Browse hierarchy: [Gastroenterology, Urology (GU)](/submissions/GU) → [Subpart F — Therapeutic Devices](/submissions/GU/subpart-f%E2%80%94therapeutic-devices) → [21 CFR 876.5860](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/876.5860) → FII — System, Dialysate Delivery, Sealed

# FII · System, Dialysate Delivery, Sealed

_Gastroenterology, Urology · 21 CFR 876.5860 · Class 2_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FII

## Overview

- **Product Code:** FII
- **Device Name:** System, Dialysate Delivery, Sealed
- **Regulation:** [21 CFR 876.5860](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/876.5860)
- **Device Class:** 2
- **Review Panel:** [Gastroenterology, Urology](/submissions/GU)
- **Life-sustaining:** yes

## Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Classification Rationale

Class II. The special controls for this device are FDA's:

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Recent Cleared Devices (5 of 5)

| Record | Device Name | Applicant | Decision Date | Decision |
| --- | --- | --- | --- | --- |
| [K030470](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FII/K030470.md) | NXSTAGE SYSTEM ONE | Nxstage Medical, Inc. | Jul 7, 2003 | SESE |
| [K881270](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FII/K881270.md) | GAMBRO MULTIPURPOSE SYSTEM MPS-10 | Gambro, Inc. | Aug 24, 1988 | SESE |
| [K872364](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FII/K872364.md) | HEMODIALYSIS (SPS), MODEL 550, CODE 5M1309 | Travenol Laboratories, S.A. | Oct 13, 1987 | SESE |
| [K862716](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FII/K862716.md) | HEMOFILTRATION SYSTEM 10 | Gambro, Inc. | Sep 9, 1986 | SESE |
| [K822386](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FII/K822386.md) | COTRAL I W/4.5 L/HR. ULTRAFILTRAT. CAPA | Hospal Medical Corp. | Oct 6, 1982 | SESE |

## Top Applicants

- Gambro, Inc. — 2 clearances
- Hospal Medical Corp. — 1 clearance
- Nxstage Medical, Inc. — 1 clearance
- Travenol Laboratories, S.A. — 1 clearance

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FII](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FII)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).