← Product Code [FIE](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIE) · K112734

# APLAN A.V. FISTULA NEEDLE SET (K112734)

_Aplan Well Enterprise Co., Ltd. · FIE · Apr 24, 2012 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIE/K112734

## Device Facts

- **Applicant:** Aplan Well Enterprise Co., Ltd.
- **Product Code:** [FIE](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIE.md)
- **Decision Date:** Apr 24, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5540
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The A.V. Fistula Needle Set is intended for temporary cannulation (non-implantable, less than 30 days) to vascular access for extracorporeal blood treatment. The device is intended for single use only.

## Device Story

A.V. Fistula Needle Set is a single-use, non-implantable device designed for temporary vascular access during extracorporeal blood treatment. The device consists of a needle attached to wings, a flexible tube, and a luer lock connector. It is used by healthcare professionals in clinical settings to facilitate blood flow to and from extracorporeal circuits. By providing a secure, temporary connection to the patient's vascular system, the needle enables necessary blood processing or treatment. The device is discarded after a single use to maintain sterility and patient safety.

## Clinical Evidence

Bench testing only. Testing included biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity per ISO 10993 series), material extract evaluation (ASTM F750-87), sterilization validation (ISO 11135-1), and tensile force testing compared to the predicate device.

## Technological Characteristics

Single-use fistula needle set; includes needle, wings, flexible tubing, and luer lock connector. Biocompatibility testing per ISO 10993-1, 4, 5, 7, 10, 11; material extract testing per ASTM F750-87. Sterilization via ethylene oxide (ISO 11135-1).

## Regulatory Identification

A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device. (2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle). (3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).

## Special Controls

*Classification.* (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
*1* ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
*2* ) Proper care and maintenance of the device and device exit site;(
*3* ) Removal of the device;(
*4* ) Anticoagulation;(
*5* ) Management of obstruction and thrombus formation; and(
*6* ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- JMS North America Corp. A.V. Fistula Needle Set for Single Use ([K990470](/device/K990470.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

APLAN WELL ENTERPRISE CO.,LTD KI/2734 P142

No.1, Aly. 81, Ln. 2, Sec. 1, Zhongxing Rd., Dali Dist., Taichung City 412, Taiwan (R.O.C.) FAX : (04)2496-7052 TEL : (04)2492-0550 E-mail:aplan.easy@msa.hinet.net

## II. 510(K) Summary of Safety and Effectiveness (Per 21 CFR 807.92)

APR 2 4 2012

3

## 2.1. General Information Establishment

- Manufacturer: APLAN Well Enterprise Co., Ltd. 1
- Address: No. 1, Alley 81, Lane 2, Sec. 1, Zhongxing Rd., Dali Dist., Taichung City, 1 Taiwan, R.O.C.
- 트 Owner Number: 3008496213

| Contact Person: | Dr. Jen, Ke-Min E-mail: ceirs.jen@msa.hinet.net<br>886-3-5208829 (Tel); 886-3-5209783 (Fax) |
|-----------------|---------------------------------------------------------------------------------------------|
| Address:        | No.58, Fu Chiun Street, Hsin Chu City, 30067, Taiwan, ROC                                   |

- Date Prepared: August 22, 2011
### Device

- A.V. Fistula Needle Set for Single Use . Proprietary Name:
- Common Name: Fistula Needle
- Classification Name: Needle, Fistula, non-implantable less than 30 days .
- FIE Class II, 876.5540 ● Product Code:

## 2.2. Safety and Effectiveness Information

#### Predicate Device: ●

Claim of Substantial Equivalence (SE) is made to JMS North America Corp. A.V. Fistula Needle Set for Single Use (K990470)

#### ● Device Description:

A.V. Fistula Needle Set is a single-use fistula needle which is inserted for temporary cannulation (non-implantable, less than 30 days) to vascular access for extracorporeal blood treatment. The device consists of a needle that is attached to wings, a flexible tube and a luer lock connector.

#### ● Performance Data:

A.V. Fistula Needle Set for Single Use completed the relevant test including:

- 1) ASTM F750-87: Standard practice for evaluating material extracts by systemic injection in the mouse;
- 2) ISO 10993-1: Biological evaluation of medical devices, Part 11.Tests for System Toxicity;
- 3) ISO 10993-4: Biological evaluation of medical devices, Part 1. Evaluation and testing with a risk management process;
- 4) ISO 10993-5: Biological evaluation of medical devices, Part 5.Tests for in vitro cytotoxicity;

{1}------------------------------------------------

# APLAN WELL ENTERPRISE CO.,LTD | Kil 2734 | Page | Pagly
12 | September Br., Ball Dist Taichung Gily 412, Taiwan (R.O.C.)

 ฝั่ง 412, Sec. 1, Zhongxing Rd., Dali Dist., Taichung City 412, Taiwan (R.O.O.)
No.1, Aly. 81, Ln. 2, Sec. 1, Zhongxing Rd., Dali Dist., Taiwan (R.O.O.).
The met E-mail:aplan.easy@msa.hinet.net FAX : (04)2496-7052 TEL : (04)2492-0550

- 5) ISO 10993-7: Biological evaluation of medical devices, Part 7. Ethylene oxide sterilization residuals;
- Stermically 10993-10: Biological evaluation of medical devices, Part 10.Tests for Irritation & sensitization;
- 11 Inchibition of consisted evaluation of medical devices, Part 11.Tests for System Toxicity;
- 8) ISO 11135-1: Sterilization of health care Ethylene oxide, Part 1. validation and routine control of a sterilization process;
- 9) Tensile Force Test and compared to the predicate device.

#### Intended Use: .

The A.V. Fistula Needle Set is intended for temporary cannulation I non-implantable, less than 30 days) to vascular access for extracorporeal blood treatment. The device is intended for single use only.

#### Substantial Equivalence (SE) .

Substantial Equivalence (SE)
A claim of substantial equivalence is made to JMS A.V. Fistula Needle Set for Single Use (K990470). Both of them have the sane indications for use, same Single Ose (1299 or e).
materials used in the blood contact components, and adopt same fundamental scientific technology as the predicate device. Bench testing was conducted to verify that the APLAN A.V. Fistula Needle Set for Single Use performs as verrly that the 11 the 11 21 21 21 21 21 11 11 they are substantially equivalent.

Ke-Miner

Dr. Jen, Ke-Min official correspondent for i APLAN Well Enterprise Co., Ltd.

4

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APLAN Well Enterprise Co., Ltd. % Dr. Jen. Ke-Min Official Correspondent ROC Chinese-European Industrial Research Society No.58. Fu Chiun Street HSIN CHU CITY 30067 TAIWAN, R.O.C.

APR 2 4 2012

Re: K112734

Trade/Device Name: A.V. Fistula Needle Set for Single Use, Nominal Capacity: 15G, 16G, 17G, Regulation Number: 21 CFR§ 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: FIE Dated: January 31, 2012 Received: February 9, 2012

Dear Dr. Ke-Min:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, ateres, provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adviteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a assist and regulations administered by other Federal agencies. You must or uny Federation the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher, Ph.D.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive. Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510 (K) Number ( If Known ):_ K | 12 7 34

Device Name: Nominal Capacity: 15G, 16G, 17G,

## Indications for Use:

The A.V. Fistula Needle Set is intended for temporary cannulation (non-implantable, less than 30 days) to vascular access for extracorporeal blood treatment. The device is intended for single use only.

## Prescription Use (21 CFR Part 801 Subpart D)

AND/OR

Over the Counter Use (21 CFR Part 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Heibert Leeman

ductive, Gastro-Renal, and

Page 1 of 1.

7 ·

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIE/K112734](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FIE/K112734)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
