← Product Code [EZL](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL) · K181346

# CystoSure Plus Catheter (K181346)

_Emmy Medical, LLC · EZL · Aug 6, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL/K181346

## Device Facts

- **Applicant:** Emmy Medical, LLC
- **Product Code:** [EZL](/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL.md)
- **Decision Date:** Aug 6, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5130
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The CystoSure® Plus catheter provides access and visualization for the female urinary bladder. The single-use CystoSure® Plus catheter provides urethral urinary catheterization and postoperative bladder irrigation/lavage with the addition of a sealed port for passage of an endoscope. It is suitable for medium- to long-term use with a maximum patient indwelling time < 30 days.

## Device Story

CystoSure® Plus is a single-use, silicone, retention-type balloon catheter for female urinary bladder access and visualization. Device features a sealed port allowing endoscope passage for internal bladder surface inspection; also facilitates urinary drainage and postoperative irrigation/lavage. Operated by clinicians in surgical or diagnostic settings. Clinician inserts catheter into urethra; endoscope is passed through the dedicated port to visualize bladder. Output is visual information for the clinician to assess bladder condition. Benefits include improved patient comfort via increased length (moving ports away from thighs) and easier insertion due to reduced shaft outer diameter and increased material stiffness. Device is intended for medium- to long-term indwelling use (< 30 days).

## Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence is based on design comparison, material specifications, and performance testing confirming the device meets all criteria of the predicate.

## Technological Characteristics

Silicone catheter with balloon; 16 Fr shaft OD; 263 mm length. Materials: silicone resin with inert filler for durometer adjustment (70A). Features a sealed endoscope port. Sterilization: Ethylene oxide. Connectivity: None. Software: None.

## Regulatory Identification

A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).

## Predicate Devices

- CystoSure Urinary Access System ([K142194](/device/K142194.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 6, 2018

Emmy Medical, LLC. % Christine Santagate Director, Northeast Regional Operations R&O Solutions, LLC. 15 Standish Road Norfolk, MA 02056

Re: K181346

> Trade/Device Name: CystoSure® Plus Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZL Dated: July 5, 2018 Received: July 9, 2018

Dear Christine Santagate:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Glenn B. Bell -S

for

Benjamin R. Fisher. Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K181346

Device Name CystoSure® Plus Catheter

Indications for Use (Describe)

The CystoSure® Plus catheter provides access and visualization for the female urinary bladder.

The single-use CystoSure® Plus catheter provides urethral urinary catheterization and postoperative bladder inrigation/ lavage with the addition of a sealed port for passage of an endoscope. It is suitable for medium- to long-term use with a maximum patient indwelling time < 30 days.

| Type of Use (Select one or both, as applicable)                                |                                                                              |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span>     Prescription Use (Part 21 CFR 801 Subpart D)   </span> </div> | <div> <span>    Over-The-Counter Use (21 CFR 801 Subpart C)   </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## SECTION 8.0 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Submitter Information

| Submitter's Name:  | Ronald Adams                           |
|--------------------|----------------------------------------|
| Address:           | 18 Hillside Drive, Holliston, MA 01746 |
| Telephone:         | 508-944-5166                           |
| Fax:               | 844-225-4600                           |
| Contact Person:    | Ronald Adams                           |
| Telephone :        | 508-944-5166                           |
| Fax :              | 844-225-4600                           |
|                    | May 1, 2018                            |
| Prepared:          |                                        |
| ce Trade Name:     | CystoSure® Plus Catheter               |
| mon/Usual Name:    | Catheter, Retention Type, Balloon      |
| latory Information |                                        |
| lation Name:       | Urological catheter and accessories    |
|                    | II                                     |
| uct Code(s):       | EZL                                    |
| lation Number(s):  | 21 CFR 876.5130                        |

Date

Devic Comn

Regul Regul Class: Produ Regul Review Panel:

Gastroenterology/Urology

Predicate Device: K142194 CystoSure Urinary Access System

{4}------------------------------------------------

#### Device Description & Comparison:

The CystoSure® and CystoSure® Plus catheters are intended for use in the diagnostic visualization of the female bladder with a port to allow for the insertion of an endoscope. They are made from silicone and they are used in both surgical and diagnostic procedures for draining the bladder and enabling the visualization of internal bladder surfaces.

The differences between Cystosure® Plus and the original Cystosure® catheter are tabulated in Table 1. Engineering drawings of the Predicate Cystosure® and Subject Cystosure® Plus catheters are provided in Attachment 1. The operational characteristics and performance criteria of the two devices are identical as both are flexible tubes used for drainage and bladder access in females.

The indications for use of the device are different by necessity, as reference to the catheter system and CystoSure® cystoscope have been removed. However, the intended use of the device remains unchanged.

#### Indications for Use:

Intended Use/Indications for Use: The CystoSure® Plus catheter provides access and visualization for the female urinary bladder.

The single-use CystoSure® Plus catheter provides urethral urinary catheterization and postoperative bladder irrigation/lavage with the addition of a sealed port for passage of an endoscope. It is suitable for medium- to long-term use with a maximum patient indwelling time < 30 days.

| Element                       | Cystosure®                        | Cystosure® Plus     | Comments                                                                                                                                                                                   |
|-------------------------------|-----------------------------------|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Shaft OD                      | 18 Fr (6.0mm)                     | 16 Fr (5.5mm)       | Shaft OD reduced for ease of insertion.                                                                                                                                                    |
| Lumen ID                      | 3.2 mm                            | 3.1 mm              | Reduction enables shaft OD decrease.                                                                                                                                                       |
| Length                        | 180 mm                            | 263 mm              | Nursing staff request to enhance patient<br>comfort – moves ports away from<br>thighs.                                                                                                     |
| Check valve color             | Red                               | Orange              | Colors are standardized depending on<br>Catheter OD. All Degania 16 Fr<br>catheters have orange check valves.                                                                              |
| Shaft material &<br>durometer | 8204503, 65A                      | 8202504, 70A        | Increased stiffness for ease of insertion.<br>*See material note below.                                                                                                                    |
| Balloon material              | 8202800 or<br>8204498<br>silicone | 8202504<br>silicone | Silicone durometer increased to support<br>shaft OD reduction.  Formulation<br>8202504 is substantially equivalent to<br>8202503 shaft material of Cystosure.<br>*See material note below. |
| Balloon port label            | N/A                               | 5cc only            | Improved recognition of recommended<br>volume.                                                                                                                                             |

### Table 1. Catheter Comparison

{5}------------------------------------------------

| Scope Port Glue  | N/A               | 8015412           | Added glue improves port attachment<br>strength and is not patient contacting. |
|------------------|-------------------|-------------------|--------------------------------------------------------------------------------|
| Sterilization    | Ethylene oxide    | Ethylene oxide    | Identical facility, chamber and cycle.                                         |
| Biocompatibility | Silicone          | Silicone          | Identical                                                                      |
| Expiration Date  | 3 Year Shelf Life | 3 Year Shelf Life | Identical                                                                      |
| Packaging        | Pouch             | Pouch             | Identical                                                                      |

*Durometer of silicone catheters is adjusted by changing the % of additional inert filler material to silicone resin. These inert filler materials do not affect the chemical formulation and therefore there is no need to repeat biocompatibility testing.

The above changes to the CystoSure Plus catheter from the predicate CystoSure Urinary Access System catheter enable the catheter to be inserted more the catheter ports away from the patient's thighs. The CystoSure Plus catheter meets all performance criteria of the predicate CystoSure catheter.

## Conclusion:

The Emmy Medical CystoSure® Urinary Access System and Cystosure® Plus catheters are substantially equivalent in materials, performance and safety characteristics.

---

**Source:** [https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL/K181346](https://fda.innolitics.com/submissions/GU/subpart-f%E2%80%94therapeutic-devices/EZL/K181346)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
