← Product Code [EZL](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL) · K982626

# URETHROPEXY SUTURE PLACEMENT KIT MODEL NUMBERS UM-300, MW-100, MINI-LAPAROTOMY MMK KIT MODEL UM-250, LAPAROSCOPIC BURCH0 (K982626)

_Medworks Corp. · EZL · Aug 20, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL/K982626

## Device Facts

- **Applicant:** Medworks Corp.
- **Product Code:** [EZL](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL.md)
- **Decision Date:** Aug 20, 1998
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

MedWorks Urinary Incontinence Surgical Devices are indicated for use in urethropexy surgical procedures for bladder neck suspension to correct female stress urinary incontinence due to urethral hypermobility.

## Device Story

Surgical procedure kits for urethropexy; used for bladder neck suspension to correct female stress urinary incontinence. Components include suture placement templates, bone anchors, bone drills, syringes, suture retrievers, and suture cutters. Used in clinical surgical settings by physicians. Device facilitates mechanical suspension of bladder neck; aids in correcting urethral hypermobility. Benefits patient by addressing stress urinary incontinence symptoms.

## Clinical Evidence

Bench testing only. Validation included pull testing, bond integrity testing, and simulated use performance testing to demonstrate safety and effectiveness.

## Technological Characteristics

Surgical kit containing mechanical components: suture placement templates, bone anchors, bone drills, syringes, suture retrievers, and suture cutters. Compliance with 21 CFR 820.30 design controls. Sterilization process validated to ensure no alteration of suture characteristics.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- Urethropexy Suture Placement Kit ([K971791](/device/K971791.md))
- Mini-Laparotomy MMK Kit ([K971796](/device/K971796.md))
- Laparoscopic Burch Set ([K971797](/device/K971797.md))
- Laparoscopic MMK Kit ([K963076](/device/K963076.md))
- Mini-Laparotomy Burch Set ([K971801](/device/K971801.md))
- Urethropexy Suture Placement Set ([K971802](/device/K971802.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MedWorks Corp. Special 510(k) Premarket Notification: Device Modification

July 24, 1998 Urinary Incontinence Surgical

# AUG 20 1998

## 510(k) Summary

Pursuant to 512(i)(3)(A) of the Food, Drug and Cosmetic Act, MedWorks Corp. is required to submit with this Premarket Notification either an "...adequate summary of any information respecting safety and effectiveness or state that information will be made available upon request of any person." MedWorks chooses to submit a summary of the safety and effectiveness information. The summary is as follows:

| Trade Name:          | MedWorks Urinary Incontinence Surgical Devices          |
|----------------------|---------------------------------------------------------|
| Sponsor:             | MedWorks Corp.                                          |
|                      | 2400 Crittenden Drive                                   |
|                      | Louisville, KY 40217                                    |
|                      | Resistration #1530618                                   |
| Device Generic Name: | Urinary incontinence surgical devices                   |
| Classification:      | According to Section 513 of the Federal Food, Drug, and |
|                      | Cosmetic Act, the device classification is Class II.    |
| Predicate Devices:   | K971791 - Urethropexy Suture Placement Kit              |
|                      | K971796 - Mini-Laparotomy MMK Kit                       |
|                      | K971797 - Laparoscopic Burch Set                        |
|                      | K963076 - Laparoscopic MMK Kit                          |
|                      | K971801 - Mini-Laparotomy Burch Set                     |
|                      | K971802 - Urethropexy Suture Placement Set              |
| Manufactured by:     |                                                         |
|                      | MedWorks Corp.                                          |
|                      | 2400 Crittenden Drive                                   |
|                      | Louisville, KY 40217                                    |

Product Description: The devices described in this 510(k) are surgical procedure kits intended for use in urethropexy procedures, and include components such as suture placement templates, bone anchors, bone drills, syringes, suture retrievers and suture cutters.

#### Indications for Use:

MedWorks Urinary Incontinence Surgical Devices are indicated for use in urethropexy surgical procedures for bladder neck suspension to correct female stress urinary incontinence due to urethral hypermobility.

#### Safety and Performance:

This submission is a Special 510(k): Device Modification as described in FDA's guidance document entitled "The New 510(k) Paradigm - Alternate Approaches to Demonstrating Substantial Equivalence in Premarket Notifications." In support of this 510(k), MedWorks has provided certification of compliance to 21 CFR 820.30 Design Control requirements, and a description of the internal Risk Analysis procedure. Validation testing including pull testing, bond integrity testing and simulated use performance testing is included in Design Validation and Verification planning.

{1}------------------------------------------------

MedWorks Corp. Special 510(k) Premarket Notification: Device Modification

July 24, 1998 Urinary Incontinence Surgical Devices

#### Conclusion:

ﭘﺎ

Based on the indications for use, technological characteristics, and comparison to predicate devices, the MedWorks Urinary Incontinence Surgical Devices have been shown to be safe and effective for its intended use.

96

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three human profiles incorporated into its design. The profiles are arranged in a way that they form the body and wings of the bird. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular fashion around the bird.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 20 1998

Medworks Corporation c/o Ms. Pamela Papineau Delphi Medical Device Consulting 50 Brewster Street Pawtucket, Rhode Island 02860

Re: K982626

Modified Foley Catheter for Medworks Urinary Incontinence Surgical Device Kits: Urethropexy Suture Placement Kit, Mini-Laparotomy Kit, Laparoscopic Burch Kit, Laparoscopic MMK Kit, and Mini-Laparotomy Burch Kit Dated: July 24, 1998 Received: July 28, 1998 Regulatory Class: II 21 CFR 876.5130/Procodes: 78 EZL

Dear Ms. Papineau:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In

{3}------------------------------------------------

### Page 2 - Ms. Pamela Papineau

addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

FDA notes that your device will contain sutures for which you have provided evidence that the suture characteristics are not altered by the sterilization process used in the device. However, you should be aware of the following additional information regarding the inclusion of a suture as a component of your device:

- The labeling, packaging and method of sterilization of the suture cannot be changed 1. without prior notification, review and clearance by FDA.
- 2. The supplier of the sutures used in your device cannot be changed without prior notification, review and clearance by FDA.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Robert R. Sather/

\$

Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal,

Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page

510(k) Number (if known):

Device Name: MedWorks Urinary Incontinence Surgical Devices: Urethropexy Suture Placement Kit, Mini-Laparotomy MMK Kit, Laparoscopic Burch Kit, Laparoscopic MMK Kit, Mini-Laparotomy Burch Kit

Indications for Use:

MedWorks Urinary Incontinence Surgical Devices are indicated for use in urethropexy surgical procedures for bladder neck suspension to correct female stress urinary incontinence due to urethral hypermobility.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Robert R Sather/

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices

510(k) Number K982626

Prescription Use (Per 21 CFR 801.109)

OR

Over-the -Counter Use _

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL/K982626](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL/K982626)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
