← Product Code [FGE](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE) · K200103

# BIOTEQ Drainage Catheter Set (One Step Type) (K200103)

_Bioteque Corporation · FGE · Oct 9, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE/K200103

## Device Facts

- **Applicant:** Bioteque Corporation
- **Product Code:** [FGE](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE.md)
- **Decision Date:** Oct 9, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5010
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The BIOTEQ Drainage Catheter Set (One Step Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and others fluids.

## Device Story

Percutaneous drainage catheter set used for abscess and fluid collection drainage; includes catheter, trocar needle, stylet, stiffening cannula, and accessories. Catheter made of radiopaque, biocompatible TPU with hydrophilic coating; distal end features pigtail, J-shape, or closed-loop configuration with drainage holes. Operator selects catheter size (5F-16F) based on fluid viscosity and placement site; utilizes Direct Access or Seldinger technique for insertion. Used in clinical settings by healthcare professionals. Output is fluid drainage; aids in managing fluid collections and patient recovery.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including sterilization validation, shelf-life testing, biocompatibility, and performance testing.

## Technological Characteristics

Materials: TPU (thermoplastic polyurethane). Features: Radiopaque, hydrophilic coating, depth printing markers. Distal configurations: Pigtail, closed-pigtail, mini-pigtail, J-shape. Sizes: 5F-16F. Sterilization: Ethylene Oxide. Connectivity: None (mechanical device).

## Regulatory Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- General Purpose Drainage Catheter ([K053245](/device/K053245.md))
- BIOTEQ® Pigtail Drainage Catheter Set (One Step Type) ([K033862](/device/K033862.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".

October 9, 2020

Bioteque Corporation Stella Hsu RA Specialist 5F-6, No. 23, Sec. 1, Chang'an E. Rd., Zhongshan Dist. Taipei City, 10441 TAIWAN

Re: K200103

> Trade/Device Name: BIOTEQ Drainage Catheter Set (One Step Type) Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE, GBO Dated: September 4, 2020 Received: September 8, 2020

Dear Stella Hsu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Glenn B. Bell, Ph.D. Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K200103

Device Name

BIOTEQ Drainage Catheter Set (One Step Type) Model Name: BT-PD1-series

Indications for Use (Describe)

The BIOTEQ Drainage Catheter Set (One Step Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and others fluids.

| Type of Use (Select one or both, as applicable)                           |                                                                          |
|---------------------------------------------------------------------------|--------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Page 1 of 5

K200103

BIOTEQUE CORPORATION BIOTEQ Drainage Catheter Set (One Step Type) Section 5 - 510 (k) Summary

# 510(k) SUMMARY

| 5.1 | Type of Submission:                                      | Traditional                                                                          |
|-----|----------------------------------------------------------|--------------------------------------------------------------------------------------|
| 5.2 | Date of Summary:                                         | January 16, 2020                                                                     |
| 5.3 | Submitter:                                               | BIOTEQUE CORPORATION                                                                 |
|     | Address:                                                 | 5F-6, No. 23, Sec. 1, Chang'an E. Rd., Zhongshan Dist.<br>Taipei City 10441, Taiwan  |
|     | Phone:                                                   | +886-2-2571-0269                                                                     |
|     | Fax:                                                     | +886-2-2536-1967                                                                     |
|     | Contact:                                                 | William Lee (General Manager)                                                        |
| 5.4 | Identification of the Device:<br>Proprietary/Trade name: | BIOTEQ Drainage Catheter Set (One Step Type)<br>Model Name: BT-PD1-series            |
|     | Classification Product Code:                             | FGE                                                                                  |
|     | Subsequent Product Code:                                 | GBO                                                                                  |
|     | Regulation Number:                                       | 876.5010; 878.4200                                                                   |
|     | Regulation Description:                                  | Biliary catheter and accessories;<br>Introduction/drainage catheter and accessories. |
|     | Review Panel:                                            | Gastroenterology/Urology;<br>General & Plastic Surgery                               |
|     | Device Class:                                            | II                                                                                   |
|     | Basis for the Submission:                                | New Device                                                                           |

#### 5.5 Identification of the Predicate Device I:

| Predicate Device Name:       | General Purpose Drainage Catheter |
|------------------------------|-----------------------------------|
| Manufacturer:                | UreSil, LLC                       |
| Classification Product Code: | FGE                               |
| Subsequent Product Code:     | GBO, GBX, LJE                     |
| Regulation number:           | 876.5010                          |
| Device Class:                | II                                |

{4}------------------------------------------------

Page 2 of 5

K200103

BIOTEQUE CORPORATION BIOTEQ Drainage Catheter Set (One Step Type)

Traditional 510(k) Section 5 - 510 (k) Summary

K053245 510(k) Number:

5.6 Identification of the Predicate Device II: BIOTEQ® Pigtail Drainage Catheter Set (One Predicate Device Name: Step Type) BIOTEQUE CORPORATION Manufacturer: Classification Product Code: LHI Subsequent Product Code: NEP Regulation number: 880.5440 Device Class: II 510(k) Number: K033862

#### 5.7 Intended Use

The BIOTEQ Drainage Catheter Set (One Step Type) is designed for percutaneous drainage of abscess fluid, cyst, gall bladders, nephrostomy, urinary, and others fluids.

#### 5.8 Device Description

The BIOTEQ Drainage Catheter Set (One Step Type), BT-PD1-series percutaneous drainage catheter with hydrophilic coating, is a percutaneous drainage catheter used for drainage of abscess and fluid collections. The catheter is made from a soft, biocompatible plastic, a material that is radiopaque for X-rays. The distal end of catheter contains a "J", a pigtail or close loop and drainage holes.

The operator can use different drainage sets according to the type of accumulated fluid and place of accumulation. These sets are classified according to the catheter size (5F~16F pertaining to various fluid viscosity) and according to the position of the accumulation. The operator can choose either Direct Access Technique or Seldinger Technique to provide access.

### 5.9 Non-clinical Testing

A series of safety and performance tests were conducted on the subject device,

{5}------------------------------------------------

K200103

BIOTEQUE CORPORATION BIOTEQ Drainage Catheter Set (One Step Type)

Traditional 510(k) Section 5 - 510 (k) Summary

BIOTEO Drainage Catheter Set (One Step Type).

- Sterilization Validation
- o Shelf-life
- Biocompatibility
- Performance

All the test results demonstrate BIOTEQ Drainage Catheter Set (One Step Type) meets the requirements of its pre-defined acceptance criteria and intended use, and is substantially equivalent to the predicate devices.

# 5.10 Clinical Testing

No clinical test data was used to support the decision of substantial equivalence.

## 5.11 Substantial Equivalence Determination

Equivalence, same and difference between the subject and predicate devices are cited as below.

| Item                            | Subject device                                                                                                                                                                                                          | Predicate device I                                                                                                                                                    | Predicate device II                                                                                                                                                                                | Substantial<br>equivalence<br>determination                                                                   |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Proprietary Name                | BIOTEQ Drainage<br>Catheter Set (One Step<br>Type)                                                                                                                                                                      | General Purpose<br>Drainage Catheter                                                                                                                                  | BIOTEQ® Pigtail<br>Drainage Catheter Set<br>(One Step Type)                                                                                                                                        |                                                                                                               |
| 510(k) No.                      | (to be assigned)                                                                                                                                                                                                        | K053245                                                                                                                                                               | K033862                                                                                                                                                                                            |                                                                                                               |
| Intended Use                    | The BIOTEQ Drainage<br>Catheter Set (One Step<br>Type) is designed for<br>percutaneous drainage of<br>abscess fluid, cyst, gall<br>bladders, nephrostomy,<br>urinary, and others fluids.                                | For percutaneous<br>drainage of abscesses,<br>nephrostomy and other<br>fluids.                                                                                        | BIOTEQ® Pigtail<br>Drainage Catheter Set<br>(One Step Type) is<br>intended to be used for<br>percutaneous drainage<br>of abscesses, cysts, gall<br>bladders,<br>nephrostomies and<br>other fluids. | Equivalent<br>All the devices are<br>used for percutaneous<br>drainage in<br>gastroenterology and<br>urology. |
| Type of use                     | Prescription Use                                                                                                                                                                                                        | Prescription Use                                                                                                                                                      | Prescription Use                                                                                                                                                                                   | Same                                                                                                          |
| Catheter Shaft<br>Material      | TPU                                                                                                                                                                                                                     | TPU                                                                                                                                                                   | TPU                                                                                                                                                                                                | Same                                                                                                          |
| Distal<br>configuration         | String Locking Pigtail,<br>Non-String Locking<br>Pigtail                                                                                                                                                                | String Locking Pigtail,<br>Non-String Locking<br>Pigtail                                                                                                              | String Locking Pigtail,<br>Non-String Locking<br>Pigtail                                                                                                                                           | Same                                                                                                          |
| Distal shape                    | Pigtail, Closed-Pigtail,<br>Mini-Pigtail, Mini-closed<br>Pigtail, J shape                                                                                                                                               | Pigtail, Closed-Pigtail,<br>Mini-Pigtail, J shape                                                                                                                     | Pigtail                                                                                                                                                                                            | Equivalent<br>All the devices have<br>pigtail, closed, mini,<br>and J shape.                                  |
| Distal Hydrophilic<br>Coating   | Yes                                                                                                                                                                                                                     | Yes                                                                                                                                                                   | Yes                                                                                                                                                                                                | Same                                                                                                          |
| Shaft Depth<br>Printing Markers | Yes                                                                                                                                                                                                                     | Yes                                                                                                                                                                   | Yes                                                                                                                                                                                                | Same                                                                                                          |
| Proximal Hub<br>Assembly        | Hub (for String Lock<br>Pigtail),<br>F.L.L. Adapter                                                                                                                                                                     | Hub (for String Lock<br>Pigtail),<br>F.L.L. Adapter                                                                                                                   | Screw Cap,<br>F.L.L Adapter                                                                                                                                                                        | Same as predicate I                                                                                           |
| Size                            | 5 Fr (Non-String Lock),<br>6, 7, 8, 10, 12, 14, 16 Fr                                                                                                                                                                   | 6, 8, 10, 12, 14, 16 Fr                                                                                                                                               | 7, 8, 9, 10, 12, 14 Fr                                                                                                                                                                             | Different but do not<br>raise new issues of<br>SE.                                                            |
| Useable Length                  | 20, 25, 30, 35, 40, 45, 50<br>cm                                                                                                                                                                                        | 12, 18, 19, 20, 25, 28,<br>40, 41, 50 cm                                                                                                                              | 20, 30, 40 cm                                                                                                                                                                                      | Different but do not<br>raise new issues of<br>SE.                                                            |
| Included Insert<br>Accessory    | • Trocar Needle (Metal<br>Stiffening Cannula)<br>• Trocar Stylet<br>• Flexible (plastic)<br>Stiffening Cannula<br>• Wire cap (for string<br>Lock Pigtail)<br>• Suture Wire<br>• Curve Straightener<br>• Radiopaque band | • Trocar Needle (Metal<br>Stiffening Cannula)<br>• Trocar Stylet<br>• Trocar<br>(Trocar)<br>• Flexible Cannula<br>• Wire-Ring<br>• Suture Wire<br>• Radiopaque marker | • Trocar Needle (Metal<br>Stiffening Cannula)<br>• Trocar Stylet<br>• Wire cap (for string<br>Lock Pigtail)<br>• Suture Wire<br>• Curve Straightener                                               | Different but do not<br>raise new issues of<br>SE.                                                            |
| Packaging                       | Tyvek/Mylar<br>(PET/LDPE) pouch                                                                                                                                                                                         | Tyvek/Mylar pouch                                                                                                                                                     | Tyvek/Mylar<br>(PET/LDPE) pouch                                                                                                                                                                    | Same as predicate II                                                                                          |
| Sterilization<br>Method         | Ethylene Oxide                                                                                                                                                                                                          | Ethylene Oxide                                                                                                                                                        | Ethylene Oxide                                                                                                                                                                                     | Same                                                                                                          |

{6}------------------------------------------------

#### K200103

# BIOTEQUE CORPORATION
BIOTEQ Drainage Catheter Set (One Step Type)

# Section 5 - 510 (k) Summary

{7}------------------------------------------------

#### K200103

#### BIOTEQUE CORPORATION BIOTEQ Drainage Catheter Set (One Step Type)

Traditional 510(k) Section 5 - 510 (k) Summary

# 5.12 Similarity and Difference

The BIOTEQ Drainage Catheter Set (One Step Type) has been compared with "General Purpose Drainage Catheter" and "BIOTEQ® Pigtail Drainage Catheter Set (One Step Type)". The subject device has same intended use, principle of operation and similar technological characteristics as the predicate devices. The subject device has undergone safety and performance tests, and the results complied with the test requests. Although there are some different specifications between these devices, the performance test has been completed to demonstrate that the differences between these parameters would not impact the safety and effectiveness of the subject device. Therefore, the difference between the subject device and the predicate devices did not raise any new issue of substantial equivalence. The subject device is substantially equivalent to the predicate devices in intended use, design and performance claims.

### 5.13 Conclusion

After analyzing non-clinical laboratory studies and safety testing data, it can be concluded that the BIOTEQ Drainage Catheter Set (One Step Type) is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE/K200103](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE/K200103)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
