K202698 · Hangzhou Primecare Medical Co., Ltd. · EZL · Dec 15, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K202698
Device Name
Prefilled Syringe
Applicant
Hangzhou Primecare Medical Co., Ltd.
Product Code
EZL · Gastroenterology, Urology
Decision Date
Dec 15, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Indications for Use
The 10cc and 30cc pre-filled syringes are intended to be used for foley catheter balloon inflation.
Device Story
Single-use 10cc and 30cc syringes pre-filled with USP purified water; gamma irradiated. Used in clinical settings to inflate urinary catheter balloons. Operator manually pushes plunger to inject water into catheter balloon. Indirect body contact < 24 hours. Benefits include sterile, ready-to-use delivery of inflation fluid, simplifying catheter maintenance.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation per ISO 10993) and performance testing (ISO 7886-1, ISO 11040-4/6/8, ISO 80369-1/7/20, USP 40) were conducted. All parameters met predefined acceptance criteria.
Technological Characteristics
Materials: Polypropylene (barrel/plunger), pharmaceutical grade latex-free rubber (stopper/cap). Contents: USP purified water. Manual operation. Sterile (gamma irradiation). Conforms to ISO 7886-1, ISO 11040-4/6/8, and ISO 80369-7.
Indications for Use
Indicated for foley catheter balloon inflation in patients requiring urinary catheterization.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Primary Care Solutions Prefilled 10cc and Prefilled 30cc Inflation Syringes with Sterile Water (K030813)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 15, 2020
Hangzhou Primecare Medical Co., Ltd. % Olivia Meng Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd. 8-9th Floor, R&D Building, No.26 Qinglan Street, Panyu District Guangzhou, Guangdong 510006 China
Re: K202698
Trade/Device Name: Prefilled Syringe Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZL Dated: September 11, 2020 Received: September 16, 2020
### Dear Olivia Meng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202698
Device Name Prefilled Syringe
Indications for Use (Describe)
The 10cc and 30cc pre-filled syringes are intended to be used for foley catheter balloon inflation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary – K202698
In accordance with 21 CFR 807.92 the following summary of information is provided:
### 1. SUBMITTER
Hangzhou Primecare Medical Co., Ltd. Room 408-409, Zancheng Center West, Shangcheng District, 310008 Hangzhou, Zhejiang Province, PEOPLE'S REPUBLIC OF CHINA Phone: +86-13357148752 Fax: +86-0571-81958625
| Primary Contact | Olivia Meng |
|--------------------|----------------------------------------------------|
| Person: | Regulatory Affairs Manager |
| | Guangzhou Osmunda Medical Device Technical Service |
| | Co., Ltd. |
| | Tel: (+86)-20-6231 6262 |
| | Fax: (+86)-20-8633 0253 |
| Secondary Contact | Xueqing Yan |
| Person: | Quality Supervisor |
| | Hangzhou Primecare Medical Co., Ltd. |
| | Phone: +86-13357148752 |
| | Fax: +86-0571-81958625 |
| Date prepared | December 15, 2020 |
| 2. DEVICE | |
| Device Name: | Prefilled Syringe |
| Common name: | Prefilled Syringe |
| Model: | PCAA01, PCAA03 |
| Regulation Number | 21 CFR 876.5130 |
| Regulation Name | Urological catheter and accessories |
| Regulatory Class: | 2 |
| Product Code: | EZL |
| Product Code Name: | Catheter, Retention Type, Balloon |
3. PREDICATE DEVICE
{5}------------------------------------------------
K030813, Primary Care Solutions Prefilled 10cc and Prefilled 30cc Inflation Syringes with Sterile Water; Catalog numbers 1010 and 1030 respectively This predicate has not been subject to a design-related recall.
#### DEVICE DESCRIPTION 4.
The device is a single-use 10cc and 30cc syringe pre-filled with USP purified water and gamma irradiated. The device has two models, PCAA01 and PCAA03. They are only different in volume; PCAA01 is 10cc, and PCAA03 is 30cc. The syringe is produced using polypropylene for the device barrel and plunger and pharmaceutical grade latex free rubber for both the plunger gasket and protect cap. The device contacts indirectly with the human body. The duration of contact is less than 24 hours. The product is used in conjunction with the urinary catheter to push the plunger of the syringe to inject the sterilized pre-filled purified water into the balloon of the urinary catheter to inflate the balloon.
# 5. INDICATIONS FOR USE
The 10cc and 30cc pre-filled syringes are intended to be used for foley catheter balloon inflation.
The intended use of the device is the same as the intended use of the predicate device.
| | Predicate Device | Proposed Device | Discussion of<br>Differences | |
|------------------------|----------------------------------------------------------------------------------------------------------|---------------------------------------------|------------------------------|------|
| Device name | Primary Care Solutions<br>Prefilled 10cc and Prefilled<br>30cc Inflation Syringes<br>with Sterile Water; | Prefilled Syringe | | |
| K number | K030813 | K202698 | | |
| Size | 10cc and 30cc | 10cc and 30cc | Same | |
| Principle of operation | Manual | Manual | Same | |
| Design | Conforms with ISO 7886.1<br>and ISO 80369-7 | Conforms with ISO 7886.1<br>and ISO 80369-7 | Same | |
| Disposable? | Disposable for single use | Disposable for single use | Same | |
| Materials<br>Of | Barrel | Polypropylene | Polypropylene | Same |
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE 6.
{6}------------------------------------------------
| Main<br>Components | Plunger | Polypropylene | Polypropylene | |
|--------------------|----------------|-------------------------------------------|-------------------------------------------|------|
| | Stopper | Pharmaceutical Grade<br>Latex Free Rubber | Pharmaceutical Grade Latex<br>Free Rubber | |
| | Protective Cap | Pharmaceutical Grade<br>Latex Free Rubber | Pharmaceutical Grade Latex<br>Free Rubber | |
| | Solution | Purified Water, USP | Purified Water, USP | |
| Sterile? | | Sterile (gamma irradiation) | Sterile (gamma irradiation) | Same |
As evidenced by the table above, the subject and predicate device have similar technological characteristics.
#### 7. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility testing
The biocompatibility evaluation for the Prefilled Syringe was conducted in accordance with the International Standard ISO 10993-1:2018, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" as recognized by FDA. The biocompatible testing included the following tests:
- Cytotoxicity - (ISO 10993-5: 2009)
- Sensitization - (ISO 10993-10:2010)
- Irritation - (ISO 10993-10:2010)
## Performance testing
Performance testing was conducted on the Prefilled Syringe according to the following standards:
- 1. ISO 7886-1: 2017 Sterile hypodermic syringes for single use Part 1: Syringes for manual use
- 2. ISO 11040-8: 2016 Prefilled Syringes Requirements and test methods for finished prefilled syringes
- 3. ISO 11040-4: 2015 Prefilled Syringes Glass barrels for injectables and sterilized subassembled syringes ready for filling
{7}------------------------------------------------
- 4. ISO 11040-6: 2019 Prefilled Syringes Plastic barrels for injectables and sterilized subassembled syringes ready for filling
- 5. ISO 80369-1: 2018 Small-bore connectors for liquids and gases in healthcare applications - Part 1: General requirements
- 6. ISO 80369-7: 2016 Small-bore connectors for liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
- 7. ISO 80369-20: 2015 Small-bore connectors for liquids and gases in healthcare applications - Part 20: Common test methods
- 8. USP 40 Purified Water <6719>
All of the tested parameters met the predefined acceptance criteria.
#### CONCLUSION 8.
From the results of non-clinical data including the performance testing described, it is concluded that the Prefilled Syringe is as safe and as effective as the predicate device. The information provided within this pre-market notification demonstrates that the device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.