K130908 · Guangdong Baihe Medical Technology Co., Ltd. · EZL · Oct 3, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K130908
Device Name
DISPOSABLE SILICONE FOLEY CATHETER
Applicant
Guangdong Baihe Medical Technology Co., Ltd.
Product Code
EZL · Gastroenterology, Urology
Decision Date
Oct 3, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Two-way Disposable Silicone Foley Catheter: Urethral catheterization for bladder drainage for urological use only; the indwell time of the proposed device is no more than 30 days. Three-way Disposable Silicone Foley Catheter: Urethral catheterization for bladder drainage and bladder irrigation for urological use only; the indwell time of the proposed device is no more than 30 days.
Device Story
Disposable Silicone Foley Catheter is a sterile, single-use urological device available in two-way and three-way configurations. Two-way version consists of a double-lumen tube, hub, balloon, and tip; three-way version includes a triple-lumen tube. Sizes range from 6Fr to 26Fr, with lengths of 310mm or 400mm. Used by clinicians for bladder drainage or irrigation. Device is inserted into the urethra; balloon is inflated to retain the catheter in the bladder. Provides mechanical drainage/irrigation path. Benefits include patient bladder management for up to 30 days.
Clinical Evidence
Bench testing only. Performance verified against ASTM F623-99 (flow rate, balloon integrity, pullout response, volume maintenance, dimensions) and biocompatibility standards (ISO 10993-5, -6, -10, -11).
Technological Characteristics
Silicone construction. Two-way (double lumen) or three-way (triple lumen) configuration. EO sterilized. Sizes 6-26Fr. Complies with ASTM F88-09, F1140-07, F1929-98, F623-99, and ISO 11135-1:2007.
Indications for Use
Indicated for pediatric, male, and female patients requiring urethral catheterization for bladder drainage (two-way and three-way) or bladder irrigation (three-way only). Indwell time limited to 30 days.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Rochester Medical Corporation All Silicone Foley Catheter (K981612)
Submission Summary (Full Text)
{0}------------------------------------------------
## E4 510(k) Summary
## Exhibit #4 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number:
- 1. Date of Submission: March 14, 2013
**OCT 03 2013**
ni Sponsor
> Guangdong Baihe Medical Technology Co., Ltd. No. 89, Taoyuan East Road, Nanhai, Foshan, Guangdong, 528225, China
Establishment Registration Number: 3006621386
Contact Person: Baoyou Ge Position: Management Representative Tel: +86-757-81207300 Fax: +86-757-81207311 Email: gby8492@163.com
3. Submission Correspondent
Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
{1}------------------------------------------------
#### Proposed Device Identification 4.
Common Name: Disposable Balloon-retention Catheter Proposed Device Name: Disposable Silicone Foley Catheter Proposed Device Type: Two-way and Three way
Regulatory Information: Classification Name: Catheter, Retention type, Balloon Classification: II Product Code: EZL Regulation Number: 21 CFR 876.5130 Review Panel: Gastroenterology/Urology Intended Use Statement:
Two-way Disposable Silicone Foley Catheter: Urethral catheterization for bladder drainage for urological use only; the indivell time of the proposed device is no more than 30 days. Three-way Disposable Silicone Foley Catheter: Urethral catheterization for bladder drainage and bladder irrigation for urological use only; the indwell time of the proposed device is no more than 30 days.
#### ડ. Predicate Device Identification
510(k) Number: K981612 Product Name: Rochester Medical Corporation All Silicone Foley Catheter Manufacturer: Rochester Medical Corporation
#### Device Description 6.
·
The Disposable Silicone Foley Catheter is available in two types, two-way disposable silicone Foley catheter and three-way disposable silicone Foley catheter.
The two-way disposable silicone Foley catheter is composed of a double lumen tube, a double lumen hub, one balloon and a tip.
The three-way disposable silicone Foley catheter is composed of a triple lumen tube, a triple lumen hub, one balloon and a tip.
The proposed device is supplied in French size ranging from 6 to 26. It is available in 310mm and 400mm length with various balloon sizes. The devices are applied for male and female; the 6Fr, 8Fr and 10Fr catheters are for pediatric, and others are for adult.
{2}------------------------------------------------
The Disposable Silicone Foley Catheter is provided EO sterilized as a single used device.
### 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ASTM F88-09. Standard Test Method for Seal Strength of Flexible Barrier Materials
ASTM F 1140-07 Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Package for Medical Applications
ASTM F1929-98(2004) Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration
ISO 11135-1:2007 Sterilization of health care products- Ethylene oxide- Part 1: Requirements for development, validation and routing control of a sterilization process for medical devices:
USP <85> Bacterial Endotoxin Limit
ASTM F 623-99 (2006) Standard Performance Specification for Foley Catheter
ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity
ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization
ISO 10993-11:2006, Biological evaluation of medical devices Part 11: Tests for systemic toxicity
ISO 10993-6:2007, Biological evaluation of medical devices- Part 6: Tests for local effects after implantation
The Disposable Silicone Foley Catheter meets the following performance requirements per testing conducted according to ASTM F623-99 (2006):
Flow Rate
Balloon Integrity
Balloon Response to Pullout
Balloon Volume Maintenance
Manufacturing Tolerances for Catheter Tip, Balloon and Shaft Diameters
Balloon Deflation Reliability
{3}------------------------------------------------
#### Substantially Equivalent Conclusion 8 .
| ITEM | Proposed Device | Predicate Device<br>K981612 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Product Code | EZL | EZL |
| Regulation No. | 21 CRF 876.5130 | 21 CRF 876.5130 |
| Class | II | II |
| Intended Use | Same | Same |
| Target<br>population | Pediatric, male and female | Same |
| Lumen | Two-way and three-way | Same |
| French size | Two-way: 6Fr, 8Fr, 10Fr, 12Fr, 14Fr, 16Fr, 18Fr, 20Fr, 22Fr, 24Fr and 26Fr<br>Three-way: 16Fr, 18Fr, 20Fr, 22Fr, 24Fr and 26Fr | Same |
| Balloon size | 1.5cc, 3cc, 5cc, 10cc, 15cc, 20cc, 30cc | Similar |
| Length | 310mm and 400mm | Similar |
| Sterile | Yes | Yes |
| Single use | Yes | Yes |
| | Tab. 3-1 Comparison between proposed device and predicated device | | | |
|--|-------------------------------------------------------------------|--|--|--|
| | | | | |
The proposed device, Disposable Silicone Foley Catheter, is determined to be Substantially Equivalent (SE) to the predicate device, Rochester Medical Corporation All Silicone Foley Catheter (K981612), in respect of safety and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a seal for the Department of Health & Human Services USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized emblem, possibly representing a symbol of health or human services. The image is in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 3, 2013
Guangdong Baihe Medical Technology Co., Ltd. % Diana Hong General Manager MID-LINK Consulting Co., Ltd. P.O. Box 120-119 Shanghai China 200120
Re: K130908
Trade/Device Name: Disposable Silicone Foley Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: August 29, 2013 Received: September 4, 2013
Dear Diana Hong.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be fourd in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{5}------------------------------------------------
Page 2 - Diana Hong
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Exhibit #3 Indications for Use
510(k) Number: K130908 Device Name: Disposable Silicone Foley Catheter
Indications for Use:
Two-way Disposable Silicone Foley Catheter. Urethral catheterization for bladder drainage for urological use only; the indwell time of the proposed device is no nore than 30 days.
Three-way Disposable Silicone Foley Catheter: Urethral catheterization for bladder drainage and bladder irrigation for urological use only; the indwell time of the proposed device is no more than 30 days.
OPRESCRIPTION USE (Part 21 CFR 801 Subpart D) OOVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
## Herbert P. Lerner -S 2013.10.03 16:31:36 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.