K080007 · Icu Medical, Inc. · EZD · May 13, 2008 · Gastroenterology, Urology
Device Facts
Record ID
K080007
Device Name
GRADUATED ADAPTER WITH CLAVE
Applicant
Icu Medical, Inc.
Product Code
EZD · Gastroenterology, Urology
Decision Date
May 13, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Indications for Use
ICU Medical Graduated Adapter with Clave® is indicated as an accessory to a urinary catheter which facilitates the transfer of fluids into or out of the bladder via connection to the graduated adapter. This device for single procedure use only. Connection to catheter not to exceed 30 days.
Device Story
Graduated Adapter with Clave is a sterile, single-use accessory for urinary catheters. Device consists of a Clave needless connector solvent-bonded to a graduated adapter. Physician manually inserts graduated adapter into open end of urinary catheter; connects syringe or ISO 594-1/-2 compliant luer to Clave. Enables infusion or withdrawal of fluids from bladder. Clave connector provides closed system; withstands back pressures up to 45 psi to prevent leakage when syringe/luer is disconnected. Used in clinical settings by physicians to simplify patient care and facilitate bladder procedures. Benefits include ease of use for caregivers and secure fluid management.
Clinical Evidence
Bench testing only. Device performance validated against international standards for connecting components. Biocompatibility confirmed per ISO 10993-1. Sterilization validated per ISO 11137-1 to achieve a Sterility Assurance Level (SAL) of 10^-6. No clinical data presented.
Technological Characteristics
Materials: ABS or Cyrex (spike body), LSR (stepped silicone plug), Polyester (body), Polycarbonate (graduated adapter). Connectivity: ISO 594-1 & -2 compliant luer. Sterilization: Gamma or E-Beam radiation. Form factor: Slip-fit graduated adapter with integrated needless connector. Mechanical principle: Manual fluid path activation via luer connection; silicone plug seal.
Indications for Use
Indicated for patients requiring urinary catheterization; facilitates fluid transfer into or out of the bladder. For use as an accessory to a urinary catheter. Single procedure use only; connection to catheter not to exceed 30 days.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Y-Clave Connector with Integrated Check Valve (K042182)
Urethral Catheter and Urine Catheter Kit (K041464)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for ICU Medical, Inc. The logo features a gray horse on the left side of the image. The text "ICU Medical, Inc." is in blue and is located to the right of the horse.
# ICU MEDICAL INC. 4455 Atherton Drive
Salt Lake City, Utah (801) 264 - 1332, Phone (801) 264 - 1755, Fax Mr. Tracy S. Best, Sr. Regulatory Affairs Specialist Preparation Date: May 06, 2008
# Traditional 510(K) - Summary of Safety and Effectiveness for the:
Trade Name: Graduated Adapter with Clave® Common Name: Needless connector with graduated adapter Accessory to Urological Catheter 21 CFR 876.5130 (EZD) Classification Name:
## Legally Marketed Predicate Devices for Substantial Equivalence:
*K915571 - Clave Connector - ICU Medical Inc.
- * K970855 Clave Connector ICU Medical, Inc.
- * K 941190 Clave Needless Connector ICU Medical, Inc.
- *K964435 Primary IV Set ICU Medical, Inc.
- *K042182 Y-Clave Connector with Integrated Check Valve ICU Medical, Inc.
- *K041464 Urethral Catheter and Urine Catheter Kit Busse Hospital Disposables
### Rationale for SE:
The Clave is a needless adapter that is used individually or in combination with other devices more than 6 million times annually around the world. The adaption of a graduated adapter is new to the clave, but accessories integrated with the Clave to make patient care easier for the caregiver is not new concept. As shown in K964435 and K941190, these submissions included adaptors for different purposes to use the Claye that fall outside the original intended use as defined in K915571. I.V. Sets and various componentry integrated the use of the Clave as requested by the physician and as needed for better patient care. ICU Medical is a custom kit manufacturer in the disposable infusion therapy market. The Graduated Adapter with Clave has been developed for the same purpose; to help make the caregiver's job easier and to simplify patient care by providing this presterilized device for those physicians that use them. The Urine Catheter Kit from Busse (K041464) is an example of how companies develop different kits for physicians for the care of patients.
### Description of Submitted Device:
The Clave with Graduated Adapter is a device that allows the Urologist a device that can "slip-fit" on a urology catheter. This is then used to infuse fluids into the bladder for procedures as the physician determines. The Clave can connect to a threaded device or any ISO 594-1 & -2 compliant luer device. When the Clave is activated, fluid may then pass through the graduated adapter. When the syringe or luer is disconnected, the Clave can withstand back pressures as much as 45 psi and thereby preclude fluid leakage coming from the bladder via the catheter and attachment.
{2}------------------------------------------------
The submitted device consists of a Clave which has a spike, stepped silicone plug and body that is then solvent-bonded to a graduated adapter using cyclohexanone, dichloride or validated equivalent.
#### Intended Uses of the ICU Medical Syringe:
ICU Medical Graduated Adapter with Clave® is indicated as an accessory to a urinary catheter which facilitates the transfer of fluids into or out of the bladder via connection to the graduated adapter. This device for single procedure use only. Connection to catheter not to exceed 30 days.
| Component: | ICU Medical<br>Graduated Adapt.<br>with Clave | ICU Medical Clave<br>Needless Connector | Urine Catheter Kit by<br>Busse Hospital<br>Disposables |
|--------------------------|-----------------------------------------------|---------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Spike Body: | ABS or Cyrex | ABS or Cyrex | N/A |
| Stepped Silicone<br>Plug | LSR – Liquid Silicone<br>Rubber | LSR – Liquid Silicone<br>Rubber | N/A |
| Body: | Polyester | Polyester | N/A |
| Graduated<br>Adapter | Polycarbonate | N/A | Polypropylene connectors,<br>PVC catheters, Polycarbonate<br>specimen containers |
| Sterilization<br>method: | Gamma / E-Beam | Gamma / E-Beam | Gamma or ETO |
| Other component: | N/A | Various: polycarbonate,<br>PVC, polypropylene, ABS,<br>Cyrex & more | PVC catheter, etc. |
| 510(k) Approval | This submission | K915571; K970855;<br>K941190; K964435;<br>K045182 | K041464 |
### Technological Characteristics and Substantial Equivalence Table:
The operational characteristics are equivalent: by manually inserting the graduated adapter into the open end of the catheter, then connecting the Clave to a syringe or an ISO 594-1; -2 compliant Luer to allow the physician to infuse or withdraw fluids.
### Safety and Performance:
ICU Medical complies with published international standards to ensure the commonality of connecting components. All standards complied with are listed in the CDRH Cover Sheet. Additionally, ICU Medical's Sterility Assurance Level, (SAL) has an established and validated history of meeting the 10° level. This single use Graduated Adapter with Clave are packaged to ensure conformity with ISO 10993-1 for biocompatible materials and ISO 11137-1 for sterilization by radiation.
### Conclusion:
The materials, performance, and operational features of both the submitted device and the predicate devices are substantially equivalent and are safe and effective for their intended use.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, which is a traditional symbol of medicine.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
MAY 13 2008
Mr. Tracy Best Sr. Regulatory Affairs Specialist ICU Medical, Incorporated 4455 Atherton Drive SALT LAKE CITY UT 84123
Re: K080007
Trade/Device Name: Graduated Adapter with Claye® Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZD and KNY Dated: April 4, 2008 Received: April 7, 2008
Dear Mr. Best:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image is a circular logo or emblem. The text "1906-2006" is at the top of the logo. The letters "FDA" are in the center of the logo, with the word "Centennial" underneath. There are three stars below the word "Centennial". The text around the circle is difficult to read.
Protecting and Promoting Public Health
{4}------------------------------------------------
#### Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Aut or any Federal statutes and regulations administered by other Federal agencies. You must to comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements s set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the ed be on product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a logally marketed predicate device results in a classification for your device and thus, cormorts your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance a t one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRF s Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 24/0276-3474. For questions regarding the reporting of device adverse events (Medical Device 2-eo-2, ing (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You portun other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2011 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
# 510(k) Number (if known): K080007
Device Name: Graduated Adapter with Clave®
#### Indications for Use:
ICU Medical Graduated Adapter with Clave® is indicated as an accessory to a urinary catheter which facilitates the transfer of fluids into or out of the bladder via connection to the graduated adapter. This device for single procedure use only. Connection to catheter not to exceed 30 days.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR .
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Signature
on Sign-Off) (Divis Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.