K082919 · Uc-Care Medical Systems, Ltd. · EZL · Oct 27, 2008 · Gastroenterology, Urology
Device Facts
Record ID
K082919
Device Name
UC-CARE POSITIONING SYSTEM
Applicant
Uc-Care Medical Systems, Ltd.
Product Code
EZL · Gastroenterology, Urology
Decision Date
Oct 27, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Indications for Use
The UC-CARE Positioning System is a mechanical positioning device, introduced into the urethra as a sterile urological catheter for visualizing the urethral course, aimed for assistance in diagnostic or therapeutic endorectal ultrasound-guided prostate procedures. The device is retained in place by inflating the balloon tip and by attaching it over an ultrasound-stepper and stabilizer with a Brachy-like template.
Device Story
The UC-CARE Positioning System is a mechanical device used during transperineal prostate procedures. It consists of a urethral catheter and a catheter holder. The catheter is inserted into the urethra and secured via an inflatable balloon tip; the holder attaches to existing commercial endorectal ultrasound steppers, stabilizers, and brachytherapy template grids. By outlining the urethral course, the device provides visual guidance to the physician during needle insertion for biopsies, brachytherapy, or cryotherapy. It functions as a positioning and imaging-visibility aid, helping to ensure safe needle placement relative to the urethra. The device is intended for use in clinical settings by healthcare professionals performing prostate interventions.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Mechanical positioning device comprising a Foley-style urological catheter and a catheter holder. Materials are identical to commercially available FDA-cleared urological catheters. Operates via physical retention (balloon inflation) and mechanical attachment to standard ultrasound steppers/stabilizers. Non-powered, manual device.
Indications for Use
Indicated for patients undergoing diagnostic or therapeutic endorectal ultrasound-guided prostate procedures (e.g., biopsy, brachytherapy, cryotherapy) to assist in visualizing the urethral course and facilitating needle insertion.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo consists of the text "UC-CARE" in a bold, sans-serif font, with the word "Medical Systems" in a smaller font below it. Above the text are five oval shapes, each with a different fill color. From left to right, the ovals are black, black, black, gray, and white.
Page 1 of 2
## 510(K) SUMMARY
### UC-CARE Positioning System
510(k) Number K 1882919
OCT 2 7 2008
- Applicant's Name: UC-CARE, Ltd. New Ind. Zone, Apollo Bldg. P.O.Box 67 Yokneam 20692 Tel: (972)4-909-7427 Fax: (972)4-993-7323
- Contact Person: Roni Zvuloni, Ph.D. CTO UC-CARE, Ltd. New Ind. Zone, Apollo Bldg., P.O.Box 67, Yokneam 20692 Tel: (972)4-909-7427, Fax: (972)4-993-7323
UC-CARE Positioning System (UCP System) Trade Name:
> Classification: Name: Urological Catheter and Accessories Product Code: EZL Regulation No: CFR 21 876.5130 Class: II Panel: Urological Catheter and Accessories
Device Description: The UCP System is a positioning device used to outline the urethral course during diagnostic or therapeutic transperineal ultrasound-guided prostate procedures.
The UCP System combines the function of a positioning and imaging-visibility enhancing device with the configuration of a urological Foley catheter. The configuration and material composition of the patient contact portion of the UCP System part are identical to a commercially available (FDA cleared) urological catheter including a balloon retention mechanism. The device is
2-2
UCP System - 510k Notification
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The logo consists of the text "UC-CARE" in a bold, sans-serif font, with the word "Medical Systems" in a smaller font below it. Above the text are five shaded circles that go from dark to light. The logo is simple and professional, and it is likely used to represent a medical organization or company.
K082919
Page 2 of 2
comprised of two main components: the Urethral Catheter and the Catheter Holder. The device is used together with any commercial endorectal ultrasound stepper and stabilizer as well as brachy-like template grid, available in the market for a transperineal prostate biopsy, brachytherapy, or cryotherapy. The use of the device facilitates simple and safe transperineal needle insertion for diagnostic and therapeutic prostate procedures.
The UCP System's Catheter is introduced into the urethra as a sterile urological catheter and is retained in place by inflating the balloon tip and by the Catheter Holder, which is mounted over the ultrasound stepper and stabilizer.
#### Intended Use Statement:
The UC-CARE Positioning System is a mechanical positioning device, introduced into the urethra as a sterile urological catheter for visualizing the urethral course, aimed for assistance in diagnostic or therapeutic endorectal ultrasound-guided prostate procedures.
The device is retained in place by inflating the balloon tip and by attaching it over an ultrasound-stepper and stabilizer with a Brachy-like template.
Predicate Devices: Substantial equivalence to the following predicate devices is claimed:
| Device Name | 510k No | Date of approval |
|---------------------------------------------------------|---------|------------------|
| BeamCath | K042110 | 25 Feb, 2005 |
| Disposable Endocavity Ultrasound needle/Biopsy<br>Guide | K970514 | 20 Jun, 1997 |
| Simplastic™ Catheter | K963993 | 17 Dec, 1996 |
#### Conclusion:
UC-CARE Ltd. believes that, based on the information provided in this submission, the UCP System is substantially equivalent to its predicate devices without raising any new safety and/or effectiveness concerns.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with three lines representing its wings or feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# OCT 2 7 2008
UC-CARE Medical Systems Ltd. % Mr. Jonathan S. Kahan Regulatory Counsel Hogan & Hartson, LLP Columbia Square, 555 Thirteenth Street, N.W. WASHINGTON DC 20004
Re: K082919
Trade/Device Name: UC-CARE Positioning System Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZL Dated: September 30, 2008 Received: September 30, 2008
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely vours.
loque M. Whang
Jovce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for UC-CARE Medical Systems. The words "UC-CARE" are in large, bold, black letters. Below that, the words "Medical Systems" are in smaller, black letters. Above the words, there are five circles that fade from black to white.
K082919
| 510(k) Number (if known): | K082919 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | UC-CARE Positioning System |
| Indications for Use: | The UC-CARE Positioning System is a mechanical positioning<br>device, introduced into the urethra as a sterile urological catheter<br>for visualizing the urethral course, aimed for assistance in<br>diagnostic or therapeutic endorectal ultrasound-guided prostate<br>procedures. |
INDICATIONS FOR USE STATEMENT
The device is retained in place by inflating the balloon tip and by attaching it over an ultrasound-stepper and stabilizer with a Brachy-like template.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-off) Division of General & Plastic Surgery Devices 510(k) Number
(Division Sign Off)
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 16 LU 510(k) Number
1-2
UCP System - 510k Notification
\\DC - 023299/000001 - 2788357 v l
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.