K200556 · H R Pharmaceuticals, Inc. · EZL · Mar 31, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K200556
Device Name
AquaFlate Pre-Filled Sterile Water Syringe 10mL
Applicant
H R Pharmaceuticals, Inc.
Product Code
EZL · Gastroenterology, Urology
Decision Date
Mar 31, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The AquaFlate Pre-Filled Sterile Water Syringe 10mL is intended to be used for inflating a Foley catheter balloon.
Device Story
Device is a 10mL syringe pre-filled with USP purified water; used for inflating indwelling urethral Foley catheter balloons. Healthcare professional removes cap, connects luer tip to catheter inflation port, and dispenses entire contents to inflate balloon; prevents catheter dislodgement. Syringe is then disconnected and discarded. Single-use only; not for injection. Provides sterile, biocompatible solution for balloon inflation in clinical settings.
Clinical Evidence
Bench testing only. Testing included small-bore connector testing (ISO 80369-7), manual syringe testing (ISO 7886-1), sterility testing (USP 42 <71>), biocompatibility (ISO 10993-5, -10), sterilization validation (ISO 11137), and packaging/shelf-life validation (ASTM F1980, ISTA 3A).
Technological Characteristics
Materials: Polypropylene barrel/plunger, latex-free synthetic rubber gasket/tip cap. Principle: Manual syringe inflation. Energy: None. Dimensions: 10mL. Sterilization: Gamma irradiation (25-40kGy). Connectivity: None. Standards: ISO 80369-7, ISO 7886-1, ISO 10993, ISO 11137, ASTM F1980.
Indications for Use
Indicated for inflating an indwelling urethral Foley catheter balloon in a healthcare facility setting.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
AMSINO INTERNATIONAL, AMSure® Prefilled Syringe for Balloon Inflation with Sterile Water (K181814)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font in blue.
March 31, 2020
HR Pharmaceuticals, Inc. % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, MN 55114
Re: K200556
Trade/Device Name: AquaFlate Pre-Filled Sterile Water Syringe 10mL Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: March 2, 2020 Received: March 3, 2020
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K200556
#### Device Name
AquaFlate Pre-filled Sterile Water Syringe 10 mL
Indications for Use (Describe)
The AquaFlate Pre-filled Sterile Water Syringe is intended to be used for inflating a Foley catheter balloon.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------|--|--|
| <table><tr><td></td><td></td></tr></table> | | |
| | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K200556 Page 1 of 6
510(k) SUMMARY 5-1
# 5.0 510(k) SUMMARY
#### 510(k) SUMMARY for AquaFlate Pre-Filled Sterile Water Syringe 10mL
#### I. SUBMITTER:
HR Pharmaceuticals, Inc. 2600 Eastern Boulevard, Suite 201 York, PA 17402
Telephone Number:717-252-1110 x104 Fax Number:717-685-2590
Contact Person: Colby Wiesman Date Prepared: November 26, 2019
II. DEVICE:
Name of Device: AquaFlate Pre-Filled Sterile Water Syringe 10mL Model Number: 250-400 Common or Usual Name: Catheter, Retention type, balloon Classification Description: Urological catheter and accessories (21 CFR 876.5130) Regulatory Class: II Product Code: EZL Review Panel: Gastroenterology/Urology
## III. PREDICATE DEVICES:
AMSINO INTERNATIONAL, AMSure® Prefilled Syringe for Balloon Inflation with Sterile Water, K181814
No reference devices were used in this submission.
## IV. DEVICE DESCRIPTION:
The proposed device, AquaFlate Pre-Filled Sterile Water Syringe 10mL (REF 250-400), is a 10mL syringe prefilled with USP purified water and sterilized via gamma irradiation. The syringe consists of a polypropylene barrel and plunger and latex free rubber for both the plunger gasket and syringe tip cap.
The AquaFlate Pre-Filled Water Syringe 10mL is intended for inflating an indwelling. urethral Foley catheter balloon. In accordance with CDC Guidelines, healthcare facility protocols, and the Foley catheter manufacturer's instructions for use, prior to catheterization, the syringe cap of the AquaFlate Pre-Filled Sterile Water Syringe 10mL is
Image /page/3/Picture/17 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H-R" in a stylized font. To the right of the letters is the text "Pharmaceuticals, Inc." The logo is simple and professional, and it is likely used on the company's website, marketing materials, and products.
{4}------------------------------------------------
removed and the luer tip of the syringe is connected to the inflation port of the catheter. Once the catheter is inserted into the bladder, the entire contents of the syringe are dispensed to inflate the catheter balloon to ensure the catheter does not slide out. A fter balloon inflation, the syringe is disconnected from the catheter inflation port and discarded.
The AquaFlate Pre-Filled Sterile Water Syringe 10mL is not intended for injection and is for single use only.
The shelf life of the AquaFlate Pre-Filled Sterile Water Syringe 10mL is one (1) year.
HR Pharmaceuticals does not intend to market this device with accessories or as part of a system.
### V. INDICATIONS FOR USE:
The AquaFlate Pre-Filled Sterile Water Syringe 10mL is intended to be used for inflating a Foley catheter balloon.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES:
The subject and predicate devices have the same technological principle whereby, the prefilled sterile water syringe is intended for inflating an indwelling, urethral Foley catheter balloon. Once the catheter is inserted into the bladder, the entire contents of the syringe are dispensed to inflate the catheter balloon to ensure the catheter does not slide out. After balloon inflation, the syringe is disconnected from the catheter inflation port and discarded. Additionally, the prefilled sterile water syringe is not intended for injection and is for single use only.
At a high level, the AquaFlate Pre-Filled Sterile Water Syringe 10mL and the predicate device are based on the following same elements:
- Same Intended Use
- Same Technological Characteristics ●
- Same Materials
- . Same Physical Characteristics
- Same Performance Specifications ●
- Packaged in the same quantity (10mL) ●
- o Provided sterile
- Biocompatible
The only identified difference between the subject and predicate devices is the shelf life. The AquaFlate Pre-Filled Sterile Water Syringe 10mL has a one-year shelf life whereby, AMSure® Prefilled Syringe for Balloon Inflation with Sterile Water has a three-year shelf life. While the AquaFlate Pre-Filled Sterile Water Syringe 10mL is currently undergoing 2 year and 3-year shelf life testing, the one-year established shelf life for the subject device is acceptable for the intended use as demonstrated by successful stability
Image /page/4/Picture/21 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet inside of a flower. The words "Pharmaceuticals, Inc." are written in a smaller font to the right of the letters.
{5}------------------------------------------------
and performance testing. Therefore, the difference in shelf life between the subject and predicate devices does not raise new questions of safety or effectiveness.
A more detailed comparison of the predicate and subject devices is presented in the table below.
| | Proposed Device<br>AquaFlate Pre-Filled Sterile<br>Water Syringe 10mL | Predicate Device<br>AMSure Prefilled<br>Syringe for Balloon<br>Inflation with Sterile<br>Water<br>(K181814) | Comparison | |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|------|
| Manufacturer | HR Pharmaceuticals, Inc. | Amsino International, Inc. | N/A | |
| Device Classification<br>Name | Catheter, Retention Type,<br>Balloon | Catheter, Retention Type,<br>Balloon | Same | |
| Regulation Description | Urological catheter and<br>accessories | Urological catheter and<br>accessories | Same | |
| Review Panel | Gastroenterology/Urology | Gastroenterology/Urology | Same | |
| Product Code | EZL | EZL | Same | |
| Reg Number | 876.5130 | 876.5130 | Same | |
| Class | II | II | Same | |
| Indications for Use | The AquaFlate Pre-Filled Sterile<br>Water Syringe 10mL is intended<br>to be used for inflating a Foley<br>catheter balloon. | Prefilled Syringe for<br>Balloon Inflation with<br>Sterile Water is intended<br>to be used in inflating<br>foley catheter balloon. | Same | |
| Patient-contacting<br>material | Indirect patient contact of sterile<br>water used to inflate a catheter<br>balloon | Indirect patient contact of<br>sterile water used to<br>inflate a catheter balloon | Same | |
| Sizes | 10mL | 10cc and 30cc | Same<br>(10mL) | |
| Materials of<br>Main<br>Components | Barrel | Polypropylene | Polypropylene | Same |
| | Plunger | Polypropylene | Polypropylene | Same |
| | Plunger<br>Gasket | Isoprene (black synthetic rubber<br>that does not contain natural<br>rubber latex) | Black pharmaceutical<br>grade,<br>synthetic rubber (Latex<br>free) | Same |
| | Tip<br>Cover | Isoprene (black synthetic rubber<br>that does not contain natural<br>rubber latex) | Same as gasket | Same |
| | Solution | Purified water, USP | Purified water, USP | Same |
| Principle of Operation | | Manual | Manual | Same |
| Technical Performance | | Conforms with ISO7886-1,<br>ISO 80369-7 and USP 42<br>Chapter <71> | Conforms with ISO7886-<br>1,<br>ISO 80369-7 and USP 40 | Same |
| Biocompatibility | | Meets ISO 10993-1:2009<br>requirements for irritation (ISO<br>10993-10), sensitization (ISO | Conforms with ISO<br>10993-5 and ISO 10993-<br>10 | Same |
| | Proposed Device<br>AquaFlate Pre-Filled Sterile<br>Water Syringe 10mL | Predicate Device<br>AMSure Prefilled<br>Syringe for Balloon<br>Inflation with Sterile<br>Water<br>(K181814) | Comparison | |
| | 10993-10), and cytotoxicity<br>(ISO 10993-5) | | | |
| Utility | Single Use | Single Use | Same | |
| Sterility (Method) | Sterile (Gamma Irradiation) | Sterile (Gamma Irradiation) | Same | |
| Environment of Use | Healthcare Facility | Healthcare Facility | Same | |
| Labeling | Prescription Use | Prescription Use | Same | |
| Shelf Life | 1 year | 3 years | Different | |
| Device Picture | Image: syringe | Image: syringes | N/A | |
Image /page/5/Picture/6 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet inside of a flower. The words "Pharmaceuticals, Inc." are written in a smaller font to the right of the "R".
{6}------------------------------------------------
*Values obtained from performance testing conducted per HR Pharmaceuticals' protocols.
# VII. PERFORMANCE DATA
There are no FDA device-specific guidance documents, special controls document, and/or requirements in a device-specific regulation for prefilled sterile water syringes. Additionally, no performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for prefilled sterile water syringes. However, the following performance data were provided in support of the substantial equivalence determination.
Bench Testing
- Small-bore connector testing ISO 80369-7:2016. Small-bore connectors for ● liquids and gases in healthcare applications - Part 7: Connectors for intravascular or hypodermic applications [FR Recognition No. 5-115]
- Manual, single-use syringe testing - ISO 7886-1:2017, Sterile hypodermic syringes for single use - Part 1: Syringes for manual use [FR Recognition No. 6-404]
- Sterility USP 42 Chapter <71> Sterility Tests [FR Recognition No. 14-532] ●
## Biocompatibility Testing
The biocompatibility evaluation for the AquaFlate Pre-Filled Sterile Water Syringe 10mL was conducted in accordance with the FDA Final Guidance "Use of International Standard ISO-10993, 'Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process, " June 16, 2016, and International Standard ISO 10993-1:2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and
Image /page/6/Picture/13 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet inside of a circle. The words "Pharmaceuticals, Inc." are written in a smaller font to the right of the letters.
{7}------------------------------------------------
Testing Within a Risk Management Process," as recognized by FDA. The biocompatibility testing included the following and yielded acceptable results:
- Cytotoxicity ANSI/AAMI/ISO 10993-5:2009 (R) 2014 Biological Evaluation of ● Medical Devices Part 5: Tests for In Vitro Cytotoxicity [FR Recognition No. 2-245]
- Irritation ANSI/AAMI/ISO 10993-10:2013—Biological Evaluation of Medical . Devices-Part 10: Tests for Irritation and Skin Sensitization [FR Recognition No. 2-1741
- . Sensitization - ANSI/AAMI/ISO 10993-10:2013—Biological Evaluation of Medical Devices-Part 10: Tests for Irritation and Skin Sensitization [FR Recognition No. 2-174]
The AquaFlate Pre-Filled Sterile Water Syringe 10mL falls into the category of limited (<24 hour) external communicating devices and is intended for indirect tissue contact.
## Sterilization
The AquaFlate Pre-Filled Sterile Water Syringe 10mL was validated for sterilization using gamma irradiation at a minimum dose of 25kGy and a maximum dose of 40kGy to achieve a Sterility Assurance Level of 10th in accordance with the following recognized standards:
- ANSI/ AAMI/ISO 11137-1:2006/(R) 2015 & A1:2013. Sterilization of health ● care products -Radiation-Part1: Requirements for development validation, and routine control of a sterilization process for medical devices. [FR Recognition No. 14-4281
- ANSI/AAMI/ISO 11137-2:2013. Sterilization of health care products-Radiation-Part 2: Establishing the sterilization dose. [FR Recognition No. 14-4091
- ANSI/AAMI/ISO 11137-3:2006 (R) 2010. Sterilization of healthcare products -Radiation-Part 3: Guidance on dosimetric aspects. [FR Recognition No. 14-510]
- ANSI/AAMI/ISO 11737-1: 2018. Sterilization of health care products, Microbiological Methods-Part 1: Determination of a population of microorganisms on product. [FR Recognition No. 14-514]
- ANSI/AAMI/ISO 11737-2: 2009 (R) 2014. Sterilization of Medical Devices-● Microbiological Methods-Part 2: Tests of sterility Performed in the definition, validation and maintenance of a sterilization Process. [FR Recognition No. 14-327]
- AAMI TIR 17-2008. Compatibility of Materials subject to sterilization. ●
- ANSI/AAMI/ISO TIR13004: 2013. Sterilization of health care products-● Radiation-Substantiation of a selected sterilization dose-Method VDmax®
Image /page/7/Picture/17 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet inside of a circle. The text "Pharmaceuticals, Inc." is to the right of the "R" and the water droplet symbol.
{8}------------------------------------------------
#### Packaging Validation and Shelf Life
The AquaFlate Pre-Filled Sterile Water Syringe 10mL underwent packaging validation testing and accelerated aging testing per the below standards to support a one-year shelf life.
- ISTA Procedure 3A (2018) Series. General Simulation Performance Test Procedure Packaged-Products for Parcel Delivery System Shipment 150lb or less.
- ASTM D4169-16. Performance Testing of Shipping Containers and Systems ● [FR Recognition No. 14-499]
- ASTM F1980-16. Standard Guide for Accelerated Aging of Sterile Barrier ● Systems for Medical Devices [FR Recognition No. 14-497]
# VIII. CONCLUSION
All necessary testing has been performed on the AquaFlate Pre-Filled Sterile Water Syringe 10mL to support the safety and effectiveness of the device per its intended use. Additionally, the performance testing concluded that the subject device is substantially equivalent to the legally marketed predicate device.
Image /page/8/Picture/10 description: The image shows the logo for H-R Pharmaceuticals, Inc. The logo is in blue and features the letters "H" and "R" in a stylized font. To the right of the "R" is a symbol that looks like a water droplet above a plant. The words "Pharmaceuticals, Inc." are written in a smaller font to the right of the "R".
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.