The 2-Way Kohli urinary drainage catheter is intended for urological bladder drainage only with a maximum patient indwelling time of <30 days.
Device Story
The Kohli Urinary Drainage Catheter is a single-use, 2-way silicone catheter designed for urethral insertion to facilitate bladder drainage. It features two ports: one for bladder drainage and one for balloon inflation (5cc capacity). The device is inserted by a healthcare professional in a clinical setting. It functions by anchoring in the bladder via the inflated balloon, allowing urine to flow through the drainage lumen. The catheter is constructed from silicone with a polypropylene check valve and a co-extruded radiopaque strip for visualization. It does not contain antimicrobial or hydrophilic coatings. The device provides a passive drainage pathway, assisting patients who cannot void naturally. Clinical benefit is derived from reliable, short-term bladder decompression and management of urinary retention.
Clinical Evidence
Bench testing only. Performance testing included inflation lumen leakage, catheter strength, connector and balloon security, flow rate, balloon volume maintenance/recovery, and resistance to traction. Biocompatibility testing was conducted per ISO 10993. Sterilization validation was performed using an Ethylene Oxide (ETO) cycle identical to the predicate devices.
Technological Characteristics
Materials: Silicone (shaft, funnel, balloon), Silicone RTV (adhesive), Polypropylene (check valve), Co-extruded silicone/barium sulfate (radiopaque strip). Dimensions: 16 FR shaft, 450 mm length, 3.4 mm drainage lumen ID, 1.0 mm inflation lumen ID. Sterilization: Ethylene Oxide (ETO). Design: 2-way lumen, 5cc balloon, no coatings.
Indications for Use
Indicated for urological bladder drainage in patients requiring an indwelling catheter for up to 30 days.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Degania Silicone All Silicone Foley Catheter, 2-Way, 3-Way and with Temperature Sensor (K063442)
Emmy Medical, LLC, Cystosure Urinary Access System (K142194)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 9, 2018
Nellie Medical, LLC % Christine Santagate Director, Boston Operations R&O Solutions 15 Standish Road Norfolk, MA 02056
Re: K172422
Trade/Device Name: Kohli Urinary Drainage Catheter Regulation Number: 21 CFR& 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZL Dated: December 20, 2017 Received: December 27, 2017
Dear Christine Santagate:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Charles Viviano -S
For Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172422
Device Name Kohli Urinary Drainage Catheter
Indications for Use (Describe)
The 2-Way Kohli urinary drainage catheter is intended for urological bladder drainage only with a maximum patient indwelling time of <30 days.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(K) SUMMARY
# Submitter Information
| Submitter's Name: | Ron Adams |
|-------------------------|----------------------------------------|
| Address: | 18 Hillside Drive, Holliston, MA 01746 |
| Telephone: | 775-800-7300 |
| Fax: | 844-225-4600 |
| Contact Person: | Ron Adams |
| Telephone: | 775-800-7300 |
| Fax: | 844-225-4600 |
| Date Prepared: | December 13, 2017 |
| Trade Name: | Kohli Urinary Drainage Catheter |
| Common/Usual Name: | Urinary drainage catheter |
| Device Name: | Catheter, retention type, balloon |
| Registration Number(s): | 21 CFR 876.5130 |
| Regulation Description: | urological catheter and accessories |
| Class: | II |
| Product Code(s): | EZL |
### Predicate Device(s):
- Primary: K063442 Degania Silicone All Silicone Foley Catheter, 2-Way, 3-Way and with . Temperature Sensor
- . Secondary: K142194 Emmy Medical, LLC, Cystosure Urinary Access System
# Device Description:
The Kohli Urinary Drainage Catheter is intended for drainage of the urinary bladder.
The Kohli Urinary Drainage Catheter provides a single use access catheter with two ports: One for bladder drainage, and one for balloon inflation. The catheter encompasses only one balloon size (5cc) and length and it does not include any hydrophilic or antimicrobial coating
# Indications for Use:
The 2-Way Kohli Urinary Drainage Catheter is intended for urological bladder drainage only with a maximum patient indwelling time <30 days.
{5}------------------------------------------------
# Substantial Equivalence
The below table demonstrates that the Kohli subject device is substantially equivalent to the predicate devices.
| Table 14-2<br>Substantial Equivalence Comparison Chart - Kohli Urinary Drainage Catheter | | | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Feature/<br>Specification | Proposed<br>Device | Comparison | Primary Predicate | Secondary<br>Predicate | |
| Manufacturer | Nellie<br>Medical | N/A | Degania Silicone, Ltd | Emmy Medical | |
| Device Trade<br>Name | Kohli<br>Urinary<br>Drainage<br>Catheter | N/A | All Silicone Foley Catheter,<br>2-Way, 3 Way and with<br>Temperature Sensor | CystoSure Urinary<br>Access | |
| General<br>Description | Urological<br>catheter | Identical | Urological catheter and<br>accessories | Urological<br>catheter | |
| 510(k) | N/A | N/A | K063442 | K142194 | |
| Product Code | EZL | Identical | EZL | EZL | |
| Device Class | II | Identical | II | II | |
| Regulation<br>Number | 21 CFR<br>876.5130 | Identical | 21 CFR 876.5130 | 21 CFR 876.5130 | |
| Intended Use | Bladder<br>drainage | Identical to<br>2-Way<br>drainage<br>catheters | Bladder irrigation and<br>drainage | Bladder irrigation<br>and drainage | |
| Indications for<br>Use | The 2-Way<br>Kohli<br>Urinary<br>Drainage<br>Catheter is<br>intended for<br>urological<br>bladder<br>drainage only<br>with a<br>maximum<br>patient<br>indwelling<br>time <30<br>days. | Identical to<br>2-Way<br>drainage<br>catheters | All Silicone Foley Catheter<br>intended for urological use<br>only. Foley Catheter 2-way:<br>for routine drainage of the<br>urinary bladder.<br>Foley Catheter 3-way: for<br>drainage of the urinary<br>bladder and bladder<br>irrigation. Foley Catheter<br>with Temperature Sensor:<br>for drainage of the urinary<br>bladder and simultaneous<br>monitoring of temperature. | The CystoSure<br>Access Catheter<br>is used to provide<br>drainage of urine<br>and irrigation<br>fluids for the<br>female bladder<br>and to provide a<br>passageway for<br>the CystoSure<br>cystoscope. | |
| Single Use | Yes | Yes | Yes | Yes | |
| Lumen | 2-way | 2-way | 2 and 3-way | 4-way | |
| Materials | | | | | |
| Shaft | Silicone* | Identical | Silicone* | Silicone* | |
| Funnel | Silicone* | Identical | Silicone* | Silicone* | |
| Table 14-2<br>Substantial Equivalence Comparison Chart – Kohli Urinary Drainage Catheter | | | | | |
| Feature/<br>Specification | Proposed<br>Device | Comparison | Primary Predicate | Secondary<br>Predicate | |
| Balloon | Silicone* | Identical | Silicone* | Silicone* | |
| Balloon Adhesive | Silicone RTV* | Identical | Silicone RTV* | Silicone RTV* | |
| Check Valve | polypropylen<br>e | Identical | Polypropylene | polypropylene | |
| Radiopaque strip | Co-Extruded<br>silicone &<br>barium<br>sulfate | Identical | Co-Extruded silicone & barium<br>sulfate | N/A | |
| Occlusion | Glue | Identical | Glue | Glue | |
| Surface<br>Modifications -<br>Pad Printing | black ink** | Identical | black ink** | black ink** | |
| Coatings | None | Identical | None | None | |
| Dimensions | | | | | |
| Shaft | 16 FR (5.3<br>mm) | Identical | 12-22 Fr (4.0-7.3mm) | 18 FR (6.0 mm) | |
| Length | 450 mm | Identical | 400 to 450 mm | 175 mm | |
| Drainage Lumen<br>(ID) | 3.4 mm | Identical | 2.5 to 3.7 mm | 3.2 mm | |
| Balloon Inflation<br>Lumen (ID) | 1.0 mm | Identical | 0.5 - 2.0 mm | 2.0 mm | |
| Manufacturing Process | | | | | |
| Extrusion | Shaft<br>formation | Identical | Shaft formation | Shaft formation | |
| Injection Molding | Funnel<br>forming | Identical | Funnel forming | Funnel forming | |
| Balloon<br>attachment | RTV glue | Identical | RTV glue | RTV glue | |
| Manual assembly | Check valve | Identical | Check valve | Check valve | |
| Surgical Approach | Urethral<br>Insertion | Identical | Urethral Insertion | Urethral Insertion | |
| Primary Material | Silicone | Identical | Silicone | Silicone | |
| Sterile | Yes | Yes | Yes | Yes | |
| Sterilization<br>Method | Ethylene<br>Oxide | Identical | Ethylene Oxide | Ethylene Oxide | |
| Table 14-2 | | | | | |
| Substantial Equivalence Comparison Chart – Kohli Urinary Drainage Catheter | | | | | |
| Feature/<br>Specification | Proposed<br>Device | Comparison | Primary Predicate | Secondary<br>Predicate | |
| Biocompatibility | Materials<br>tested per<br>ISO 10993 | Materials<br>tested per<br>ISO 10993 | Materials tested per ISO<br>10993 | Materials tested per<br>ISO 10993 | |
| *Both VMQ and RTV silicones are identical to those of the predicates. Detailed VMQ silicone<br>formulations and silicone RTV are proprietary to manufacturer. | | | | | |
| ** Proprietary formulation of the black ink was developed by Degania Silicone and is included as part<br>of the biocompatibility testing assessments. | | | | | |
{6}------------------------------------------------
{7}------------------------------------------------
## Non-Clinical Performance testing
The bench testing performed verifies that the performance of the substantially equivalent in terms of critical performance characteristics to the predicate device. Testing included inflation lumen leakage, catheter strength, connector and balloon security, flow rate, balloon volume maintenance/recovery and integrity, and resistance to traction.
## Biocompatibility
A complete battery of biocompatibility tests was conducted on silicone catheters using the same materials and same processes as the Kohli catheter. The results of the tests demonstrate that the subject device is biocompatible.
### Sterilization
The ETO sterilization process is identical to the cycle employed for both predicate devices. A parametric validation was performed on silicone catheters whose length and lumen diameter challenge far exceed those of the proposed Kohli catheter. Both the proposed device and the predicate can be sterilized using the same cycle.
Based on the performance testing, biocompatibility evaluation, the Kohli Urinary Drainage Catheter is substantially equivalent to the primary predicate device for its intended use.
### Conclusion:
The Kohli Urinary Drainage Catheter is substantially equivalent in design, materials, intended use, and manufacturing processes to the predicate devices. The geometric and mass properties are the same. Based on the results of the testing and the comparability to the predicate devices, we believe that the Kohli urinary drainage catheter does not present new concerns of safety or efficacy and is thus substantially equivalent to the legally marketed predicate devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.