The InnoCare Specialty Foley Catheter is intended for urethral catheterization for urological bladder drainage. The device is indicated for adult use only with a maximum indwell time of no more than 30 days.
Device Story
InnoCare Specialty Foley Catheter is a 16 French, two-way silicone catheter for urethral bladder drainage. Device features straight or Coude tip configurations and a retention balloon at the distal end. Key innovation is a secondary balloon deflation mechanism within the inflation lumen, providing a fluid escape pathway to facilitate balloon deflation during forced withdrawal or catheter pullout. Used in clinical settings by healthcare professionals for urine drainage. Device is provided sterile for single use. Secondary deflation mechanism enhances safety by mitigating risks associated with traumatic catheter removal.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), dimensional verification, functional/performance testing per ASTM F623:2019, simulated use of the secondary balloon deflation mechanism, and MR compatibility. Sterilization validated per ISO 11135:2014. All acceptance criteria met.
Technological Characteristics
Silicone construction; two-way lumen; 16 French size; 17-inch length; 10 mL inflation volume. Features Luer-activated valve for balloon inflation/deflation and a secondary fluid escape pathway for emergency deflation. Sterilized via ethylene oxide. Complies with ASTM F623:2019 and ISO 10993-1:2018.
Indications for Use
Indicated for adult patients requiring urethral catheterization for urological bladder drainage; maximum indwell time 30 days.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 18, 2024
InnoCare Urologics, LLC Patrick Mccain VP, QA/RA 7175 SW 47th Street, Suite 207 Miami, Florida 33155
Re: K241424
Trade/Device Name: InnoCare Specialty Foley Catheter Regulation Number: 21 CFR 876.5130 Regulation Name: Urological Catheter and accessories Regulatory Class: II Product Code: EZL, SCT Dated: August 21, 2024 Received: August 21, 2024
Dear Patrick Mccain:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}------------------------------------------------
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sharon M. Andrews -S
for Jessica N. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology, and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K241424
Device Name InnoCare Specialty Foley Catheter
Indications for Use (Describe)
The Inno Care Specialty Foley Catheter is intended for urological bladder drainage. The device is indicated for adult use only with a maximum indwell time of no more than 30 days.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for InnoCare Urologics. The logo is in blue and features the letters "iu" in a stylized font. The word "InnoCare" is written in a larger font above the word "Urologics". There is a trademark symbol to the right of the word "InnoCare".
#### Traditional 510(k) Summary
### 1. Submitter Information
| Applicant Name: | InnoCare Urologics, LLC |
|--------------------|-----------------------------------------------------------------|
| Applicant Address: | 7175 SW 47th Street, Suite 207<br>Miami FL 33155, United States |
| Contact Telephone: | (855) 466-6305 |
| Applicant Contact: | Patrick McCain, VP, QA/RA |
| Contact Email: | pmccain@innocareurologics.com |
| Date Prepared: | May 17, 2024 |
#### 2. DEVICE NAME
| Trade or Proprietary Name: | InnoCare Specialty Foley Catheter |
|----------------------------|-------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Foley Catheter |
| Classification Regulation: | 21 CFR 876.5130 |
| Device Class: | II |
| Product Code: | EZL (Catheter, Retention Type, Balloon),<br>SCT (Foley Catheters and Accessories with Additional Safety Features) |
| Panel: | Gastroenterology/Urology |
#### 3. Predicate Device Identification
Predicate Device: Medline Silicone Foley Catheter (K142635)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for InnoCare Urologics. The logo features the letters "iu" in blue on the left side of the image. To the right of the letters is the text "INNOCARE" in a larger font, and below that is the text "UROLOGICS".
#### 4. Device Description
The InnoCare Specialty Foley Catheter is a 16 French, two-way silicone Foley catheter for urethral catheterization to drain urine from the bladder. The catheter is retained in the bladder by use of a retention balloon attached to the distal end of the catheter. The device is offered in a straight tip configuration and a Coude tip configuration.
The inflation lumen of the catheter contains a secondary balloon deflation mechanism. This mechanism is designed to provide a fluid escape pathway to allow for balloon deflation in the event of a forced withdrawal or catheter pullout.
The device is provided sterile and for single use.
#### 5. Indications For Use
The InnoCare Specialty Foley Catheter is intended for urethral catheterization for urological bladder drainage. The device is indicated for adult use only with a maximum indwell time of no more than 30 days.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for InnoCare Urologics. The logo consists of the letters "iu" in blue, with the "i" being a darker shade of blue than the "u". To the right of the letters are the words "INNOCARE" and "UROLOGICS", stacked on top of each other, in a dark blue font. There is a trademark symbol to the right of the word "INNOCARE".
#### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Comparison Element | Subject Device:<br>InnoCare Specialty<br>Foley Catheter | Predicate Device:<br>Medline Silicone Foley Catheter (K142635) |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The InnoCare Specialty Foley Catheter<br>is intended for urethral catheterization<br>for urological bladder drainage. The<br>device is indicated for adult use only<br>with a maximum indwell time of no<br>more than 30 days. | Medline silicone Foley catheters are intended to be<br>placed in the bladder, through the urethra, to drain<br>urine into a collection device. Two-way catheters<br>are intended for urological bladder drainage only.<br>Three-way catheters are intended for urological<br>bladder drainage and bladder irrigation. |
| Product Code | EZL, SCT | EZL |
| Regulation | 21 CFR 876.5130 | 21 CFR 876.5130 |
| Prescription use (Rx)/<br>Over the counter (OTC) | Rx | Rx |
| Indwell Time | No more than 30 days | No more than 30 days |
| Lumen | Two-Way | Two-Way or Three- Way |
| Primary Shaft Material | Silicone | Silicone |
| Catheter Size | 16 Fr | 12 Fr - 24 Fr |
| Catheter Length | 17" | 17" |
| Inflation Volume | 10 mL | 10 mL – 30 mL |
| Balloon Deflation | Via syringe through Luer-activated<br>valve. | Via syringe through Luer-activated valve. |
| Secondary Balloon<br>Deflation Mechanism | The inflation lumen of the catheter<br>contains a secondary balloon deflation<br>mechanism. This mechanism is<br>designed to provide a fluid escape<br>pathway to allow for balloon deflation<br>in the event of a forced withdrawal or<br>catheter pullout. | No secondary balloon deflation mechanism. |
| Tip Shapes | Straight,<br>Coude | Straight,<br>Coude |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
As evidenced by the above table, both the subject and the predicate devices have intended use, but the subject and the predicate devices have different technological characteristics.
Performance testing was conducted on the subject device, and it was established that the differences in technological characteristics between the subject and the predicate do not raise different questions of safety or effectiveness.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for InnoCare Urologics. The logo features the letters "iu" in a stylized font, with the "i" in dark blue and the "u" in light blue. To the right of the letters, the words "INNOCARE" and "UROLOGICS" are stacked on top of each other in dark blue. The letters "TM" are in the upper right corner of the word "INNOCARE".
#### 7. SUMMARY OF NON-CLINICAL TESTING
Device performance data was provided in support of the substantial equivalence determination.
#### BIOCOMPATIBILITY TESTING:
Biocompatibility testing was conducted per ISO 10993-1:2018. The following tests were conducted:
- Cytotoxicity
- Sensitization
- Irritation
- Acute Systemic Toxicity
- Subacute Systemic Toxicity
- Genotoxicity
- Implantation
The results of this testing demonstrate that the subject device is biocompatible for its intended use.
#### Bench Testing:
Bench testing of the subject device included dimensional verification, functional and performance testing per ASTM F623:2019 requirements, simulated use testing of the secondary balloon deflation mechanism and Magnetic Resonance Compatibility testing per FDA Guidance (2023), "Testing and Labeling Medical Devices for Safety in the Maqnetic Resonance (MR) Environment | FDA".
All pre-determined acceptance criteria were met. Performance testing demonstrates that the differences in technological characteristics between the subject and the predicate do not raise different questions of safety or effectiveness.
#### Sterility, Packaging and Distribution:
The InnoCare Specialty Foley Catheter was validated to be sterilized using ethylene oxide per ISO 11135:2014. To support the claimed shelf life, performance testing and package integrity testing was conducted following environmental conditioning, simulated distribution and accelerated aging.
#### 8. CONCLUSION
Based on the information presented in this submission, it can be concluded that the subject device is substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.