LITHOSTAR MODULARIS WITH SHOCKWAVE SYSTEM CPLUS

K033335 · Siemens Medical Solutions USA, Inc. · MMZ · Nov 25, 2003 · Gastroenterology, Urology

Device Facts

Record IDK033335
Device NameLITHOSTAR MODULARIS WITH SHOCKWAVE SYSTEM CPLUS
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeMMZ · Gastroenterology, Urology
Decision DateNov 25, 2003
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.4890
Device ClassClass 2
AttributesTherapeutic

Intended Use

LITHOSTAR Modularis with Shockwave System Cplus is a urologic procedures system which can be used for a wide range of applications. It is a transportable system. The system is capable of both diagnostic and therapeutic procedures of urolithiasis in the kidney and upper ureter. The primary use of this device is for the fragmentation of urinary tract stones, i.e. renal calyceal stones, renal pelvis stones, and upper ureteral stones by extracorporeal shock wave lithotripsy (ESWL). In addition to ESWL use, the unit is designed for patient placement in positions which facilitate urological and diagnostic procedures.

Device Story

Transportable urologic system for extracorporeal shockwave lithotripsy (ESWL) and diagnostic positioning. Modification of predicate device increases shockwave head penetration depth from 120mm to 140mm to accommodate obese patients. Device delivers shockwaves to fragment urinary tract stones (renal calyceal, renal pelvis, upper ureteral). Operated by clinical staff in a medical setting. Healthcare providers use the system to position patients and target stones for fragmentation, facilitating stone removal and potentially reducing the need for invasive surgery.

Clinical Evidence

Confirmatory clinical study conducted to evaluate safety and effectiveness. Results suggest the device is safe and effective for treating urinary tract stones and is comparable to the predicate device (P870018, Supplement 21).

Technological Characteristics

Extracorporeal shockwave lithotripter; transportable urologic table system. Modification increases shockwave penetration depth to 140mm. Fundamental technology unchanged from predicate. Shockwave parameters documented per FDA guidance.

Indications for Use

Indicated for diagnostic and therapeutic procedures of urolithiasis in the kidney and upper ureter, specifically fragmentation of renal calyceal, renal pelvis, and upper ureteral stones via ESWL. Contraindicated in pregnancy, coagulation abnormalities, patients on anticoagulants, arterial calcification/aneurysm in shockwave path, distal urinary tract obstruction, or anatomy precluding focusing (e.g., severe obesity, excessive spinal curvature).

Regulatory Classification

Identification

A urological table and accessories is a device that consists of a table, stirrups, and belts used to support a patient in a suitable position for endoscopic procedures of the lower urinary tract. The table can be adjusted into position manually or electrically.

Special Controls

*Classification.* (1) Class II (special controls) for the electrically powered urological table and accessories. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.(2) Class I for the manually powered table and accessories, and for stirrups for electrically powered table. The device subject to this paragraph (b)(2) is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOV 2 5 2003 LITHOSTAR MODULARIS Special 510(k) Siemens Medical Solutions, Inc. K033335 Page 1 of 2 ### SECTION 3 - 510 (K) SUMMARY ## Submitter's information Name Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway Malvern, PA 19355 Contact Person: Debbie Peacock Name: Technical Specialist. Requlatory Affairs Title: (610) 448-1773 Phone: Fax: (610) 448-1787 debra.peacock@siemens.com E-mail: Date summary prepared 10/13/03 #### Device names Trade Name LITHOSTAR MODULARIS with Shockwave System Colus #### Common or Classification Name LITHOSTAR MODULARIS #### A. Legally marketed device to which the device is substantially equivalent The predicate device is the Siemens LITHOSTAR MODULARIS, P870018, Supplement 21. #### B. Description of device Siemens LITHOSTAR MODULARIS with Shockwave System Collis is a modification to our predicate device. The shockwave head has been modified to provide an increased penetration depth for a more obese patient population. The penetration depth has been increased from 120mm to 140mm.. #### C. Indications for Use LITHOSTAR MODULARIS with Shockwave System Calle is a urologic procedures system which can be used for a wide range of applications. It is a transportable system. The system is capable of both diagnostic and therapeutic procedures of urolithiasis in the kidney and upper ureter. The primary use of this device is for the fragmentation of urinary tract stones, i.e. renal calvceal stones, renal pelvis stones, and upper ureteral stones by extracorporeal shockwave lithotripsv (ESWL). In addition to ESWL use, the unit is designed for patient placement in positions which facilitate urological and diagnostic procedures. {1}------------------------------------------------ LITHOSTAR MODULARIS Special 510(k) Siemens Medical Solutions, Inc. K033335 Page 2 of 2 ## D. Contraindications Do not use the LITHOSTAR MODULARIS in patients with: - Confirmed or suspected pregnancy. ● - Coaqulation abnormalities (as indicated by abnormal prothrombin time, . partial thromboplastin time, or bleeding time) or those currently receiving anticoagulants (including aspirin). - Arterial calcification or vascular aneurysm in the lithotripter's shockwave . path. - Urinary tract obstruction distal to the stone. . - Anatomy which precludes focusing the device at the target stone, such as . severe obesity or excessive spinal curvature. ## E. Warnings and Precautions A listing of Warnings and Precautions are included in labeling. ## F. Technological characteristics The modification has not altered the fundamental technology of the predicate device. ## G. Assessment of non-clinical performance data Device shockwave parameters were measured and documented according to Guidance for the Content of Premarket Notifications (510(k)s) for Extracorporeal Shockwave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.. ### H. Assessment of clinical performance data The confirmatory clinical study suggests that treatment of urinary tract stones with the LITHOSTAR MODULARIS with Shockwave System Colls is safe and effective and comparable to the currently approved LITHOSTAR MODULARIS (P870018, Supplement 21). {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that form the wings and body of the bird. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # NOV 25 2003 Ms. Debra A. Peacock Technical Specialist, Regulatory Affairs Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway MALVERN PA 19355 Re: K033335 Trade/Device Name: Lithostar Modularis with Shockwave System Colus Regulation Number: 21 CFR§ 876.5990 Regulation Name: Extracorporeal Shock Wave Lithotripters, Urological Regulatory Class: II Product Code: 78 LNS Regulation Number: 21 CFR\$ 876,4890 Regulation Name: Urological Table and Accessories Regulatory Class: II (Exempt) Product Code: 78 MMZ Dated: November 17, 2003 Received: November 18, 2003 Dear Ms. Peacock: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {4}------------------------------------------------ #### INDICATIONS FOR USE SECTION 2 510(k) Number (if known): K033335 Device Name: LITHOSTAR MODULARIS WITH SHOCK WAVE SYSTEM Cplus Indications For Use: LITHOSTAR Modularis with Shockwave System Coplis is a urologic procedures system which can be used for a wide range of applications. It is a transportable system. The system is capable of both diagnostic and therapeutic procedures of urolithiasis in the kidney and upper ureter. The primary use of this device is for the fragmentation of urinary tract stones, i.e. renal calyceal stones, renal pelvis stones, and upper ureteral stones by extracorporeal shock wave lithotripsy (ESWL). In addition to ESWL use, the unit is designed for patient placement in positions which facilitate urological and diagnostic procedures. (Please do not write below this line - continue on another page if needed) ============================================================================================================================================================================== Concurrence of the CDRH, Office of Device Evaluation (ODE) V Prescription Use OR Over-The-Counter Use (Per 21 CFR 801.109) David A. Lozano (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number K033335 CONFIDENTIAL
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...