K223890 · Chess Medical, Inc. · FGD · Jun 16, 2023 · Gastroenterology, Urology
Device Facts
Record ID
K223890
Device Name
NAJA Gastrointestinal Catheter
Applicant
Chess Medical, Inc.
Product Code
FGD · Gastroenterology, Urology
Decision Date
Jun 16, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Indications for Use
The NAJA Gastrointestinal Balloon Catheter is intended to administer contrast medium to the small intestine.
Device Story
Single-use, over-the-wire, double-balloon catheter; designed for temporary occlusion of small intestine lumen to facilitate contrast medium infusion during enteroclysis imaging. Device tracks over 0.035" guidewire through gastroscope working channel. Features detachable hub connector allowing gastroscope removal while catheter remains in situ. Balloons inflated independently with air via 30cc syringe. Used in clinical settings by physicians. Output is localized contrast delivery; enables visualization of small intestine for diagnostic purposes.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-5, -10, -23), package integrity, mechanical tensile/shear strength, balloon inflation/deflation/burst performance, and simulated use in gastroscope channels.
Indicated for administration of contrast medium to the small intestine for enteroclysis imaging studies.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square, the words "U.S. FOOD & DRUG" are stacked on top of the word "ADMINISTRATION".
June 16, 2023
Chess Medical Inc. % Christopher Sloan President Sloan Regulatory Consulting 322 Hart Road Gaithersburg, MD 20878
Re: K223890
Trade/Device Name: NAJA Gastrointestinal Catheter Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: Class II Product Code: FGD Dated: May 16, 2023 Received: May 16, 2023
#### Dear Christopher Sloan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling
{1}------------------------------------------------
(21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medicaldevice-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advicecomprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sivakami Venkatachalam -S
for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K223890
Device Name NAJA Gastrointestinal Balloon Catheter
Indications for Use (Describe)
The NAJA Gastrointestinal Balloon Catheter is intended to administer contrast medium to the small intestine.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary NAJA Gastrointestinal Balloon Catheter
#### Submitter
| Name | Chess Medical |
|----------------|-----------------------------------------------------------------------------|
| Address | 1650 Sherbrooke Street West, Suite 9E<br>Montreal, Quebec H3H 1C9<br>Canada |
| Contact Person | Dr. Yen-I Chen |
| Phone | (514) 291-3154 |
| Email Address | cyen33@gmail.com |
| Date Prepared | December 23, 2022 |
#### Device
| Trade Name | NAJA Gastrointestinal Balloon Catheter |
|-----------------------|----------------------------------------|
| Common Name | Gastrointestinal Balloon Catheter |
| Classification Name | Gastrointestinal Tube and Accessories |
| Classification Number | 21 CFR 876.5980 |
| Product Code | FGD |
| Regulatory Class | II |
## Predicate Device
| Name [510(k) Number] | Maglinte Enteroclysis Balloon Catheter [K884379] |
|----------------------|--------------------------------------------------|
|----------------------|--------------------------------------------------|
#### Device Description
The NAJA Gastrointestinal Balloon Catheter ("NAJA device") is a single-use, single-operator, over the-wire, double balloon catheter that is compatible with 0.035" guidewires. The device is introduced orally and designed for temporary occlusion of the gastrointestinal (GI) lumen (small intestine), allowing infusion of fluids for enteroclysis studies (i.e., imaging tests of the small intestine). The device is designed to fit inside the working channel of a gastroscope and reaches its intended location by tracking over a 0.035" guidewire.
The NAJA device features a separate hub connector that is attachable and detachable from the catheter. This feature allows removal of the gastroscope from the catheter once the balloons are in the correct location and before the hub connector is attached for balloon inflation and infusion of fluids into the GI lumen. The hub connector is supplied with 1-way stopcocks on the inflation
{4}------------------------------------------------
and infusion ports. The NAJA device is also provided with a 30cc syringe for balloon inflation and infusion of fluids. The balloons are inflated independently using air.
## Indications for Use
The NAJA Gastrointestinal Balloon Catheter is intended to administer contrast medium to the small intestine.
## Comparison of Intended Use and Technological Characteristics with the Predicate Device
The proposed NAJA device has the same intended use and similar technological characteristics as the Maglinte Enteroclysis Balloon Catheter ("Maglinte Catheter"). The differences in the design between the devices do not raise different questions of safety and effectiveness. The table below summarizes the comparison between the NAJA device and Maglinte Catheter.
| | Attribute | NAJA Device<br>(Subject Device) | Maglinte Catheter<br>(Predicate Device) | | |
|--------------|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|-----|
| | 510(k) Number | To be assigned | K884379 | | |
| | Applicant | Chess Medical | Lafayette Pharmacal | | |
| | Classification Regulation;<br>Device Class | 876.5980 (gastrointestinal tube and<br>accessories); class II | 876.5980 (gastrointestinal tube and<br>accessories); class II | | |
| | Product Code<br>(Device Type) | FGD<br>(Catheter, Retention, Barium Enema<br>With Bag) | FGD<br>(Catheter, Retention, Barium Enema<br>With Bag) | | |
| | Intended Use / Indications for Use | ...intended to administer contrast<br>medium to the small intestine. | ...intended to administer contrast<br>medium to the small intestine. | | |
| | Sterile / Single Use | Yes | Yes | | |
| | Sterilization Method | Sterile (Ethylene Oxide) | Sterile (Ethylene Oxide) | | |
| | Principle of Operation | Inflation of one or two compliant<br>balloons for occlusion of the small<br>intestine, followed by infusion of<br>contrast medium in the small<br>intestine, distal to the proximal<br>balloon or in the inter-balloon<br>space. | Inflation of one compliant balloon for<br>occlusion of the small intestine,<br>followed by infusion of contrast<br>medium in the small intestine, distal to<br>the balloon. | | |
| | | Method of Delivery | Oral insertion, through a compatible<br>working channel (≥ 3.7 mm) of a<br>gastroscope. | Nasal or oral insertion. | |
| Balloons | | | Number of Balloons | Two | One |
| | | | Balloon Inflation Media | Air | Air |
| | | Balloon Diameter | 20mm - 50mm | 30 mm | |
| | | Attribute | NAJA Device<br>(Subject Device) | Maglinte Catheter<br>(Predicate Device) | |
| | | Balloon Length | 36 mm – 48 mm | 28 mm | |
| | Inflation Volume | 8 ml – 76 ml | 15 ml | | |
| | Balloon Material | Polyurethane | Latex | | |
| Marker Bands | Number of Marker<br>Bands | Five<br>(1) Distal tip, (4) proximal and<br>distal ends of the two balloons | One<br>(1) Distal tip | | |
| | Marker Band Length | Distal tip - 4 mm<br>Proximal/Distal balloon ends – 2 mm | Distal tip - 5 mm | | |
| | Marker Band Material | Polyether Block<br>Amide (Pebax®) 5533 SA 01 MED<br>w/ Tungsten 80% | Metallic (material unknown) | | |
| | Working Length | 260 cm | 155 cm | | |
| | Number of Layers | Two<br>Interior – Multi-Lumen Extrusion<br>Exterior – Braided Shaft Extrusion | One<br>Bi-Lumen Extrusion | | |
| | Number of Lumens | Four<br>(1) Guidewire lumen<br>(1) Infusion lumen<br>(2) Individual balloon inflation<br>lumens | Two<br>(1) Guidewire lumen,<br>doubles as infusion lumen<br>(1) Balloon inflation lumen | | |
| | Number of Inflation Exit<br>Ports | Two<br>One inflation exit port per balloon | Two<br>Two inflation exit ports to inflate a<br>single balloon | | |
| | Number of Infusion Exit<br>Ports | Three<br>In between the two balloons | Four<br>Distal to the single balloon | | |
| Shaft | Materials | Interior – Multi-Lumen Extrusion<br>Polyether Block Amide (Pebax®)<br>Exterior – Braided Shaft Extrusion<br>Polyether Block Amide (Pebax®) | Radiopaque Polyvinyl Chloride | | |
| | Outer Diameter | Distal 30cm (Multi-Lumen<br>Extrusion)<br>Outer Diameter: 7.5 Fr<br>Proximal 206cm (Braided Shaft<br>Extrusion)<br>Outer Diameter: 10 Fr | Outer Diameter: 13 Fr | | |
| | Distal Tip | Open lumen at distal tip,<br>guidewire can advance | Closed distal tip,<br>guidewire cannot advance | | |
| | | | | | |
| | Proximal End | Enclosed within a removable hub | Enclosed within a permanent hub | | |
| | Attribute | NAJA Device<br>(Subject Device) | Maglinte Catheter<br>(Predicate Device) | | |
| Proximal Hub | Detachable | Yes<br>Able to attach/detach proximal hub<br>to/from proximal end of shaft | No<br>Permanently fixed to proximal end of<br>shaft | | |
| | Number of Ports | Four<br>(1) Guidewire port<br>(1) Infusion port<br>(2) Individual balloon inflation ports | Two<br>(1) Guidewire port, doubles as infusion<br>port<br>(1) Balloon inflation port | | |
| | Mechanism to maintain<br>inflation volume | One-way Stopcock | Check Valve | | |
| | Guidewire Compatibility | 0.035in | 0.065in | | |
#### Comparison of NAJA Device to Maglinte Catheter
{5}------------------------------------------------
{6}------------------------------------------------
# Biocompatibility Testing
The following biocompatibility testing was performed on the NAJA device.
| Biocompatibility<br>Test | Acceptance Criteria<br>Origin | Result | Conclusion |
|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Cytotoxicity<br>Study Using the<br>ISO Elution<br>Method | ISO 10993-5: 2009,<br>Biological Evaluation of<br>Medical Devices -<br>Part 5: Tests for in vitro<br>cytotoxicity | The test article extract showed no evidence of<br>causing cell lysis or toxicity. The test article<br>extract met the requirements of the test since<br>the grade was less than a grade 2 (mild<br>reactivity). | Pass |
| ISO<br>Intracutaneous<br>Study | ISO 10993-23: 2021,<br>Biological Evaluation of<br>Medical Devices -<br>Part 23: Tests for<br>irritation | The test article met the requirements of the<br>test since the difference between each test<br>article extract overall mean score and<br>corresponding control extract overall mean<br>score was 0.0 and 0.0 for the SC and SO test<br>article extracts, respectively. | Pass |
| ISO<br>Maximization<br>Sensitization<br>Study | ISO 10993-10: 2021,<br>Biological Evaluation of<br>Medical Devices -<br>Part 10: Tests for skin<br>sensitization | The test article extracts showed no evidence<br>of causing delayed dermal contact<br>sensitization in the guinea pig. The test article<br>was not considered a sensitizer in the guinea<br>pig maximization test. | Pass |
## Bench and Performance Testing
The following bench and performance testing were conducted on the NAJA device.
| Design Verification Test | Specification | Test Result |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Package Performance Testing | | |
| Legibility of Markings | IFU, pouch label, and carton label are legible. | Pass |
| Design Verification Test | Specification | Test Result |
| Seal Width | Pouch seal width is $\ge$ 0.19". | Pass |
| Pouch Seal Visual<br>Inspection | Pouch seals are free of channels or other<br>damages that could impair sterility. | Pass |
| Bubble Emission Test | No evidence of leakage through any surface of<br>pouch. | Pass |
| Pouch Seal Strength | Pouch seal can withstand a force $\ge$ 3.35N over<br>a seal length of 25.4mm. | Pass |
| Pouch Ease of Opening | Pouch is easily opened with gloved hands,<br>while keeping internal contents stable.<br>Components inside pouch are easily<br>removable from backboard. | Pass |
| Component Visual<br>Inspection | Device components are free of damages that<br>could affect performance of product, such as:<br>kinks, cracks, separated components, etc. | Pass |
| Surface Defects | All surfaces and open channels of device are<br>free of loose foreign matter, pores, cracks, and<br>remainders of tooling agents. | Pass |
| Atraumatic Tip | Distal tip is smooth, rounded, and free from<br>sharp edges. | Pass |
| Atraumatic Infusion Holes | Infusion holes are smooth and free from sharp<br>edges. | Pass |
| Radiopaque Markers | Five (5) markers are present on device. | Pass |
| 0.035" GW Loading -<br>Front/Back | Device accommodates front- and back-loading<br>of a 0.035" GW. | Pass |
| Proximal Extrusion<br>Dimensions | Dimensions suitable for insertion into hub<br>connector. | Pass |
| 10.5Fr Channel<br>Compatibility | Distal 30cm of device fits through a 10.5Fr<br>channel. | Pass |
| Working Length | Device working length is 2360 $\pm$ 20mm. | Pass |
| Tip Length | Tip length is 20 $\pm$ 2mm. | Pass |
| Scaffold Length | Scaffold length is 120 $\pm$ 2mm. | Pass |
| Catheter Outer Diameter | Outer diameter is $\le$ 3.5mm for the catheter<br>shaft, balloon welds, distal tip. | Pass |
| Simulated Use | | |
| Number of Balloons<br>and Size | Device has two balloons that can be inflated to<br>a diameter of 50mm. | Pass |
| Hub Connector<br>Detachability | Hub connector is detachable from device and<br>can be attached for fluid delivery. | Pass |
| Shaft Insertion Feature | Shaft marker is no longer visible once inserted<br>into hub connector. | Pass |
{7}------------------------------------------------
{8}------------------------------------------------
| Design Verification Test | Specification | Test Result | | | | | | | | | | | |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|
| Gastroscope Compatibility | Device fits through working channel of<br>gastroscope. | | | | | | | | | | | | |
| Simulated Use Testing | The device remains functional throughout the<br>lifecycle of one clinical procedure. | | | | | | | | | | | | |
| Unique Hub Connector<br>Identification | Proximal terminations of hub assembly are<br>appropriately identified. Proximal<br>terminations of the hub assembly are<br>appropriately identified. | | | | | | | | | | | | |
| Legibility of Device<br>Markings | Wording on hub assembly labels is legible. | | | | | | | | | | | | |
| Functional Testing | | | | | | | | | | | | | |
| Hub Connector and Hub<br>Assembly Pressure | Pressure decay specification is met. | Pass | | | | | | | | | | | |
| Balloon Pressure and<br>Crosstalk | Pressure decay specification is met. | Pass | | | | | | | | | | | |
| Balloon OD/L Ratio –<br>unconfined | Unconfined balloon shape specification is met. | Pass | | | | | | | | | | | |
| Balloon Diameter and<br>Inflation Volume<br>Relationship | Relationship between balloon diameter and<br>inflation volume shall be determined. | Balloon ComplianceVolume [mL]OD [mm]920203044407650 (MAX) | | | | | | | | | | | |
| Balloon OD/L Ratio –<br>confined | Confined balloon shape specification is met. | Pass | | | | | | | | | | | |
| Inflation Time | Injection of 30mL of air in $\le$ 4s. | Pass | | | | | | | | | | | |
| Deflation Time | Removal of 30mL of air in $\le$ 17s. | Pass | | | | | | | | | | | |
| Infusion Time | Infusion of 30mL of water in $\le$ 19s. | Pass | | | | | | | | | | | |
| Channel Occlusion | Device is able to occlude channels 20-50mm. | Pass | | | | | | | | | | | |
| Destructive Testing | | | | | | | | | | | | | |
| Rated Burst Volume | Balloons have a rated burst volume (RBV)<br>$\ge$ 81.5mL. | Pass | | | | | | | | | | | |
| Bend Radius | No shaft kink and/or fracture when wrapped<br>180° around a 20mm bend radius. | Pass | | | | | | | | | | | |
| Shaft Buckling Force | Shaft buckle force specification is met. | Pass | | | | | | | | | | | |
| Hub to Shaft Slip Force | Hub connector slip force specification is met. | Pass | | | | | | | | | | | |
| Tensile – Proximal Shaft to<br>Extrusion | Tensile force of proximal shaft to extrusion is<br>$\ge$ 15N. | Pass | | | | | | | | | | | |
| Tensile - Distal Shaft to<br>Extrusion | Tensile force of distal shaft to extrusion is<br>$\ge$ 15N. | Pass | | | | | | | | | | | |
{9}------------------------------------------------
| Design Verification Test | Specification | Test Result |
|----------------------------------------------|--------------------------------------------------------------------|-------------|
| Tensile -Extrusion at<br>Scaffold | Tensile force of scaffold shall be ≥ 15N. | Pass |
| Shear – Balloon Distal Neck | Shear force of distal balloon neck to extrusion<br>bond is ≥ 15N. | Pass |
| Peel - Balloon Proximal<br>Neck | Peel force of proximal balloon neck to<br>extrusion bond is ≥ 15N. | Pass |
| Tensile - Extension Tube to<br>Hub Connector | Tensile force of extension tube to hub<br>connector bond is ≥ 15N. | Pass |
| Tensile - Luer to Extension<br>Tube | Tensile force of luer to extension tube bond is<br>≥ 15N. | Pass |
## Conclusions
The NAJA Gastrointestinal Balloon Catheter (subject device) is substantially equivalent to the legally marketed Maglinte Enteroclysis Balloon Catheter (predicate device) for administering contrast medium to the small intestine. The two devices have intended use and similar technological characteristics. The differences in the design between the devices do not raise different questions of safety and effectiveness. The results of biocompatibility testing confirm that the materials of construction of the NAJA device are safe, and the results of bench and performance (design verification) testing demonstrate that the NAJA device can function adequately in accordance with its intended use.
Chess Medical Inc. concludes that the NAJA Gastrointestinal Balloon Catheter (subject device) is substantially equivalent to the Maglinte Enteroclysis Balloon Catheter (predicate device).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.