← Product Code [FGE](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE) · K053404

# LIFESTENT FLEXSTART BILIARY STENT SYSTEM (K053404)

_Edwards Lifesciences, LLC · FGE · Dec 21, 2005 · Gastroenterology, Urology · SESU_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE/K053404

## Device Facts

- **Applicant:** Edwards Lifesciences, LLC
- **Product Code:** [FGE](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE.md)
- **Decision Date:** Dec 21, 2005
- **Decision:** SESU
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5010
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Edwards Lifesciences LifeStent FlexStar Biliary Stent System is indicated for use in the palliation of malignant strictures (neoplasms) in the biliary tree.

## Device Story

LifeStent FlexStar Biliary Stent System is a permanently implanted, self-expanding nitinol stent; laser-cut open lattice design. Stent is pre-loaded into an over-the-wire delivery catheter. Used by physicians to maintain patency of bile ducts obstructed by malignant tumors. Device is deployed in the biliary tree to provide palliation of strictures. Benefits include restoration of bile flow through obstructed ducts. No software or electronic components.

## Clinical Evidence

Bench testing only. Testing included deployment, dimensional, and tensile strength testing on delivery system joints. Results demonstrated performance substantially equivalent to predicate devices.

## Technological Characteristics

Self-expanding stent; nitinol tube; laser-cut open lattice design. Dimensions: 20mm-80mm length, 6mm-10mm diameter. Delivery system: over-the-wire catheter. Non-electronic, mechanical device.

## Regulatory Identification

A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- LifeStent NT18 Self-Expanding Biliary Stent System ([K024303](/device/K024303.md))
- LifeStent NT35 Self-Expanding Biliary Stent System ([K042985](/device/K042985.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Edwards Lifesciences. The logo consists of a square graphic to the left of the company name. The text "Edwards Lifesciences" is written in a serif font. 

Ko5340f

CONFIDENTIAL May not be reproduced without written permissio Edwards Lifesciences

# 510(k) Summary

# 1. Submitter's Name and Address:

DEC 2 1 2005

Edwards Lifesciences LLC One Edwards Way Irvine, CA 92614

# 2. Contact:

Kevin Drisko Sr. Regulatory Manager Edwards Lifesciences LLC One Edwards Way Irvine, CA 92614 Phone: 949-250-2416 FAX: 949-250-2303 E-Mail: kevin drisko@edwards.com

# 3. Date Prepared:

December 6, 2005

# 4. Device Trade Name:

LifeStent FlexStar Biliary Stent System

### 5. Device Common Name:

Biliary Stent

# 6. Device Classification Name:

Biliary Catheter (78 FGE), Class II

### 7. Predicate Devices:

LifeStent NT18 Self-Expanding Biliary Stent System (K024303) and LifeStent NT35 Self-Expanding Biliary Stent System (K042985).

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for Edwards Lifesciences. The logo consists of a graphic on the left and the text "Edwards Lifesciences" on the right. The text is in a serif font and is black. The graphic on the left is a complex design that is difficult to discern.

# 510(k) Summary (continued)

# 8. Device Description:

The LifeStent FlexStar Biliary Stent System consists of a self-expanding stent that is provided loaded into an over-the-wire catheter that acts as a delivery system. The stent is a permanently implanted device used to maintain patency of a major bile duct obstructed by tissue of an impinging turnor. The flexible, self-expanding stent is made by laser cutting an open lattice design into a nitinol tube. The subject device is supplied in lengths of 20mm to 80mm and diameters of 6mm to 10mm.

### 9. Intended Use:

The LifeStent FlexStar Biliary Stent System is indicated for use in the palliation of malignant strictures (neoplasms) in the biliary tree.

## 10. Technological Characteristics:

Comparisons of the subject and predicate devices show that the technical characteristics such as materials, performance properties, biocompatibility, method of sterilization, and packaging are identical or substantially equivalent.

## 11. Performance Data:

Edwards Lifesciences completed bench testing such as deployment testing, dimensional testing, as well as tensile strength testing on applicable joints of the delivery system. The results indicate that the system performed in a manner substantially equivalent to the predicate devices cited in item 7 above.

### 12. Conclusion:

Since the LifeStent FlexStar Biliary Stent System has the same intended use, similar materials, similar performance properties, packaging and sterilization methods, it may be considered substantially equivalent to the predicate devices cited in item 7 above.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle symbol, composed of three curved lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle symbol. The logo is in black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 1 2005

Mr. Kevin Drisko Senior Manager, Regulatory Affairs Edwards Lifesciences LLC One Edwards Way IRVINE CA 92614

Re: K053404

Trade/Device Name: Edwards Lifesciences LifeStent FlexStar Biliary Stent System Regulation Number: 21 CFR §876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: December 6, 2005 Received: December 7, 2005

Dear Mr. Drisko:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{3}------------------------------------------------

### Page 2 - Mr. Kevin Drisko

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4616. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International, and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Sincerely yours,

Qatula

Donna-Bea Tillman, Ph.D., M.P.A. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page 1 of 1

510(k) Number: K053404

Device Name: Edwards Lifesciences LifeStent FlexStar Biliary Stent System

FDA's Statement of the Indications For Use for device:

The Edwards Lifesciences LifeStent FlexStar Biliary Stent System is indicated for use in the palliation of malignant strictures (neoplasms) in the biliary tree.

Opmil G. Seguram
(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ____ 8053404

Prescription Use
(Per 21 CFR 801.109)

OR

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE/K053404](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/FGE/K053404)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
