SHELLY PROTECTED AV FISTULA NEEDLE

K992653 · Diasol, Inc. · FIE · Oct 27, 1999 · Gastroenterology, Urology

Device Facts

Record IDK992653
Device NameSHELLY PROTECTED AV FISTULA NEEDLE
ApplicantDiasol, Inc.
Product CodeFIE · Gastroenterology, Urology
Decision DateOct 27, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5540
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Fistula needle is part of the Hemodialysis system that provides access to the patient's blood. It is also used in plasmapheresis for the same purpose.

Device Story

The Shelly AV Fistula Needle is a medical device designed to provide vascular access for hemodialysis and plasmapheresis. It functions as a conduit to facilitate the withdrawal and return of blood between the patient and the extracorporeal circuit. The device is intended for use by trained healthcare professionals in clinical settings. By establishing this access, the needle enables the necessary blood processing required for dialysis or plasmapheresis, directly supporting patient treatment for renal failure or other conditions requiring blood component separation.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

The device is a fistula needle used for vascular access. Specific materials, dimensions, and sterilization methods are not detailed in the provided documentation.

Indications for Use

Indicated for patients requiring vascular access for hemodialysis or plasmapheresis procedures.

Regulatory Classification

Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided. (ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated. (B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided. (C) Priming volumes must be established. (D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use. (E) Air leakage testing and liquid leakage testing must be conducted. (F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage. (G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern. (H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established. (iii) Performance data must demonstrate the sterility of the device. (iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing. (v) Labeling of implanted blood access devices for hemodialysis must include the following: (A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated. (B) Labeling must specify the forward and reverse recirculation rates. (C) Labeling must provide the arterial and venous priming volumes. (D) Labeling must specify an expiration date. (E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device. (F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient. (G) Labeling must include a patient implant card. (H) The labeling must contain comprehensive instructions for the following: ( *1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;( *2* ) Proper care and maintenance of the device and device exit site;( *3* ) Removal of the device;( *4* ) Anticoagulation;( *5* ) Management of obstruction and thrombus formation; and( *6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following: (A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port. (B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following: (A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied. (B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties. (C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens. (D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness. (viii) The following must be included for A-V shunt cannulae (with vessel tips): (A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply. (B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation. (C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use. (2) Class II (performance standards) for the nonimplanted blood access device. (3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section. (4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 27 1999 Ms. Monica Abeles President Diasol Inc. 13212 Ravmer Street North Hollywood, CA 91605 Dear Ms. Abeles: Re: K992653 Shelly A V Fistula Needle Dated: August 6, 1999 Received: August 6, 1999 Regulatory Class: Il 21 CFR §876.5540/Procode: 78 FIE We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have delemined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicated in interstate commerce proctories for the enclosive) to legally to legally Medical Device Amendments on to devices that the the hay 28, 1976, the enactmen Medical Device Amendments, or todevies that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic have been recordance with the provisions of the provisions of the Act (Act (Act (Act (Act (Act (Acelore, market the device, subject to the g provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing provisions of the Act findder registering on annual regist devices, good manufacturing practice, labeling, and prohibitions against misbranding a If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations of dass ill (Premarket Approval), it may Federal Requlations, Title 21, Parts 800 to 805 - A eubetstible a Federal Regulational Californials (1895). A substantially guivalent device can be found in the Code of the Current Good Manufacturing Practice as a substantialy equivalent de the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation in and Medical Devices: General requirence as seriorn in the Quality System Regulation (QS) for Drug Administration (EDA) will verify such and that through periodic QS inspections, (the Food and the Program) (e) of t r art ozo, and that, through periode CS inspections, the Food and requlatory action In addino, FDA may publish the GMP regulation may regulatory action. ( ( ( Register. Please note: this response to your prematic concerning your device in the Federal might have under sections 53 through F 3 a f t a f f r a f f might have under sections of your plication submission does not affect any obligation your successor and announces on online and provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device as described in your 50(kg) pemarket notition). The FDA your device and this permits your device to assocceded in a classi your device and thus, permits your device to a legally market. your device and thus, permits your device to proceed to the market. lf you desire specific advice for your labeling regulation (21 CFR Pat 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compilance at (CHC Part of Stationally 809.10 for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation of the contract to continue of Complance at (301) 594-4639. Also, general information on vour responsibilities understand of (21 CFR 807.97). Oth general information on your responsibilities under the Act may be obtains. general information on your responsibilities under the Act may be obtained from the Division of Sma Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {1}------------------------------------------------ Page ____ of ## 510 (k) NUMBER: K992653 + DEVICE NAME: Shelly AV Fistula Needle INDICATION FOR USE: Fistula needle is part of the Hemodialy sis system that INDICATION FOR OBL. Therault provides sceess to the patient's blood. It is also used in plasmapheresis for the same purpose. ! 【 ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUED ON ANOTHER PAGE IF NEEDED.) Concurrence of CDHR, Office of Devices Evaluation (ODE) Prescription Use (Per21 CFR 801.109) OR Over -The-Counter Use (Optional Format 1-2-96) ![image](image) (Division Sign-Off) Division of Reproductive, Abdominal, BNT, and Radiological Devices 510(k) Number.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...