LLC. The logo features a stylized droplet above the word "ULTRAPURE" in a bold, sans-serif font. Below that is the text "ULTRAPURE & INDUSTRIAL SERVICES, LLC" in a similar font, but slightly smaller in size. The overall design is simple and professional.
#### 7. Comparison to Predicate: AmeriWater Exchange Tanks (K991519)
Ultrapure & Industrial Services, LLC exchange tanks for dialysis are substantially equivalent to the currently marketed AmeriWater Dialysis tanks and have not altered the fundamental scientific technologies used in the predicate device. The intended use of the exchange tanks for dialysis is the same as the intended use of the AmeriWater predicate device (K991519).
| (Predicate Device)<br>Ameriwater Dialysis Deionizer<br>Exchange Service (K991519) | Ultrapure & Industrial Services, LLC<br>Dialysis Deionizer<br>Exchange Service |
|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Carbon Filtration. Calgon's Centaur | *Carbon Filtration. Medical Grade Resin<br>Tech (AGC-40 MG) carbon. |
| FRP tanks manufactured by Park<br>International (Now Pentair), brand "Park" | **FRP tanks manufactured by Pentair<br>under brand name Park / Structure |
| MBD-10 Resin manufactured by ResinTech | MBD-10 Resin manufactured by<br>ResinTech |
| Resi-Lite 1 megohm audio/visual alarm<br>positioned between worker and polisher | Resi-Lite 1 megohm audio/visual alarm<br>positioned between worker and polisher |
| Interconnecting Tubing. High Purity PVC. | Interconnecting Tubing. High Purity PVC. |
| Connectors; Made of glass filled Noryl | Connectors; Made of glass filled Noryl |
| Heads and fill plugs: PVC schedule 80<br>machined head with PVC<br>PVC schedule 80<br>fill plug. | ***Heads and fill plugs: PVC schedule 80<br>machined head with PVC schedule 80 fill<br>plug or Glass filled Noryl head and fill<br>plug. |
| PVC schedule stand pipe and distributor<br>basket | PVC schedule stand pipe and distributor<br>basket |
#### Substantial Equivalence
#### MINOR DIFFERENCE
* For carbon media, the predicate uses Calgon's Centaur while Ultrapure & Industrial Services device uses ResinTech Medical Grade Activated Carbon AGC-40 MG. It is much superior premium grade curbon especially prepared for use in medical industry and is fully tested to AAMI standards. Both medias meet AAMI/RD62:2006 standards.
** Park International is now owned by Pentair Water Treatment. Tanks are sold under the nume of Park and Structure. Predicute AmeriWater uses the same manufacturer.
> Ultrapure & Industrial Services, LLC 4429 Mint Way, Dallas, Texas 75236 Telephone: 972.432.9951 Fax: 972.432.9954 www.ultrapure.com
{5}------------------------------------------------
# INTRAPURE & INDU
K123282 page 6 of 9
હ
*** Heads and Fill plugs: Ultrapure and Industrial Services, LLC uses both PVC schedule 80 machined head with PVC schedule 80 fill plug (predicate AmeriWater R991519) and glass filled Noryl head and fill plug (predicates Aqua Systems K092481).
Note: All components used meet or exceed ANSVAAMI RD62-2006 standards.
Ultrapure and Industrial Services exchange tanks for both Deionization (DI) and Carbon (GAC) are fiberglass reinforced polypropylene (FRP) tanks. These tanks are 100% non-corrosive with inner-shell material of Seamless Polyethylene. The design parameters meet and exceed NSF, ASME & ANSI/AAMI RD62-2006 standards. Tanks are dedicated either for deionization or carbon and are not interchangeable. The tanks and sizes are similar to those used by the predicate AmeriWater (K991519) [Also predicates Aqua Sciences K102892]. Sizes are common and widely used in the dialysis industry and tank size utilization is based on water volume required for each individual clinic. Ultrapure & Industrial Services purchase all tanks from the same manufacturer as the predicate AmeriWater.
Deionization tanks from both Ultrapure & Industrial Services and AmeriWater are utilized to remove dissolved solids from the water. Both companies utilize mixed bed resin, consisting of cations and anions resins to remove the charged particles in the water. Both utilize parts and materials that are NSF and/or FDA approved.
Activated carbon tanks/filtration is utilized by both Ultrapure & Industrial Services and AmeriWater to filter out chlorine and chloramines from the water. Ultrapure utilizes the "bighest quality Medical Grade activated carbon that meets NSF and AAMI standards.
Both companies use two (2) carbon tanks in a series configuration. Both Ultrapure & Industrial Services and AmeriWater recommend that chlorine and chloramines are checked before each patient shift. Both companies utilize activated carbon with an iodine number of 1000 or greater. Ultrapure & Industrial Services recommends using dual carbon filtration in series in every dialysis water system installed, including single patient systems.
#### 8. Performance Testing
A system performance test was run by simulating operational conditions. A system consisting of Carbon, and DI resins was set in place and operated. Quality testing of chlorine, chloramines and resistivity levels were monitored to ensure performance.
To qualify performance six (06) water samples were taken for validation purposes. Three samples of the raw feed water were drawn for use as a beginning foundation of the need for treatment. Three samples were taken from the final DI unit to validate performance meeting AAMI RD62 standards.
{6}------------------------------------------------
### ULTRAPURE & IN IL SERVICES. LLC
K123282 page 7 of 9
7
The water samples were sent to three separate labs. To test for AAMI standards, a raw water feed sample and a treated water feed sample was sent to AmeriWater and Culligan Laboratories. To test for TOC, a raw water feed sample and a treated water feed sample was sent to Oxidor Testing Laboratory.
Analysis results from AmeriWater and Culligan show AAMI standard performance. Results from Oxidor Testing Laboratories show a reduction of TOC to non detectable limits. Analysis results and testing facility contacts are submitted along with this submission. {See Lab tests reports enclosed]
#### Test Labs Used
- V AmeriWater Laboratories (Test conformance to AAMI Standards) 1303 Stanley Ave, Dayton, OHIO 45404
- A Culligan Water Testing Laboratory (Test conformance to AAMI Standurds) 9399 W.Higgins Road, Suite#1100 Rosemont, IL 60018
- A OXIDOR Laboratories, LLC (Test for TOC) 1825 E. Plano Pkwy #160 Plano, TX 75074
{7}------------------------------------------------
# ULTRAPURE & INDÜSTRIAL SERVICES,LLC
K123282 page 8 of 9
8
#### LAB ANALYSIS (Testing Lab: AmeriWater)
၊
ﻤﻴﺮﺳ ﺴﻴﺮﺳ ﺍﻟ
| | Feed<br>Water | Treated | AAMI Suggested<br>Maximum Level | Meets or Exceeds |
|----------------|---------------|------------|---------------------------------|------------------|
| Component | mg/L | Water mg/L | mg/L | AAMI Standard |
| Aluminum | 0.001 | <0.001 | 0.010 | Yes |
| Antimony | <0.001 | <0.001 | 0.006 | Yes |
| Arsenic | <0.001 | <0.001 | 0.005 | Yes |
| Barium | 0.013 | <0.01 | 0.100 | Yes |
| Beryllium | <0.0001 | <0.0001 | 0.0004 | Yes |
| Cadmium | <0.0001 | <0.0001 | 0.0010 | Yes |
| Calcium | 30.09 | <0.001 | 2.000 | Yes |
| Chromium | <0.001 | <0.001 | 0.014 | Yes |
| Copper | 0.107 | <0.001 | 0.100 | Yes |
| Fluoride | 0.07 | <0.01 | 0.200 | Yes |
| Iron | 0.027 | <0.001 | N/A | Yes |
| Lead | 0.004 | <0.001 | 0.005 | Yes |
| Magnesium | 13.22 | <0.001 | 4.000 | Yes |
| Mercury | <0.0001 | <0.0001 | 0.0002 | Yes |
| Nitrate (as N) | 0.74 | <0.01 | 2.000 | Yes |
| pH | 7.91 | 7.00 | N/A | Yes |
| Potassium | 3.736 | <0.001 | 8.000 | Yes |
| Resistivity | <0.001 | 0.699 | N/A | Yes |
| Selenium | <0.001 | <0.001 | 0.090 | Yes |
| Silver | <0.001 | <0.001 | 0.005 | Yes |
| Sodium | 26.64 | 0.289 | 70.000 | Yes |
| Sulfate | 79.60 | <0.01 | 100.00 | Yes |
| Thallium | <0.001 | <0.001 | 0.002 | Yes |
| TDS | 180.50 | 0.64 | N/A | Yes |
| Zinc | 0.141 | <0.001 | 0.100 | Yes |
Ultrapure & Industrial Services, LLC 4429 Mint Way, Dallas, Texas 75236 Telephone: 972.432.9951 Fax: 972.432.9954 www.ultrapure.com
{8}------------------------------------------------
# ULTRAPURE & IND
K123282 page 9 of 9
9
#### 9. Summary/Conclusion
The Ultrapure & Industrial Services, LLC. water purification components and the AmeriWater predicate device components are substantially equivalent to one another. All the water purification components and technology in this submission are comparable. Nonclinical tests conducted on the product water from a replicated exchange tank configuration verify product complies with AAMI-RD62 Standard. We can confidently state that the performance testing demonstrates that the exchange tank devices perform to the same standard as that of the predicate.
{9}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/9/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or waves, representing the department's focus on health and human well-being.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 25, 2013
Ultrapure & Industrial Services, LLC % Mr. Greg Franks Quality Manager 4429 Mint Way DALLAS TX 75236
Re: K123282
Trade/Device Name: Deionizer and Carbon Exchange Tanks for Hemodialysis Regulation Number: 21 CFR§ 876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: FIP Dated: February 25, 2013 Received: March 22, 2013
Dear Mr. Franks:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{10}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
## Herbert P. Merner -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
#### Indications For Use
#### 510(K) Number: K123282
Device Name: Deionizer and Carbon Exchange Tanks for Hemodialysis
#### Indications for Use
The Ultrapure & Industrial Services Deionization Tanks are exchangeable/rechargeable mix bed tanks intended to remove ions from the water to a sufficient level to allow safe treatment of Hemodialysis patients. These deionization tanks are not to be used alone, but are intended to be a part of a larger water treatment system employing adequate pre-treatment and post treatment. Upon, exhaustion, these tanks will be replaced with other tanks containing newly regenerated resin, or new resin altogether.
| Deionizer Models | |
|------------------|---------------|
| Model Number | Cubic<br>Feet |
| UIS-618-MB-D | 0.27 |
| UIS-818-MB-D | 0.48 |
| UIS-844-MB-D | 1.25 |
| UIS-1047-MB-D | 1.90 |
| UIS-MB-1447-D | 3.60 |
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
# Herbert | Payerner -S
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K123282 510(k) Number
Page 1 of 2
{12}------------------------------------------------
#### Indications For Use
#### 510(K) Number: K123282
Device Name: Deionizer and Carbon Exchange Tanks for Hemodialysis
#### Indications for Use
The Ultrapure & Industrial Services Carbon Exchange Tanks are activated carbon tanks intended to remove chlorine and chioramines from the water to allow safe treatment of Hemodialysis patients. These carbon tanks are not to be used alone, but are intended to be part of a larger water treatment system employing adequate pre-treatment and post-treatment. Upon exhaustion, these tanks will be replaced with other tanks containing new activated carbon.
| UIS Carbon Exchange<br>Tank Models | |
|------------------------------------|------------|
| Model Number | Cubic Feet |
| UIS-618-GAC | 0.27 |
| UIS-818-GAC | 0.48 |
| UIS-844-GAC | 1.25 |
| UIS-1047-GAC | 1.90 |
| UIS-1447-GAC | 3.60 |
All Ultrapure & Industrial Services Deionizer and Carbon Exchange Tank Service for hemodialysis are intended to be used in a hemodialysis facility according to ANSI/AAMI-RD62:2006 standards to supply purified water for use in hemodialysis.
Prescription Use __ × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Herbert P. Lerner -S
Page 2 of 2
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K123282 510(k) Number __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.