K232969 · Ningbo Xinwell Medical Technology Co., Ltd. · PKL · Jun 7, 2024 · Gastroenterology, Urology
Device Facts
Record ID
K232969
Device Name
Disposable Hemostatic Clips
Applicant
Ningbo Xinwell Medical Technology Co., Ltd.
Product Code
PKL · Gastroenterology, Urology
Decision Date
Jun 7, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of : 1) Endoscopic marking; 2) Hemostasis for: A) Mucosal/sub-mucosal defects < 3cm. B) Bleeding ulcers. C) Polyps < 1.5cm in diameter. D) Diverticula in the colon. E) Arteries < 2 mm. F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. 3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively 4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel
Device Story
Disposable Hemostatic Clips consist of a delivery system (slide bar, rotation head, spring tube) and a stainless steel clip assembly. Operated by a physician via a flexible endoscope; the device allows for clip opening/closing (up to 5 times) for lesion repositioning before final deployment. Used for endoscopic marking, hemostasis of GI defects/ulcers/polyps/arteries, closure of small perforations, and anchoring feeding tubes. The device is provided sterile for single use. Clinical benefit includes effective management of GI bleeding and tissue defects through minimally invasive endoscopic intervention.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including appearance, rotation performance, release force, clamping strength, relocation, tensile strength, mechanical integrity, scope compatibility, and three-ring handle strength. Biocompatibility testing performed per ISO 10993-1.
Indicated for adult patients requiring endoscopic clip placement in the GI tract via flexible endoscope for marking, hemostasis (mucosal/sub-mucosal defects <3cm, bleeding ulcers, polyps <1.5cm, colonic diverticula, arteries <2mm, post-resection prophylaxis), closure of GI perforations <20mm, or anchoring jejunal feeding tubes.
Regulatory Classification
Identification
A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.
Special Controls
*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 7, 2024
Ningbo Xinwell Medical Technology Co., Ltd % Diana Hong General Manager Mid-Link Consulting Co, Ltd. P.O. Box 120-119 Shanghai. 200120 CHINA
Re: K232969
> Trade/Device Name: Disposable Hemostatic Clips Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: Class II Product Code: PKL Dated: September 21, 2023 Received: May 9, 2024
Dear Diana Hong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Sivakami Venkatachalam -S
for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K232969
Device Name
## Disposable Hemostatic Clips
Indications for Use (Describe)
The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of : 1) Endoscopic marking; 2) Hemostasis for: A) Mucosal/sub-mucosal defects < 3cm. B) Bleeding ulcers. C) Polyps < 1.5cm in diameter. D) Diverticula in the colon. E) Arteries < 2 mm. F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection. 3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively 4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K232969
- 1. Date of Preparation: 06/03/2024
- 2. Sponsor Identification
Ningbo Xinwell Medical Technology Co., Ltd. No.188 Binjiang Road, Cixi High-tech Industrial Development Zone, Cixi City, Zhejiang,P.R. China
Establishment Registration Number: 3013526170
Contact Person: Lei Xi Position: RA manager Tel: +86-021-34693382 Email: xilei@xinhaigroup.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Tingting Su (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
- 4. Identification of Proposed Device
Trade Name: Disposable Hemostatic Clips Common Name: Hemostasis Clipping Device
Regulatory Information Classification Name: Hemorrhoidal Ligator;
{5}------------------------------------------------
Classification: II, Product Code: PKL, Regulation Number: 21CFR 876.4400 Review Panel: Gastroenterology/Urology,
#### Indications for Use:
The Disposable Hemostatic Clips is indicated for Endoscopic clip placement within the Gastrointestinal tract in adult populations only via a straight or side viewing flexible endoscope for the purpose of:
- 1) Endoscopic marking;
- 2) Hemostasis for:
- A) Mucosal/sub-mucosal defects < 3cm.
- B) Bleeding ulcers.
- C) Polyps < 1.5cm in diameter.
- D) Diverticula in the colon.
E) Arteries < 2 mm.
F) Prophylactic clipping to reduce the risk of delayed bleeding post lesion resection.
3) As a supplementary method, closure of GI tract luminal perforations <20mm that can be treated conservatively
4) Anchoring to affix jejunal feeding tubes to the wall of the small bowel.
- 5. Device Description
The proposed device is consisted of a clip section. The delivery section. The delivery section include: slide bar, slip ring, protective sleeve, rotation head, coated spring tube, slip ring holder, traction assembly, connecting tube, decoupled piece, decoupled shrapnel, rotation sleeve. And the clip section include: clip, sliding block, shaft and rocker arm. The clip is made of stainless steel. The clip can be opened and closed more than five times prior to deployment, aiding in repositioning of the lesion site. The proposed device is available in single ring and three ring models, and each case is divided into rotatable and non-rotatable clip. The effective length of the proposed device includes 1250mm, 1650mm, 2350mm and the opening width is 9mm, 11mm. 13mm and 16mm. The proposed devices are provided in sterile and single use.
- Identification of Predicate Device 6.
Predicate Device 510(k) Number: K202333 Product Name: Lockado™ Repositionable Hemostasis Clip
{6}------------------------------------------------
#### Non-Clinical Test Conclusion 7.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ♪ ISO 10993-3:2014 Biological evaluation of medical devices Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- > ISO 10993-5:2009 Biological evaluation of medical devices- Part 5: Test for in vitro cytotoxicity
- > ISO 10993-6:2016 Biological evaluation of medical devices -- Part 6: Tests for local effects after implantation
- > ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene Oxide Sterilization Residuals
- > ISO 10993-10:2021 Biological evaluation of medical devices Part 10: Tests for skin sensitization
- > ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- > ISO 10993-23:2021 Biological evaluation of medical devices Part 23: Tests for irritation
- > ASTM F1929-15 Standard test method for detecting seal leaks in porous medical packaging by dye penetration.
- > ASTM F88/F88M-15 standard method for seal strength of flexible barrier materials
- > ASTM F1886 / F1886M-16, Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- > USP <85> Bacterial Endotoxins Test
- A ASTM D4169-22 Standard Practice for Performance Testing of Shipping Containers and Systems
#### Performance testing
The following bench tests were performed on the Disposable Hemostatic Clips:
- . Appearance
- Rotation performance •
- Release force
- . Clamping strength
- . Relocation
- . Tensile strength
- . Dimension
- . Mechanical integrity of clip assembly
- Scope compatibility/usability
- . Endoscope damage
- . Torque
- Clip approach .
- Three ring handle strength
Above all tests were passed and demonstrated the result can meet the product requirements.
{7}------------------------------------------------
### Biocompatibility testing
The biocompatibility evaluation for the proposed device was conducted in accordance with FDA Guidance, Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" issued on September 4, 2020. The following tests were conducted for the clip component of Disposable Hemostatic Clips:
- . Cytotoxicity
- Irritation .
- Skin Sensitization .
- Acute Systemic Toxicity
- . Sub-acute Systemic Toxicity
- . Genotoxicity
- . Pyrogenicity
- Implantation Test
The following tests were conducted for the delivery component of Disposable Hemostatic Clips:
- . Cytotoxicity
- Irritation ●
- . Skin Sensitization
- . Acute Systemic Toxicity
- . Pyrogenicity
#### Sterility, Shipping, and Shelf -life
The proposed device sterilization process using Ethylene Oxide (EO) has been validated in accordance with ISO 11135:2014 to achieve a sterility assurance level (SAL) of 106. EO and Ethylene Chlorohydrin (ECH) residuals were below the limits specified in ISO 10993-7:2008. Bacterial Endotoxin Levels were below the level of 2.15 EU/device in accordance with USP <85>. Both baseline and accelerated shelf life testing were conducted demonstrating the device will perform as intended to support the proposed 3 year shelf-life.
- . Package integrity test - after environmental conditioning, simulated transportation testing in accordance to ASTM D4169-22 on final, packaged, and sterile device.
- Sterile Barrier Packaging performed on the proposed device: .
- Seal Strength ASTM F88/F88-15 o
- Dye penetration ASTM F1929-15 O
- Visual Inspection ASTM F1886/F1886M-16 O
- . Shelf-life of 3-years is validated using FDA recognized standard ASTM F1980 -16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
{8}------------------------------------------------
#### Clinical Test Conclusion 8.
No clinical study is included in this submission.
- 9. Substantially Equivalent (SE) Comparison
| Table 1. General Comparison | | | |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Item | Proposed Device | Predicate Device | Remark |
| Product Code | PKL | PKL | Same |
| Regulation Number | 21CFR 876.4400 | 21CFR 876.4400 | Same |
| Indications for Use | The Disposable<br>Hemostatic Clips is indicated<br>for Endoscopic clip placement<br>within the Gastrointestinal tract<br>in adult populations only via a<br>straight or side viewing flexible<br>endoscope for the purpose of :<br>1) Endoscopic marking;<br>2) Hemostasis for:<br>A) Mucosal/sub-mucosal<br>defects < 3cm.<br>B) Bleeding ulcers.<br>C) Polyps < 1.5cm in<br>diameter.<br>D) Diverticula in the<br>colon.<br>E) Arteries < 2 mm.<br>F) Prophylactic clipping to<br>reduce the risk of delayed<br>bleeding post lesion resection.<br>3) As a supplementary<br>method, closure of GI tract<br>luminal perforations <20mm<br>that can be treated<br>conservatively<br>4) Anchoring to affix jejunal<br>feeding tubes to the wall of the<br>small bowel; | The Lockado™ Repositionable<br>Hemostasis Clip is indicated for<br>Endoscopic clip placement<br>within the Gastrointestinal tract<br>in adult patients only via a<br>straight or side viewing flexible<br>endoscope for the purpose of:<br>(1) Endoscopic marking,<br>(2) Hemostasis for<br>(a)Mucosal / sub-mucosal<br>defects < 3cm,<br>(b) Bleeding ulcers,<br>(c) Polyps < 1.5cm in diameter,<br>(d) Diverticula in the colon,<br>(e) Arteries < 2 mm<br>(f) Prophylactic clipping to<br>reduce the risk of delayed<br>bleeding post lesion resection;<br>(3) As a supplementary method,<br>closure of GI tract luminal<br>perforations <20mm that can be<br>treated conservatively<br>(4) Anchoring to affix jejunal<br>feeding tubes to the wall of the<br>small bowel; | Same |
| Configuration | Delivery system and clip | Delivery system and clip | Same |
| | assembly | assembly | |
| Open width | 9mm, 11mm, 13mm and 16mm | 8mm, 11mm and 16mm, 22mm | Different |
| Minimal working channel | 2.8mm | 2.8mm | Same |
| Working length | 1250mm, 1650mm, 2350mm and 2700mm | 1650mm 1950mm, 2350mm and 2700mm | Different |
| Ring number | Single ring and three rings | Single ring | Different |
| Single Use | Single Use | Single Use | Same |
| Labeling | Comply with 21 CFR Part 801 | Comply with 21 CFR Part 801 | Same |
| Material | | | |
| Clip assembly | Stainless Steel (SUS316L)<br>Stainless Steel (SUS631)<br>Stainless Steel (SUS316F) | | |
| Traction assembly | Stainless Steel (SUS304) | | |
| Outer tube assembly | Stainless Steel (SUS304)<br>Stainless Steel (SUS316F)<br>Polyethylene (PE) | | |
| Slide bar | Acrylonitrile Butadiene Styrene (ABS) | Unknown | Different |
| Slip ring | Thermoplastic Elastomer (TPE) | | |
| Rotation head | Acrylonitrile Butadiene Styrene (ABS) | | |
| | Thermoplastic Elastomer (TPE) | | |
| Protective sleeve | High pressure-low density polyethylene (LDPE) | | |
| Biocompatibility for Delivery system | | | |
| Cytotoxicity | No cytotoxicity | | |
| Skin Sensitization | No skin sensitization | | Same |
| Irritation | No irritation | Comply with ISO 10993 standards | |
| Acute Systemic Toxicity | No acute toxicity | | |
| Sub-acute Systemic Toxicity | No sub-acute toxicity | | |
| Biocompatibility for clip assembly | | | |
| Cytotoxicity | No cytotoxicity | | |
| Skin Sensitization | No skin sensitization | | Same |
| Irritation | No irritation | Comply with ISO 10993 standards | |
| Acute Systemic Toxicity | No acute toxicity | | |
| Sub-acute Systemic Toxicity | No sub-acute toxicity | | |
| Implantation | No abnormal reaction | | |
| Genotoxicity | No genotoxicity | | |
| Sterilization | | | |
| Method | Ethylene oxide | Ethylene oxide | Same |
| SAL | $10^{-6}$ | $10^{-6}$ | Same |
| Endotoxin Limit | 20EU | 20EU | Same |
Table 1. General Comparison
{9}------------------------------------------------
{10}------------------------------------------------
#### Different - Open width
The open width for proposed device is available in 9mm, 11mm, 13mm and 16mm four different specifications. The predicate device is available in 8mm, 11mm 16mm and 22mm four different specifications. The 9mm and 13mm specification is different from predicate device. However, these specifications can be covered in the open width range of predicate device. Different working length devices will be selected by physician per patient's condition. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.
### Different - Working length
The working length for proposed device is available in 1250mm, 2350mm and 2700mm. The working length of predicate device is available in 1650mm 1950mm and 2700mm. Different working length devices will be selected by physician per patient's condition. The length for proposed device of 1650mm, 2350mm and 2700mm can be covered by predicate device length range. Although 1250mm effective length device are not included in the effective length range of the predicate device, it is very close to the minimum length of predicate device. Therefore, this difference will not raise new questions on safety and effectiveness of the proposed device.
#### Different- Ring number
The proposed device has two models of singl…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.