← Product Code [FIP](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP) · K013677

# MODIFICATION TO: MED-RO REVERSE OSMOSIS SYSTEM (K013677)

_Usfilter/Ionpure, Inc. · FIP · Mar 8, 2002 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP/K013677

## Device Facts

- **Applicant:** Usfilter/Ionpure, Inc.
- **Product Code:** [FIP](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP.md)
- **Decision Date:** Mar 8, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5665
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The USFilter/lonpure Med-RO™ Reverse Osmosis System is intended to supply water for use in the preparation of dialysate for hemodialysis patients and for other hemodialysis-related procedures such as cleaning, rinsing of equipment, and reprocessing of dialyzers. The water produced will meet the minimum water quality requirements as specified by ANSI/AAMI American National Standard for Water Treatment Equipment for Hemodialysis Applications RD 62:2001 in conjunction with other components of a water treatment system as necessary with compatible input and output water requirements.

## Device Story

Med-RO™ Reverse Osmosis System purifies water for hemodialysis applications. Device functions as component of larger water treatment system; processes input water to meet ANSI/AAMI RD 62:2001 quality standards. Used in clinical settings to supply water for dialysate preparation, equipment rinsing, and dialyzer reprocessing. Operated by clinical staff. Output water quality ensures safety for hemodialysis procedures; benefits patient by providing water meeting regulatory standards for dialysis treatment.

## Clinical Evidence

No clinical data.

## Technological Characteristics

Reverse osmosis water purification system. Meets ANSI/AAMI RD 62:2001 standards for hemodialysis water treatment equipment.

## Regulatory Identification

A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- USFilter/Ionpure, Inc. M-Series Reverse Osmosis Water Purification System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### MAR 8 2002

# 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted by USFilter/lonpure, Inc. in accordance with the requirements of 21 CFR§807.92.

2

The assigned 510(k) number is:

- Original Submitter's Name: 1 USFilter/Continental Water Systems
- 2 Submitter's Identification USFilter/lonpure, Inc. 725 Wooten Road Colorado Springs, CO 80915 719-570-9600

Date Summary Prepared 15 October 2001

- 3 Official Correspondent/Contact David A. Dentino 10 Technology Drive Lowell, MA 01851 978-614-7359
- 4 Name of the Device Med-RO™ Reverse Osmosis System
- 5 Predicated Device Information USFitter/lonpure, Inc. M-Series Reverse Osmosis Water Purification System

#### ర్ Device Description

The Med-RO™ Reverse Osmosis System is a machine used to purify water.

#### Intended Use 7

The Med-RO™ Reverse Osmosis System series is intended to be used as a component of a water purification system for Hemodialysis.

#### Comparison to Predicated Devices 8

The Med-RO™ Reverse Osmosis System unit contains the same fundamental scientific technology as the predicated device. Device modification, that will not negatively affect performance, safety , or effectiveness are detailed in Exhibit A section 1.

### Discussion of Non-Clinical Test Performed for Determination of Substantial တ Equivalence is as follows:

Based on the testing performed as well as the information provided, there are no changes in the performance, safety or effectiveness of the modified device.

{1}------------------------------------------------

013677 PAGE 2 2 OF

## Discussion of Clinical Test Performed: 10 N/A

#### 11 Conclusions:

As was true for the original USFilter/lonpure, Inc. M-Series Reverse Osmosis Water Purification Systems, the intended use of the Med-RO™ Reverse Osmosis system is
intended to purify water for Hemodialysis when used as a component of a complete water purification system.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### MAR 8 2002

U.S. Filter/Ionpure, Inc. C/o Mr. David A. Dentino Regulatory Compliance Manager USFilter/S&P 10 Technology Drive LOWELL MA 01851

Re: K013677 Trade/Device Name: Med-RO™ Reverse Osmosis System (Modified) Regulation Number: 21 CFR §876.5665 Regulation Name: Water purification system for hemodialysis Regulatory Class: II Product Code: 78 FIP Dated: February 5, 2002 Received: February 6, 2002

Dear Mr. Dentino:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your been wined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 8xx.1xxx                         | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number (if known):___K013677

Device Name: Med-RO™ Reverse Osmosis System

Indications For Use:

The USFilter/lonpure Med-RO™ Reverse Osmosis System is intended to supply water for use in the preparation of dialysate for hemodialysis patients and for other hemodialysis-related procedures such as cleaning, rinsing of equipment, and reprocessing of dialyzers. The water produced will meet the minimum water quality requirements as specified by ANSI/AAMI American National Standard for Water Treatment Equipment for Hemodialysis Applications RD 62:2001[RD-5: 1992 if Treatment Equipment for Temodiarysle Applications RD & LES T.(RE ST 1000 in conjunction with other components of a water treatment system as necessary with compatible input and output water requirements.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C Hiradon

| Prescription Use<br>(Per 21 CFR 801.109) |
|------------------------------------------|
|------------------------------------------|

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP/K013677](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIP/K013677)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
