14Fr Colon Decompression Set, Macon Colon Decompression Set
Applicant
Cook Ireland, Ltd.
Product Code
FEG · Gastroenterology, Urology
Decision Date
Aug 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
This device is used for treatment of acute non-toxic megacolon, (Ogilvie's syndrome) and colonic strictures.
Device Story
Colon decompression sets (CDSG/CDSM) facilitate colonic decompression; used for acute non-toxic megacolon (Ogilvie's syndrome) and colonic strictures. CDSG (14Fr) includes decompression tube, guiding catheter, wire guide; CDSM (7Fr, 8.5Fr, 10Fr) includes decompression tube with distal pigtail feature and wire guide. Both feature side ports for drainage/flushing; radiopaque for fluoroscopic visualization. Operated by trained healthcare professionals; placed via colonoscope. Output is decompression of colon via drainage collection bag attachment. Benefits include relief of colonic obstruction/distension.
Clinical Evidence
Bench testing only. Performance evaluation included biocompatibility (ISO 10993-1), simulated use, dimensional testing, resistance to collapse, flow rate, tensile strength, leakage tests, MR safety, radiopacity, and shelf-life testing.
Indicated for patients requiring treatment of acute non-toxic megacolon (Ogilvie's syndrome) and colonic strictures.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 1, 2017
Cook Ireland Ltd Laura Graham Regulatory Affairs Specialist O'Halloran Road, National Technology Park Limerick Ireland
Re: K171619 Trade/Device Name: 14Fr Colon Decompression Set (CDSG) Marcon Colon Decompression Set (CDSM) Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: FEG Dated: May 31, 2017 Received: June 2, 2017
Dear Laura Graham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Charles Viviano -S
Benjamin R. Fisher, Ph.D. For Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171619
### Device Name
14Fr Colon Decompression Set (CDSG) Marcon Colon Decompression Set (CDSM)
Indications for Use (Describe)
This device is used for treatment of acute non-toxic megacolon, (Ogilvie's syndrome) and colonic strictures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### I. SUBMITTER
Cook Ireland Ltd. O' Halloran Road National Technology Park Limerick Ireland
Phone: +353 61 334440 Fax: +353 61 239293
Contact Persons: Laura Graham, Regulatory Affairs Specialist Jacinta Kilmartin, Regulatory Affairs Manager
Phone: +353 61 334440 Fax: +353 61 239293
Date Prepared: May 31, 2017
#### II. DEVICE
Trade Name of Device:
- . 14Fr Colon Decompression Set (CDSG)
- . Marcon Colon Decompression Set (CDSM)
The model numbers are CDSG-14-175, CDSM-7, CDSM-8.5 and CDSM-10.
Common or Usual Name: Colon Decompression Set
Classification Name: Gastrointestinal Tubes and Accessories (21 CFR 876.5980)
Regulatory Class: II
Product Code: FEG
{4}------------------------------------------------
#### III. PREDICATE DEVICE
Predicate: Cook Colon Decompression Set, K900035 cleared on March 07, 1990.
#### IV. DEVICE DESCRIPTION
# 14Fr Colon Decompression Set (CDSG)
The CDSG device consists of a colon decompression tube, guiding catheter and wire guide. The colon decompression tube is 14Fr in diameter with a length of 175cm. The device design includes side ports which allow for colon decompression. The device has connections to allow attachment to drainage collection bag or for the device to be flushed. The colon decompression tube, guiding catheter and wire guide can be observed fluoroscopically. The device is supplied sterile, intended for single use only. Use of this device is restricted to a trained healthcare professional.
# Marcon Colon Decompression Set (CDSM)
The CDSM device consists of a colon decompression tube and wire guide. The colon decompression tube is supplied in a range of diameter options, 7Fr. 8.5Fr and 10Fr. The colon decompression tube has a length of 350cm. The device design includes side ports which allow for colon decompression. The device has connections to allow attachment to drainage collection bag or for the device to be flushed. The colon decompression tube and wireguide can be observed fluoroscopically. The colon decompression tube has a design feature (pigtail) at the distal end of the device which helps reduce migration. The device is supplied sterile, intended for single use only. Use of this device is restricted to a trained healthcare professional.
# Wire Guide
The same wire guide is supplied with both the CDSG and CDSM devices; the wire guide has a flexible distal ball tip.
#### INDICATIONS FOR USE V.
This device is used for treatment of acute non-toxic megacolon, pseudo-obstruction (Ogilvie's syndrome) and colonic strictures.
The subject and predicate device are intended to be used for colon decompression.
The intended use is identical; however, the indications for use (patient population) are limited for the subject device compared to the original clearance for the predicate device.
{5}------------------------------------------------
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH A PREDICATE DEVICE
The subject device is substantially equivalent to the currently marketed device, Colon Decompression Sets, K900035 cleared on March 07, 1990.
In brief, the subject device has the same technological characteristics as the predicate Colon Decompression Sets, with respect to the following:
- Both devices have the same intended use for colon decompression. —
- | Both devices are for use in the colon.
- -Both devices are intended for single use only.
- -Both devices are sold sterile and are sterilised using ethylene oxide.
- -Both devices are visible under fluoroscopy (radiopaque).
- Both devices have side ports. -
- Both devices are placed over the 0.035in wire guide. -
- -Both devices are supplied with a wire guide with the same material and dimensions.
The following technological differences exist between the subject device and the predicate device, Cook Colon Decompression Sets;
- Method of placement (Via a Colonoscope) and supplied with of guiding catheter. |
- । Colon decompression tube dimensions (Diameter and Length) and sideport configuration
- Pigtail (For CDSM models only) |
- | Material
#### PERFORMANCE DATA VII.
The biocompatibility evaluation for the Colon Decompression Sets, was conducted in accordance with ISO 10993-1: 2009 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" and FDA's biocompatibility guidance, Use of International Standard ISO-10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" (June 16, 2016). Performance testing included simulated use, dimensional testing, resistance to collapse, flow rate, tensile strength testing, dimensional and leakage tests, MR safety testing, radiopacity and shelf life testing.
{6}------------------------------------------------
#### VIII. CONCLUSIONS
The subject device has indications for use and technological characteristics that are similar to the predicate device. The results of the non-clinical testing demonstrates that the Colon decompression sets met the design input requirements based on the intended use, and do not raise new questions of safety or effectiveness. The results of these tests support a determination of substantial equivalence of the colon decompression sets to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.