COMEG ENDOSCOPY GRASPING FORCEPS
Device Facts
| Record ID | K970489 |
|---|---|
| Device Name | COMEG ENDOSCOPY GRASPING FORCEPS |
| Applicant | Comeg Endoscopy |
| Product Code | FFL · Gastroenterology, Urology |
| Decision Date | Mar 19, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 876.4680 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
The intended use for the foreign body grasping forceps is in urological procedures to endoscopically remove, manipulate or grasp calculi (i.e. stones) and other foreign objects.
Device Story
COMEG Endoscopy Grasping Forceps; stainless steel surgical instruments; 7 and 9 French diameters; 40 cm length; alligator jaw with double action mechanism. Used in urological procedures by physicians to endoscopically grasp, manipulate, or remove calculi and foreign bodies. Device provides mechanical assistance for stone retrieval; enhances procedural efficiency; minimizes tissue trauma during foreign body extraction.
Clinical Evidence
Bench testing only; no clinical data provided. Substantial equivalence established through comparison of physical and functional characteristics.
Technological Characteristics
Materials: Stainless steel. Configuration: 7 and 9 French diameters, 40 cm length. Mechanism: Double action alligator jaw. Energy source: Manual/mechanical. Sterilization: Not specified.
Indications for Use
Indicated for use in urological procedures for endoscopic removal, manipulation, or grasping of calculi (stones) and foreign objects in patients requiring such intervention.
Regulatory Classification
Identification
A ureteral stone dislodger is a device that consists of a bougie or a catheter with an expandable wire basket near the tip, a special flexible tip, or other special construction. It is inserted through a cystoscope and used to entrap and remove stones from the ureter. This generic type of device includes the metal basket and the flexible ureteral stone dislodger.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
- Karl Storz Grasping forceps for removal of foreign bodies
Related Devices
- K191900 — Single Use Grasping Forceps · Zhejiang Chuangxiang Medical Technology Co., Ltd. · Mar 27, 2020
- K962378 — OLYMPUS FG-45/46-1 GRASPING FORCEPS · Olympus America, Inc. · Aug 2, 1996
- K970121 — STONE DISLODGER, BASKET · Boston Scientific Corp · Apr 2, 1997
- K152802 — Grasping Forceps · Micro-Tech (Nanjing) Co., Ltd. · Jan 21, 2016
- K962485 — OLYMPUS FG SERIES OF RAT TOOTH GRASPING FORCEPS (ENT) · Olympus America, Inc. · Aug 20, 1996