P180046 · Boston Scientific Corporation · EZW · Nov 13, 2019 · Gastroenterology, Urology
Device Facts
Record ID
P180046
Device Name
Axonics Sacral Neuromodulation System
Applicant
Boston Scientific Corporation
Product Code
EZW · Gastroenterology, Urology
Decision Date
Nov 13, 2019
Decision
APPR
Regulation
21 CFR 876.5270
Device Class
Class 3
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
P180046 · Nov 13, 2019
Axonics Sacral Neuromodulation System
Boston Scientific Corporation
Published clinical literature (original research articles and systematic reviews); Medline/PubMed database
The FDA used a systematic literature review of 1,277 patients treated with the predicate InterStim SNM system to provide clinical evidence of safety and effectiveness for the Axonics SNM System, given the similarities in design, technology, and indications.
Systematic literature review; Meta-analysis; Retrospective data; Clinical cohort studies
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Systematic review of InterStim SNM System clinical studies; Systematic review and meta-analysis of published clinical studies; Follow-up/Duration: Various (up to 5 years); Study Period: Literature published through January 15, 2019
Patients with urinary retention (UR) and overactive bladder (OAB) symptoms (urinary urgency-frequency and/or urinary urgency incontinence); Sample Size: 1,277 patients (1,111 for safety, 1,075 for effectiveness); Number of Sites: Not applicable
Not applicable for this study
Safety (adverse events) and Effectiveness (responder rates for UUI, UF, and UR)
Indications for Use
The Axonics Sacral Neuromodulation Therapy for urinary control is indicated for the treatment of urinary retention and the symptoms of overactive bladder, including urinary urge incontinence and significant symptoms of urgency-frequency alone or in combination, in patients who have failed or could not tolerate more conservative treatments.
Device Story
Axonics Sacral Neuromodulation System is a rechargeable, implantable nerve stimulation system for urinary retention and overactive bladder. System includes an Implantable Pulse Generator (IPG), tined lead, trial stimulator, clinician programmer, and patient remote control. Procedure involves intraoperative test stimulation of sacral nerve (typically S3) via foramen needle, followed by a trial period (up to 14 days) using a temporary lead or percutaneous extension. If successful, IPG is implanted subcutaneously in the upper buttock and connected to the tined lead. Clinician programs stimulation parameters (frequency, pulse width, amplitude) via tablet-based programmer. Patient uses remote control to adjust amplitude, check battery, or turn stimulation on/off. Device uses current-controlled stimulation to maintain constant voltage. Rechargeable battery supports 15-year life. Therapy aims to improve bladder control, reduce urgency/frequency, and improve quality of life. System is MR Conditional.
Clinical Evidence
Evidence includes a systematic literature review of 7 articles (1,111 patients) on the InterStim system and the prospective, multicenter ARTISAN-SNM pivotal study (129 patients). Primary effectiveness endpoint (responder rate ≥50% reduction in UUI symptoms) was 89.9% at 6 months (p<0.0001). Secondary endpoints showed significant reductions in urgency leaks, urgency episodes, and voids, with clinically meaningful improvements in ICIQ-OABqol scores. Safety data showed no serious device- or procedure-related AEs in the ARTISAN-SNM study; common AEs included implant site pain and procedural pain.
Technological Characteristics
Rechargeable IPG (Model 1101) with 15-year life; 4-electrode platinum-iridium tined lead; current-controlled stimulation (0-12.5 mA, 2.1-130 Hz, 60-450 μs). Materials: Platinum-iridium electrodes, polyurethane lead body, MP35N conductor wires. Connectivity: RF induction for charging and wireless communication with remote/programmer. Sterilization: Ethylene oxide. MR Conditional (1.5T/3T head, 1.5T body).
Indications for Use
Indicated for patients 18+ years with urinary retention or overactive bladder (urinary urge incontinence, urgency-frequency) who failed or cannot tolerate conservative treatments. Contraindicated if patient fails test stimulation or cannot operate the system.
Regulatory Classification
Identification
An implanted electrical urinary device is a device intended for treatment of urinary incontinence that consists of a receiver implanted in the abdomen with electrodes for pulsed-stimulation that are implanted either in the bladder wall or in the pelvic floor, and a battery-powered transmitter outside the body.
Axonics Sacral Neuromodulation System for bowel control (P190006)
Submission Summary (Full Text)
{0}
SUMMARY OF SAFETY AND EFFECTIVENESS DATA (SSED)
I. GENERAL INFORMATION
Device Generic Name: Implantable Electrical Stimulator for Incontinence
Device Trade Name: Axonics Sacral Neuromodulation System
Device Procode: EZW
Applicant’s Name and Address: Axonics Modulation Technologies, Inc.
26 Technology Drive
Irvine, CA 92618
Date(s) of Panel Recommendation: None
Premarket Approval Application (PMA) Number: P180046
Date of FDA Notice of Approval: November 13, 2019
II. INDICATIONS FOR USE
The Axonics Sacral Neuromodulation Therapy for urinary control is indicated for the treatment of urinary retention and the symptoms of overactive bladder, including urinary urge incontinence and significant symptoms of urgency-frequency alone or in combination, in patients who have failed or could not tolerate more conservative treatments.
III. CONTRAINDICATIONS
Implantation of the Axonics Sacral Neuromodulation System is contraindicated for the following patients:
- Patients who have not demonstrated an appropriate response to test stimulation; or
- Patients who are unable to operate the Axonics SNM System.
IV. WARNINGS AND PRECAUTIONS
The warnings and precautions can be found in the Axonics Sacral Neuromodulation System labeling.
V. DEVICE DESCRIPTION
A. Overview of Device Use
The Axonics Sacral Neuromodulation (SNM) System is a rechargeable battery-powered, implantable nerve stimulation system that applies electrical stimulation to the sacral nerve
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 1
{1}
(typically S3, sometimes S2 or S4) for the purpose of treating urinary retention and the symptoms of overactive bladder.
Prior to being permanently implanted with the device, the patient first undergoes a brief period of intraoperative test stimulation of the sacral nerve using a foramen needle and a temporary stimulator. After the nerve is located and if an acceptable response is elicited, the patient next proceeds to a period of at-home trial stimulation. The trial period is used to evaluate the effects of the therapy on the patient's symptoms (via a bladder diary) and to assess possible side effects. This information is used to determine if the patient is a candidate for long-term treatment with the permanently implanted Axonics SNM System.
Trial Stimulation Phase: Trial stimulation is delivered by an external trial stimulator that is connected either to a partially implanted temporary lead that is removed following the trial period, or to a timed lead (via a temporary percutaneous extension) that remains implanted following a successful trial. Trial stimulation with the temporary lead may last up to 7 days, or with the timed lead up to 14 days. During the trial period, changes in bladder control symptoms are tracked using a bladder diary. If the bladder diary demonstrates an acceptable improvement in symptoms compared to baseline over the trial period (e.g., $\geq 50\%$ reduction in urinary symptoms), the patient may proceed to have the temporary test stimulation components removed and surgically replaced with the permanently implanted system components for long-term therapy. If, however, the patient does not have an acceptable response to test stimulation, the lead and cable will be removed and the patient will not receive the permanent implant for long-term therapy.
Permanent Implant Phase: For patients experiencing a successful response to trial stimulation, if a temporary lead was used for the trial, the permanent timed lead is implanted in its place, again typically targeting S3. The proximal portion of the timed lead is tunneled to the upper buttock where it is securely connected to the neurostimulator. The neurostimulator (also referred to as the implantable pulse generator or IPG) is implanted subcutaneously in the upper buttock. After the patient recovers from the surgery, the neurostimulator is programmed by a clinician using the clinician programmer. Based on patient feedback, programming adjustments (including changes to stimulation parameters and/or active electrodes) can be made during clinic visits. Additionally, the physician will program the patient remote control to allow the patient to make a limited degree of adjustments to the pulse amplitude. At any time, the patient can turn the stimulator ON or OFF using the remote control.
## B. Device Components
The components of the Axonics SNM System used for urinary control are identical to those of the system of the same name used for bowel control (approved under P190006). Many of the Axonics system components are also similar to those used in another approved SNM System, the Medtronic® InterStim® Therapy System, also indicated urinary and bowel control (approved under P970004 and P080025, respectively, and further modified in subsequent PMA supplements).
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 2
{2}
The Axonics Sacral Neuromodulation System consists of the following device components:
- Implantable Pulse Generator (IPG), Model 1101: A rechargeable, battery-powered implanted device that provides electrical pulses to stimulate the S2-S4 sacral nerve (typically S3). The IPG includes current-controlled stimulation to maintain constant voltage regardless of local tissue impedance. The IPG is controlled via radiofrequency with the remote control, the charger, and/or the clinician programmer, so that stimulation parameters, programming, and charging can be managed wirelessly. The IPG's stimulation output parameters and battery characteristics are listed in Table 1.
Table 1. IPG Stimulation Output Parameters and Battery Characteristics
| Stimulation Output Parameters | |
| --- | --- |
| Frequency | 2.1-130 Hz |
| Pulse Width | 60-450 μs |
| Amplitude | 0-12.5 mA |
| Stimulation Output | Current Controlled |
| Stimulation Modes | Unipolar and bipolar |
| Cycling Mode | Yes |
| Ramp feature | Yes |
| Battery Characteristics | |
| Battery capacity (nominal voltage) | 50 (3.6 V) mAh |
| Battery Type | Rechargeable |
| Device Life (at moderate energy) | 15 Years |
- Tined Lead: A stimulation cable with four (4) independent platinum-iridium ring electrodes used to deliver stimulation pulses from the neurostimulator to the sacral nerve. The lead's distal tip is implanted through the applicable sacral foramen near the sacral nerve with the proximal end connected to the neurostimulator. A set of 16 tines facilitate fixation of the lead just posterior to the sacral foramen. The tined lead is packaged as a component within the Model 1201 Tined Lead Kit that also includes two lead stylets (straight tip and curved tip), and a tined lead test stimulation cable. The Model 2201 Tined Lead Kit additionally includes a Percutaneous Extension for use of the tined lead during trial stimulation. Key specifications of the tined lead are listed in Table 2.
PMA P180046: FDA Summary of Safety and Effectiveness Data
{3}
Table 2. Tined Lead Specifications
| Feature | Specification |
| --- | --- |
| Physical Attributes | |
| Electrodes | 4 |
| Electrode shape | Cylindrical ring |
| Electrode size | 3 mm |
| Electrode spacing | 3 mm |
| Lead length | 30 cm |
| Lead impedance | 135 Ohms (Max) |
| Lead shape | Straight |
| Lead diameter | 1.3 mm, 5-French compatible |
| Retention feature | Anchoring Tines (16) |
| Connector | In-line coil, 4 filar |
| Number of conductor wires | 4 |
| Materials | |
| Proximal contacts | Platinum- Iridium |
| Electrode material | Platinum-Iridium |
| Fixation material | Polyurethane |
| Jacket tubing | Polyurethane |
| Conductor wires | MP35N(35NLT) |
| Retention sleeve | MP35N |
| Conductor wire insulation | Fluoropolymer |
- Trial Stimulator or External Pulse Generator (EPG): The Axonics Trial Stimulator (EPG) is part of the Axonics SNM Trial System. The EPG is a single-use, non-rechargeable, external device, that provides electrical pulses to stimulate the sacral nerve either by a PNE Lead (via a Basic Trial Cable) for up to 7 days or by a Tined Lead (via a Percutaneous Extension Cable) for up to 14 days. Also known as an External Pulse Generator or EPG, it is placed in an EPG belt around the patients' abdomen hip area.
- Peripheral Nerve Evaluation (PNE) Lead: A percutaneous, temporary monopolar trial lead that allows the electrical pulses from the EPG to be delivered to the sacral nerve during trial stimulation.
- Clinician Programmer (CP): A rechargeable tablet-style handheld device used to provide test stimulation during lead implantation and to program the neurostimulator.
- Patient Remote Control (RC): A battery-operated key-fob style device that uses Radio-Frequency (RF) signals to communicate with the implanted or external neurostimulator. The RC allows the patient to adjust the stimulation level within limits set by the clinician, check the stimulator battery charge level, and turn stimulation on or off.
PMA P180046: FDA Summary of Safety and Effectiveness Data
{4}
- **Charging System, Model 1401**: The Charging System includes the Charging Device, Charging Station, Charger Belt, and Charger Adhesive Carrier.
- **Charging Device (Charger or CD)**: A portable device powered by a rechargeable battery. The charging device is used for transcutaneous wireless charging of the neurostimulator through RF induction and can either be adhered to the patient’s skin or it can be held in place using a belt. The Charging Device is also referred to as a Charger or CD.
- **Charging Station (Dock or CS)**: A device that connects to a wall outlet and is used to recharge the Charging Device before the Charging Device is used to wirelessly charge the implanted neurostimulator. The Charging Station is also referred to as a Dock or CS.
- **Surgical Tool Kit (Lead Implant Kit)**: This kit consists of the following tools used during the implant procedure.
- **Foramen needle with needle stylet**: A needle used for acute intraoperative stimulation testing to locate the correct sacral nerve and sacral foramen for lead implant and nerve stimulation.
- **Directional guide**: A metal rod that holds the position in the sacral foramen as determined by using the foramen needle for the subsequent placement of the introducer sheath and dilator.
- **Introducer sheath and dilator**: A tool that increases the diameter of the hole through the foramen to allow introduction of the tined lead.
- **Lead stylet (straight or curved tip)**: A stiff wire that is inserted into the lead to increase its firmness and stability during lead placement.
- **Torque wrench**: A small wrench used to tighten the setscrew that locks the lead into the neurostimulator.
- **Tunneling tool**: A stiff device with a sharp end that creates a subcutaneous tunnel, allowing the lead to be placed along a path under the skin.
- **Needle test stimulation cable**: A cable provided to connect the CP to the foramen needle to deliver test stimulation during the lead placement procedure.
- **Lead test stimulation cable**: A 4-channel cable provided to connect the CP to the tined lead to deliver test stimulation during the lead placement procedure.
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 5
{5}

Figure 1. Main components of the Axonics Sacral Neuromodulation System
# VI. ALTERNATIVE PRACTICES AND PROCEDURES
There are several other alternatives for the treatment of urinary retention and the symptoms of overactive bladder. Each alternative has its own advantages and disadvantages. A patient should fully discuss these alternatives with his or her physician to select the treatment that best meets their expectations and lifestyle.
# Overactive Bladder
Behavioral interventions are the first choice in helping manage overactive bladder (OAB). These include pelvic floor exercises, maintaining a healthy weight, fluid consumption management, scheduled voiding, and bladder training. When these measures fail or are inadequate for symptom resolution, medication that promotes relaxation of the bladder can
PMA P180046: FDA Summary of Safety and Effectiveness Data
{6}
be used. However, these medications may fail to resolve symptoms or may have side effects that can lead to non-compliance. If a patient cannot tolerate drugs or does not experience adequate symptom relief, third line therapies may be prescribed, including SNM, posterior tibial nerve stimulation, and Botox injections. Posterior tibial nerve stimulation is an in-office procedure involving stimulation of the tibial nerve using a percutaneous needle; this technique requires multiple, on-going office visits and may not be as effective as the other third-line therapies. Patients may also receive injections of Botox into the bladder wall. This treatment lasts only a few months and can lead to urinary retention and the need for self-catheterization.
## Urinary Retention
Chronic, idiopathic, non-obstructive urinary retention can be managed by emptying the bladder with a catheter, often multiple times per day. The most commonly prescribed approach is clean intermittent self-catheterization; however, self-catheterization induces risks of urinary tract infection and is burdensome for patients. Surgical interventions include augmentation cystoplasty, urinary diversion, and sacral neuromodulation.
## VII. MARKETING HISTORY
The Axonics Sacral Neuromodulation System has received marketing approval for the treatment of fecal incontinence in the United States (September 2019) and for the treatment of urinary urge incontinence and fecal incontinence in the European Union (June 2016), Canada (December 2016), and Australia (December 2017). The device has not been withdrawn from marketing for any reason related to its safety or effectiveness.
## VIII. POTENTIAL ADVERSE EFFECTS OF THE DEVICE ON HEALTH
Below is a list of the potential adverse effects (e.g., complications) associated with the use of the device, which are risks beyond those normally associated with surgery, some of which may necessitate surgical intervention:
- Adverse change in voiding function (bowel and/or bladder)
- Allergic or immune system response to the implanted materials that could result in device rejections
- Change in sensation or magnitude of stimulation which has been described as uncomfortable (jolting or shocking) by some patients
- Device fracture/failure
- Device migration
- Electrical shock
- Infection
- Pain or irritation at Neurostimulator and/or lead site
- Seroma, hemorrhage, and/or hematoma
- Suspected lead or Neurostimulator migration or erosion
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 7
{7}
- Suspected nerve injury (including numbness)
- Suspected technical device malfunction
- Transient electric shock or tingling
- Unintended nerve activation
- Heating or burn at Neurostimulator site
- Lack of effectiveness
- Reoperation/Revision
- Undesirable change in pelvic function
For the specific adverse events that occurred in the clinical studies, please see Section X below.
## IX. SUMMARY OF NONCLINICAL STUDIES
## A. Laboratory Studies
### 1. Implantable Pulse Generator (IPG)
Testing was conducted on the Model 1101 IPG, including the following: mechanical design verification, electrical/firmware design verification, electromagnetic compatibility, and medical procedure compatibility. Key testing on the IPGs is summarized in Table 3 below. Testing demonstrated that the IPGs operated according to specifications after exposure to the tested conditions (i.e., passed testing).
Table 3. Summary of key testing performed on the IPG
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Dimensional Requirements | To demonstrate IPGs meet shape and profile requirements. | IPG samples shall meet size specifications for IPG width, height, thickness, volume, mass, and radius. |
| Mechanical Requirements | To demonstrate IPGs meet various mechanical requirements (Insertion and removal forces). | Lead removal force with setscrew disengaged shall be ≤ 5 N.
Lead insertion force with setscrew fully disengaged shall be ≤ 8 N. |
PMA P180046: FDA Summary of Safety and Effectiveness Data
{8}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Lead Insertion and Withdrawal Forces | To demonstrate that the IPG and lead meet specified interface requirements for insertion force, retention, and withdrawal force (without setscrew engaged). | Lead insertion force shall be ≤ 8 N (1.8 lbf). Lead retention force with setscrew engaged shall be ≥ 16 N (3.6 lbf). Lead withdrawal force shall be ≤ 5 N (1.1 lbf). |
| Measurement of Output Pulses | The characteristics of the output pulses shall be measured as described in ISO 14708-3. Verify proper output (amplitude, Pulse width, frequency, etc.) of the IPG function are within specified tolerances. | Amplitude, pulse width, frequency, and inter pulse delay shall be within output specifications. |
| Leakage Current | To verify the leakage current is in an acceptable range. | The maximum leakage current shall be < 9 μA. |
| Hermetic Leak Test | To demonstrate that the IPG (including feedthroughs) maintains hermeticity after exposure to environmental testing. | Device enclosure shall be hermetically sealed, helium leak shall be < 2 x 10-9atm-cc/s. |
| IPG Enclosure Deflection | To demonstrate the IPG remains mechanically intact and capable of normal operation following exposure to an enclosure deflection load. | The IPG shall remain mechanically intact and operate within specifications following the application of 20 lbf to the center of the device enclosure. |
| Battery | Capacity verification (longevity). | The battery performance is tested by performing 1,000 charge/discharge cycles. |
| | Electrical, visual, dimensional, hermeticity, short circuit testing, environmental, and forced discharge tests. | Meets specifications to have at least 15-year usable life. |
| Accelerated Aging | To demonstrate the IPGs meets 27 months of shelf life. | The IPG shall remain functional and at after 27 month shelf life. |
| Particulate Matter | To verify there is no unacceptable release of particulate matter when the device is used as intended. | Particle counts shall not exceed 6,000 particles ≥ 10 μm and 600 particles ≥ 25 μm. |
PMA P180046: FDA Summary of Safety and Effectiveness Data
{9}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Environmental Conditions | Atmospheric Pressure Exposure: To expose each IPG to pressure extremes the device may encounter during storage and distribution. | The IPG shall meet the pressure requirements of ISO 14708-3 encountered during storage and shipment. |
| | Operating Temperature: To demonstrate the IPG remains mechanically intact and capable of normal operation during exposure to low and high temperatures. | Testing per ISO 14708-3. The IPG shall remain mechanically intact and capable of normal operation during exposure to low (20°C) and high temperatures (45°C) for 2 hours. |
| | Mechanical Forces: To verify the device conforms to functional requirements and is not damaged by mechanical forces that may occur during conditions of use. | The IPG shall meet the mechanical force requirements of ISO 14708-3. |
| Biocompatibility | To test biocompatibility of the IPG in compliance with 10993 standard. | IPG shall prove to be biologically safe in accordance with ISO 10993 (as described below). |
| EMC | To demonstrate compliance with EMC requirements per ISO 14708-3 standard. | IPG shall continue to function as intended in presence of electromagnetic interference (EMI), shall present no interference to other devices, and shall not result in an unacceptable risk due to EMI. |
# 2. Tined Lead
Testing was conducted on the Model 1201 and 2201 Tined Lead, including dimensional verification, electrical safety, and environmental and mechanical conditions. Key testing of the Tined Lead is summarized in Table 4 below. Testing demonstrated that the Tined Lead operated according to specifications after exposure to the tested conditions (i.e., passed testing).
Table 4. Summary of key testing performed on the Tined Lead
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Dimensional | To demonstrate the leads meet dimensional requirements for overall lead length, lead body diameter, electrode dimensions, lead tip length, crossing profile. | The lead shall meet the specified dimensional requirements. |
PMA P180046: FDA Summary of Safety and Effectiveness Data
{10}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| DC Resistance | To demonstrate the DC resistance of the lead is within specification. | The DC resistance from each conductor contact to its corresponding electrode shall be < 135 ohms. No two (2) conductors shall be shorted with each other. |
| Stylet Interactions – Insertions / Removal | To demonstrate the number of cycles to fully insert or remove each stylet into the lead. | The Lead shall be able to withstand 5 cycles of stylet insertion and removal. |
| Tensile Strength | To demonstrate the lead remains electrically and mechanically intact after a tensile load. | The lead shall withstand 5 N tensile force between the proximal and distal most section of the lead for a minimum of 1 minute. No conductor failure shall be observed after the testing. |
| Lead Body Flex Fatigue | To test the ability of the lead body to withstand loading during long term use after conditioning. | The lead shall withstand flexure cycling (± 90 degrees) at a rate of 2 Hz for a minimum of 100,000 cycles and maintain a DC resistance of < 135 ohms. |
| Connector End Flex Fatigue | To test the ability of the lead connector to withstand repeated flexure. | The proximal region of the lead at the IPG connector junction lead connector shall withstand flexure cycling (± 45 degrees) at a rate of 2 Hz for a minimum of 175,000 cycles and maintain a DC resistance of < 135 ohms. |
| Lead Anchor Testing | To verify the lead anchoring tines provide appropriate retention force for clinical usage. | The lead shall meet the specified retention force requirement. |
| Dielectric Withstand Test | To test that the lead has effective functional electrical insulation between conductors. | After immersion in 9.0 g/L saline for a minimum of 10 days at 37 °C, the leakage current between all conductors and the reference electrode shall be < 9 μA when tested to a minimum of 200 volts DC. |
| IPG Interaction | To demonstrate the number of connection cycles with the IPG. | The lead shall withstand five (5) connection cycles to the IPG without damage. |
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 11
{11}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Percutaneous Extension (PE) Interaction | To demonstrate the number of connection cycles with the PE. | The lead shall withstand five (5) connection cycles to the PE without damage. |
| Particulate Matter | To test particulate release from the leads. | Particle counts shall not exceed 6,000 particles ≥ 10 μm and 600 particles ≥ 25 μm. |
| Biocompatibility | To test biocompatibility of the lead in compliance with 10993 standard. | Lead shall be biologically safe in accordance with ISO 10993 (as described below). |
# 3. Surgical Toolkit / Lead Implant Kit
Testing was conducted on the Model 1801 Surgical Tool Kit, including dimensional verification, environmental and mechanical conditions. Key testing on the tools in the kit is summarized in Table 5 below. Testing demonstrated that the Tools operated according to specifications after exposure to the tested conditions (i.e., passed testing).
Table 5. Summary of key testing performed on the Surgical Tool Kit (Lead Implant Kit)
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Dimensional | To demonstrate the surgical tool kit components meet dimensional requirements (e.g., overall length, outer and inner diameters, and visual marker locations). | The surgical tools shall meet the specified dimensional requirements. |
| Insertion and Removal Cycles | To demonstrate the mating components (e.g., Directional Guide and Dilator/Sheath) withstand cycles of insertion and removal without damage. | The applicable surgical tools shall meet the specifications for insertion and removal cycles. |
| Tensile Strength | To demonstrate that applicable surgical tool kit components meet minimum tensile force requirements at relevant junctions. | The applicable surgical tools shall meet the specifications for tensile strength. |
| Foramen Needles Dielectric Withstand Test | To demonstrate the foramen needles have effective functional electrical insulation between the proximal and distal conductive regions. | After immersion in 9.0 g/L saline for a minimum of 10 days at 37 °C, the insulated region shall withstand 500 volts DC for 60 seconds minimum without dielectric breakdown. |
PMA P180046: FDA Summary of Safety and Effectiveness Data
{12}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Simulated Tissue Insertion and Removal Cycles | To demonstrate the foramen needles, introducer sheath and dilator, and tunneling tool withstand insertion and removal from simulated tissue without damage. | The components shall withstand 5 cycles of insertion and removal from simulated tissue. |
| Foramen Needle Maximum Stimulation | To demonstrate the foramen needle is capable of delivering maximum stimulation parameters. | The foramen needle shall be able to deliver up to 12.5 mA stimulation pulse amplitude, 450 μs pulse width, and 130 Hz pulse frequency. |
| Introducer Sheath Radiopacity | To demonstrate the distal tip of the sheath is radiopaque and visible under fluoroscopy. | The distal dip of the introducer sheath shall be visible under fluoroscopy. |
| Torque Wrench Functionality | To demonstrate the torque wrench is limited to 4 oz-in. | The torque wrench shall be torque limited to 4 oz-in ± 10%. |
## 4. PNE Lead Implant Kit
Testing was conducted on the Model 1701 PNE Kit, including environmental conditions and transit testing. Key testing on the kit is summarized in Table 6 below. Testing demonstrated that the PNE Lead operated according to specifications after exposure to the tested conditions (i.e., passed testing).
Table 6. Summary of key testing performed on the PNE Kit
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Component Contents Verification | To ensure the PNE Kit contains the specified contents in sub-kits including the Accessories Kit, Components Kit, Cables Kit, and a PNE Lead. | The PNE Lead Implant Kit shall meet the specifications. |
| Environmental Conditioning | To demonstrate the PNE Kit components remain operational while exposed to minimum and maximum allowed temperature, pressure, and humidity. | The PNE Kit components shall remain operational at specified minimum and maximum temperature, pressure, and humidity. |
| Transit Testing | To demonstrate the PNE Kit is compliant with ASTM D4169-16 transit specification. | The PNE kit shall be compliant with ASTM D4169-16. |
The PNE Kit also includes a PNE Lead and Basic Trial Cable (BTC).
PMA P180046: FDA Summary of Safety and Effectiveness Data
{13}
Testing was conducted on the Model 1901 PNE Lead, including dimensional verification, electrical safety, environmental and mechanical conditions. Key testing on PNE Lead is summarized in Table 7 below. Testing demonstrated that the PNE Lead operated according to specifications after exposure to the tested conditions (i.e., passed testing).
Table 7. Summary of key testing performed on the PNE Lead
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Dimensional | To ensure the PNE lead meets dimensional requirements for overall lead length, electrode dimensions, lead tip length, crossing profile, and visual marker locations. | The PNE lead shall meet the specified dimensional requirements. |
| DC Resistance | To demonstrate the DC resistance of the PNE lead is within specification. | The DC resistance from proximal tip to the distal electrode shall be ≤ 150 ohms. |
| PNE Stylet Compatibility | To demonstrate the PNE lead is compatible with the PNE stylet. | The force to remove the PNE stylet from the PNE Lead shall be ≤ 1 N. |
| Basic Trial Cable (BTC) Compatibility – Insertion and Removal | To demonstrate the PNE lead is compatible with the BTC and meets specified interface requirements for insertion and removal cycles and forces. | The PNE lead shall be able to withstand three (3) cycles of insertion and removal from the BTC. The force to insert the PNE lead in to the BTC shall be ≤ 5 N. The force to remove the PNE lead from the BTC shall be ≤ 4 N. |
| Foramen Needle Compatibility – Insertion and Removal | To demonstrate the PNE lead is compatible with the foramen needles and can withstand multiple insertion and removal cycles. | The PNE lead shall be able to withstand three (3) cycles of insertion and removal from the foramen needle. |
| Tensile Strength and Elongation | To demonstrate the PNE lead remains electrically and mechanically intact after sustained elongation and the proximal tip is adequately secured to the coiled lead body. | The PNE lead shall withstand 20% elongation for ≥ 1 minute. DC resistance shall be ≤ 150 ohms after the testing. The proximal tip shall be secured to the coiled lead body with a tensile force of ≥ 5 N. |
| PNE Lead Body Flex Fatigue | To demonstrate the PNE lead body can withstand repeated flexure. | The lead shall withstand flexure cycling (± 90 degrees) at a rate of 2 Hz for ≥ 1,000 cycles without mechanical or electrical damage. |
PMA P180046: FDA Summary of Safety and Effectiveness Data
{14}
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 15
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Connector End Flex Fatigue | To demonstrate the lead connector can withstand repeated flexure. | The proximal region of the PNE lead at the proximal contact junction shall withstand flexure cycling (± 45 degrees) at a rate of 2 Hz for a minimum of 2,000 cycles without mechanical or electrical damage. |
| Dielectric Withstand Test | To demonstrate the PNE lead has effective electrical insulation between the distal electrode and proximal tip. | After immersion in saline for a minimum of 10 days at 37 °C, the leakage current between the conductor and a reference electrode shall be < 9 μA when tested to a minimum of 500 volts DC. |
| Maximum Stimulation | To demonstrate the PNE lead is capable of delivering maximum stimulation parameters. | The PNE lead shall be able to deliver up to 12.5 mA stimulation pulse amplitude, 450 μs pulse width, and 130 Hz pulse frequency. |
| Corrosion Resistance | To demonstrate the PNE lead is corrosion resistant per ASTM F2129-17 and does not release excess metal ions. | Per ASTM F2129-17 and applicable literature guidelines for acceptable levels of metal ion release. |
| Biocompatibility | To test biocompatibility of the PNE Lead in compliance with 10993 standard. | PNE shall be biologically safe in accordance with ISO 10993 (as described below). |
Testing was conducted on the BTC, including dimensional verification, electrical safety, environmental and mechanical conditions. Key testing on BTC is summarized in Table 8 below. Testing demonstrated that the BTC operated according to specifications after exposure to the tested conditions (i.e., passed testing).
Table 8. Summary of key testing performed on the Basic Trial Cable
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Dimensional | To ensure the BTC meets dimensional requirements including overall length, width, and cable diameter. | The BTC shall meet the specified dimensional requirements. |
| DC Resistance | To demonstrate the DC resistance of the BTC is within specification. | The DC resistance from each conductor proximal connector to distal connector shall be < 10 ohms. |
{15}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Tensile Strength | To demonstrate the BTC is able to withstand the specified minimum tensile force at each junction. | The BTC shall meet the specifications for tensile strength. |
| BTC Cable Body Flex Fatigue | To demonstrate the ability of the BTC cable body to withstand repeated flexure. | The BTC body shall withstand flexure cycling (± 90 degrees) for ≥ 1,000 cycles without mechanical or electrical damage. |
| Connectors Flex Fatigue | To demonstrate the ability of the BTC connectors to withstand repeated flexure. | The proximal and distal connectors BTC shall withstand flexure cycling (± 45 degrees) for ≥ 2,000 cycles without mechanical or electrical damage. |
| Dielectric Withstand Test | To demonstrate the BTC has effective functional electrical insulation from conductor to conductor and conductor to externally applied voltage. | Each conductor pair shall withstand 500 volts DC for ≥ 15 seconds without dielectric breakdown. All conductors to shall withstand 1500 volts AC for 60 seconds applied externally to the surface of the BTC without dielectric breakdown. |
| PNE Lead Compatibility – Insertion/Removal and Retention | To demonstrate the BTC is compatible with the PNE lead and meets specified interface requirements for insertion and removal cycles and forces. | The BTC shall be able to withstand three (3) cycles of PNE lead insertion and removal. |
| Ground Pad Compatibility – Insertion/Removal and Retention | To demonstrate the BTC is compatible with the ground pad and meets specified interface requirements for insertion and removal cycles and forces. | The BTC shall be able to withstand three (3) cycles of ground pad insertion and removal. The ground pad shall be retained within the BTC when subjected to a 5 N tensile force. |
| EPG Compatibility – Insertion and Removal | To demonstrate the BTC is compatible with the EPG connector and meets specified interface requirements for insertion and removal cycles and forces. | The BTC shall be able to withstand 25 cycles of insertion and removal from the EPG connector. |
| Stimulation Capability | To demonstrate the BTC is capable of delivering maximum stimulation parameters. | The BTC shall be able to deliver up to 12.5 mA stimulation pulse amplitude, 450 μs pulse width, and 130 Hz pulse frequency. |
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 16
{16}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Safety Compliance | To demonstrate the BTC complies with IEC 60601-1 requirements. | The BTC shall be compliant with applicable clauses of IEC 60601-1 standard. |
| Biocompatibility | To demonstrate compliance to applicable 10993 standards. | BTC shall comply with all biocompatibility requirements of the applicable 10993 standards. |
## 5. Percutaneous Extension
Testing was conducted on the Model 9009 Percutaneous Extension (PE), which is included as part of the Model 2201 Tined Lead Kit. The testing included dimensional verification, electrical safety, environmental and mechanical conditions. Key testing on the PE is summarized in Table 9 below. Testing demonstrated that the PE operated according to specifications after exposure to the tested conditions (i.e., passed testing).
Table 9. Summary of key testing performed on the Percutaneous Extension
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Dimensional | To demonstrate the PE meets dimensional requirements including overall lead length, lead body diameter, stopper position, and crossing profile. | The PE shall meet the specified dimensional requirement. |
| DC Resistance | To demonstrate the DC resistance of the PE is within specification. | The DC resistance from each conductor contact to its corresponding electrode shall be < 85 ohms. |
| Tensile Strength | To demonstrate the PE is able to withstand the specified minimum tensile force at each junction. | The PE shall meet the specifications for tensile strength. |
| PE Body Flex Fatigue | To demonstrate the ability of the PE body to withstand repeated flexure. | The PE body shall withstand flexure cycling (± 90 degrees) for ≥ 2,000 cycles without mechanical or electrical damage. |
| Connectors Flex Fatigue | To demonstrate the ability of the PE connectors to withstand repeated flexure. | The proximal and distal connectors PE shall withstand flexure cycling (± 45 degrees) for ≥ 4,000 cycles without mechanical or electrical damage. |
PMA P180046: FDA Summary of Safety and Effectiveness Data
{17}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Dielectric Withstand Test | To demonstrate the PE has effective functional electrical insulation between conductors and to a reference electrode in saline bath. | After immersion in saline for a minimum of 10 days at 37 °C, the leakage current between all conductors shall be < 9 μA when tested to a minimum of 200 volts DC. When all conductors are tested to a reference electrode in saline bath, the insulation shall withstand 1500 volts AC for ≥ 60 seconds without dielectric breakdown. |
| Tined Lead Compatibility – Insertion and Removal | To demonstrate the PE is compatible with the Tined Lead and meets specified interface requirements for insertion and removal cycles and forces. | The PE shall be able to withstand five (5) cycles of Tined Lead insertion and removal. |
| Tined Lead Retention | To demonstrate the PE is capable of retaining the Tined Lead within the connector when the set screw is engaged. | The Tined Lead shall remain functional and electrically connected to the PE after applying 16 N tensile force with final assembly resistance within 10% of pre-test value. |
| EPG Compatibility – Insertion and Removal | To demonstrate the PE is compatible with the EPG and meets specified interface requirements for insertion and removal cycles and forces. | The PE shall be able to withstand 25 cycles of insertion and removal from the EPG connector. |
| Stimulation Capability | To demonstrate the PE is capable of delivering stimulation for the duration of the 45 day operating life and can deliver maximum stimulation parameters. | The PE shall pass functional testing after delivering 45 days. The PE shall be able to deliver up to 12.5 mA stimulation pulse amplitude, 450 μs pulse width, and 130 Hz pulse frequency. |
| Corrosion Resistance | To demonstrate the PE is corrosion resistant per ASTM F2129-17 and does not release excess metal ions. | The PE shall meet the requirements of ASTM F2129-17 and applicable literature guidelines for acceptable levels of metal ion release. |
| Biocompatibility | To test biocompatibility of the PE in compliance with 10993 standard. | PE shall be biologically safe in accordance with ISO 10993 (as described below). |
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 18
{18}
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 19
# 6. Trial Stimulator
Testing was conducted on the Model 1601 Trial Stimulator (EPG), including dimensional verification, electrical safety, EMC, environmental and mechanical conditions. Key testing on the EPG is summarized in Table 10 below. Testing demonstrated that the EPG operated according to specifications after exposure to the tested conditions (i.e., passed testing).
Table 10. Summary of key testing performed on the Trial Stimulator (EPG)
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| Measurement of Output Pulse Characteristics | To demonstrate output (amplitude, pulse width, frequency, etc.) of the EPG function are within specified tolerances. | The EPG stimulation output shall meet the specified requirements. |
| Dimensional Requirements | To demonstrate EPGs meet shape and profile requirements. | EPG samples shall meet size specifications for EPG width, height, thickness, volume, mass, and radius. |
| Environmental Conditions | Atmospheric Pressure Exposure: To expose each EPG to pressure extremes the device may encounter during storage and distribution. | The EPG shall continue to function post exposure to atmospheric pressure per ISO 14708-3. |
| | Operating Temperature: To demonstrate the EPG remains mechanically intact and capable of normal operation during exposure to low and high temperatures. | Testing per ISO 14708-3. The EPG shall remain mechanically intact and capable of normal operation during exposure to low (5 °C) and high temperatures (40 °C). |
| | Mechanical Forces: Verify device conforms to functional requirements and is not damaged by mechanical forces that may occur during conditions of use. | The EPG shall meet the mechanical force requirements of ISO 14708-3. |
| EPG Enclosure Deflection | To demonstrate the EPG remains mechanically intact and capable of normal operation following exposure to an enclosure deflection load. | The EPG shall remain mechanically intact and operate within specifications following the application of 200 pounds of force to the center of the device enclosure. |
{19}
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| BTC or PE Insertion and Withdrawal Forces | To demonstrate that the EPG and the interfacing cable meet specified interface requirements for insertion and withdrawal force. | Mating cable insertion force shall be ≤ 25 N.
Mating cable withdrawal force shall be between 4 N and 20 N. |
| Battery | Electrical, visual, dimensional, short circuit testing, environmental, and forced discharge tests. | The EPG battery shall meet the specified requirements. |
| Operational and Shelf Life | To demonstrate operating and shelf life of EPG. | EPG shall be operable for a minimum of 28 days. |
| Software | To demonstrate the EPG meets functional and software requirements of specification documents. | EPG shall meet functional requirements as defined in product requirements, software requirements, and stimulation specification documents. |
| EMC | To demonstrate compliance with EMC requirements per IEC 60601-1-2 and AIM 7351731 standard. | EPG shall continue to function as intended in presence of electromagnetic interference (EMI) and present no interference to other devices and shall not result in an unacceptable risk due to EMI. |
| Biocompatibility | To demonstrate compliance to applicable 10993 standards. | EPG shall be biologically safe in accordance with ISO 10993 (as described below). |
7. Device Accessories
The Axonics System accessories include a Patient Remote Control, Charging System, and Clinician Programmer (Models 1501 and 2501). The software associated with these new devices was tested in accordance with the FDA Guidance Document entitled, "Guidance for the Content of Pre-market Submission for Software Contained in Medical Devices" (May 11, 2005) and all requirements were met. All these accessories were also tested for electrical functionality, mechanical, shipping, environmental (storage and operational), product safety testing (per IEC 60601-1), EMC testing (IEC 60601-1-2), and FCC parts 95 and 15. All test articles met defined acceptance criteria for the defined verification tests.
8. System Testing
Testing was conducted to verify that system-level design requirements for interactions between Axonics System components were met. All test articles met defined acceptance criteria for the system integration tests conducted. System validation testing demonstrated that the system operated as expected and has been validated for safe and effective use.
PMA P180046: FDA Summary of Safety and Effectiveness Data
{20}
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 21
# 9. System (IPG + Lead) Medical Compatibility Testing (Models 1101, 1201, and 2201)
Testing was performed on the Axonics System for compatibility with external defibrillation, high power electric fields, diagnostic ultrasound, and diagnostic x-ray exposure. Key system testing is summarized in Table 11 below. All test articles met all functional requirements of the testing after exposure to medical therapy conditions, verifying that the IPG and Tined Lead meet requirements for compatibility with these therapies. Testing demonstrated that the Axonics System operated according to specifications after exposure to the tested conditions (i.e., passed testing).
The Axonics System was tested for compatibility with magnetic resonance imaging (MRI) using a head coil with 1.5T and 3T scanners and using a body coil with 1.5T scanners under certain conditions. The device is labeled as MR Conditional.
Table 11. List of Tests and Acceptance Criteria
| Test | Test Purpose | Acceptance Criteria |
| --- | --- | --- |
| External Defibrillator | To demonstrate IPG functionality after defibrillation. | The Axonics System shall meet the functional electrical test requirements after exposure to external defibrillation per ISO 14708-3. |
| Electrocautery (High Power Electrical Fields) | To demonstrate IPG functionality after electrocautery. | The Axonics System shall withstand exposure to high power electrical fields according to standard ISO 14708-3. |
| Diagnostic Ultrasound | To demonstrate IPG functionality after ultrasound. | The Axonics System shall withstand exposure to ultrasound specified in ISO 14708-3. |
| X-Ray Compatibility | To demonstrate IPG functionality after X-ray. The Axonics ID shall be visible under X-ray. | The Axonics System shall withstand exposure to x-ray; radiographic marker shall be visible in x-ray; and there shall be minimal to no distortion of anatomical features adjacent to device. |
| MRI Compatibility with 1.5T and 3T Head Coil | MRI can be performed safely. | MRI shall be safely performed under the specific conditions listed for the Head Coil MRI. |
| MRI Compatibility with 1.5T Full Body Coil | MRI can be performed safely. | MRI shall be safely performed under the specific conditions listed for the Full Body Coil MRI. |
{21}
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 22
## B. Animal Studies
Animal studies were conducted to validate the safety and functionality of the Axonics Sacral Neuromodulation System. The testing demonstrated that the system met the safety and functionality endpoints. Key information from the animal studies is summarized in Table 12 below.
Table 12. Summary of In vivo Animal Studies
| Test | Test Objective | Results/Conclusion |
| --- | --- | --- |
| 30 and 60 day GLP
hound dog study
(106-0116-001) | To evaluate local tissue safety of the IPG and lead with and without stimulation. | Histology showed no significant unexpected device-related adverse events: Tissue changes were typical of an IPG and lead encapsulation site and no significant difference were noted between active and control devices. Some migrations occurred due to high activity of animal. |
| 30 day porcine study
(106-0077-002) | To evaluate the IPG/lead implant, stimulation of the S2 nerve root and recharging feasibility. | IPG’s functioned for 4 weeks and were successfully transcutaneously recharged at least once during the study. Animals tolerated implants and were successfully explanted. No histology was completed with this testing given the objectives. Migration in this testing was less remarkable than in canine testing, likely due to porcine anatomy. |
## C. Biocompatibility
Biocompatibility testing was performed for all patient-contacting components of the Axonics Sacral Neuromodulation System in accordance with ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, on the finished sterilized devices. All biocompatibility studies were conducted in compliance with Good Laboratory Practices (GLP), 21 CFR Part 58.
The IPG and Lead components of the Axonics Sacral Neuromodulation System are considered permanent (> 30 days) implants in contact with tissue/bone. The following biocompatibility endpoints were assessed for these device components:
- Cytotoxicity
- Irritation/Intracutaneous Reactivity
- Sensitization
- Acute Systemic Toxicity
- Genotoxicity
- Material – mediated pyrogenicity
- Implantation (13 weeks)
{22}
- Toxicological Risk Assessment of compounds extracted from the device to evaluate chronic systemic toxicity and carcinogenicity
The surgical tools for the implantation of the device are considered externally communicating devices, in contact with breached mucosal tissue for limited duration (≤ 24 hrs). The following biocompatibility endpoints were assessed for the surgical tools:
- Cytotoxicity
- Irritation/Intracutaneous Reactivity
- Sensitization
- Acute Systemic Toxicity
- Material – mediated pyrogenicity
The external charging system is considered a surface device, in contact with intact skin for a limited duration (≤ 24 hrs). The following biocompatibility endpoints were assessed for the charging system:
- Cytotoxicity
- Irritation/Intracutaneous Reactivity
- Sensitization
All pre-specified test acceptance criteria were met and all tests passed.
## D. Sterility
The Axonics components that are provided sterile are terminally sterilized using a 100% ethylene oxide (EO) sterilization cycle. Validation of the sterilization process demonstrates a Sterility Assurance Level (SAL) of 10-6 and is in compliance with ANSI/AAMI/ISO 11135-1:2007 Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices. Sterilant residuals conform to the maximum allowable limits of EO and Ethylene Chlorohydrin (ECH) residuals specified in ISO 109937: 2008 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals. The product bacterial endotoxin limits were chosen based on FDA’s Guidance for Industry - Pyrogen and Endotoxins Testing: Questions and Answers (June 2012) and were verified using Limulus Amebocyte Lysate (LAL) testing.
## E. Packaging and Shelf Life
Packaging and shelf-life validation tests were completed in compliance with ISO 11607:2006 Packaging for Terminally Sterilized Medical Devices. Shelf life is the term or period during which a commodity remains suitable for the intended use. An expiration date is the termination of shelf life, after which a percentage of the commodity (e.g., medical devices, may no longer function as intended). Please see FDA’s Guidance Document entitled “Shelf Life of Medical Devices.”
PMA P180046: FDA Summary of Safety and Effectiveness Data
{23}
Packaging materials were able to withstand the rigors of shipping and distribution to maintain product sterility.
Table 13. Summary of shelf life and operating life for device components
| Device Component | Operating Life | Shelf Life |
| --- | --- | --- |
| IPG | 15 years | 27 Months |
| Tined Lead | 15 years | 27 Months |
| Surgical Tool Kit | 2 years | 25 Months |
| Patient Remote | 5 years | 12 Months |
| Charging System | 5 years | N/A |
| Clinical Programmer | 5 years | N/A |
F. Additional Studies
1. System Usability Testing
Patient and clinician usability testing were conducted to verify the following tasks:
- Tasks with high risks of failure
- Tasks required for the overall safe and effective use of the device, but not posing serious risk to the user.
System usability testing was completed successfully with no critical user errors identified in any of the use environments.
2. Perfusion Phantom Temperature Study
In vitro testing was conducted on IPG and Charger to demonstrate that while charging the IPG, unsafe temperature rise does not occur. Testing was conducted using a perfusion phantom model to simulate the thermal environment of an IPG implanted into a human fat layer (fat presents worst case thermal environment). All test cases passed, and no unsafe conditions were observed, and no temperature readings exceeded the acceptance criteria during any of the testing.
X. SUMMARY OF PRIMARY CLINICAL STUDIES
A. Study Design
The safety and effectiveness of the Axonics SNM System for urinary control was based on a systematic review of published clinical studies that evaluated the safety and/or effectiveness of the fully implantable Medtronic InterStim SNM System, and on a prospective, multicenter clinical study designed to evaluate the safety and effectiveness of the Axonics SNM System (IDE number G170100).
PMA P180046: FDA Summary of Safety and Effectiveness Data
{24}
The Axonics SNM System is similar in design, technology, performance, indications for use, output characteristics, and patient population to the InterStim system evaluated in the studies. The literature review strategy was conducted according to the guidelines and methods suggested by Egger, Smith, and Altman in their book "Systematic Reviews in Health Care."²
The result of the systematic review and meta-analysis included seven articles, representing a total of 1,277 patients implanted with SNM systems. Safety data were reported in a total of 1,111 patients that had SNM system implants, and effectiveness data were reported in a total of 1,075 implanted patients that had SNM system implants. The articles included in the systematic review and meta-analysis included patients with urinary retention (UR) and OAB. The OAB patients had symptoms of urinary urgency-frequency (UF) and/or urinary urgency incontinence (UUI).
Additionally, safety and effectiveness data for the Axonics SNM System were reviewed from the ARTISAN-SNM study, which was an investigational device exemption (IDE) pivotal study in which 129 patients with urinary urgency incontinence (UUI) were treated with the Axonics SNM System.
Taking these two sources of data together, safety data were evaluated in a total of 1,240 patients that had SNM system implants, and effectiveness data were evaluated in a total of 1,204 patients with SNM system implants.
Based on nonclinical studies that demonstrated that the Axonics neurostimulator has comparable output characteristics to the InterStim system reported in the literature, the objective of the systematic literature review was to use published clinical literature to provide clinical evidence of the safety and effectiveness of the device for the improvement of UUI, UF, and UR symptoms. In addition, inclusion of safety and effectiveness data from the ARTISAN-SNM study provides direct evidence of the safety and effectiveness of the Axonics SNM System in the treatment of UUI.
Safety was demonstrated by a review of the following sources, which totaled 1,259 patients:
- Review of incidence of complications of the InterStim System from seven literature articles for urinary dysfunction indications. These consisted of two review articles and five original clinical research articles.
- Review of all Adverse Events (AE) from the ARTISAN-SNM study, the IDE pivotal study for the Axonics SNM System, which was conducted in 15 US clinical sites and 5 sites in Western Europe under G170100. The study enrolled 153 patients, of which 129 were implanted with the Axonics SNM System.
Effectiveness of the Axonics device was evaluated using the responder rate endpoint (obtained from the literature specific to the improvement of urinary dysfunction with the use of SNM systems and from the ARTISAN-SNM study). Responder rate was defined as:
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 25
{25}
- For UUI: Proportion of patients that obtained at least a 50% reduction in the number of leaks per day (analyses included all leaks or only urgency leaks)
- For UF: Proportion of patients that obtained at least a 50% reduction in the number of voids per day or less than 8 voids per day
- For UR: Proportion of patients that obtained at least a 50% reduction in the volume per catheterization
## B. Literature Search Strategy
The objective of the literature review was to systematically identify, select, collate and review relevant studies to support the marketing application of the Axonics SNM System. A summary of the literature search strategy and inclusion/exclusion criteria is provided below.
The scientific literature database Medline/PubMed was used by Axonics and duplicated by FDA to perform a search for published data relevant to the clinical evaluation of the Axonics SNM System. The search was conducted for literature published through January 15, 2019.
All articles from the published literature were triaged for inclusion based on their suitability prior to full review. Studies were selected for inclusion in this review if the methods section clearly indicated that the equivalent neurostimulation system (InterStim) was used in the treatment of urinary and/or bowel dysfunction. These studies were initially selected by Axonics based on the studied endpoints and the safety and effectiveness criteria selected. Systematic meta-analysis reviews, randomized clinical trials, and prospective clinical studies were included by Axonics because, according to Axonics, these were deemed "to be of the highest data quality." Individual cohort studies published less than 15 years ago were included, or if the cohort studies were published over 15 years ago and had more than 100 patients, the studies were also included in this search.
The literature search strategy from Axonics, and duplicated by FDA, consisted of the following three steps. FDA added one more step to select articles focused on urinary dysfunction that had a clearly defined study design:
1. The Medline database was searched for indexed articles using 21 MeSH terms (Medical Subject Headings, National Library of Medicine) and broad relevant terms for pelvic neurostimulation systems and treatment of fecal and urinary incontinence. After eliminating duplicates, there were 923 articles.
2. The abstract of each article was reviewed and categorized according to the same rigorous inclusion/exclusion criteria used by Axonics. Exclusions eliminated 896 articles, resulting in the selection of 27 articles for full review.
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 26
{26}
Exclusions included: n < 100 pts non-randomized (42 articles), n < 100 pts, > 15 years (83 articles), > 10 years, non-randomized (1 article), animal data (3), technical note/clinician technique (66 articles), case report/series (38 articles), cost assessment (20 articles), disease state (17 articles), dissimilar medical area (7 articles), dissimilar patient population (64 articles), dissimilar device (e.g., tibial) (151 articles), dissimilar indication (53 articles), excluded study type (e.g., bench, retrospective study) (123 articles), intra-device comparison (2 articles), medicinal substance (16 articles), no abstract (53 articles), no author (4 articles), no clinical data (98 articles), no device evaluation/no device identification (32 articles), patient care management (30 articles), and articles that only included patient physiology/anatomy/demographics (54 articles). Of note, the exclusion numbers above add to 957, because some excluded articles fit in more than one category.
3. Three additional articles were selected from other sources including two articles identified from meta-analysis reviews and one more that was found by cross reference (i.e., it was cited in the most current study publication). This step brought the review to a total of 30 articles for full assessment.
4. FDA performed an additional step to exclude articles that focused on bowel dysfunction. FDA also excluded articles on urinary dysfunction that either reported results in a study cohort already included in the literature review or articles that did not have adequate details on study design methodology. In the case of the InSite study, two articles were included (Siegel 2015⁷, and Siegel 2018⁹), which reported on two phases of this study. Phase 1 was a randomized, controlled trial (RCT) comparing SNM to standard medical therapy (SMT) at 6 months. Phase 2 was a prospective evaluation of the safety and effectiveness of SNM for 5 years. Overall, a total of seven articles were deemed appropriate for inclusion by the FDA. Out of the seven included articles, all seven had endpoints appropriate for the assessment of safety, and six of seven articles provided long-term effectiveness endpoints appropriate assess improvements in urinary dysfunction.
C. Safety and Effectiveness Results
1. Safety Results
FDA evaluated the safety of the Axonics SNM System based on two sources of data, namely the published articles on the use of the InterStim System for urinary dysfunction and a review of any AE from the ARTISAN-SNM study (the IDE study for the Axonics SNM System).
A total of seven published articles on urinary dysfunction were evaluated. These consisted of two review articles (Herbison 2009³ and Siddiqui 2008⁶) and five original clinical research articles (Amundsen 2018¹, Siegel 2015, Siegel 2018, White 2009¹², van Kerrebroeck 2007¹¹). Since patients from Siegel 2015 (InSite Phase 1) were rolled over to Siegel 2018 (InSite Phase 2), only the number of patients from Siegel 2018 are used for calculations of the total number of implanted patients. These
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 27
{27}
articles presented safety data in a total of 1,111 patients that had SNM system implants.
The ARTISAN study was conducted in 15 US clinical sites under IDE G170100 and evaluated 129 implanted patients. Taking these two sources of data together, a total of 1,240 patients that had SNM system implants were evaluated for safety.
## Safety Results from Literature Sources
The literature provided strong evidence to support low serious AE (SAE) rates for the use of the InterStim System to treat urinary dysfunction. A total of 1,111 patients had SNM system implants.
All AEs and SAEs reported per article are provided in Table 14 below.
Table 14. Adverse Events Reported in the Literature for the InterStim System.
| Article Reference | Follow up duration | Adverse Events | SAE |
| --- | --- | --- | --- |
| Amundsen 2018^{1}
(139 subjects) | 2 years | • Device revision 3%
• Device removal 8.6%
• Infection 2.9%
• Pain 1.4%
• Procedural pain 6.0% | NR^{L} |
| Herbison 2009^{3*}
(219 subjects) | 12 months | • Pain at implant site 15.3%
• Pain, new 9%
• Suspected lead migration 8.4%
• Infection 6.1%
• Transient sensation of electrical shock** 5.5%
• Pain, lead site 5.4%
• Surgical revision 33.3% | NR^{L} |
| Siddiqui 2010^{6***}
(Spinelli 2005^{10}: 127 subjects) | 13.8 months | • Lead migration 7%
• Lead revision performed 3% | NR^{L} |
| Siegel 2015^{7 €}
(InSite study – Phase 1)
(59 subjects with test stimulation, 51 subjects with full system implant) | 6 months | • Change in stimulation, undesirable 10.2%
• Pain, implant site 8.5%
• Lead migration/dislodgement 3.4%
• Infection, implant site 3.4%
• Surgical intervention† 3.9% | 0% |
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 28
{28}
| Article Reference | Follow up duration | Adverse Events | SAE |
| --- | --- | --- | --- |
| Siegel 2018^{9}
(InSite study – Phase 2)
(272 subjects) | 5 years | • Surgical intervention related to tined lead 22.4% (primary safety endpoint)
• Undesirable change in stimulation 22%
• Implant site pain 15%
• Therapeutic product ineffective 13%
• Implant site erosion 0.4%
• Other AEs 6%
• Surgical interventions ***
o Due to AE 30.9%
o Due to Battery replacement 33.5%
o Due to Lack or loss of effectiveness 33.5%
o Permanent explant 19.1% | Implant site erosion 0.4%^{§} |
| van Kerrebroeck 2007^{11 ¥}
(152 subjects) | 5 years | • New pain/undesirable change in stimulation 28.3%
• Pain at neurostimulator site 19.8%
• Pain at lead site 7.9%
• Infection at lead or neurostimulator site 7.9%
• Sensation of electric shock** 7.9%
• Undesirable change in voiding function 7.2%
• Lead migration 8.6%
• Technical problems during implant (surgery) 5.3%
• Device problem 10.6%
• Other AE 33.6%
• Surgical intervention 39.5%
• Device explant 10.5%
• Device exchange 23.7% | NR^{L} |
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 29
{29}
| Article Reference | Follow up duration | Adverse Events | SAE |
| --- | --- | --- | --- |
| White 2009^{12} €
(221 subjects with test stimulation, 202 subjects with full system implant) | 36.9 months | • Pain, implant site 2.9%
• Device malfunction, secondary to trauma 8.9%
• Infection 3.5%
• Post-operative hematoma requiring intervention 1.5%
• Lead migration 5.9%
• Explant due to lack of effectiveness 3.5%
• Revision due to battery depletion 2%
• Elective removal 5%
• Overall surgical intervention 30.3% | NR^{L} |
NR¹ Rates are not reported by the authors or not meaningful due to small sample size (n < 30).
* Only AEs with > 5% occurrence rate were reported by the authors.
** Typically classified as Uncomfortable sensation or stimulation.
*** Review article referencing multiple original clinical articles; Only one original article (Spinelli 2005¹⁰) met the inclusion/exclusion criteria set for literature review, and data from this article is provided.
€ Authors reported AE rates in subjects receiving SNM test stimulation.
† Authors reported this AE rate in subjects with full system SNM implant.
*** The sub-categories of Surgical interventions are not mutually exclusive.
§ This SAE occurred in 1 subject and was resolved.
¥ Device- and therapy-related AE rates are combined and are not mutually exclusive.
As stated earlier, the Siegel 2015 and Siegel 2018 articles reported results from the InSite study. The InSite study was Medtronic’s post-approval study as required by the FDA at the time of approval of a Premarket Approval (PMA) to help assure continued safety and effectiveness of the approved device. Post-approval studies (PAS) are conditions of device approval.
More information on the InSite study for P970004 can be found on FDA’s website: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpma/pma_pas.cfm?t_id=101911&c_id=335
The enrollment across 38 sites included a total of 571 subjects with a diagnosis of OAB as demonstrated by greater than or equal to eight voids per day and/or a minimum of two involuntary leaking episodes on a 3-day voiding diary. Subjects must have failed or were not candidates for more conservative medical treatments and
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 30
{30}
were 18 years of age or older. Additional inclusion/exclusion criteria can be found in Siegel (2015).
As stated above, the InSite study was conducted in two phases. Phase 1 was a prospective, multicenter RCT comparing SNM to SMT at 6 months. Phase 2 of the InSite study was a prospective evaluation of the safety and effectiveness of SNM for 5 years. Siegel (2015) reported results on Phase 1 of the InSite study, and Siegel (2018) reported results on Phase 2 of the InSite study.
The InSite Phase 1 study (Siegel et al, 2015) included 147 randomized subjects (70 to SNM and 77 to SMT). Adverse event data from a total of 59 subjects assigned to the SNM group were available at the 6-month follow-up. There were no unanticipated adverse device effects. Device-related AEs (related to surgery, therapy, device, or implant site) occurred in 30.5% (18/59) of subjects. None of the device-related AEs was serious. The most common device-related AEs in SNM subjects were undesirable change in stimulation 10.2% (6/59), implant site pain 8.5% (5/59), lead migration/dislodgment 3.4% (2/59), and implant site infection 3.4% (2/59). For the 51 SNM subjects with full system implant, the 6-month post-implant surgical intervention rate was 3.9% (2/51).
The InSite Phase 2 study (Siegel et al, 2018) included 340 subjects who completed the test stimulation, of which 272 received a full system implant. The primary safety objective of the study was to demonstrate that the upper bound of the 95% confidence interval for the cumulative 5-year rate of AEs related to the tined lead requiring surgery was less than 33%. The 5-year cumulative rate of surgical intervention related to tined lead was 22.4% (95% CI 16.6-27.7), which fulfilled the primary safety objective. There were no unanticipated device-related AEs. In subjects with a fully implanted system, an undesirable change in stimulation was the most common AE, which occurred in 60 of 272 subjects (22%), followed by implant site pain in 40 subjects (15%) and therapeutic product ineffectiveness in 36 subjects (13%). All other device related AEs, which developed upon or after implantation, were reported in fewer than 6% of subjects. One event, implant site erosion, was classified as serious but it resolved. Surgical interventions were also reported, including revision, replacement, and permanent explant of any device component. A subject could have experienced multiple types of surgical interventions and an intervention could have been due to multiple reasons, such as an AE, subject request, lack or loss of effectiveness or battery replacement. Surgical intervention was performed in 84 subjects (30.9%) due to an AE and 91 (33.5%) underwent a surgical intervention due to battery replacement. In all 272 implanted subjects, the permanent explant rate was 19.1% (95% CI 14.1-23.9) at 5 years. The top reason reported by investigators for permanent explant was an AE in 30 of the 272 subjects (11.0%), which was most often an ineffective therapeutic product (7 of 272 or 2.6%). Other reasons included subject need for magnetic resonance imaging, lack or loss of effectiveness and withdrawal of subject consent. Of the permanent explants, 23 (8.5%) were associated with a lack or loss of effectiveness. Surgical intervention was performed in 91 subjects (33.5%) due to lack or loss of effectiveness after full system implantation.
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 31
{31}
van Kerrebroeck et al (2007) conducted a prospective, single-arm, multicenter study initiated after FDA approval of InterStim therapy. A total of 163 subjects were enrolled and 152 subjects received the full system implant. Safety data through 5-year follow-up were presented in all implanted subjects, and relatedness to device or therapy was provided. Table 14 above provides AE rates combined across device-related and therapy-related AEs, and as such, an AE may be either device-related or therapy-related or both. There were 102 (67%) subjects who had at least one device- or therapy-related AE. Of the AEs, 31 were device-related (24 subjects, 15.8%) and 240 were therapy-related (97 subjects, 63.8%). Most AEs (96%) were resolved by the time the data were analyzed. A total of 60 (39.5%) subjects experienced an AE requiring surgical intervention, with 36 (23.7%) requiring device exchange. The system was explanted from 16 subjects due to adverse event or lack of effectiveness.
Amundsen et al (2018) conducted a multicenter, open-label, RCT in 386 women with more than six episodes of UUI over 3 days and inadequately managed by medications. Subjects were assigned to the SNM arm (n=194) or the Botox arm (n=192). Of the 194 subjects assigned to SNM, 139 received full implants, and safety data are reported in these subjects. At 2 years, device revisions occurred in 4/139 (3%) because of decreased effectiveness. Device removal occurred in 12/139 (8.6%) (infection 2.8%, decreased effectiveness 2.8%, subject desire 1.4%, and pain 1.4%). One participant was re-implanted after a resolved surgical site infection. Post-procedure pain was reported in 6% of subjects. Additional analysis compared all AEs between Botox and SNM groups, and the only observed clinical difference was an increased rate of urinary tract infections in subjects treated with Botox.
White et al (2009) conducted a prospective, longitudinal study in 221 subjects who received test stimulation, of which 202 received full system SNM implants. Subjects had refractory urinary urgency and frequency (n=121), urge incontinence (n=63), or urinary retention (n=37). At a mean follow-up of 36.9 months, 67 subjects (30.3%) had experienced AEs that required surgical interventions at the lead and neurostimulator site. The complications included pain at the site of the neurostimulator in six subjects (2.97%), device malfunction secondary to trauma in 18 (8.9%), infection in seven (3.5%), postoperative hematoma requiring re-exploration in three (1.5%), and lead migration in 12 subjects (5.9%). An additional seven subjects (3.5%) underwent device removal for lack of effectiveness, four subjects (2.0%) required revision secondary to battery expiration, and 10 subjects (5.0%) underwent elective removal.
Herbison et al (2009) reported safety data from three articles (Hassouna 2000; Jonas 2001; Schmidt 1999) with 219 implanted subjects at 12 months. Only AEs with more than 5% prevalence were reported by the authors. These AEs included pain at the implant site (15.3%), new pain (9.0%), suspected lead migration (8.4%), infection (6.1%), transient sensation of electric shock (5.5%), and pain at the lead site (5.4%). Surgical revision of the implant or leads had to be carried out in 33.3% of the subjects.
PMA P180046: FDA Summary of Safety and Effectiveness Data
Page 32
{32}
Siddiqui et al (2010) was a review article that summarized safety data from six original articles (five full-text, one abstract only). Only one of the articles (Spinelli 2005) met Axonics' literature review inclusion/exclusion criteria, and AE data from this study are summarized in Table 14. This article reported AEs in 127 subjects followed up for an average duration of 13.8 months. Lead migration rate as reported at 6 months was $7\%$ , and lead revision was performed in $3\%$ of the cases.
# Safety Results from Axonics Clinical Study
The ARTISAN-SNM Study was a single arm, prospective, multicenter, unblinded, pivotal study with the primary objective of evaluating the safety and effectiveness of the Axonics Sacral Neuromodulation System for the treatment of Urinary Urgency Incontinence (UUI), a subtype of OAB. The study was conducted at 15 US investigational sites (with 97 patients implanted) and at five sites in Western Europe (with 32 patients implanted). In this study, patients were tested intraoperatively for responses suggestive of lead placement near the target sacral nerve, and were then implanted with the permanent implant, rather than undergoing the typical SNM trial period (with external stimulator and percutaneous lead). FDA used the outcomes of this study for their evaluation of the safety of the Axonics SNM System at 6 months post-implantation and therapy activation. In McCrery et. al $(2019)^{5}$ , additional study design details are provided.
The primary safety endpoint was the rate of AEs reported in the study.
A total of 181 AEs were reported among 80 subjects across the entire study experience. One hundred eighty (180) of the 181 AEs occurred in implanted subjects, and one AE occurred in a subject that was enrolled in the study, but not implanted. Of the 180 AEs, seven were SAEs; no SAEs were procedure-related or device-related. Out of the 173 non-serious AEs, 13 were related to the device and 15 were related to the procedure (as shown in Tables 15 and 16 below). One death occurred from complications following multiple perforated diverticulum of the large intestine. The death was not related to the device or procedure. None of the reported AEs was unanticipated.
The total number and percentage of AEs by event category, seriousness, and relatedness to the device or procedure is presented in Tables 15 and 16 below.
Table 15. Device-Related AEs and SAEs Reported in the ARTISAN-SNM Study
| | Device Related | | Serious Device Related | |
| --- | --- | --- | --- | --- |
| AE Type | Events (n) | Subjects (n/N) (%) | Events (n) | Subjects (n/N) (%) |
| Proctalgia | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Pain | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Medical device discomfort | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Implant site pain | 2 | 2 (1.6) | 0 | 0 (0.0) |
| Incision site infection | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Pain at extremity | 2 | 2 (1.6) | 0 | 0 (0.0) |
PMA P180046: FDA Summary of Safety and Effectiveness Data
{33}
| | Device Related | | Serious Device Related | |
| --- | --- | --- | --- | --- |
| AE Type | Events (n) | Subjects (n/N) (%) | Events (n) | Subjects (n/N) (%) |
| Groin Pain | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Dysesthesia | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Lead dislodgement | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Vulvovaginal pain | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Vulvovaginal discomfort | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Total | 13 | 13 (10.1) | 0 | 0 (0.0) |
Table 16. Procedure-Related AEs and SAEs Reported in the ARTISAN-SNM Study
| | Procedure Related | | Serious Procedure Related | |
| --- | --- | --- | --- | --- |
| AE Type | Events (n) | Subjects (n/N) (%) | Events (n) | Subjects (n/N) (%) |
| Vomiting | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Implant site pain | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Hypersensitivity | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Allergy to chemicals | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Incision site infection | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Fungal infection | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Procedural Pain | 4 | 4 (3.1) | 0 | 0 (0.0) |
| Incision site pain | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Paresthesia | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Keloid scar | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Dermatitis papillaris capillitii | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Suture insertion | 1 | 1 (0.8) | 0 | 0 (0.0) |
| Total | 15 | 13 (10.1)** | 0 | 0 (0.0) |
** Note: A total of 15 events occurred in a total of 13 subjects.
The most common device-related AEs were implant site pain $(n = 2)$ , extremity pain $(n = 2)$ , and vulvovaginal pain/discomfort $(n = 2)$ . No other device-related AE occurred more than once. The most common procedure-related AE was procedural pain $(n = 4)$ . No other procedure-related AE occurred more than once.
There were no device- or procedure-related SAEs.
The time course and resolution status of device-related and procedure-related AEs from the ARTISAN-SNM study are provided in Tables 17 and 18, respectively, below. All AEs and their resolution status are reported as of the data lock date of January 18, 2019.
PMA P180046: FDA Summary of Safety and Effectiveness Data
{34}
Table 17. Summary and Time-Course of Device-Related Adverse Events
| Number of implanted subjects = 129 | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| AE Type | Implant to 2 Weeks | 2 weeks to 1 Month | 1 to 3 Months | 3 to 6 Months | 6 to 12 Months | Beyond 12 Months | Status Resolved*/Ongoing |
| Total events | 1 | 4 | 2 | 3 | 3 | 0 | 13/0 |
| Proctalgia | 0 | 0 | 0 | 1 | 0 | 0 | 1/0 |
| Pain | 0 | 1 | 0 | 0 | 0 | 0 | 1/0 |
| Medical device discomfort | 0 | 0 | 0 | 0 | 1 | 0 | 1/0 |
| Implant site pain | 1 | 0 | 1 | 0 | 0 | 0 | 1*/0 |
| Incision site infection | 0 | 1 | 0 | 0 | 0 | 0 | 1/0 |
| Pain in extremity | 0 | 1 | 0 | 1 | 0 | 0 | 1/0 |
| Groin pain | 0 | 0 | 1 | 0 | 0 | 0 | 1/0 |
| Dysesthesia | 0 | 0 | 0 | 0 | 1 | 0 | 1/0 |
| Lead dislodgement | 0 | 1 | 0 | 0 | 0 | 0 | 1/0 |
| Vulvovaginal pain | 0 | 0 | 0 | 0 | 1 | 0 | 1/0 |
| Vulvovaginal discomfort | 0 | 0 | 0 | 1 | 0 | 0 | 1/0 |
* Includes events that were resolved with sequelae.
Table 18. Summary and Time-Course of Procedure-Related Adverse Events
| Number of implanted subjects = 129 | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| AE Type | Implant to 2 Weeks | 2 weeks to 1 Month | 1 to 3 Months |…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.