This intermittent catheter is a flexible tubular device that is inserted through the urethra by female and female pediatric patients who need to drain urine from the bladder.
Device Story
The Female IC is a sterile, single-use, hydrophilic-coated intermittent catheter designed for urinary incontinence management. It features drainage eyelets and is constructed from thermoplastic elastomer (TPE). The device is packaged in a polypropylene/polyethylene case with a 'first use' indicator to verify if the package was previously opened. It is intended for use by female and female pediatric patients in hospital, home, or public settings. The catheter is inserted through the urethra to drain urine from the bladder. The device utilizes direct hydration for the hydrophilic coating, differing from the water-vapor hydration of the predicate. It is sterilized via E-Beam radiation. The device provides a discrete, easy-to-handle solution for patients to perform self-catheterization, facilitating bladder emptying and improving quality of life for those with urinary retention or incontinence.
Clinical Evidence
Bench testing only. Testing included physical performance properties per BS EN ISO 20696:2018, comprehensive biocompatibility testing per ISO 10993 series (genotoxicity, cytotoxicity, irritation, sensitization, systemic toxicity, chemical characterization), sterilization validation per ISO 11137 and ISO 11737, packaging integrity, and transportation testing.
Technological Characteristics
Materials: Thermoplastic Elastomer (TPE), phthalate-free, PVC-free. Design: Hydrophilic-coated, 5.5-inch length, 2 smooth eyelets, color-coded funnel. Sterilization: E-Beam radiation. Standards: BS EN ISO 20696:2018 (catheter performance), ISO 10993 (biocompatibility), ISO 11137/11737 (sterilization).
Indications for Use
Indicated for female and female pediatric patients requiring urinary bladder drainage via intermittent catheterization.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 13, 2022
Hollister Incorporated Michelle Schiltz-Taing Regulatory Affairs Manager 2000 Hollister Drive Libertyville, IL 60048
Re: K213575
Trade/Device Name: Female IC (not Finalized) Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZD Dated: August 9, 2022 Received: August 9, 2022
Dear Michelle Schiltz-Taing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K213575
#### Device Name
Female IC (not finalized)
#### Indications for Use (Describe)
This internittent catheter is a flexible tubular device that is inserted through the urethra by female pediatic patients who need to drain urine from the bladder.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> </span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
# 510(k) Summary
| Applicant: | Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60048 |
|----------------------|------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Michelle Schiltz-Taing<br>Hollister Incorporated<br>2000 Hollister Drive<br>Libertyville, IL 60018<br>(t) 224-864-0431 |
| Date Prepared: | 13 September 2022 |
| Trade Name: | Female IC (Not Finalized) |
| Common Name: | Catheter, Urethral |
| Product Code/Class: | EZD/Class II |
| Classification Name: | Catheter, Straight |
| Regulation Number: | 21 CFR 876.5130 |
#### Predicate Device:
Infyna Chic K191633 by Hollister Incorporated. The predicate has not been subject to a design-related recall.
#### Indications for Use:
This intermittent catheter is a flexible tubular device that is inserted through the urethra by female and female pediatric patients who need to drain urine from the bladder.
#### Description of Applicant Device:
The Female IC (not finalized) is a sterile, hydrophilic-coated, single use catheter with drainage eyelets that is used to manage urinary incontinence. The Female IC is inserted into the urethra to drain urine from the bladder. This directly hydrated catheter is packaged in a polypropylene/polyethylene case which was designed to be easy to open, close and dispose of, with a discrete consumer design. A "first use label" on the case indicates to the end user if the product was previously opened.
#### Technological Characteristics:
The table below summarizes the technological characteristics of the Female IC as compared to the predicate.
{4}------------------------------------------------
| | Predicate Device<br>Infyna Chic<br>(K191633) | Female IC | Same or Different/Rationale<br>for No impact to Efficacy or Safety |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | This intermittent catheter is a<br>flexible tubular device that is<br>inserted through the urethra by<br>female patients who need to<br>drain urine from the bladder. | This intermittent catheter is a<br>flexible tubular device that is<br>inserted through the urethra by<br>female and female pediatric<br>patients who need to drain<br>urine from the bladder | No change-additional<br>verbiage was used to clarify<br>that "female" includes female<br>pediatric patients. No impact<br>to the existing Female IC risk<br>file was identified and no<br>additional safety concerns<br>were added due to the<br>clarification of the patient<br>population. |
| Condition of Use | Single Use | | No Change |
| Pre-lubricated | Water vapor hydration | Direct Hydration | Although a different method<br>of hydration is utilized<br>between the predicate and<br>subject device; equivalency of<br>catheter lubricity is<br>demonstrated within this<br>submission |
| Ready to use | Yes | | No Change |
| End Design | Color coded Funnel | | No Change |
| Sterile | Gamma Radiation | E-Beam Radiation | Gamma and E-Beam<br>sterilization methods are both<br>forms of radiation<br>sterilization. The required<br>Sterility Assurance Level<br>(SAL) has been validated for<br>both methods. |
| Hydrophilic Coating | PVP Based (polyvinylpyrrolidone) Coating | | No Change |
| Catheter Eyelets | 2 smooth catheter eyelets | | No Change |
| Not Made with<br>Natural Rubber Latex | Yes | | No Change |
| Catheter Material | Polyvinyl Chloride (PVC);<br>phthalate free | Thermo-plastic Elastomer<br>(TPE); phthalate free, PVC<br>free) | Catheter material changes has<br>no impact as both the<br>predicate and subject device<br>conform to BS EN 20696. |
| | | | "Sterile urethral catheters for<br>single use". Biocompatibility<br>has been assessed and all<br>requirements have been met |
| Catheter Length | 5 inches | 5.5 inches | The subject device has a<br>length of 0.5 inches more<br>than the predicate. No impact<br>as both the predicate and<br>subject device conform to BS<br>EN 20696- "Sterile urethral<br>catheters for single use"<br>effective shaft length. |
| Fr Size | Fr 8, 10, 12, 14 | | No Change |
| Environment of Use | Hospital<br>Home Setting<br>Public Places | | No Change |
{5}------------------------------------------------
## Brief Description of Non-Clinical Testing:
The physical performance properties of the Intermittent Catheter met all applicable requirements of BS EN ISO 20696:2018, Sterile urethral catheters for single Use.
Biocompatibility testing met the requirements of the following standards:
- . ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
- ISO 10993-3:2014. Biological evaluation of medical devices Part 3:Tests for genotoxicity. ● carcinogenicity, and reproductive toxicity.
- 10993-5:2009, Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.
- 10993-10:2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- 10993-11:2017, Biological evaluation of medical devices Part 11: Tests for systemic toxicity ●
- 10993-12:2021, Biological evaluation of medical devices Part 12: Sample preparation and ● reference materials.
- 10993-17:2002, Biological evaluation of medical devices – Part 17: Establishment of allowable limits for leachable substances.
- 10993-18:2020, Biological evaluation of medical devices - Part 18: Chemical characterization of medical device materials within a risk management process.
Sterilization met all requirements of the following FDA-recognized standards:
- ISO 11137-1: 2006, Sterilization of health care products-Radiation Part 1: Requirements for ● development, validation and routine control of a sterilization process for medical devices.
- ISO 11137-2:2013, Sterilization of health care products Radiation Part 2: Establishing the ● sterilization dose.
- ANSI/AAMI/ISO 11737-1:2018, Sterilization of health care products Microbiological methods ● - Part 1: Determination of a population of microorganisms on products.
{6}------------------------------------------------
- ANSI/AAMI/ISO 11737-2:2019, Sterilization of medical devices Microbiological methods -● Part 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process.
Packaging integrity testing was conducted to verify the maintenance of the sterile barrier through shelf life. Transportation testing was conducted in order to verify that there is no impact to the device safety or efficacy of the catheter performance due to the hazards associated with the transportation environment.
### Conclusion:
It is concluded that the information supplied in this submission has demonstrated that the Female IC is substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.